H1 CAL
jeffrey.kissel@LIGO.ORG - posted 13:59, Wednesday 11 January 2017 (33153)
New DARM Loop Model for ER10 and O2a (2016-11-12 to 2016-12-22) Including Analog AA/AI Filter Bug Fixes
J. Kissel

I've completed creating an updated parameter set for the matlab DARM loop model that  contains the updates from the fit of the 2016-11-12 calibration reference measurements (see LHO aLOG 31433 for original measurement set, and LHO aLOG 32989 for updated analysis). 

I've compared the model output against those measurements using
${CalSVN}/aligocalibration/trunk/Runs/ER10/H1/Scripts/DARMOLGTFs/compareDARMOLGTFs.m  (r4123, ls: r4123)
Results are attached. The agreement is excellent.

I also attached the sensing and actuation function fit parameter results, just for completeness -- including nlinfit results (and because LHO aLOG 32989 has some typos, only corrected in its comment, LHO aLOG 33090).

                                                       Former Value        MAP     (95% C.I.)  MAP       (95% C.I.)
                                                       2016-11-12          2016-11-12 (nlinfit)      2016-11-12 (MCMC)

 Optical Gain                K_C         [ct/m]        1.15e6              1.166e6 (0.0003e6)        1.164e6   (0.0002e6)
 Couple Cav. Pole Freq.      f_c         [Hz]          346.7               347.2   (5.8)             346.7     (2.7)
 Residual Sensing Delay      tau_C       [us]          2.3                 2.12    (5.5)             2.3       (2.3)
 SRC Detuning Spring Freq.   f_s         [Hz]          7.4                 7.38    (0.3)             7.38      (0.16)
 Inv. Spring Qual. Factor    1/Q_s       [ ]           0.05                0.046   (0.013)           0.046     (0.008)

 UIM/L1 Actuation Strength   K_UIM       [N/ct]        8.164e-8            n/a                       8.104e-8  (0.2%)
 PUM/L2 Actuation Strength   K_PUM       [N/ct]        6.841e-10           n/a                       6.773e-10 (0.04%)
 UIM/L3 Actuation Strength   K_TST       [N/ct]        4.389e-12           n/a                       4.346e-12 (0.05%)

Not only is the model in excellent agreement with the measurement, the nlinfit fit parameter values are in excellent agreement with the MCMC values with the sensing function. As with the updated / new 2017-01-03 model (LHO aLOG 33004), I've used the nlinfit values to update the sensing function parameter set. For the actuator coefficients, I've again used 
${CalSVN}/aligocalibration/trunk/Runs/ER10/H1/Scripts/FullIFOActuatorTFs/
    getActValuesForParamsFile.m   (r4126, lc: r4126)
and the clunky process described in LHO aLOG 33092.
Also, I've run the fitting algorithms on *only* the reference measurement data set (i.e. I did not run on multiple measurements worth, assuming no and/or compensating for time-dependence).

I've committed the following model pieces to the Cal SVN repo:
(r = svn revision, lc = last changed revision)

/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/
O2/DARMmodel/src/
    DARMmodel.m        (r4093, lc: r3511)
    computeSensing.m   (r4093, lc: r4025)
    computeActuation.m (r4093, lc: r4003)

Param Files
O2/Common/params/
    IFOindepParams.conf (r4042, lc: r4002)

ER10/H1/params/
    H1params.conf       (r4121, lc: r4121)

ER10/H1/params/2016-11-12/
    H1params_2016-11-12.conf                      (r3944, lc: 3786)
    measurements_2016-11-12_sensing.conf          (r4122, lc: r4122)
    measurements_2016-11-12_ETMY_L1_actuator.conf (r3944, lc: r3810)
    measurements_2016-11-12_ETMY_L2_actuator.conf (r3944, lc: r3810)
    measurements_2016-11-12_ETMY_L3_actuator.conf (r3944, lc: r3810)
 

