Displaying reports 53761-53780 of 83242.Go to page Start 2685 2686 2687 2688 2689 2690 2691 2692 2693 End
Reports until 11:13, Friday 07 October 2016
H1 TCS
betsy.weaver@LIGO.ORG - posted 11:13, Friday 07 October 2016 (30306)
TCSY chiller water level

Attached is a rough trend of the amount of water we have added to the TCSY chiller over the last week in order to keep it "topped off".  There is ~kinda a downward trend, which may indicate we are finally getting to the point where the system is "full".  Recall, Wed Sept 28th the system ran dry due to a leak at the chiller.  At that time the chiller was refilled with ~5L of water.  We have since added another 5L of water (~300mL at a time over the last week).  The reservoir holds 7.2L and the piping probably takes a few more L so, maybe we are still clearing out air leaks in the whole stsyem...

We'll just slowly filling and watching for another few days and replot.

Images attached to this report
H1 IOO
daniel.sigg@LIGO.ORG - posted 11:11, Friday 07 October 2016 - last comment - 09:23, Thursday 13 October 2016(30305)
RAM measurements: Take 3

Evan, Daniel

17:12:30 UTC Oct 7 2016:

17:16:30 UTC Oct 7 2016:

17:18:30 UTC Oct 7 2016:

17:24:30 UTC Oct 7 2016:

17:32:00 UTC Oct 7 2016:

17:34:30 UTC Oct 7 2016:

18:06:30 UTC Oct 7 2016:

Comments related to this report
evan.hall@LIGO.ORG - 11:16, Friday 07 October 2016 (30308)

Spectra attached.

Images attached to this comment
daniel.sigg@LIGO.ORG - 15:12, Friday 07 October 2016 (30313)

Coherence (modulation on)

Non-image files attached to this comment
evan.hall@LIGO.ORG - 09:22, Monday 10 October 2016 (30369)

Using 2600 V/W for the demod gain and transimpedance, and 29 mW of dc PD power, this implies the following AM depths:

  I Q
9 MHz 0.95×10−4 2.4×10−4
45 MHz 1.9×10−4 8.2×10−4

Using 0.22 rad and 0.28 rad for the 9 MHz and 45 MHz modulation depths, this implies the following AM/PM ratios:

  I Q
9 MHz 0.43×10−3 1.1×10−3
45 MHz 0.67×10−3 2.9×10−3
Non-image files attached to this comment
evan.hall@LIGO.ORG - 11:01, Wednesday 12 October 2016 (30450)

The attachment contains a budget of the expected CARM residual. The in-loop error point is taken from the CM board control signal, as was done previously. Here I used 2600 V/W for the transimpedance and demod gain.

The other measured traces are taken from the REFL9I readback (not from the CM board), so in principle there could be some extra dark noise at the error point from the summing node board or CM board. However, based on the O1 level this is of the same order as the shot noise (so we are not missing a huge amount of extra noise in this estimate).

Non-image files attached to this comment
evan.hall@LIGO.ORG - 17:05, Wednesday 12 October 2016 (30469)

Attaching earlier RAM plot, this time with informative labels

Images attached to this comment
evan.hall@LIGO.ORG - 09:23, Thursday 13 October 2016 (30493)

Here is a time series of REFL LF during the modulation depth reductions that happen during lock acquistion.

During the 9 MHz depth reduction (from 0.22 rad to 0.11 rad), the dc power changes from 4.83(3) mW to 4.27(3) mW. That means that after the modulation depth reduction, 4.08(4) mW of the dc light is from the carrier and 0.19(2) mW of the dc light is from the 9 MHz sideband (this assumes the 45 MHz contribution is negligible).

Note that the dc level is still settling to its final value of ~3.7 mW, so it's possible that these power ratios are evolving during the lock.

Images attached to this comment
H1 General
jeffrey.bartlett@LIGO.ORG - posted 09:50, Friday 07 October 2016 (30301)
08:30 Meeting Minutes
SEI: All OK – Working on ESs sensor correction
SUS: No Report
CDS Ele: All OK
CDS Swt: All OK – Fixed problem with bad commissioning frame effecting NDS
PSL: All OK – PSL has been running since the Diode controller swap
TCS: No Report
VAC: All OK – Kyle planning for the 5 day RGA bake out at End-X
H1 ISC (DetChar, ISC, PSL)
gabriele.vajente@LIGO.ORG - posted 09:28, Friday 07 October 2016 (30300)
Time-domain subtraction of jitter noise: it works!

