Displaying reports 54841-54860 of 83134.Go to page Start 2739 2740 2741 2742 2743 2744 2745 2746 2747 End
Reports until 22:46, Friday 12 August 2016
H1 ISC
sheila.dwyer@LIGO.ORG - posted 22:46, Friday 12 August 2016 - last comment - 23:58, Friday 12 August 2016(29082)
misalignment causes problems with carm offset reduction

Today we have had some difficulties with alignment in the process of recovering the IFO. 

This morning we moved ETMX by 20 urad in yaw, Patrick checked trends and the top mass osems roughly agree with the alingment sliders that we moved by about 20 urad, while the optical lever indicates that this was moving the test mass back to the position it was in last time we were locking, on August 1st. 

We also saw that even after doing initial alingment twice, the arms atill had a large relative misliagnment which caused a low recycling gain and therefore a lower cam offset than expected while CARM is locked  which cause random locklosses at different points.  Earlier tonight we we seeing a 5% drop in the refl power when the arm transmission was at 5 times the single arm power, while the same arm transmitted power level on August 1st coresponded to only a 1% drop in reflected power.  Koji and Stefan worked on solving this problem, so I will leave it to them to log their solution. 

Comments related to this report
stefan.ballmer@LIGO.ORG - 23:58, Friday 12 August 2016 (29084)

Koji, Stefan

We co-aligned the y-arm for IR and green by comanding the guardian in an y-arm IR lock, and simultaneously engaging the green arm alignment. Then we moved the ITM green camera set point around, and kept the IR maximized with the BS. THis resulted in a nice co-alignment.

WIth that setting we had much less truble during the CARM reduction sequence, but once we were on resonance, we kept loosing the lock in a fast transient. Our guess is that there is a problem with the CARM loop, but we didn't have the energy to check that.

At least the fast sutter is reliable triggering on every lock-loss.

H1 General
sheila.dwyer@LIGO.ORG - posted 22:35, Friday 12 August 2016 (29081)
LOCKLOSS shutter checking guardain

Today I started a draft of a gaurdain to identify locklosses in which there was significant power build up in the arms, so that a check of the fast shutter should be run.  This is on the medm screens and is running, but there is no shutter check implemented yet.  The idea is that the state of this guardian this will be checked in our watchdog checker, and the ISC_LOCK guardian will require manual intervention to reset the state of this gaurdian before relocking the IFO.  This will be a reminder for the operator to run Stefan's shutter check ploting tool which is accessible from the bottom of the ISC guardian overview before manully resetting the state of this guardian so that you can relock.  Once we have an automated check this can all be automated. 

Since I didn't get a chance to check the arm power thresholds tonight, I will not integrate this into ISC_LOCK, so we are still relying on humans to check that the shutter closed.  This nominally means that if we get the interferometer locked we should set the request of ISC_LOCK to DOWN which will not unlock the interferometer but will preventing it from trying to relock automatically.  Hopefully this will be integrated soon. 

Images attached to this report
LHO FMCS
kyle.ryan@LIGO.ORG - posted 19:56, Friday 12 August 2016 (29080)
Moved pumping+heating components from Y-end RGA to Vertex RGA
Hope to repeat the success of the Y-end RGA bake at the Vertex using same settings/durations etc...  Maybe can piggyback on the already running pump carts (a.k.a. noise generators) that will continue to be running near HAM5/HAM6 this week?  One more running pump cart should just get buried in the noise - no?  
LHO VE
kyle.ryan@LIGO.ORG - posted 19:32, Friday 12 August 2016 - last comment - 14:06, Wednesday 24 August 2016(29079)
Y-end RGA captures IFO locking effects on partial pressures
After decoupling the pumping components used during the recent bake out of the Y-end RGA, I exposed the RGA to the Y-end vacuum volume, energized the filament and let it come into equilibrium for an hour or more.  I then let the RGA scan continuously with the multiplier (SEM) on for an additional hour or so while I gathered up my mess(es).  I periodically checked the scanning as I walked past the screen.  At one point, I noticed that the spectrum was changing rapidly towards the "dirty".  I monitored the scanning and noted that after reaching a temporary maximum, the amus which had increased then returned to near their original values.  

