Displaying reports 57361-57380 of 85646.Go to page Start 2865 2866 2867 2868 2869 2870 2871 2872 2873 End
Reports until 17:36, Friday 12 August 2016
LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 17:36, Friday 12 August 2016 - last comment - 18:06, Friday 12 August 2016(29076)
H1 Output Septum Viewport Glass

HAM6 Crew, vacuum group and others.

During the vent last week, the glass for the septum viewport was inspected by Betsy W. and she found a rather large smudge, a photo was taken of the glass (see attached) and such photo was circulated among the vacuum group and others.
At first, the intent was to clean the contamination by using first contact, but after careful consideration and following procedure it was determined that, in order for us to clean up the window we would have to vent the other side of the viewport, which would have added more time to the HAM6 incursion.
Betsy checked the beam location at the glass, and the beam is far away from the smudge (the beam is center on the glass), that made the decision somewhat easy to postpone the cleaning of the smudge for the next vent event.

FRS ticket  6053

Images attached to this report
Comments related to this report
kyle.ryan@LIGO.ORG - 18:06, Friday 12 August 2016 (29077)
Good log entry Gerardo.  I think the procedure that was referenced was LIGO-T080145 "Viewport Safety Protocols"
H1 SEI
hugh.radkins@LIGO.ORG - posted 16:15, Friday 12 August 2016 - last comment - 10:05, Sunday 14 August 2016(29075)
WHAM6 ISI Trips with Fast Shutter Closing

The attached is just one instance and is fairly representative of the trips.  Most of the time, the ACTuator is called to blame, but as I reported in 29056, I'm not sure the Last Trip is absolutely robust.

The 2 seconds of full data show the ISI in Full Isolation (GRD STATE) when Stefan throws the shutter (FASTSHUTTER A MODE.)  You can see a fraction of a second later, the Actuator MASTER DRIVEs going very large.  All the actuators are contributing to the saturation counting as the threshold is only 32760 counts.  The watch dog trips near the end of the two seconds (WD MON STATE.)  It was going to happen anyway but I can't be sure why the WD tripped because none of the counters (iii SAT COUNT) actually reached the trip threshold.  This is a mystery to me, still.  The SAFE COUNT is 8192, 4096, and 10 for the ACT, GS13, and CPS.  None of the counters reach the SAFE COUNT level....  Regardless, the Actuators where going to reach it so that is another problem to figure out.

Unless the shutter softens its blow on the ISI, the trips are going to happen almost everytime.  For now, we are running with HAM6 in the DAMPED state.  There the HAM6 sees a few saturations on the GS13s when the Shutter triggers but literally just a few.  The HAM6 will be transitioned to ISOLATED by the ISC after the FAST SHUTTER is tested.

Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 10:05, Sunday 14 August 2016 (29094)

We could have the HAM6 ISI in damped while we are testing the shutter, but we want it to beisolated when we are running the IFO, so when we loose lock and the shutter triggers we will stil be frequently tripping the ISI. 

H1 CAL (CAL, INJ)
evan.goetz@LIGO.ORG - posted 16:03, Friday 12 August 2016 (29072)
Pcal inverse actuation filter has correct sign
Summary
I checked that the sign of the Pcal inverse actuation filters is correct by injecting a known sine wave into the TRANSIENT filter bank and measuring at TRANSIENT_OUT and also the calibrated Tx PD signal, H1:CAL-PCALX_TX_PD_OUT_DQ. The signals are consistent and no minus sign is missing.

Details
To check the sign consistency, I made a 100 Hz sine wave with zero phase in Matlab and saved to an ASCII text file. Then I used awgstream to inject the waveform into H1:CAL-PINJX_TRANSIENT_EXC. I wanted to make sure that the output of H1:CAL-PCALX_TX_PD_OUT_DQ was correctly calibrated (not whitened) so I turned on the [:1,1] filter module, FM7. I then injected the waveform using awgstream and then measured the output in DTT, using a bandpass filter to see the 100 Hz signal.

I expect the Tx PD signal to be in phase with the injected waveform, while the TRANSIENT_OUT signal should be 180 degrees out of phase. The attached plot shows the output of H1:CAL-PCALX_TX_PD_OUT_DQ and H1:CAL-PINJX_TRANSIENT_OUT_DQ. The two signals are indeed 180 degrees out of phase, with the Tx PD output rising on the integer second as it should be, while the TRANSIENT_OUT is falling on the integer second (with some small delay, ~240 usec).