There was only a short, not very good lock, after the DBB signal started to be acquired. I used ten minutes of data from GPS 1159862777 to compute the coherence between DARM (CAL-DELTAL) and the DBB QPD signals.

The first plot shows that the most coherent signal is Q1Y.

The second plot shows the transfer function from Q1Y (in arbitrary units, not sure about the calibration) to CAL-DELTAL (properly calibrated in meters, including the de-whitening filter). The shape is quite smooth, and it looks like a monothonic increase like f for most of the range. The low frequency noise of the IFO was quite bad, so I could only measure coherence above 100 Hz. Hopefully this will be improved in future locks.

However, I could fit the measured transfer function between Q1Y and CAL-DELTAL with a 3rd order model. The fit is shown in the third plot: the result is reasonably good.

I then converted the model into a IRR filter and computed the time-domain subtraction of the Q1Y signal from DARM. The result is quite good, most of the bump has disappeared, see the 4th plot for a comparison of the spectra and the 5th plot for the residual (basically null) coherence of the subtracted DARM with DBB signals.

Of course, we'll have to check in future locks how much this coupling changes over time and how hard it is to compensate for this change.

Images attached to this report
H1 PSL
peter.king@LIGO.ORG - posted 08:31, Friday 07 October 2016 (30262)
laser power
Trying to check where the laser power went to.  From the attached plots, the drop in laser power
is not easily explained by a power drop in the laser diodes.  So my suspicion is that there's a
small resonator alignment change.  Perhaps not all that surprising given the temperature change
of the laser due to being off for a few days.
Images attached to this report
H1 ISC
kiwamu.izumi@LIGO.ORG - posted 01:43, Friday 07 October 2016 - last comment - 18:59, Friday 07 October 2016(30297)
Brief mid-power lock

Just a quick report. Broadband noise in 200-1000 Hz was also visible at a mid-high input power of 27 W.

I did not insert the cutoff filters in the hard ASC loops. ISS 2nd loop was fully engaged with a gain of 19 dB and the boost on. TCS was held at the lock acquisition settings, i.e. [CO2X CO2Y] = [500 mW 1000 mW]. The period when the interferometer was low-ish noise is in 8:15 - 8:25 UTC which was followed by lockloss due to me failing to handle PI mode 27.

Images attached to this report
Comments related to this report
kiwamu.izumi@LIGO.ORG - 18:59, Friday 07 October 2016 (30319)

Synopsis -- Not surprisingly, the broadband noise in terms of RIN at the DCPDs seems to grow proportionally to the carrier field amplitude in the interferometer.


[Cross correlated noise in DCPDs]

Here is a plot of cross-correlated noise between OMC DCPDs A and B. DARM loop suppression is removed by a post process.

For comparison, I overlaid the cross correlated spectrum of a 50 W lock which is from Sep 30th (30115). For both data, the OMC DCPD sum current was held at 20 mA by the DARM loop. It is clear that the mid-power lock (27 W in blue) has a slightly lower noise level in 200 Hz - 2 kHz. The high power lock has a much higher noise level above 2 kHz. I attach the plot in fig format as well. The pcal calibration line at 331.9 Hz for the mid-power lock was smaller than the high power lock by 20 %. Qualitatively, this is due to the higher optical gain in the high power lock although the ratio ideally should be sqrt( 27 W / 50 W ) = 36% instead of 20%. Probably this discrepancy can be partly due to the smaller recycling gain for the high power lock.

Now, if one scales the mid-power spectrum so that both the pcal lines have a same height at 331.9 Hz, it gives you the following plot.

The two curves overlap more in 200 Hz- a few kHz. This simply means that the broadband noise scales with the field amplitude of the carrier light circulating the arm cavities. Because the DARM optical gain also scales with the carrier field amplitude in the same way, this unfortunately means that the calibrated displacement noise does not change regardless of the laser power level. This rules out some local electronics pickup/cross-talks, but does not rule out laser noise couplings (jitter, intensity, frequency) or displacement noise.

