Displaying reports 49321-49340 of 83212.Go to page Start 2463 2464 2465 2466 2467 2468 2469 2470 2471 End
Reports until 16:06, Thursday 16 March 2017
LHO General
patrick.thomas@LIGO.ORG - posted 16:06, Thursday 16 March 2017 (34874)
Ops Day Shift Summary
TITLE: 03/16 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 68Mpc
INCOMING OPERATOR: Corey
SHIFT SUMMARY: Remained locked entire shift. TCS ITMY CO2 guardian took us out of observing multiple times at the beginning of the shift (accepted SDF differences attached). Bubba ran the fire pumps and I ran a2l when LLO lost lock.
LOG:

16:04 - 16:31 UTC: TCS ITMY CO2 guardian periodically taking us out of observing.
16:37 UTC LLO down. Bubba running fire pumps. Out of observing.
16:41 UTC running a2l.
16:46 UTC Bubba done running fire pumps. a2l done.
16:47 UTC Back to observing.
17:03 UTC GRB. LLO down.
18:02 UTC Betsy and Vern to both mid stations.
18:12 UTC Karen to optics lab
18:48 UTC Karen out of optics lab
19:33 UTC Cheryl to optics lab
19:47 UTC Elisabeth has car parked by H2 enclosure to move cables from enclosure to car and then mid Y.
20:00 UTC Elisabeth to mid Y
20:19 UTC Bubba to mid X to drop off wire
20:27 UTC Jeff B. driving to receiving, opening outside rollup door, putting ladder inside.
20:29 UTC Chandra to mid Y to check on thermocouples in discharge lines
20:35 UTC Jeff B. done
20:36 UTC Elisabeth back from mid Y
20:45 UTC Chandra back
21:29 UTC Gave Carlos remote access to check whatsup
21:52 UTC Betsy and Hugh to mid X
22:09 UTC Marc and Elisabeth to mid Y
22:24 UTC Cheryl done in optics lab
22:33 UTC Marc and Elisabeth back
Images attached to this report
H1 SEI
hugh.radkins@LIGO.ORG - posted 10:38, Thursday 16 March 2017 (34869)
BRSY on Board T240 -- Leg Insulation has helped!

Attempting to get the best tilt subtracted ground motion for sensor correction to the ISI has a seismometer piggy backing on the BRS Platform.  However, seismometer difficulties lead to testing other STS2s.  Still, our 'extra' STSs were not doing as well as we would like.  So, we pulled our one spare T240 from storage and ran it on the floor nearby. It was then moved into the BRS enclosure replacing the STS there but low frequencies were noisier than when on the floor.  On Tuesday, the Legs of the Table bolted to the BRS platform, on which the T240 sits, were insulated with 3/4" closed cell foam.

Krishna has looked and reports a factor of 3 improvement in the ASD and the coherence with the BRS is better.  Here attached is before and after spectra comparing the ground seismometer X & Y with the T240 (ADC channels).  The bottom plot has references from 0600utc 13 March before the Tuesday maintenance insulation addition.  The upper plot with current traces are from 1140utc on 15 March (Cool! I did not know DDT would display the T0 of the reference traces, I guess as long as you have no current measurement traces displayed.  19 years and still something to learn!)  We haven't had a low wind time since Maintenance basically so the comparison is not ideal.

Even with the larger ground motion, the Y axis of the now insulated T240 is quieter than before everywhere below 100mHz and closer to the well insulated ISI_STS nearby.  The X axis is noisier but matches the GND STS down to 20mHz, much better than before.  The winds have been about South 50degrees West (hitting the SW crane corner and that may rile up the X axis much more than the Y direction.  Since this temperature sensitivity was apparently a large part of our low frequency noise problem, we'll look at other items in the BRS enclosure and the enclosure itself for thermal improvements.