After consulting with the Jeff B. (the operator on shift), I feel that the observed changes in partial pressures were likely the result of IFO locking attempts as they coincide closely in time.  Perhaps something gets hot when the IFO is locked or when mirrors are steered?  

See attached scans 
Non-image files attached to this report
Comments related to this report
michael.zucker@LIGO.ORG - 04:46, Saturday 13 August 2016 (29086)

If true that could be kind of scary (!)  Can we set an RGA in MID (stripchart) mode and run time series following the main peaks through a locking attempt?

 I could imagine baking the adsorbed water off the ETM and perhaps nearby baffles. But this should not persist (or repeat) after the first good cavity buildup. 

 

 

 

joshua.smith@LIGO.ORG - 15:39, Saturday 13 August 2016 (29087)DetChar
It's not clear whether this effect could be seen on the overall vacuum pressure, but if it could be, we don't see it. Here is 8 hours (yesterday/today) ofthe y-end vacuum pressure together with the y-arm transmitted power. I don’t see anything in the vacuum pressure channels that relates to the locking attempts, though that may be my untrained eye. 
 
1. H1:ASC-Y_TR_A_NSUM_OUT_DQ, raw,start: 2016-08-13 00:00:00 (1155081617) len: 8:20:00. 
https://ldvw.ligo.caltech.edu/ldvw/view?act=getImg&imgId=137580
 
2. H0:VAC-EY_Y3_PT410B_PRESS_TORR, raw,start: 2016-08-13 00:00:00 (1155081617) len: 8:20:00. 
https://ldvw.ligo.caltech.edu/ldvw/view?act=getImg&imgId=137573
 
3. H0:VAC-EY_Y2_PT424B_PRESS_TORR, raw,start: 2016-08-13 00:00:00 (1155081617) len: 8:20:00. 
https://ldvw.ligo.caltech.edu/ldvw/view?act=getImg&imgId=137576
 
...and here is 5 hours from the day we think the lock losses damaged the OMC (this includes the 5 lock losses reported in https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=28684 ). 
 
4. H1:ASC-Y_TR_A_NSUM_OUT_DQ, raw,start: 2016-07-27 03:40:00 (1153626017) len: 5:00:00.
https://ldvw.ligo.caltech.edu/ldvw/view?act=getImg&imgId=137591
 
5. H0:VAC-EY_Y2_PT424B_PRESS_TORR, raw,start: 2016-07-27 03:40:00 (1153626017) len: 5:00:00. 
https://ldvw.ligo.caltech.edu/ldvw/view?act=getImg&imgId=137590
 
6. H0:VAC-EY_Y3_PT410B_PRESS_TORR, raw,start: 2016-07-27 03:40:00 (1153626017) len: 5:00:00. 
https://ldvw.ligo.caltech.edu/ldvw/view?act=getImg&imgId=137589
 
I don’t see any obvious relationship between these high power locks and vacuum pressure in the Y-end. 
 
The last plot shows the first set of data with the vacuum channels highpassed 3rd order at 1Hz. Nothing visible. 
 
Let me know if you’d like to chase this another way. 
Images attached to this comment
kyle.ryan@LIGO.ORG - 20:36, Saturday 13 August 2016 (29090)
Mike - 
Chandra's stated goal is to eventually continuously trend 7 AMUs (max allowed by software) at each building.  The observance cited in this aLOG obviously would have been missed while in Faraday.  Too bad that the RGAs don't live long with their SEMs on 24/7.  As we install/commission the RGAs and as she works out the issues with the CDS and/or GC folks this trending will eventually be happening.  