Images attached to this comment
Non-image files attached to this comment
H1 ISC
kiwamu.izumi@LIGO.ORG - posted 01:28, Friday 07 October 2016 - last comment - 09:12, Friday 07 October 2016(30296)
change in AS36->SRM loop; AS WFS DC signals and shutter; step in AS90

Terra, Travis, Kiwamu,

Here are some unexpected issues that we addressed tonight.

Comments related to this report
daniel.sigg@LIGO.ORG - 07:04, Friday 07 October 2016 (30298)

The ASC DC SUM channels have been split into to SUM and NSUM. The SUM channel has a steep low pass, so it can be used to normalize PIT and YAW. NSUM is the new fast channel. See alog 30214.

kiwamu.izumi@LIGO.ORG - 09:12, Friday 07 October 2016 (30299)

Thanks, Daniel. I have just edited the FAST_SHUTTER guardian so that it looks at AS_A(B)_DC_*NSUM_OUTPUT* instead of *SUM_OUTPUT*. Also, I found my original statement in the above entry a bit wrong. AS_A(B)_DC_SUM did not have a low pass at all before Oct.5th as Jenne reported in 30214

The FAST_SHUTTER is now at rev14405.

H1 General
travis.sadecki@LIGO.ORG - posted 23:06, Thursday 06 October 2016 (30295)
OPS Eve Shift Summary

TITLE: 10/07 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Lock Aquisition
INCOMING OPERATOR: None
SHIFT SUMMARY:  Started shift with an initial alignment.  After IA, spent the entire shift working with Jenne and Kiwamu diagnosing issues with ASC loops and fast shutter.  Only made it past DRMI a few times.
LOG:

5:30 Filled TCSY chiller

6:00 Shift end due to early start for 3IFO meeting
 

H1 ISC
terra.hardwick@LIGO.ORG - posted 21:39, Thursday 06 October 2016 (30294)
PI Mode2 Open Loop Transfer Function

Kiwamu, Matt E, Terra

We broke our overnight lock this morning from Mode2 (ITMX 15520.7 Hz) ringing up; usually this is easily damped back down with a slight phase change but there was no operator in the chair at the time. This mode has been especially prone to ring up with very minor (a few degrees) phase drifts as the frequency drifts over ~3 hours lock across a static bandpass. 

I measured this mode's open loop transfer function last night (with the same settings it later rang up with). Width of loop is 24 mHz; cursors at UGFs shown on transfer function plots. Measured Q of mode is ~20 M, giving a half-width of 0.776 so the gain on resonance of our loop is about 31. Phase differential across this fairly wide frequency range is ~160 deg, so we're close to instability. 

To this end, I've halved the gain of the damping drive. I also shifted the bandpass filter: previously the resonant frequency range was sitting slightly off center picking up ~30 deg and now it shifts around within ~2 deg phase window (for a five hour lock at least). We had trouble locking tonight and are now working at low power, so I'll have to remeasure another time. 

Images attached to this report
H1 CAL (CAL)
darkhan.tuyenbayev@LIGO.ORG - posted 20:17, Thursday 06 October 2016 (30293)
Front-end kappas from October 6 locks

Joe B, Darkhan T,

From the past day's lock stretches it seems that now the DARM time-dependent parameters (kappas) calculated in the CAL-CS model are mostly within expected ranges. The optical gain seems 10% lower compared to what the DARM model suggests.

Related alogs (kappas from September): 30219, 29992.

Cal. line coherence averaging:

Coherence calculations in the CAL-CS front-end model outputs overestimated coherences for the calibration lines between Sept. 20 and now. The issue will be fixed by replacing the averaging C code, BUFFER_AND_AVERAGE.c with a previous version (modification to the script proposed in LHO alog 29744 will be reverted).

Images attached to this report
H1 PSL
robert.schofield@LIGO.ORG - posted 18:33, Thursday 06 October 2016 (30290)
A 9% increase in crystal chiller water flow increased the HPO jitter signal at the DBB and in DARM by 25% at 450 Hz.