Images attached to this report
LHO General
patrick.thomas@LIGO.ORG - posted 08:16, Thursday 16 March 2017 (34867)
Ops Day Shift Transition
TITLE: 03/16 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 67Mpc
OUTGOING OPERATOR: Jim
CURRENT ENVIRONMENT:
    Wind: 10mph Gusts, 6mph 5min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.17 μm/s 
QUICK SUMMARY:

All of mid X and some corner station FMCS channels are not being seen from CDS.
H1 General
jim.warner@LIGO.ORG - posted 07:55, Thursday 16 March 2017 (34866)
Shift Summary

TITLE: 03/16 Owl Shift: 07:00-15:00 UTC (00:00-08:00 PST), all times posted in UTC
STATE of H1: Observing at 65Mpc
INCOMING OPERATOR: Cheryl
SHIFT SUMMARY:
LOG:
Another quiet night. 

LHO General
corey.gray@LIGO.ORG - posted 00:00, Thursday 16 March 2017 (34859)
EVE Operator Summary

TITLE: 03/15 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 68Mpc
INCOMING OPERATOR: Jim
SHIFT SUMMARY:

Windy night with H1 Range all over the place, but have been locked for almost 8hrs.
LOG:

H1 ISC (ISC)
hang.yu@LIGO.ORG - posted 22:59, Wednesday 15 March 2017 - last comment - 11:58, Thursday 16 March 2017(34863)
SRM sensing vs. differetial ITM lens

We tried to simulate the effects of differential ITM thermal lens on SRM sensing. The idea of differential ITM thermal lensing is motivated by alog 34853

In the attachments we considered 4 different SRC geometries corresponding to G={0.76, 0.80, 0.83, 0.86} or one-way gouy phase={20.1, 18.6, 17.0, 15.3}, respectively. Note that here G=(A+D)/2, which is different from the stability g=arccos(gouy phase) = sqrt((A+D+2)/4). 

*In each plot, the top panel showed the recycling gain relative to nominal (we defined the gain as the ratio of E_{-9}^ast E_{9} inside/outside PRC).

*The left two panels showed the AS36 signal as function of AS port gouy phase, with the upper panel showing the abs of AS36 to SRM pitch response, and the lower panel the difference of SRM and BS demodulation phase. The orange trace is for the nominal setting (based on the finesse input file T1300904), and the blue trace when differential ITM lensing introduced. 

*The right two panel show an anatomy of the AS36 signal, i.e. the beat note of <9_00, 45_01>, <9_02, 45_01>, <9_01, 45_00>, <9_01, 45_02>. For conciseness we do not show the contribution from the negative sidebands. 

==================================================================

Some thing might be interesting to note:

1. When the ifo is at its nominal setting, the dominating AS36 signal comes from the beat between <9_00, 45_01>. This pair stays roughly constant w.r.t. different ITM diff lensing. However, because 9_00 is weak, when we add differential lensing, both <9_02, 45_01> and <9_01, 45_00> can be as large as, or even larger than <9_00, 45_01>, making the AS36 signal very messy. 

2. If SRC gouy phase >~ 18 deg, if we happen to put the two WFS at AS gouy phase around 35 deg and 125 deg, we would basically have no srm signal at all. If this is indeed the cause, increasing SR2-SR3 distance alone might not be sufficient to solve the sensing problem. 

3. If SRC gouy phase <~ 18 deg, the SRM signal becomes stronger as we adding diff lensing. However, it also becomes more degenerate with BS signal whose response is ~ 100 greater than SRM. If non of the WFS are placed at AS gouy phase [75, 140] deg, the SRM signal, despite higher amplitude, will also be lost due to degenerate demod phase with BS. 

4. The diff lensing can reduce recycling gain. For ~75 km extra lens added to ITMX, the recycling gain decreases by ~5%. 

5. Bad centering loops might not be necessary in explaining the bad srm sensing. As argued in T0810007, the carrier doesn't see ITM RoC mismatch to 1st order whereas the sidebands do. The effects due to carrier junk might not be as significant as sideband homs. 

Images attached to this report
Comments related to this report
kiwamu.izumi@LIGO.ORG - 05:24, Thursday 16 March 2017 (34864)

Thank you, Hang, for doing this. A few questions/suggestions.

