Displaying reports 57221-57240 of 85498.Go to page Start 2858 2859 2860 2861 2862 2863 2864 2865 2866 End
Reports until 14:26, Friday 12 August 2016
H1 General (AOS, CAL, CDS, COC, CSWG, DAQ, DCS, DetChar, GRD, INJ, INS, IOO, ISC, Lockloss, OpsInfo, PEM, PSL, SEI, SUS, SYS, TCS, VE)
vernon.sandberg@LIGO.ORG - posted 14:26, Friday 12 August 2016 (29066)
Return to normal operations, commissioning at 50W ifo input power

The WHAM6 "Fast Shutter" system and OMC were repaired and replaced, respectively, during the recent vent.  The logic for the fast shutter has been augmented,  reviewed, and tested (see recent alogs for test results).  A TRB review of the OMC-shutter incident took place this morning and their recommendation was to return to normal operations.  The new fast shutter monitoring and control software (ECR E1600246) should now be in place and in routine use. Commissioining may resume at the 50W input power level of interferometer operation.

This relieves LHO alog #29010, "Input power to be held to be < 25 W until technical review of fast shutter
Notice from LIGO Systems".
 

H1 PEM (SEI)
edmond.merilh@LIGO.ORG - posted 13:48, Friday 12 August 2016 (29065)
Ultrasonic Anemometer Installed at End-X

This morning I installed the Young Model 81000 Ultrasonic Anemometer next to the wind fence at End-X. It's being powered by a sand alone 18VDC supply. All connections are being routed through two SOLA STC-DRS-038 Isolated Loop Circuit Protectors located in the rear of the electronics rack  in the VEA. There are 3 BNC pigtails awaiting connection through the PEM patch panel. Unfortunately, I didn't take a picture of the actual anemometer.

Images attached to this report
H1 DAQ
james.batch@LIGO.ORG - posted 13:29, Friday 12 August 2016 (29064)
h1nds1 reconfigured, restarted
WP 6069

The copying of raw minuted trend files from h1tw1 is complete and the copied files have been verified.  The nds1 service on h1nds1 has been restarted with a new configuration to read the copied files from their new location.
H1 ISC
koji.arai@LIGO.ORG - posted 12:25, Friday 12 August 2016 - last comment - 13:59, Friday 12 August 2016(29060)
Timing test for the OMC PZT shutter and the fast mechanical shutter

[Stefan, Koji]

Quick Summary

Shutter timing test has been performed for the OMC PZT shutter and the fast mechanical shutter. Everything seems reasonable.

- The trigger logic fires at the PD input of +2V.

- The trigger logic reacts at 1.5us since the PD input thresholding.

- The PZT shutter detunes the OMC cavity at 4.8us.

- The fast mechanical shutter start blocking the beam at 1.7ms, reaches half blocking at 1.85ms, complete block at 2.0ms.

More about the test comes later.

Comments related to this report
rich.abbott@LIGO.ORG - 13:02, Friday 12 August 2016 (29061)ISC
The fast shutter timing numbers agree well with the values seen during the initial installation, so I agree with the health status.
koji.arai@LIGO.ORG - 13:59, Friday 12 August 2016 (29062)

Shutter Test Setup (Attachment1)

- The PD signal from AS_C QPD was temporarily disconnected from the left PD port. Instead, a function generator (SRS DS345) was connected. The static offset of the func gen was used to mimic the PD signal.

- The output of the shutter logic unit (D1102312) was branched by a BNC-T.

- A thorlabs PD (PDA55, Si DCPD, BW ~10MHz) was placed on the ISCT6 to monitor the power at OMCT and OMCR for testing the PZT shutter and the fast mechanical shutter, respectively.

- The monitor signals were connected to an oscilloscope (TDS3034B). The yellow, blue, and magenta traces corresponds to the FG offset, trigger signal, and the optical power at OMCT/R.

Shutter Test Results

- The offset voltage of DS345 was manually swept from 0V to a positive voltage until the trigger fired. We confirmed that the trigger fired at 2V.

- The offset voltage was changed from 0V to 5V on DS345. This made the actual output voltage changed from 0V to ~9V. The output voltage was not so clean because the relay seemed to be switched upon the change. However this essentially does not affect the test.