The first figure shows the results of 2.5 cycles of crystal chiller total flow variation (24.2 l/m vs. 22.2 l/m) in DARM and in the 1QX signal of the DBB quad diodes. The diode signal shows a detectable increase with flow from about 60 to about 6000 Hz, with the biggest increase in the 1000 Hz region. DARM increased from about 200 to 2000 Hz, with the biggest increase around 450 Hz.

The second plot shows other DBB diode signals.  The peaks do not change as much as the rest of the jitter spectrum (see 1000 Hz). The lower percentage changes at some of the peaks are more consistent with the changes in table accelerometer signals. The water signal is apparently a less dominant driver of the peaks than of the smooth regions, and the source of the peaks may therefore be easier to access and damp should need be. 

Robert, Jason

Non-image files attached to this report
H1 DAQ (CDS)
david.barker@LIGO.ORG - posted 18:20, Thursday 06 October 2016 - last comment - 10:28, Friday 07 October 2016(30291)
DAQ nds saw a corrupted frame file from h1fw1, problem is in the NFS exporting/mounting of this file

Jonathan, Jim, Dan, Dave

Jim W reported data errors from 06:14PDT this morning in the full frame. We found that the commissioning frame file for this time is correct on the LDAS QFS file system, but corrupt when NFS exported by the h1ldasgw2 NFS server machine to the nds machines. This is a relatively new solaris server installed this summer to take read-only exports away from h1ldasgw0 when h1fw0 was unstable.

I've opened an FRS for this #6370

Investigation is continuing, including a full characterization of the problem.

Comments related to this report
jonathan.hanks@LIGO.ORG - 18:25, Thursday 06 October 2016 (30292)
I have a new build of daqd frame writer running h1fw2.  Now when we write raw frame files we also write a checksum file next to the file.

This will give us a view of what the daqd says was written out so that LDAS/DCS can verify they receive the file CDS produces.

Tomorrow I will do this on the trend files, and reflect some of this information into EPICS so that we can get a graphical alert when frame writers produce different output.
jonathan.hanks@LIGO.ORG - 10:28, Friday 07 October 2016 (30302)

I now have a build of daqd frame writer running on h1fw2 (and the test stand) that provides checksum files for all frames being written.

In addition it adds four EPICS channels that give 32bits of the checksum to help with medm monitors for Dave.

Prefix each with IFO:DAQ-FW[012]_   (in general, currently only deployed for H1:DAQ-FW2_)

FRAME_CHECK_SUM_TRUNC

SCIENCE_FRAME_CHECK_SUM_TRUNC

SECOND_FRAME_CHECK_SUM_TRUNC

MINUTE_FRAME_CHECK_SUM_TRUNC

H1 TCS
nutsinee.kijbunchoo@LIGO.ORG - posted 18:11, Thursday 06 October 2016 (30289)
New HWS python code now running at both end stations

Dave, Nutsinee

There are still some issues (several white channels on the medm screen, unclear where the files are being written) but so far the new code runs on Ubuntu 14 at both end stations without complains.

H1 PSL (OpsInfo, PSL)
corey.gray@LIGO.ORG - posted 10:52, Sunday 02 October 2016 - last comment - 10:50, Friday 07 October 2016(30150)
PSL Recovery From Flow Sensor Trip (and Being down/cold for 10+hrs)

(Corey, Jason on phone)

Last night, Nutsinee came in early for her OWL shift & just before midnight the PSL tripped.  We opted to leave it tripped for the night to wait for Jason to walk us through recovery procedure.  Atleast for now, we always want to run through this procedure with one of our PSL people (Jason or Peter K).  Here are my rough notes:

Now on to "Aligning" on Observatory Mode!

PSL Status NOTE:

Comments related to this report
dennis.coyne@LIGO.ORG - 12:21, Sunday 02 October 2016 (30153)

The Xtal chiller has tripped 3 times in the last week:

  • 9/25 (29964) - added 200 mL water - "normal due to known slow leak"

then 4 days later:

  • 9/29 (30063) - added a "few 100 mL" water - water sprayed onto door

now 3 days later:

  • 10/2 (30150) - "cap blew off" - added 375 mL

The last two events imply an overpressure condition. The problem seems more significant than replacing lost fluid from a slow leak.