  • Have you looked at the carrier recycling gain as well?
    • I am asking this because the H1 input file assumes ultra low loss on test masses.
  • What is the definition of the angle shown in the anatomy panel?
  • Why does the demod phase difference vary as a function of the Gouy phase? I naively thought that the demod and Gouy phases were independent...
hang.yu@LIGO.ORG - 11:58, Thursday 16 March 2017 (34872)ISC

Thanks Kiwamu for the comments. 

* According to the simulation, the carrier recycling gain (which we approximated by the total DC power in PRC/ total DC before entering the ifo) does not vary significantly w/ extra ITM thermal lensing, which might be a consequence that the carrier doesn't see RoC mismatch to first order? The simulation is attached. The recycling gain doesn't care that much about the SRC so I just used the nominal 20 deg gouy SRC for plotting. 

And you are right, the Finesse input file assumes an extremely loss ifo; for RF9 the build up is essentially the same as the build up inside an ideal FP cav w/ input T=0.03 and output T=0. We haven't touched the losses for now, however, as adding loss somewhere to reduce the recycling gain can be a very degenerate process. But as suggested in Aidan's log the extra absorption is very localized, meaning a thermal lens should not capture all the physical effects. Our next step will be adding some extra ITMX loss to our parameter searching space and explore what effect that may cause. 

* For angles in the anatomy panel, we defined (w/ abused notation) <f1_{TEM nm}, f2_{TEM n'm'}> = (geometrical pre-factor) E*(f1)_{nm} ·E(f2)_{n'm'}, where E represent the complex-valued field. Therefore the beatnote at (f2-f1) is also complex. And the angle of this complex value is what we plotted. Or in other words, this angle is the phase difference of the two fields that beat to 36. 

* We defined the AS response as AS_I + 1j * AS_Q, and we called the demodulation phase as the phase angle of this complex quantity. Or more specifically it is the phase at a given gouy phase that maximizes the signal in the I-quadrature. Because AS_I and AS_Q sees the AS gouy phase differently, angle(AS_I + 1j*AS_Q) is thus gouy phase dependent. 

Images attached to this comment
LHO General
corey.gray@LIGO.ORG - posted 20:22, Wednesday 15 March 2017 (34862)
Mid Shift Status

Wee bit of a windy night.  Right now we are riding through 30mph sustained winds here at the Corner Station (can hear gusts in Control Room).  

H1 is still locked, but range definitely following the winds.

H1 TCS (TCS)
corey.gray@LIGO.ORG - posted 18:09, Wednesday 15 March 2017 - last comment - 11:04, Thursday 16 March 2017(34861)
TCS ITMx CO2 Channels Dropping H1 From OBSERVING

Recent Laser Setpoint Values Different From Lock To Lock:  Keep ACCEPTING or NOT MONITOR?

Over the last few days it has been noted that there have been TCS Diffs from lock to lock & they have simply been ACCEPTED in SDF.  These Differences have been tiny changes (e^-15) for the (H1:TCS-ITMY_CO2_LASERPOWER_SETPOINT & ...CO2_LSRPWR_SET_POINT_OFFSET).  

TCSx Servo Changes Knocking Us Out of OBSERVING Tonight:  Known Issue

Tonight, we have been bumped out of OBSERVING twice to some automatated servo changes being made to the TCSx.  Drops from Observing were:

Basically it looks like the Input, a couple of filters, & output for two TCS ITMX CO2 filter banks are toggled, and then the offset for the TCSx PZT setpoint is ramped.  

(Currently messaging Nutsinee & she mentions this is a known issue & has been brought to Aidan/Alastair's attention.)

Images attached to this report
Comments related to this report
alastair.heptonstall@LIGO.ORG - 11:04, Thursday 16 March 2017 (34871)

Yes this was seen before.  Changes to values in these channels should not be a reason for the IFO to be taken out of observation.