- The oscilloscope traces showed that the trigger logic reacted 1.5us after the PD input crossed the threshold. (Attachment 2)

- The OMCT optical power started to drop 4.8us after the thresholding (Attachment 3) and reach 1/10 of the power after ~60us (Attachment 4). The trace for the OMCT power showed a decay curve. The half life was 10~20us depending on which part of the curve is taken. This corresponds to the cutoff freq of 5.5kHz ~ 11kHz, which is way to low and unlikely to be the cavity pole of the OMC (330kHz). It is more likely to be determined by the actual motion speed of the PZT.

- The OMCR light was used for the test of the fast mechanical shutter. During this test we didn't try to lock the OMC cavity to make the observation of the light drop easier to interpret. The shutter started blocking the beam at 1.7ms, reached half blocking at 1.85ms, completed blocking at 2.0ms. (Attachment 5). The observation in a longer time scale (~0.5s) showed that there was no boucing effect of the mechanical shutter any more. (Attachment 6)

Images attached to this comment
H1 SEI
hugh.radkins@LIGO.ORG - posted 12:19, Friday 12 August 2016 - last comment - 08:12, Tuesday 23 August 2016(29056)
SEI response to 7.2 EQ in SW Pacific (New Caledonia)

HEPI BS Tripped few minutes before ITMX ISI.  This is the only HEPI that tripped in the neighborhood of the large quake.

ITMY ISI tripped--timing (H1:ISI-ITMY_ST2_WD_MON_GPS_TIME) indicates stage2 tripped on ACTuators 1 second before Stage1 on T240s but looking at the plots, the Actuators have only registered a few counts, nothing near saturation/trip level.  But the T240s hit their rail almost instantly.  It seems the Stage2 Last Trip (H1:ISI-ITMY_ST2_WD_MON_FIRSTTRIG_LATCH) should be indicating ST1WD rather than Actuator. On ETMY, the Trip Time is the same for the two stages and Stage2 notes it is an actuator trip but again, there are only a few counts on the MASTER DRIVE; seems this too should have been a ST1WD trip[ indication trip on Stage2--I'll look into the logic.

On the BS ISI, the Stage1 and Stage2 trip times are the same, and the Last Trip for Stage2 indicates ST1WD.  The Stage2 sensors are very rung up after the trip time but not before unlike the T240s which are ramping to to the rail a few seconds before trip.  ETMX shows this same logical pattern in the trip sequence indicators.

On the ITMX ISI, Stage1 Tripped 20 minutes before the last Stage2 trip. This indicates the Stage1 did not trip at the last Stage2 trip.

No HAM ISI Tripped on this EQ.

Bottom line: the logical output of the WDs are not consistent from this common model code--needs investigating. Maybe I should open an FRS...

Attachment 1) Trip plots showing Stage2 trip time 1 second before the stage1 trip where the stage2 actuators do not go anywhere near saturation levels.

Attachment 2) Dataviewer plot showing the EQ on the CS ground STS and the platform trip times indicated.

Images attached to this report
Comments related to this report
hugh.radkins@LIGO.ORG - 15:20, Monday 22 August 2016 (29232)

It seems this is not a problem with the watchdog but a problem with the plotting script.  It seems for ST2 Actuators, it misses a multiplier on the Y axis.  It works correctly for ST1 Actuators and all the sensors; it does not work for other chambers as well for ST2 ACT.  FRS 6072.

hugh.radkins@LIGO.ORG - 08:12, Tuesday 23 August 2016 (29241)

Actually, the plotting script is working fine.  When the spike is so large that the plotting decides to switch to exponential notation, the exponent is hidden by the title until you blow up the plot to very large size. 

H1 ISC (ISC, Lockloss)
richard.mccarthy@LIGO.ORG - posted 11:04, Friday 12 August 2016 - last comment - 08:17, Monday 15 August 2016(29059)
Modifications to ISC rack and Fast Shutter near Ham6
In an effort to eliminate the need to connect and disconnect shutter cables during vent we have moved the shutter logic controler to the rack from the table.  Also replaced the BNC cables with Yellow BNC cables so they can be easily identified.
See attachment.
Non-image files attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 08:17, Monday 15 August 2016 (29099)

Drawings D1200196 and D1201500 updated.