Was the cap which "blew off" in today's event the "bleeding cap" (section 4.2.2), the "filter sleeve cap" (section 6.2) or the "filler pipe cap" (section 6.3 of T1100374-v1, "200 W Laser Crystal Chiller Manual")?

 

BTW, LRA = Long Range Actuator; see section 7.2 of  T0900641-v5, "Under Manual 200 W Laser".

matthew.heintze@LIGO.ORG - 12:51, Sunday 02 October 2016 (30155)PSL

From my quick look through the alog I think the PSL has tripped more than this (for LHO to concur). From what I can see it has gone off 11 times in the last week (thats as far as I looked back).

 

I'm not sure if all the same problems, but alogs reporting the PSL laser off are LHO alogs:

10/2/2016

LHO alog 30160

LHO alog 30154

10/1/2016

LHO alog 30146

LHO alog 30143

9/30/2016

LHO alog 30118

LHO alog 30086  (due to power glitch)

9/29/2016

LHO alog 30076

LHO alog 30063 (this alog reports two different  instances of the laser going off)

9/27/2016

LHO alog 30016

9/25/2016

LHO alog 29964

 

The filler pipe cap popping out is a known thing (happens all the time at LLO) when the chiller turns OFF. At LLO at least this has not shown to be due to any problem (just a consequence for whatever reason when the chiller is turned off).  Its why we try to not turn these chillers OFF if can help it as they "burp" water over the floor and pop these fill caps even when trying to restart

 

I have as one of the main agenda items of this Wednesdays PSL meeting to discuss this problem and see if we can work out whats going on. Some statistics on how many times happened in say the last month or two, what the PSL trip was attributed to, and how many times happened before and after the chiller swap (to see if accelerating or at the same rate), would help this. Im not sure if FRS fault reports have been made for each laser trip to make this search easy for us remote to the site to do and work out how much lost observatory time we have had due to this issue.

corey.gray@LIGO.ORG - 12:58, Sunday 02 October 2016 (30157)PSL

Yes, Dennis, as Matt says it's the Filler Pipe Cap that pops off.  We have had (3) trips since Friday evening (so Matt is probably right about there being more trips over the whole week).  OH, and I should correct ourselves here because we just had a 4th Weekend PSL trip (this was just after I had H1 at NLN for 15min.  This time the cap was blown off and there was a puddle on the floor.

Able to get back in 30min this time (vs 60min this morning).

OK, back to locking.  

corey.gray@LIGO.ORG - 13:42, Sunday 02 October 2016 (30158)

Another Note:

Something I wanted to add about the chiller was that when filling it, I noticed quite a bit of turbulence in the fill pipe.  And you could see a air bubble vortex/tornado in there.  Something we probably don't want if air bubbles are postulated as a trigger for flow sensor trips.  

matthew.heintze@LIGO.ORG - 13:48, Sunday 02 October 2016 (30159)

We see these bubbles as well in the chiller fill pipe at LLO with no chiller trips due to it (hopefully haven't jinxed myself). Perhaps post a movie of it so can see if looks the same as here

corey.gray@LIGO.ORG - 14:49, Sunday 02 October 2016 (30163)

Jason & crew will be investigaing tomorrow.  We should ask them to record a video of it.

matthew.heintze@LIGO.ORG - 05:21, Monday 03 October 2016 (30165)

Movie of the water turbulence of LLO's crystal chiller fill tube posted at LLO alog 28397

jason.oberling@LIGO.ORG - 10:50, Friday 07 October 2016 (30304)

I took a video, but I can't post it; my phone only takes video in .mp4 format, which is apparently not a valid file type for upload to the alog.  Huh.

I attached a still from the video to give you some idea of what we've been seeing here for the last few weeks.  It's appearing to fluctuate though; the video was taken on Monday, 10/3/2016, but this morning our fill port looks very similar to what's seen in Matt's video of the LLO crystal chiller fill port.