More notes:
- I did not update the export of foton implementation of the inverse sensing function. This is because
    (a) DCS "offline" calibration filters DARM_ERR and DARM_CTRL directly with FIR filters. For the DARM_ERR filter, the inverse sensing function is computed directly from the matlab model and does not suffer from IIR filter warping, and does not need low- or high- frequency cutoffs that the front-end pipeline must have for causality.
    (b) If we *were* to regenerate GDS calibration, which correct the front-end for its flaws -- including this inverse sensing function filter -- then we would need to use exactly what filter was in place, and the 
${CalSVN}/aligocalibration/trunk/Runs/
    ER10/H1/Measurements/Foton/
        2016-12-07_H1CALCS_InverseSensingFunction_Foton_SRCD-2N_Gain_tf.txt (r3921, lc: r3921)
file is what was used, at least from 2016-12-07 to 2016-12-22, until it was *actually* updated on 2017-01-03.

What to do from here:
- Use the loop model to push new EPICs records that document model values at calibration line frequencies
- Use the new loop model to generate DCS FIR filters to recalibrate the all data from the pre-winter break from raw DARM_ERR and DARM_CTRL.
- Push hard for uncertainty estimations for both the first part of O2 and data post-winter break.
Non-image files attached to this report
H1 General
vernon.sandberg@LIGO.ORG - posted 13:21, Wednesday 11 January 2017 (33154)
Work Permit Summary for 2017 January 10
Work Permit Date Description alog/status
6428.html 2017-01-10 09:19 Replace TCSY Laser Controller D1200745. This work is related to the flow sensor alarms/glitches. The in-line flow sensor was replaced on Dec. 20, but glitches were seen after swap. See Alog 32776. Planning on doing work this or next Tues.  33129
6427.html 2017-01-10 07:55 Investigate Condenser Fan VFD fault on the end station chiller. End Y work deferred. Blocked by snow.
6426.html 2017-01-10 07:50 Shut down AHU-4 to investigate air flow issues. This air handler supplies the OSB office area. Apollo will need to cut into the duct to be able identify the blockage. This has also been reported to HFD because of the smoke sensors in the duct, therefore the fire system for the OSB will be down for this work.  
6425.html 2017-01-09 17:57 Investigate apparent clipping discovered by J. Kissel recently. We'll need to unlock the MC temporarily to eliminate the 1064 nm light for Pcal camera images. We expect the work to take less than two hours. End Y work deferred. Blocked by snow.
6424.html 2017-01-09 15:01 Upgrade BRS analysis software to reduce auto-collimator non-linearity. See last paragraph of SWG aLOG 11414.  33128
6423.html 2017-01-09 15:00 Install PEM STS on to Modified BRS Y Mounting Plate to explore for better coherence and/or subtraction of tilt at EY. Installation on Tue during maintenance. STS will remain there for the foreseeable future. No new digital infrastructure should be needed. For details, see SEI aLOG 1090.  33128 End Y work deferred. Blocked by snow.
6422.html 2017-01-09 11:51 Remove connected auxiliary pumps from PT180 hardware*Requires venting small turbo pump which is currently bolted to but isolated from PT180's pump port, removing connected pump and replacing with the nominal 1.5" redundant O-ring valve*Will then pump O-ring valve to rough vacuum and isolate*Work requires use of ladder leaning against BSC8  33137
6421.html 2017-01-09 11:27 Replace fried ion pump #5 controller (located in mech. room) and update CDS signal cabling.  33133 & 33105
6420.html 2017-01-09 07:51 Increases OMC dither line amplitude for 1 hour test. (https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=33037) I will do this when LLO is down but may try to do it remotely.  33104
6419.html 2017-01-06 16:11 Test adding a boost to OMC length loop, because of evidence that OMC length noise is showing up in DARM from alog 33037  33104
6418.html 2017-01-05 08:02 Move trend writer 1 raw minute files from h1tw1 SSD RAID to h1fw1 SATAboy RAID. Will require a reconfiguration and restart of the nds system on h1nds0 and h1nds1 at the start and end of the process. This is a maintenance task that needs to be performed about every 6 months. Expected duration 3-5 days.  33127
6417.html 2017-01-04 11:20 Retroactive work permit; discussed and approved a prior to work done with Keita and JRPC. Update the PUM/TST or L2/L3 actuation heirarchy filters to restore new and improved design from LHO aLOG 28746. See discussion in LHO aLOG 32933. Update front-end calibration to compensate for it. Also update front-end to include bug fix for missing 1% from analog AA and AI filters. See LHO aLOG 32907 for details. Run calibration measurement suite to confirm new results. (Already done, see LHO aLOG 32942).  
6416.html 2017-01-03 16:25 Similar to LLO upgrade, upgrade LHO CW hardware injection code to version 1.15.0.1. Requires restart of ps-inject on h1hwinj1.  
6415.html 2017-01-03 08:03 Install HVAC Controls Upgrade. This will involve accessing VAV boxes and pulling new thermostat wire in the OSB. Out building work will be scheduled on a "as we can get access to them" basis.  
6414.html 2016-12-29 11:36 Tour of LVEA & VEAs (while performing holiday inspection).  
6413.html 2016-12-22 07:18 This work permit is mostly to inform that I will be on site beginning Tuesday the 27th with various contractors working at the LSB to try and minimize additional disruption that would occur after the holiday break. Some of this work may last through the week of the break.  
6412.html 2016-12-21 17:26 Add all cryo-pump discharge line pressures to the cell phone alarm system. Alarms will be raised if the pressure exceeds 2.0 PSI for an extended time, or the Beckhoff error channels show errors.  
H1 General
patrick.thomas@LIGO.ORG - posted 12:12, Wednesday 11 January 2017 (33152)
Ops Day Mid Shift Summary
In observing, but range has degraded quickly from start of lock. Just started climbing back. Now around 60 MPc. Front end loader is parked.
LHO VE
logbook/robot/script0.cds.ligo-wa.caltech.edu@LIGO.ORG - posted 12:10, Wednesday 11 January 2017 (33151)
CP3, CP4 Autofill 2017_01_11
Starting CP3 fill. LLCV enabled. LLCV set to manual control. LLCV set to 50% open. Fill completed in 42 seconds. TC B did not register fill. LLCV set back to 14.0% open.
Starting CP4 fill. LLCV enabled. LLCV set to manual control. LLCV set to 70% open. Fill completed in 108 seconds. TC A did not register fill. LLCV set back to 35.0% open.
Images attached to this report
H1 GRD (CAL)
thomas.shaffer@LIGO.ORG - posted 11:55, Wednesday 11 January 2017 (33150)
Added test to DIAG_MAIN for PCAL RX PD