Thus, I conclude the sign of the inverse actuation filter (and, by extension, the actuation functions provided for the CW injections) is correct. If there is a sign flip, perhaps it is on the output side? This needs to be investigated.
Non-image files attached to this report
H1 General
jeffrey.bartlett@LIGO.ORG - posted 16:01, Friday 12 August 2016 (29074)
Ops Day Shift Summary
Title:  08/12/2016, Day Shift 15:00 – 23:00 (08:00 –16:00) All times in UTC (PT)
State of H1: IFO is unlocked.  
Commissioning: Recovering the IFO.
Outgoing Operator:  N/A
 
Activity Log: All Times in UTC (PT)

14:56 (07:56) Chris – Going down Y-Arm to deliver tools to Ed 
15:00 (08:00) Start of shift
15:19 (08:19) Jim & Krishna – Going the End-Y to reset BRS
15:50 (08:50) Stephen – Going into the Optics lab
16:00 (09:00) Karen – Going into the LVEA to cleanup around HAM6
16:01 (09:01) Kiwamu – Transitioning End-X to Laser Hazard
16:11 (09:11) Gerardo – Checking Ion pumps at HAM5/6
16:20 (09:20) Gerardo – Out of the LVEA
16:25 (09:25) Jim – Restarting NDS1
16:33 (09:33) Ed – Going to End-X to work on wind instrument install
17:12 (10:12) Krishna & Jim – Back from End-Y
17:13 (10:13) Krishna – Going into the Biergarten to work on Compact BRS
17:58 (10:58) Koji & Stefan – Going to ISCT6 to test shutter timing
19:05 (12:05) Koji & Stefan – Out of the LVEA
19:06 (12:06) Krishna – Out of the LVEA
19:14 (12:14) Ed – Back from End-X
20:04 (13:04) Kyle – Going to End-Y for RGA work
20:57 (13:57) Gerardo – Going to both Ends to check on test stands
22:14 (15:14) Kiwamu – Going out to check on PDs
22:15 (15:15) Gerardo – Going into LVEA 
22:25 (15:25) Kiwamu – Out of the LVEA
22:30 (15:30) Kyle – At End-Y doing RGA sweeps – Seeing contamination spikes in vac
22:58 (15:58) Kyle – Back from End-Y

Title: 08/12/2016, Day Shift 15:00 – 23:00 (08:00 – 16:00) All times in UTC (PT)
Support:  Kiwamu, Stefan, Sheila, 
Incoming Operator: Patrick

Shift Detail Summary: Commissioners testing fast shutter operation in the morning. Around 12:00 started to align the IFO. Took several tries and a lot of work by the commissioning team to complete the alignment. Started locking. Commissioners are working through various challenges. 
IFO locked at DRMI_LOCKED. Commissioners continuing working on the fast shutter and locking.
Dale – brought a small tour (his last) into the control room. He was given a hearty round of applause by all present. Good luck to Dale, from his LIGO family, as he starts this next journey.        
H1 ISC
stefan.ballmer@LIGO.ORG - posted 15:22, Friday 12 August 2016 (29071)
FAST_SHUTTER guardian now check ASC-AS_A and ASC-AS_B light levels during shutter test

Updated the FAST_SHUTTER guardian to  check ASC-AS_A and ASC-AS_B light levels during shutter test.

The "light on diodes" level is scaled from the trigger PD levels from beckhoff (H1:SYS-MOTION_C_SHUTTER_G_TRIGGER_LOW and H1:SYS-MOTION_C_SHUTTER_G_TRIGGER_HIGH), using a relative gain specified in the FAST_SHUTTER guardian.
The dark limits are specified in in the FAST_SHUTTER guardian.

The changes include:

- A check thet the downstream PD levels are OK when the shutter is open on every lock
- During a fast-shutter test:
    - A check that the downstream PD levels are OK with the shutter open.
    - A check that the downstream PD levels are dark for 9 seconds, starting 0.5 sec after the fast shutter closed.
    - A check that the downstream PD levels are OK after the sutter reopens.

 

We also tested compatibility with the HAM6 ISI - we seem to trip it a lot on shutter firings when isolated. For now we will leave it damped until after the test.
The guardian was checked into svn as version 14028.

LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 15:19, Friday 12 August 2016 - last comment - 15:20, Friday 12 August 2016(29069)
Over-filled CP3 at 21:30 utc

Over-filled CP3 with the exhaust bypass valve fully open and the LLCV bypass valve 1/2 turn open.