J. Smith - 
The partial pressures that are changing are too small and not expected to show up on the total pressure gauges. From the graphic scans and knowing that the total pressure at the Y-end is 2 x 10-9 torr, we see that the partial pressures that changed are small (10-12 torr) - but still interesting because they are measurable and even more interesting if the changes can be shown to be tied to some IFO locking activity.  (Science interesting?  Who knew?)
kyle.ryan@LIGO.ORG - 09:17, Monday 15 August 2016 (29101)
Doh!!!  Here are the .txt versions of the ASCII data
Non-image files attached to this comment
kyle.ryan@LIGO.ORG - 14:06, Wednesday 24 August 2016 (29283)
The indicated currents for these scans are typical of the SEM @ 1300 volts (which is the factory default). I have noticed in the past that setting the SEM voltage value in the EDIT tab does not change the value displayed in the device status screen or vice versa - so, though I set this to 1500 volts in one of those two fields, it may not have taken effect.
H1 ISC
stefan.ballmer@LIGO.ORG - posted 19:03, Friday 12 August 2016 - last comment - 15:38, Saturday 13 August 2016(29078)
ASC-AS_C calibrated in W into HAM6

Using the calibration of alog 28133 (3.0554e+04 cts / Watt from SRM, or 3.273e-05Watt/ct) I added FM10 filters to the segments of ASC-AS_C to cast the output into Watt entering HAM6. I also updated H1:ASC-AS_C_SUM_OFFSET (which avoid astronomic PIT and YAW values).

SDF is updated.

Comments related to this report
koji.arai@LIGO.ORG - 23:47, Friday 12 August 2016 (29083)

[Stefan Koji]

LHO ALOG 28953 and LHO ALOG 28133 show descripancy of factor of two.

When we have 10W input this should give us 24mW

10W (P_in) * 0.9 (T_IMC) * 0.03 (T_PRM) * 1/2 (R_BS) * 0.985 (ITM) * 1/2 (T_BS) * 0.37 (T_SRM) * 0.97 (T_FI) = 24mW

This is closer to the measurement on LHO ALOG 28953. (We still don't know the origin of this difference of x2.)

We decided to use the number in this log (0.0162mW/cnt) to calibrate H1:ASC-AS_C_SUM_OUT to indicate incident power to HAM6.
Also the calibration factor for H1:LSC-ASAIR_B_LF_OUT was entered to match those two.

peter.fritschel@LIGO.ORG - 15:38, Saturday 13 August 2016 (29088)

It looks like Stefan's earlier log (28133) missed a factor of 2 gain in the QPD transimpedance amplifier (D1001974) due to the balanced differential output of this amplifier (which feeds the Whitening/VGA). With that factor of 2, Stefan's estimate would be very close to the measured value of 0.0162 mW/count.

LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 17:36, Friday 12 August 2016 - last comment - 18:06, Friday 12 August 2016(29076)
H1 Output Septum Viewport Glass

HAM6 Crew, vacuum group and others.

During the vent last week, the glass for the septum viewport was inspected by Betsy W. and she found a rather large smudge, a photo was taken of the glass (see attached) and such photo was circulated among the vacuum group and others.
At first, the intent was to clean the contamination by using first contact, but after careful consideration and following procedure it was determined that, in order for us to clean up the window we would have to vent the other side of the viewport, which would have added more time to the HAM6 incursion.
Betsy checked the beam location at the glass, and the beam is far away from the smudge (the beam is center on the glass), that made the decision somewhat easy to postpone the cleaning of the smudge for the next vent event.

FRS ticket  6053

Images attached to this report
Comments related to this report
kyle.ryan@LIGO.ORG - 18:06, Friday 12 August 2016 (29077)
Good log entry Gerardo.  I think the procedure that was referenced was LIGO-T080145 "Viewport Safety Protocols"
H1 SEI
hugh.radkins@LIGO.ORG - posted 16:15, Friday 12 August 2016 - last comment - 10:05, Sunday 14 August 2016(29075)
WHAM6 ISI Trips with Fast Shutter Closing

The attached is just one instance and is fairly representative of the trips.  Most of the time, the ACTuator is called to blame, but as I reported in 29056, I'm not sure the Last Trip is absolutely robust.