H1 SEI (SEI)
krishna.venkateswara@LIGO.ORG - posted 10:31, Friday 12 August 2016 - last comment - 13:39, Friday 12 August 2016(29057)
NN Array L4C 7 and 8 swapped temporarily

Krishna,

Jenna and I had noticed that L4C 7 didn't appear to be working. L4C is close to the c-BRS, so it would have been good for comparison. In order to test if it was the instrument or the electronics, I temporarily swapped the cables with those of L4C 8 going in to the SUS rack at the LVEA. Now L4C "7" appears to be working and L4C "8" is not, indicating that the problem must be on the instrument side (or the pre-amp, I guess?). I'll swap it back in the afternoon.

Comments related to this report
krishna.venkateswara@LIGO.ORG - 13:39, Friday 12 August 2016 (29063)

Cables replaced in original configuration.

After looking at a different working L4C output (5), I think I was wrong about the instrument being faulty. It looks like the L4C was working when I swapped the cables, suggesting that the problem must be on the electronics side.

H1 AOS
jeffrey.bartlett@LIGO.ORG - posted 10:21, Friday 12 August 2016 (29055)
TCS-Y Laser Power vs Temp 45 Day Trend
  Posted below is a 45 day trend of the TCS laser power compared to operating temperature. There is a drop in power and temperature on 07/19/16 (aLOG #28506). During the HAM6 vent the RF driver was replaced (aLOG 28837) but was never tested due to lack of a beam. After the vent TCS-Y came back at ~16W with a temp of 22.   
Images attached to this report
LHO General
richard.oram@LIGO.ORG - posted 09:35, Friday 12 August 2016 (29053)
LLO site closed today and over weekend due to bad weather

Dear Colleagues,

 

As of now, LLO is closed until further notice due to widespread flooding in the area that prevents access, including emergency access.  I expect we will be able to reopen on Monday.

Any personnel or visitors on site now should immediately leave, taking US-190.  Explicit permission from either myself or Richard Oram is now required to be on site.

More details will follow, but I did not want this message delayed.  As always, liaisons should be in touch with your visitors to convey this message.

 

Joe

 

Joseph A. Giaime, Observatory Head, LIGO Livingston (Caltech); Professor of Physics & Astronomy (LSU).

phone at LIGO: +1 225 686-3169.

H1 General
jeffrey.bartlett@LIGO.ORG - posted 09:04, Friday 12 August 2016 (29052)
08:30 Meeting Minutes
SEI: All back to normal after vent
SUS: All Ok
CDS: Looking into several read-back issues. Will need to do a restart of NDS1 this morning.
PSL: OK. Looking into the low power from the TCS-Y laser.
VAC: No report
FMC: No report

Relocking and commissioning work today.
H1 ISC
kiwamu.izumi@LIGO.ORG - posted 00:20, Friday 12 August 2016 (29047)
OMC R, ASAIR images recorded

I have taken camera images of the OMC R and ASAIR beam. This was primarily to compare the beam quality before and after the OMC swap. However, I did not find any obvious or significant change in the beam shapes. See the attached and compare them to the ones in 28718. Also the data is attached as well.

Images attached to this report
Non-image files attached to this report
LHO General
patrick.thomas@LIGO.ORG - posted 00:13, Friday 12 August 2016 - last comment - 10:54, Friday 12 August 2016(29046)
Ops Evening Summary
Stefan and Sheila working on shutter control logic. I helped resolve problems with h1ecatc1. Had to restart EPICS IOC and copy updated PLC2 to target.

23:32 UTC Gerardo to HAM6 area to check on vacuum equipment
23:39 UTC Gerardo back
03:05 UTC Joe to Y2-5 to watch meteor shower with telescope
04:30 UTC Stefan and Sheila done working on shutter control logic. Kiwamu starting work.
06:17 UTC Kiwamu to end X to investigate green laser electronics
07:00 UTC Kiwamu back

I was able to get the Y arm green power to around 1 with LOCKED_NO_SLOW_NO_WFS, but if I put it on INITIAL_ALIGNMENT it would start to be driven away. Kiwamu was investigating the X arm.
Comments related to this report
kiwamu.izumi@LIGO.ORG - 00:25, Friday 12 August 2016 (29048)

Patrick, Kiwamu,

We are unable to engage the QPD servo for the X green injection beam. It seems to be an issue with the PZTs. We are leaving this issue until tomorrow.