Flow was noted after 10 minutes 23 seconds, closed LLCV valve, and 3 minutes later the exhaust bypass valve was closed.

Comments related to this report
gerardo.moreno@LIGO.ORG - 15:20, Friday 12 August 2016 (29070)
I raised CP3's LLCV from 18% to 19%, due to the amount of time it took to overfill today.
H1 General (AOS, CAL, CDS, COC, CSWG, DAQ, DCS, DetChar, GRD, INJ, INS, IOO, ISC, Lockloss, OpsInfo, PEM, PSL, SEI, SUS, SYS, TCS, VE)
vernon.sandberg@LIGO.ORG - posted 14:26, Friday 12 August 2016 (29066)
Return to normal operations, commissioning at 50W ifo input power

The WHAM6 "Fast Shutter" system and OMC were repaired and replaced, respectively, during the recent vent.  The logic for the fast shutter has been augmented,  reviewed, and tested (see recent alogs for test results).  A TRB review of the OMC-shutter incident took place this morning and their recommendation was to return to normal operations.  The new fast shutter monitoring and control software (ECR E1600246) should now be in place and in routine use. Commissioining may resume at the 50W input power level of interferometer operation.

This relieves LHO alog #29010, "Input power to be held to be < 25 W until technical review of fast shutter
Notice from LIGO Systems".
 

H1 PEM (SEI)
edmond.merilh@LIGO.ORG - posted 13:48, Friday 12 August 2016 (29065)
Ultrasonic Anemometer Installed at End-X

This morning I installed the Young Model 81000 Ultrasonic Anemometer next to the wind fence at End-X. It's being powered by a sand alone 18VDC supply. All connections are being routed through two SOLA STC-DRS-038 Isolated Loop Circuit Protectors located in the rear of the electronics rack  in the VEA. There are 3 BNC pigtails awaiting connection through the PEM patch panel. Unfortunately, I didn't take a picture of the actual anemometer.

Images attached to this report
H1 DAQ
james.batch@LIGO.ORG - posted 13:29, Friday 12 August 2016 (29064)
h1nds1 reconfigured, restarted
WP 6069

The copying of raw minuted trend files from h1tw1 is complete and the copied files have been verified.  The nds1 service on h1nds1 has been restarted with a new configuration to read the copied files from their new location.
H1 ISC
koji.arai@LIGO.ORG - posted 12:25, Friday 12 August 2016 - last comment - 13:59, Friday 12 August 2016(29060)
Timing test for the OMC PZT shutter and the fast mechanical shutter

[Stefan, Koji]

Quick Summary

Shutter timing test has been performed for the OMC PZT shutter and the fast mechanical shutter. Everything seems reasonable.

- The trigger logic fires at the PD input of +2V.

- The trigger logic reacts at 1.5us since the PD input thresholding.

- The PZT shutter detunes the OMC cavity at 4.8us.

- The fast mechanical shutter start blocking the beam at 1.7ms, reaches half blocking at 1.85ms, complete block at 2.0ms.

More about the test comes later.

Comments related to this report
rich.abbott@LIGO.ORG - 13:02, Friday 12 August 2016 (29061)ISC
The fast shutter timing numbers agree well with the values seen during the initial installation, so I agree with the health status.
koji.arai@LIGO.ORG - 13:59, Friday 12 August 2016 (29062)

Shutter Test Setup (Attachment1)

- The PD signal from AS_C QPD was temporarily disconnected from the left PD port. Instead, a function generator (SRS DS345) was connected. The static offset of the func gen was used to mimic the PD signal.

- The output of the shutter logic unit (D1102312) was branched by a BNC-T.

- A thorlabs PD (PDA55, Si DCPD, BW ~10MHz) was placed on the ISCT6 to monitor the power at OMCT and OMCR for testing the PZT shutter and the fast mechanical shutter, respectively.

- The monitor signals were connected to an oscilloscope (TDS3034B). The yellow, blue, and magenta traces corresponds to the FG offset, trigger signal, and the optical power at OMCT/R.

Shutter Test Results

- The offset voltage of DS345 was manually swept from 0V to a positive voltage until the trigger fired. We confirmed that the trigger fired at 2V.

- The offset voltage was changed from 0V to 5V on DS345. This made the actual output voltage changed from 0V to ~9V. The output voltage was not so clean because the relay seemed to be switched upon the change. However this essentially does not affect the test.