The 2 seconds of full data show the ISI in Full Isolation (GRD STATE) when Stefan throws the shutter (FASTSHUTTER A MODE.)  You can see a fraction of a second later, the Actuator MASTER DRIVEs going very large.  All the actuators are contributing to the saturation counting as the threshold is only 32760 counts.  The watch dog trips near the end of the two seconds (WD MON STATE.)  It was going to happen anyway but I can't be sure why the WD tripped because none of the counters (iii SAT COUNT) actually reached the trip threshold.  This is a mystery to me, still.  The SAFE COUNT is 8192, 4096, and 10 for the ACT, GS13, and CPS.  None of the counters reach the SAFE COUNT level....  Regardless, the Actuators where going to reach it so that is another problem to figure out.

Unless the shutter softens its blow on the ISI, the trips are going to happen almost everytime.  For now, we are running with HAM6 in the DAMPED state.  There the HAM6 sees a few saturations on the GS13s when the Shutter triggers but literally just a few.  The HAM6 will be transitioned to ISOLATED by the ISC after the FAST SHUTTER is tested.

Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 10:05, Sunday 14 August 2016 (29094)

We could have the HAM6 ISI in damped while we are testing the shutter, but we want it to beisolated when we are running the IFO, so when we loose lock and the shutter triggers we will stil be frequently tripping the ISI. 

H1 CAL (CAL, INJ)
evan.goetz@LIGO.ORG - posted 16:03, Friday 12 August 2016 (29072)
Pcal inverse actuation filter has correct sign
Summary
I checked that the sign of the Pcal inverse actuation filters is correct by injecting a known sine wave into the TRANSIENT filter bank and measuring at TRANSIENT_OUT and also the calibrated Tx PD signal, H1:CAL-PCALX_TX_PD_OUT_DQ. The signals are consistent and no minus sign is missing.

Details
To check the sign consistency, I made a 100 Hz sine wave with zero phase in Matlab and saved to an ASCII text file. Then I used awgstream to inject the waveform into H1:CAL-PINJX_TRANSIENT_EXC. I wanted to make sure that the output of H1:CAL-PCALX_TX_PD_OUT_DQ was correctly calibrated (not whitened) so I turned on the [:1,1] filter module, FM7. I then injected the waveform using awgstream and then measured the output in DTT, using a bandpass filter to see the 100 Hz signal.

I expect the Tx PD signal to be in phase with the injected waveform, while the TRANSIENT_OUT signal should be 180 degrees out of phase. The attached plot shows the output of H1:CAL-PCALX_TX_PD_OUT_DQ and H1:CAL-PINJX_TRANSIENT_OUT_DQ. The two signals are indeed 180 degrees out of phase, with the Tx PD output rising on the integer second as it should be, while the TRANSIENT_OUT is falling on the integer second (with some small delay, ~240 usec).

Thus, I conclude the sign of the inverse actuation filter (and, by extension, the actuation functions provided for the CW injections) is correct. If there is a sign flip, perhaps it is on the output side? This needs to be investigated.
Non-image files attached to this report
H1 General
jeffrey.bartlett@LIGO.ORG - posted 16:01, Friday 12 August 2016 (29074)
Ops Day Shift Summary
Title:  08/12/2016, Day Shift 15:00 – 23:00 (08:00 –16:00) All times in UTC (PT)
State of H1: IFO is unlocked.  
Commissioning: Recovering the IFO.
Outgoing Operator:  N/A
 
Activity Log: All Times in UTC (PT)