The test takes an average of the last 120s of H1:CAL-PCAL(X/Y)_RX_PD_OUTPUT and compares it to nominal values that I found by taking many different averages over times without clipping. If the percent difference is greater than 1% it will notify.

H1 General
patrick.thomas@LIGO.ORG - posted 10:54, Wednesday 11 January 2017 - last comment - 10:57, Wednesday 11 January 2017(33148)
Observing
Set to observing at 18:50 UTC. Had to revert one SDF difference: Changed H1:SYS-MOTION_C_PICO_D_CURRENT_ENABLE from Enabled to Disabled. See attached screenshot.
Images attached to this report
Comments related to this report
patrick.thomas@LIGO.ORG - 10:57, Wednesday 11 January 2017 (33149)
From the camera, John is moving snow with the front end loader at mid Y.
LHO General
patrick.thomas@LIGO.ORG - posted 08:05, Wednesday 11 January 2017 (33147)
Ops Day Shift Start
TITLE: 01/11 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: None
CURRENT ENVIRONMENT:
    Wind: 15mph Gusts, 11mph 5min avg
    Primary useism: 0.05 μm/s
    Secondary useism: 0.27 μm/s 
QUICK SUMMARY:
Richard M. was attempting to lock ALS when I arrived. Aligned the ETM mirrors and then the BS in CHECK_MICH_FRINGES. Attempting to lock PRMI. May move to initial alignment. John is clearing snow by CP1 (WP 6429).
LHO General
corey.gray@LIGO.ORG - posted 20:29, Tuesday 10 January 2017 - last comment - 16:51, Wednesday 11 January 2017(33143)
EVE Operator Summary

TITLE: 01/11 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 60.9258Mpc
INCOMING OPERATOR: Jim may or may not decide to venture in (I have been in contact with him this evening)
SHIFT SUMMARY:
H1 has been in OBSERVING for almost 6hrs at a fairly steady 60Mpc.