Also, the readback of the POPX whitening settings don't match to the requested. This also needs to be fixed.

kiwamu.izumi@LIGO.ORG - 10:54, Friday 12 August 2016 (29058)

For the QPD loop issue -- after a through investigation today, it tuend out that it was not the PZTs but it was the QPD signals that were not functioninng. I drove to EX again today and found that the whitening chassis for this particular signal was physically switched off. So I switched it on by latching the switch to the on position on the back side of the chasis. The QPD servo now runs. Also, during the investigation I confirmed that the pzt mirros were functioning by sending some sinusoidal excitation and looking at the beam spots on the table.

H1 TCS
kiwamu.izumi@LIGO.ORG - posted 00:11, Friday 12 August 2016 - last comment - 06:20, Friday 12 August 2016(29045)
CO2Y laser power too low

After CO2Y was powered on today, we noticed that the laser power was as low as 16 W which should be nominally 60-ish W. See the attached. This is the one with the RF driver recently swapped (28837). Needs more investigation.

Images attached to this report
Comments related to this report
richard.mccarthy@LIGO.ORG - 06:20, Friday 12 August 2016 (29049)
Jason is aware of this and plans to look at it Friday.
H1 SUS
jeffrey.kissel@LIGO.ORG - posted 16:48, Thursday 11 August 2016 - last comment - 15:59, Friday 12 August 2016(29035)
HAM6 SUS Final Health Assessment -- SUS Dynamics Look Great After Pump-Down
J. Kissel

I took an opportunity between shutter testing and beckhoff problems to gather the final at-vacuum measurements of the HAM6 SUS. I'll process the data in complete detail tomorrow (I attach a screenshot of an example DOF from the OMC), but the message is that the TFs look just as good as they did before the vent: free, clear of rubbing, and all resonances are at the same frequency with roughly the same Q. 

Excellent!

Watch this space for full analysis and comparison with previous results.

/ligo/svncommon/SusSVN/sus/trunk/OMCS/H1/OMC/SAGM1/Data/
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_L_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_P_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_R_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_T_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_V_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_Y_0p02to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM1/SAGM1/Data/
2016-08-11_2249_H1SUSOM1_M1_WhiteNoise_L_0p03to50Hz.xml
2016-08-11_2249_H1SUSOM1_M1_WhiteNoise_P_0p03to50Hz.xml
2016-08-11_2249_H1SUSOM1_M1_WhiteNoise_Y_0p03to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM2/SAGM1/Data/
2016-08-11_2151_H1SUSOM2_M1_WhiteNoise_L_0p03to50Hz.xml
2016-08-11_2151_H1SUSOM2_M1_WhiteNoise_P_0p03to50Hz.xml
2016-08-11_2151_H1SUSOM2_M1_WhiteNoise_Y_0p03to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM3/SAGM1/Data/
2016-08-11_2257_H1SUSOM3_M1_WhiteNoise_L_0p03to50Hz.xml
2016-08-11_2257_H1SUSOM3_M1_WhiteNoise_P_0p03to50Hz.xml
2016-08-11_2257_H1SUSOM3_M1_WhiteNoise_Y_0p03to50Hz.xml
Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 15:15, Friday 12 August 2016 (29068)
J. Kissel

After looking at the results of OMC transfer functions in detail -- using the traditional Matlab plotting software -- I can doubly confirm that the OMC is as healthy as the previous installation of the OMC breadboard. In addition to the cross coupling information which looks quite normal (see 2016-08-11_2305_H1SUSOMC_M1_ALL_TFs.pdf ), a comparison between these measurements, the measurements taken just prior to the vent with the former OMC breadboard, plus a few measurements of LLO's OMC reveals the following:
- For the Vertical, Yaw, Roll, and Transverse DOFs, either all measurements agree with the model, or where they do not, the measured resonant frequencies are consistently offset from from the model implying the model is not accurately representing the physical system.
- For Longitudinal and Pitch, the first pitch mode modeled to be at 1.9 Hz, is different from the model and different from L1 to H1. However, the H1 OMC suspended with the former and current OMC breadboard show the same dynamics, so the discrepancy is likely not due this past week's work or re-suspension.
- Although it appears as though the Q is lower that previous results, it is an artifact of the measurement: in order to expedite testing, the frequency resolution has been reduced from 0.01 Hz to 0.02 Hz. 
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 15:59, Friday 12 August 2016 (29073)
J. Kissel