Images attached to this comment
H1 ISC
keita.kawabe@LIGO.ORG - posted 19:06, Friday 30 September 2016 - last comment - 11:15, Friday 07 October 2016(30124)
Jitter coupling model via BS wedge: Plausible but unclear

BS horizontal wedge doesn't have any couterpart Y path. Any horizontal beam shift on the BS will cause MICH path difference because of this wedge.

(CPY has a horizontal wedge but it's a mirror image of CPX about BS HR surface. ITMY has a vertical wedge but again it's a mirror image of ITMX.)

This effect doesn't look small, and a rough calculation shows that the HPO jitter peaks from DBB (but not broad humps seen in DBB) are consistent with what we see in DARM.

----

In the attached cartoon, when the beam on the BS is shifted in Y direction by y (blue line), the optical single-trip distance from the beam spot on the BS to ITMX gets shorter than to ITMY by:

EQN1: MICH = n*yW*sqrt(2)/cos(theta) - yW*(1+tan(theta)) ~ 0.7896yW = 1.0E-3*y

where W=1.3E-3 rad is the wedge, n=1.4496 is the refractive index and theta=29.2deg=0.5096rad is the angle of refraction.

DARM will only see about (1-rI)/(1+rI) of MICH where rI is the amplitude reflectivity of the ITMs, which is sqrt(0.986):

EQN2: DARM=MICH*(1-rI)/(1+rI) = 3.5E-3*MICH ~ 3.5E-6 * y.

Because of this, DARM should have a flat coupling to YAW displacement on BS.

Instead of the displacement y, we'll use the displacement A which is just the displacement normalized by the beam radius on the BS (53.4mm):

EQN3: DARM=3.5E-6*53.4[mm]*A = 1.9E-7 [m] * A.

----

Now, let B be a normalized misalignment parameter out of HPO (i.e. HPO jitter), e.g. B=disp/waistRadius+ i* angle/divAngle. Note that abs(B) is conserved unless 01/00 mode ratio is altered.

PMC gives us 1.6% HOM01 suppression in amplitude (T0900616).

For IMC there's some difference for PIT and YAW due to additional sign flip on top of Gouy shift per round trip, but PIT suppression is about 0.5% and YAW is 0.4% in amplitude assuming that the common round trip Gouy shift is 102 deg.

For PRC, assuming carrier recycling gain of 30, PRM reflectivity of 97% and the round trip Gouy shift of 32.5 deg, HOM01 suppression is 5.7% in amplitude.

PMC, IMC and PRC combined, HOM01 suppression is 3.7e-6 for YAW, 4.6e-6 for PIT.

Though A cannot be known without knowing the Gouy shift from HPO to BS, since |B| is conserved except through cavities, we can set the upper bound on |B| using the HOM suppression through PMC, IMC and PRC as

EQN4: |A| < 3.7e-6 * |B|

therefore upper bound on DARM

EQN5: DARM < 7E-13 [m] * |B|.

DBB measurements of B are available in alog, e.g. alog 29754 (direct link to the HPO jitter plot is here).

If you look at 1kHz peak which does not appear in DBB RIN measurement, the peak height is somewhere between 5E-6 to 1E-5, and using EQN5 and these numbers,

EQN6: DARM < 3.5 to 7E-18 [m/sqrtHz] at 1kHz peak.

In DARM spectrum, 1kHz peak is visible at or somewhat above 1E-19 [m/sqrtHz], so it seems consistent with EQN5.

----

Caveats:

IF the broad noise in DBB jitter measurement is actually jitter,  assuming the same jitter coupling as the above, you should see the huge broad thing in DARM, which we do not.

This coupling only applies to YAW. As for PIT, there could be similar coupling mechanism if ITM wedges don't cancel with each other, but the imbalance is only 0.001 degree.

MICH is not the only thing that is modulated by this, SRCL and PRCL are simultaneously modulated. Full simulation might be useful.

Images attached to this report
Comments related to this report
keita.kawabe@LIGO.ORG - 11:15, Friday 07 October 2016 (30307)

Update: Wrong.

If you do the correct calculation there's no differential phase change (see attached). In the above entry I ignored the fact that BS transmission is deflected relative to the incident.

Images attached to this comment
Displaying reports 53761-53780 of 83242.Go to page Start 2685 2686 2687 2688 2689 2690 2691 2692 2693 End