Continue to experience non-normal winter weather with forecast for possibly more snow overnight & freezing temperatures.  Have also had fairly blizzard like conditions with winds just below 20mph.  Have been in contact with OWL shift operator with regards to coverage & they will do what feels safe.  

I am leaving NOW to brave roads home & so H1 will be flying on automatic pilot for a few hours or until the morning. 

I have talked with William Parker (LLO operator) and aprised him of our situation.

LOG:

DIAG_MAIN message:   PCAL:  Y RX PD is 0.01 OFF

Comments related to this report
corey.gray@LIGO.ORG - 00:46, Wednesday 11 January 2017 (33144)OpsInfo

~5:30utc (9:30pmPST) 6:37utc (10:37pmPST) H1 had a lockloss.  Looking at the seismic trends showed nothing (my guess is this was a PI mode ringing up).

Sheila had remote access and tried to relock, but H1 kept dropping at ALS LOCKING.  She mentioned that the Diff beatnote was low.  I made an attempt to drive back in, but Rt10 was impassable for my vehicle.  

Observatory Mode was left at OBSERVING through the lockloss and upt to around 12:40am.  Then I asked Sheila to switch it to UNAVOIDABLE (we don't have a SNOW STORM option).

This is where H1 will be left until the morning.

terra.hardwick@LIGO.ORG - 08:00, Wednesday 11 January 2017 (33146)

Looking at the summary pages (PI Overview under ISC and then I looked at the last aka broadband monitoring; the blue spike right at lockloss is from signals saturating during lockloss, not from a ringing up PI), this lockloss didn't appear to be from PI.

corey.gray@LIGO.ORG - 16:51, Wednesday 11 January 2017 (33167)

Thanks, Terra!  Yeah, also checking the VERBAL_ALARM log, we have a lockloss at 6:36:55utc & no PI Mode notifications.

LHO General
corey.gray@LIGO.ORG - posted 16:36, Tuesday 10 January 2017 (33142)
Transition To EVE

TITLE: 01/11 Eve Shift: 00:00-08:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Observing at 59.2008Mpc
OUTGOING OPERATOR: Travis (handoff over phone & thumbs up on road) & TJ who gave me a verbal handoff in hallway
CURRENT ENVIRONMENT:
    Wind: 18mph Gusts, 15mph 5min avg
    Primary useism: 0.09 μm/s
    Secondary useism: 0.27 μm/s

Hanford Site issued an early DAY shift ending at 12:45pmPST.  Drive in was snowy & not terrible, but Route10's inbound lane (which I was a sole occupier of) was not plowed much & had some drifts of a 1' which caused me to throttle it down to 40mph (mostly less than that).  Also saw one sad stranded BIG 4x4 truck on the side of the road on SR240 (which I was on from Twin Bridges to Route 10).  It probably only took me 30-40min to come in on my usual 25min route (but this included me stopping to take some photos).

Winds look to be picking up and are hitting 20mph (more so on the Y-arm).


QUICK SUMMARY:
As I entered the building TJ, Kissel, & Dick were leaving.  There was mention of pesky PI Modes, but other than that, I was handed an OBSERVING H1 with a range hovering around 60Mpc.  We had just completed (at 0:14utc) the STAND DOWN time for a GRB at the end of the last shift.

I will carry on with the shift, but I am hesitant about the whole shift and will issue an email to Warner about our wintry conditions for his OWL shift.