Attached are the full results for the OMs. Similar to the OMC, I can double confirm after complete analysis that the OMs are as healthy as they were prior to the vent. Nothing to see here!
Non-image files attached to this comment
H1 SEI (CDS, PEM, SEI)
krishna.venkateswara@LIGO.ORG - posted 15:14, Thursday 11 August 2016 - last comment - 13:29, Friday 12 August 2016(29029)
Compact-BRS relevant channels and latest spectra

Krishna

Here are the recorded c-BRS associated channels and their calibrations (I think it is accurate to <20%):

1. H1:NGN-CS_CBRS_RY_PZT1_CTRL_IN1_DQ  :   PhotoDiode 1 output  :  1.6e-10 rad/count

2. H1:NGN-CS_CBRS_RY_PZT1_MASTER_OUT_DQ : PZT1 output  :  2.8e-10 rad/count

3. H1:NGN-CS_CBRS_RY_PZT2_CTRL_IN1_DQ  :   PhotoDiode 2 output  :  1.23e-10 rad/count

4. H1:NGN-CS_CBRS_RY_PZT2_MASTER_OUT_DQ : PZT2 output  :  2.8e-10 rad/count


I've also attached more spectra recorded this morning after the clean room fans were turned off. Many of the noise peaks went away, especially above 100 Hz, but some lines remain which are common to the seismometers and c-BRS. The DIFF signal above 100 Hz, drops to ~10 picorad/rt(Hz), indicating that it's noise floor is indeed well below the SUM channel. I expect that the current SNR at ~10 Hz is between 10-30. I've attached Matlab code showing the analysis.

 

Non-image files attached to this report
Comments related to this report
krishna.venkateswara@LIGO.ORG - 13:29, Friday 12 August 2016 (29050)

Spectra from night time are attached.

Non-image files attached to this comment
H1 ISC
koji.arai@LIGO.ORG - posted 18:16, Tuesday 09 August 2016 - last comment - 09:01, Friday 12 August 2016(28969)
HAM6 vent work - ISC part completed

[Corey, Betsy, Fil, Peter, Calum, Stefan, Koji]

Summary

- The ISC portion of the HAM6 vent work has been completed.

- Shield isolation of the in-vacuum cables was confirmed.

- The fast shutter was reinstalled on the table after some modifications.

=> Ready for the SEI mass balancing and other exiting procedures. OMC PZT HV is still on. Remeber to turn it off before pumping down.


Some details

- Shield isolation: It is always confusing to check the shiels isolation on HAM6 because of several reasons.

  1. All the OMC cables (OMC DCPDs, OMC QPDs, and OMC PZT) are sharing their shield on the OMC breadboard. Therefore one needs to remove the related three DB25 cables from the flange when the shield shorting is checked for them.
  2. The beam diverters connects the shields of their cables shorted to the ISI table. Therefore the shorting needs to be checked with the beam diverter disconnected at the mighty mouse connectors. When the test is done, the connectors need to be restored and the function of the beam diverters should be checked again.
  3. The picomotors don't short the shields on the table. Check them as usual.

We initially had several cables shorting to the chamber but all of them but one happened at the slack of the cables right inside of the flange. The last one happened on the DB25 cable before it climb up to the vertical wall of the ISI.

- The fast shutter modification / reinstallation

(Photos are supposed to come later.)

  1. The new bobbin came in for replacing the moving part of the shutter.
  2. The wire stays were modified to have the bobbin wires more relaxed. Also it should allow us to have more clearance of the wires from the beam towards OM1.
  3. To allow this modification, the direction of a coil wire (shorter, negative, lower slot) in the terminal block was flipped (Corey's attachment 1).
  4. The shutter was placed on the ISI table. Then we found that clearance of the wire is in fact poor. We bent one of the wire clamps (Corey's attachment 2: The clamp at the right side of the shutter) and also adjusted the wire length to have 8~9mm clearance from the beam now. (Corey's attachment 3: The clearance does not look great in the photo but in reality the beam is clearly away from the wire.)
  5. The shutter function was tested with the shutter test controller. No HV was used during the test.
  6. The beam alignment from the shutter to the gold coated tube was checked. The beam is on the tube but close to the top edge of the tube aperture when the shutter mirror wobbles. We believe this does not create any issue even if the beam falls off from the dump as there is nothing sensitive behind the tube beam dump.
  7. Related to the entry  LHO ALOG 28958, we found the shutter height was too low. The distance between the top of the shutter mirror and the beam was ~9mm. The nominal distance is 6mm. We adjusted the height of the shutter to have the nominal height. (Corey's attachment 4)
Comments related to this report
corey.gray@LIGO.ORG - 12:30, Wednesday 10 August 2016 (28984)