14:56 (07:56) Chris – Going down Y-Arm to deliver tools to Ed 
15:00 (08:00) Start of shift
15:19 (08:19) Jim & Krishna – Going the End-Y to reset BRS
15:50 (08:50) Stephen – Going into the Optics lab
16:00 (09:00) Karen – Going into the LVEA to cleanup around HAM6
16:01 (09:01) Kiwamu – Transitioning End-X to Laser Hazard
16:11 (09:11) Gerardo – Checking Ion pumps at HAM5/6
16:20 (09:20) Gerardo – Out of the LVEA
16:25 (09:25) Jim – Restarting NDS1
16:33 (09:33) Ed – Going to End-X to work on wind instrument install
17:12 (10:12) Krishna & Jim – Back from End-Y
17:13 (10:13) Krishna – Going into the Biergarten to work on Compact BRS
17:58 (10:58) Koji & Stefan – Going to ISCT6 to test shutter timing
19:05 (12:05) Koji & Stefan – Out of the LVEA
19:06 (12:06) Krishna – Out of the LVEA
19:14 (12:14) Ed – Back from End-X
20:04 (13:04) Kyle – Going to End-Y for RGA work
20:57 (13:57) Gerardo – Going to both Ends to check on test stands
22:14 (15:14) Kiwamu – Going out to check on PDs
22:15 (15:15) Gerardo – Going into LVEA 
22:25 (15:25) Kiwamu – Out of the LVEA
22:30 (15:30) Kyle – At End-Y doing RGA sweeps – Seeing contamination spikes in vac
22:58 (15:58) Kyle – Back from End-Y

Title: 08/12/2016, Day Shift 15:00 – 23:00 (08:00 – 16:00) All times in UTC (PT)
Support:  Kiwamu, Stefan, Sheila, 
Incoming Operator: Patrick

Shift Detail Summary: Commissioners testing fast shutter operation in the morning. Around 12:00 started to align the IFO. Took several tries and a lot of work by the commissioning team to complete the alignment. Started locking. Commissioners are working through various challenges. 
IFO locked at DRMI_LOCKED. Commissioners continuing working on the fast shutter and locking.
Dale – brought a small tour (his last) into the control room. He was given a hearty round of applause by all present. Good luck to Dale, from his LIGO family, as he starts this next journey.        
H1 ISC
stefan.ballmer@LIGO.ORG - posted 15:22, Friday 12 August 2016 (29071)
FAST_SHUTTER guardian now check ASC-AS_A and ASC-AS_B light levels during shutter test

Updated the FAST_SHUTTER guardian to  check ASC-AS_A and ASC-AS_B light levels during shutter test.

The "light on diodes" level is scaled from the trigger PD levels from beckhoff (H1:SYS-MOTION_C_SHUTTER_G_TRIGGER_LOW and H1:SYS-MOTION_C_SHUTTER_G_TRIGGER_HIGH), using a relative gain specified in the FAST_SHUTTER guardian.
The dark limits are specified in in the FAST_SHUTTER guardian.

The changes include:

- A check thet the downstream PD levels are OK when the shutter is open on every lock
- During a fast-shutter test:
    - A check that the downstream PD levels are OK with the shutter open.
    - A check that the downstream PD levels are dark for 9 seconds, starting 0.5 sec after the fast shutter closed.
    - A check that the downstream PD levels are OK after the sutter reopens.

 

We also tested compatibility with the HAM6 ISI - we seem to trip it a lot on shutter firings when isolated. For now we will leave it damped until after the test.
The guardian was checked into svn as version 14028.

LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 15:19, Friday 12 August 2016 - last comment - 15:20, Friday 12 August 2016(29069)
Over-filled CP3 at 21:30 utc

Over-filled CP3 with the exhaust bypass valve fully open and the LLCV bypass valve 1/2 turn open.

Flow was noted after 10 minutes 23 seconds, closed LLCV valve, and 3 minutes later the exhaust bypass valve was closed.