When the TCS lasers are locked the power output is monitored and fed back to the PZT and chiller temperature.  Because we want to lock the lasers to the nearest stable point at their current temperature, the Guardian script scans the PZT and maps the power output, then it calculates the gradient and uses this to set the gain for the feedback to the PZT.  There are a couple of entries in matrices that are altered as well, though we could in theory have this all in one value.  The gist of this is that there are values that are meant to change every time the laser is locked.

I think we should remove the monitoring of changes to these channels as a reason for dropping out of observation mode.

H1 TCS (TCS)
aidan.brooks@LIGO.ORG - posted 17:39, Wednesday 15 March 2017 - last comment - 10:54, Thursday 16 March 2017(34853)
H1-ITMX_HWS measurement indicates a strong point source thermal lens - need to check for any excess 1064nm

Following Betsy and Nutsinee's efforts to clean up the alignment of the ITMX_HWS probe beam yesterday, we reinitialized the HWSX code. The code has been running continuously since then. Last night there was a lock-loss in the IFO (at 1173593679) after being locked for about 2 hours. 

I pulled the HWS gradient field data from that time and have plotted it in the attached movie. I have rezeroed the HWS so that the reference wavefront was taken about 600s before the lock-loss, i.e when the IFO is hot. All subsequent gradient fields are plotted in reference to this initial state.

I then plot the difference in the gradient field from this reference time at intervals of 60s. Therefore, any thermal lenses you see in the gradient field appear as the negative of what would appear during a lock-acquisition (and when the cold state of the IFO is used for a reference wavefront). However, the heat flow is a linear process with regards to the input power, so analyzing the lens decay is fine for determining absorption.

From the point at which the lock-loss occurs:

Notes:

  1. there are few errant spots out near the edges of the HWS FOV. These are noise - just ignore them. I've not had an opportunity to remove them from the data
  2. At the time of the lock-loss, you can also see a bulk pitch towards the top left - this represents a tilt change in the optic.
  3. By and large, the gradient field shows no systematic noise on it, indicating that this measurement is quite clean.

The gradient field is shown above. The numerically integrated wavefront is shown below. There is some residual tilt in this wavefront.

The lens scale is about 60-70nm over a radius of about 20mm.

The next step is to compare this to COMSOL models of thermal lensing for heat sources of different sizes. This is to determine if this data matches a physical model.

Other possible explanations:

So far, I can only come up with one possible other interpretation. If there is excessive 1064nm light leaking onto the HWS, and the HWS doesn't filter it out enough with the bandpass filter (or this isn't in place), this might cause a systematic error in the HWS measurement. We can test to see if we're getting excess contamination of 1064nm light by leaving the HWS code running and turning off the SLED. The HWS code will record images on the HWS camera and when the IFO lock is lost or reacquired, we'll be able to see if there is excess 1064nm light on the CCD in the same position as the apparent thermal lens.

 

Images attached to this report
Non-image files attached to this report
Comments related to this report
kiwamu.izumi@LIGO.ORG - 05:35, Thursday 16 March 2017 (34865)

Looks familiar to me -- see the animations at the bottom of 29712.

aidan.brooks@LIGO.ORG - 10:01, Thursday 16 March 2017 (34868)

Based on fiducial reference points (baffle edges and earthquake stops) in the image of the ITMX HWS return beam, I estimate that the center of the ITM optic is [-6mm, +9mm] in the HWS coordinate system (as magified up to the ITM).This is illustrated in the attached image.

Orientation: the image is inverted vertically, the top of the optic is at the bottom of the image. We don't have the left-right orientation determined.

Best estimate: the displacement of the point source is about [20mm,25mm] from the optic center - toward the lower part of the optic.

A RH measurement would provide a more precise estimate.

 

Images attached to this comment
aidan.brooks@LIGO.ORG - 10:54, Thursday 16 March 2017 (34870)

Here is a zoom of the OPD. The tilt has been removed and the contours are labeled (they're every 10nm).