Just before submitting this alog: 

H1 General
travis.sadecki@LIGO.ORG - posted 15:52, Tuesday 10 January 2017 (33139)
Ops Day Shift Summary

TITLE: 01/10 Day Shift: 16:00-00:00 UTC (08:00-16:00 PST), all times posted in UTC
STATE of H1: Observing at 58.2484Mpc
INCOMING OPERATOR: Corey
SHIFT SUMMARY:  A bit of a rough IA after Maintenance Day, but after pushing through that, we have had no real issues (other than weather).
LOG:

17:03 Karen to LVEA

17:12 Peter done is PSL

17:57 Kyle to BSC8

17:59 JeffK starting charge measurements

18:11 TCSy chiller controller swap

18:44 TCSy chiller back on

19:04 Charge meas. done

19:29 Kyle out

19:43 Bubba to LVEA to check 3IFO storage

19:58 Bubba out

20:02 Karen out

Note:  Bubba has been plowing snow around the corner station with a new LARGE front end loader since ~23:00UTC.
 

H1 General
travis.sadecki@LIGO.ORG - posted 15:52, Tuesday 10 January 2017 (33141)
Out of Observing 23:44-23:48

TCS_ITMX_CO2 guardian node kicked us out of Observe, then fixed itself. 

H1 General
vernon.sandberg@LIGO.ORG - posted 15:48, Tuesday 10 January 2017 (33140)
DOE Weather Alert

Update: Early Release due to Adverse Weather Conditions (Staggered Release)

   

From the DOE Hanford Operations Office:

Due to the snow currently falling on the Hanford Site and surrounding communities, non-essential Hanford employees will be released from work early today. Only essential employees needed to maintain minimum safe operations are to remain at work through the remainder of the shift.

To alleviate traffic congestion, please adhere to the staggered schedule below.

• 200 East Area and all Areas South of the Wye Barricade, including employees working in Richland, 12:45 pm • 200 West Area and 100 Areas, 1:30 pm

The Rattlesnake Barricade will remain open until 6:00 pm.

Employees must check with their manager prior to leaving, are urged to use caution when driving home, and be prepared for adverse weather or road conditions. Employees are advised to stay tuned to area radio stations for possible road closures, which could affect travel.

In addition, swing shift and graveyard shift has been cancelled tonight for non-essential employees. Only essential employees needed to maintain minimum safe operations are to report to work at the usual time.

For updated information call the Hanford Hotline at 376-9999 or log-on to the Hanford Home Page at www.hanford.gov.

 

Weather

									Station# 12  (WYEB) at 550 ft above MSL     

 Time: 15:30:00 PST      Date: 1/10/2017

 Ave Wind Direction = 316
 Ave Wind Speed     =  15
 Max Wind Speed     =  21
                         Celsius
 Ave Temp = 25.9          -3.42
 Max Temp = 26.0          -3.31
 Min Temp = 25.4          -3.64

 Wind Chill  =  13.8

 Rain Gauge = 0.00



H1 General
travis.sadecki@LIGO.ORG - posted 15:14, Tuesday 10 January 2017 (33138)
GRB alert

GRB alert at 23:13 UTC.  Begin 1 hour standdown.

LHO VE
kyle.ryan@LIGO.ORG - posted 15:02, Tuesday 10 January 2017 (33137)
~1030 - 1130 hrs. local -> Removed pump hardware from PT180 at BSC8


			
			
H1 General
travis.sadecki@LIGO.ORG - posted 14:41, Tuesday 10 January 2017 - last comment - 14:43, Tuesday 10 January 2017(33134)
Ops Day Mid Shift Summary

After Maintenance Day ended at noon, we struggled for a bit to recover arm alignment to get past ALS locking.  Although the stability of the green arm locking was a bit questionable, JeffK, Jenne, and I decided to move on through IA.  After completing IA, we have finally made it to NLN as of 22:35 UTC.

Comments related to this report
travis.sadecki@LIGO.ORG - 14:43, Tuesday 10 January 2017 (33135)

Observing at 22:43 UTC.

LHO General
thomas.shaffer@LIGO.ORG - posted 12:44, Tuesday 10 January 2017 - last comment - 14:57, Tuesday 10 January 2017(33131)
LVEA Swept

I went through the checklist and didn't find anything too out of place or worrysome. The two things that I thought worth mentioning are:

Images attached to this report
Comments related to this report
thomas.shaffer@LIGO.ORG - 14:57, Tuesday 10 January 2017 (33136)

To clarify, it is not a computer propped up on the wall, it is a power supply. The noise of the fans from it is what drew me to this area.