Comments related to this report
gerardo.moreno@LIGO.ORG - 15:20, Friday 12 August 2016 (29070)
I raised CP3's LLCV from 18% to 19%, due to the amount of time it took to overfill today.
H1 General (AOS, CAL, CDS, COC, CSWG, DAQ, DCS, DetChar, GRD, INJ, INS, IOO, ISC, Lockloss, OpsInfo, PEM, PSL, SEI, SUS, SYS, TCS, VE)
vernon.sandberg@LIGO.ORG - posted 14:26, Friday 12 August 2016 (29066)
Return to normal operations, commissioning at 50W ifo input power

The WHAM6 "Fast Shutter" system and OMC were repaired and replaced, respectively, during the recent vent.  The logic for the fast shutter has been augmented,  reviewed, and tested (see recent alogs for test results).  A TRB review of the OMC-shutter incident took place this morning and their recommendation was to return to normal operations.  The new fast shutter monitoring and control software (ECR E1600246) should now be in place and in routine use. Commissioining may resume at the 50W input power level of interferometer operation.

This relieves LHO alog #29010, "Input power to be held to be < 25 W until technical review of fast shutter
Notice from LIGO Systems".
 

H1 PEM (SEI)
edmond.merilh@LIGO.ORG - posted 13:48, Friday 12 August 2016 (29065)
Ultrasonic Anemometer Installed at End-X

This morning I installed the Young Model 81000 Ultrasonic Anemometer next to the wind fence at End-X. It's being powered by a sand alone 18VDC supply. All connections are being routed through two SOLA STC-DRS-038 Isolated Loop Circuit Protectors located in the rear of the electronics rack  in the VEA. There are 3 BNC pigtails awaiting connection through the PEM patch panel. Unfortunately, I didn't take a picture of the actual anemometer.

Images attached to this report
H1 ISC
koji.arai@LIGO.ORG - posted 12:25, Friday 12 August 2016 - last comment - 13:59, Friday 12 August 2016(29060)
Timing test for the OMC PZT shutter and the fast mechanical shutter

[Stefan, Koji]

Quick Summary

Shutter timing test has been performed for the OMC PZT shutter and the fast mechanical shutter. Everything seems reasonable.

- The trigger logic fires at the PD input of +2V.

- The trigger logic reacts at 1.5us since the PD input thresholding.

- The PZT shutter detunes the OMC cavity at 4.8us.

- The fast mechanical shutter start blocking the beam at 1.7ms, reaches half blocking at 1.85ms, complete block at 2.0ms.

More about the test comes later.

Comments related to this report
rich.abbott@LIGO.ORG - 13:02, Friday 12 August 2016 (29061)ISC
The fast shutter timing numbers agree well with the values seen during the initial installation, so I agree with the health status.
koji.arai@LIGO.ORG - 13:59, Friday 12 August 2016 (29062)

Shutter Test Setup (Attachment1)

- The PD signal from AS_C QPD was temporarily disconnected from the left PD port. Instead, a function generator (SRS DS345) was connected. The static offset of the func gen was used to mimic the PD signal.

- The output of the shutter logic unit (D1102312) was branched by a BNC-T.

- A thorlabs PD (PDA55, Si DCPD, BW ~10MHz) was placed on the ISCT6 to monitor the power at OMCT and OMCR for testing the PZT shutter and the fast mechanical shutter, respectively.

- The monitor signals were connected to an oscilloscope (TDS3034B). The yellow, blue, and magenta traces corresponds to the FG offset, trigger signal, and the optical power at OMCT/R.

Shutter Test Results

- The offset voltage of DS345 was manually swept from 0V to a positive voltage until the trigger fired. We confirmed that the trigger fired at 2V.

- The offset voltage was changed from 0V to 5V on DS345. This made the actual output voltage changed from 0V to ~9V. The output voltage was not so clean because the relay seemed to be switched upon the change. However this essentially does not affect the test.

- The oscilloscope traces showed that the trigger logic reacted 1.5us after the PD input crossed the threshold. (Attachment 2)

- The OMCT optical power started to drop 4.8us after the thresholding (Attachment 3) and reach 1/10 of the power after ~60us (Attachment 4). The trace for the OMCT power showed a decay curve. The half life was 10~20us depending on which part of the curve is taken. This corresponds to the cutoff freq of 5.5kHz ~ 11kHz, which is way to low and unlikely to be the cavity pole of the OMC (330kHz). It is more likely to be determined by the actual motion speed of the PZT.