Images attached to this comment
Non-image files attached to this comment
LHO General
corey.gray@LIGO.ORG - posted 16:50, Wednesday 15 March 2017 (34857)
Transition To EVE

TITLE: 03/15 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 69Mpc
OUTGOING OPERATOR: Ed
CURRENT ENVIRONMENT:
    Wind: 28mph Gusts, 22mph 5min avg
    Primary useism: 0.07 μm/s
    Secondary useism: 0.23 μm/s

Winds are starting to pick up above 20mph.  And we recovering from a recent 5.4 earthquake in Alaska.
QUICK SUMMARY:

Ed was returning H1 to OBSERVING as I walked in (due to an AK EQ lockloss).  While transitioning, PI Mode#28 immediately & quickly rung up within 2min after we locked---luckily Ed was on it and he was able to adjust the phase to turn it around.

Just finished off going through the Shift Check Sheet.

H1 General
edmond.merilh@LIGO.ORG - posted 16:07, Wednesday 15 March 2017 (34843)
Shift Summary

TITLE: 03/15 Day Shift: 15:00-23:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 0Mpc
INCOMING OPERATOR: Corey
SHIFT SUMMARY:
LOG:
15:!5 Karen and Christrina to Mid stations. Chris Soike to EX Chiller yard

16:06 Chris back from X and heading to Y.

16:55 Vacuum alarm EX_INSTAIR_PT599_PRESS_PSIG. There was  an earlier one as well for EY PT499 " ". This is probably due to Chris visiting chiller yards at each end.

17:33 Chris back 

 

H1 SEI
edmond.merilh@LIGO.ORG - posted 15:49, Wednesday 15 March 2017 - last comment - 16:50, Wednesday 15 March 2017(34854)
Earthquake Report: 5.4 Alaska

EQ seen at: 
USGS: Listed as occurring at 19:43UTC
Seis BLRMS:  ~22:41utc (peaked at 0.4um/s for 0.03-0.1Hz)
Terramon:  Not sure when this posted, but it was at/before arrival time of 3:30utc
Tidal signals start seeing around 22:37utc
Magnitude: 5.4
Location: Semisopochnoi Island, Alaska
Start time: 3:30utc
Lock Status: Lockloss

Comments related to this report
corey.gray@LIGO.ORG - 16:24, Wednesday 15 March 2017 (34858)

H1 & L1 eventually lost lock.

krishna.venkateswara@LIGO.ORG - 16:50, Wednesday 15 March 2017 (34860)SEI

I've attached a plot showing the ETMY seismometer Z signal and a proxy for the interferometer's lock state. Looks like the lockloss happened a few minutes after the peak of the ground motion, which might suggest they are not related. It would be helpful to look at a lockloss plot to assess if this earthquake was responsible or not.

Images attached to this comment
H1 PSL (PSL)
peter.king@LIGO.ORG - posted 13:22, Tuesday 14 March 2017 - last comment - 15:20, Thursday 16 March 2017(34811)
Flow sensors replaced
All four turbine flow sensors for heads 1-4 were replaced with new ones.  The signals from
all four sensors look okay but the one from head 3 is the noisiest.  Some air bubbles were
seen in the flow sensors but seemingly have worked their way out of the sensors by the time
the oscillator lid was put back.

    Going by the froth in the crystal chiller spout, we might still have some air bubbles
in the system.

    A large metal, copper coloured chip came out of the head 1 flow sensor.  The chip was
approximately 2 mm x 2 mm.

    The vortex flow sensors for the power meter and front end laser were re-attached and
released in Beckhoff - they were previously force written.  If problems come up again, we
can force write them again.

    The bypass valve is fully open.



Jason / Peter
Images attached to this report
Comments related to this report
peter.king@LIGO.ORG - 15:20, Thursday 16 March 2017 (34873)
Given the colour of the chip, it is likely that it came from one of the following:
 i.   the pump light block
 ii.  the ASE block, or
 iii. the pump light monitoring block
Displaying reports 49321-49340 of 83212.Go to page Start 2463 2464 2465 2466 2467 2468 2469 2470 2471 End