Photos From Yesterday's Fast Shutter Work

Photos are on Resourcespace here:

https://ligoimages.mit.edu/?c=1705

Images attached to this comment
corey.gray@LIGO.ORG - 12:58, Wednesday 10 August 2016 (28986)

Cable Grounding Check

(Corey, Fil, Koji)

Documents of interest:

  • Overall HAM6 Cabling document (D1300122)
  • HAM6 Flange Feedthru document (D1002877)
  • Rich Abbott's Grounding & Sheild document (T1200131)

We went through as many of the ISC/SUS cables as we could, which is mainly all the DB25 connectors (most were on a single flange, D6).  The non D6 cables [i.e. SUS OMC & WFS Heads] were checked at the CDS rack (east of HAM6) & the SUS rack (south west of HAM6).  We did not check the RF WFS cables at D5 (not sure how to make check on these connectors, and Koji was worried about disconnecting the connectors since it could make things worse.  We also did not check SEI since those cables were not touched and are nicely clamped & out of the way.

Some notes:

OMC Cables (D6:  F1, F2, & F3):  These cables all go to a harness on the OMC breadboard.  Their sheilds are all tied to each other.  Since we checked these cables at the flange, You must disconnect all three cables at the same time.  Then you can do the ground loop check (otherwise, if you do one at a time, you will trick yourself into seeing ground loops, but this is because their sheilds are all connected).

For the Picomotor cables, I thought Koji said we should disconnect the cable from the picomotor and then check for grounding, but I can't remember if that is really necessary.

Bird's Nest!:  We found some shorts.  When one finds a short, the job is to go in chamber and then "wiggle" the cable in question until you no longer see the short.  I believe we were able to do this in-chamber right next to the D6 flange.  This is where there is a "bird's nest" of ISC cables (unfortunate...I reckon we could clean this up by carefully clamping ISC cables down to HAM6's Stage0.).  

At any rate, whenver one fixes a cable with a ground loop, one must then re-check all cables!  This is because all these cables are in this "bird's nest".  And if you remedied one by moving it around, you don't know whether you made things worse for a neighboring cable in the "nest".  

At the end of the day, we were happy with our ground loop checks for these cables.

OH, a To Do Item:

At the D6 Flange, it was easier for Fil to plug/unplug cables when he removed some protective bars were removed from the flange protector.  At the nearest convenience, those bars should get re-installed.

betsy.weaver@LIGO.ORG - 13:13, Wednesday 10 August 2016 (28987)

Images attached to this comment
hugh.radkins@LIGO.ORG - 14:11, Wednesday 10 August 2016 (28991)

Re the To Do Item--removal of the 'protective bars' aka strain relief, to make it easier to unplug cables.  These should be replaced and returned to relieving strain by the remover.  There is plenty of instances where this has not been done and I expect is the SOP, sadly.  The job is always so much easier if you don't have to return to it later, after you've remembered it.

koji.arai@LIGO.ORG - 16:14, Wednesday 10 August 2016 (28998)

> For the Picomotor cables, I thought Koji said we should disconnect the cable from the picomotor and then check for grounding,
> but I can't remember if that is really necessary.

No. It turned out that the picomotors have no shields at the mighty mouse connectors and have no shorting the the table.

betsy.weaver@LIGO.ORG - 09:01, Friday 12 August 2016 (29051)

HAM6 CC wafer placement - it is now back in the same place as the last one was pulled from.

Images attached to this comment
Displaying reports 57221-57240 of 85498.Go to page Start 2858 2859 2860 2861 2862 2863 2864 2865 2866 End