- The OMCR light was used for the test of the fast mechanical shutter. During this test we didn't try to lock the OMC cavity to make the observation of the light drop easier to interpret. The shutter started blocking the beam at 1.7ms, reached half blocking at 1.85ms, completed blocking at 2.0ms. (Attachment 5). The observation in a longer time scale (~0.5s) showed that there was no boucing effect of the mechanical shutter any more. (Attachment 6)

Images attached to this comment
H1 SUS
jeffrey.kissel@LIGO.ORG - posted 16:48, Thursday 11 August 2016 - last comment - 15:59, Friday 12 August 2016(29035)
HAM6 SUS Final Health Assessment -- SUS Dynamics Look Great After Pump-Down
J. Kissel

I took an opportunity between shutter testing and beckhoff problems to gather the final at-vacuum measurements of the HAM6 SUS. I'll process the data in complete detail tomorrow (I attach a screenshot of an example DOF from the OMC), but the message is that the TFs look just as good as they did before the vent: free, clear of rubbing, and all resonances are at the same frequency with roughly the same Q. 

Excellent!

Watch this space for full analysis and comparison with previous results.

/ligo/svncommon/SusSVN/sus/trunk/OMCS/H1/OMC/SAGM1/Data/
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_L_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_P_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_R_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_T_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_V_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_Y_0p02to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM1/SAGM1/Data/
2016-08-11_2249_H1SUSOM1_M1_WhiteNoise_L_0p03to50Hz.xml
2016-08-11_2249_H1SUSOM1_M1_WhiteNoise_P_0p03to50Hz.xml
2016-08-11_2249_H1SUSOM1_M1_WhiteNoise_Y_0p03to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM2/SAGM1/Data/
2016-08-11_2151_H1SUSOM2_M1_WhiteNoise_L_0p03to50Hz.xml
2016-08-11_2151_H1SUSOM2_M1_WhiteNoise_P_0p03to50Hz.xml
2016-08-11_2151_H1SUSOM2_M1_WhiteNoise_Y_0p03to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM3/SAGM1/Data/
2016-08-11_2257_H1SUSOM3_M1_WhiteNoise_L_0p03to50Hz.xml
2016-08-11_2257_H1SUSOM3_M1_WhiteNoise_P_0p03to50Hz.xml
2016-08-11_2257_H1SUSOM3_M1_WhiteNoise_Y_0p03to50Hz.xml
Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 15:15, Friday 12 August 2016 (29068)
J. Kissel

After looking at the results of OMC transfer functions in detail -- using the traditional Matlab plotting software -- I can doubly confirm that the OMC is as healthy as the previous installation of the OMC breadboard. In addition to the cross coupling information which looks quite normal (see 2016-08-11_2305_H1SUSOMC_M1_ALL_TFs.pdf ), a comparison between these measurements, the measurements taken just prior to the vent with the former OMC breadboard, plus a few measurements of LLO's OMC reveals the following:
- For the Vertical, Yaw, Roll, and Transverse DOFs, either all measurements agree with the model, or where they do not, the measured resonant frequencies are consistently offset from from the model implying the model is not accurately representing the physical system.
- For Longitudinal and Pitch, the first pitch mode modeled to be at 1.9 Hz, is different from the model and different from L1 to H1. However, the H1 OMC suspended with the former and current OMC breadboard show the same dynamics, so the discrepancy is likely not due this past week's work or re-suspension.
- Although it appears as though the Q is lower that previous results, it is an artifact of the measurement: in order to expedite testing, the frequency resolution has been reduced from 0.01 Hz to 0.02 Hz. 
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 15:59, Friday 12 August 2016 (29073)
J. Kissel

Attached are the full results for the OMs. Similar to the OMC, I can double confirm after complete analysis that the OMs are as healthy as they were prior to the vent. Nothing to see here!
Non-image files attached to this comment
Displaying reports 54841-54860 of 83134.Go to page Start 2739 2740 2741 2742 2743 2744 2745 2746 2747 End