- The oscilloscope traces showed that the trigger logic reacted 1.5us after the PD input crossed the threshold. (Attachment 2)

- The OMCT optical power started to drop 4.8us after the thresholding (Attachment 3) and reach 1/10 of the power after ~60us (Attachment 4). The trace for the OMCT power showed a decay curve. The half life was 10~20us depending on which part of the curve is taken. This corresponds to the cutoff freq of 5.5kHz ~ 11kHz, which is way to low and unlikely to be the cavity pole of the OMC (330kHz). It is more likely to be determined by the actual motion speed of the PZT.

- The OMCR light was used for the test of the fast mechanical shutter. During this test we didn't try to lock the OMC cavity to make the observation of the light drop easier to interpret. The shutter started blocking the beam at 1.7ms, reached half blocking at 1.85ms, completed blocking at 2.0ms. (Attachment 5). The observation in a longer time scale (~0.5s) showed that there was no boucing effect of the mechanical shutter any more. (Attachment 6)

Images attached to this comment
H1 SEI
hugh.radkins@LIGO.ORG - posted 12:19, Friday 12 August 2016 - last comment - 08:12, Tuesday 23 August 2016(29056)
SEI response to 7.2 EQ in SW Pacific (New Caledonia)

HEPI BS Tripped few minutes before ITMX ISI.  This is the only HEPI that tripped in the neighborhood of the large quake.

ITMY ISI tripped--timing (H1:ISI-ITMY_ST2_WD_MON_GPS_TIME) indicates stage2 tripped on ACTuators 1 second before Stage1 on T240s but looking at the plots, the Actuators have only registered a few counts, nothing near saturation/trip level.  But the T240s hit their rail almost instantly.  It seems the Stage2 Last Trip (H1:ISI-ITMY_ST2_WD_MON_FIRSTTRIG_LATCH) should be indicating ST1WD rather than Actuator. On ETMY, the Trip Time is the same for the two stages and Stage2 notes it is an actuator trip but again, there are only a few counts on the MASTER DRIVE; seems this too should have been a ST1WD trip[ indication trip on Stage2--I'll look into the logic.

On the BS ISI, the Stage1 and Stage2 trip times are the same, and the Last Trip for Stage2 indicates ST1WD.  The Stage2 sensors are very rung up after the trip time but not before unlike the T240s which are ramping to to the rail a few seconds before trip.  ETMX shows this same logical pattern in the trip sequence indicators.

On the ITMX ISI, Stage1 Tripped 20 minutes before the last Stage2 trip. This indicates the Stage1 did not trip at the last Stage2 trip.

No HAM ISI Tripped on this EQ.

Bottom line: the logical output of the WDs are not consistent from this common model code--needs investigating. Maybe I should open an FRS...

Attachment 1) Trip plots showing Stage2 trip time 1 second before the stage1 trip where the stage2 actuators do not go anywhere near saturation levels.

Attachment 2) Dataviewer plot showing the EQ on the CS ground STS and the platform trip times indicated.

Images attached to this report
Comments related to this report
hugh.radkins@LIGO.ORG - 15:20, Monday 22 August 2016 (29232)

It seems this is not a problem with the watchdog but a problem with the plotting script.  It seems for ST2 Actuators, it misses a multiplier on the Y axis.  It works correctly for ST1 Actuators and all the sensors; it does not work for other chambers as well for ST2 ACT.  FRS 6072.

hugh.radkins@LIGO.ORG - 08:12, Tuesday 23 August 2016 (29241)

Actually, the plotting script is working fine.  When the spike is so large that the plotting decides to switch to exponential notation, the exponent is hidden by the title until you blow up the plot to very large size. 

H1 ISC (ISC, Lockloss)
richard.mccarthy@LIGO.ORG - posted 11:04, Friday 12 August 2016 - last comment - 08:17, Monday 15 August 2016(29059)
Modifications to ISC rack and Fast Shutter near Ham6
In an effort to eliminate the need to connect and disconnect shutter cables during vent we have moved the shutter logic controler to the rack from the table.  Also replaced the BNC cables with Yellow BNC cables so they can be easily identified.
See attachment.
Non-image files attached to this report
Comments related to this report
daniel.sigg@LIGO.ORG - 08:17, Monday 15 August 2016 (29099)

Drawings D1200196 and D1201500 updated.

H1 SEI (SEI)
krishna.venkateswara@LIGO.ORG - posted 10:31, Friday 12 August 2016 - last comment - 13:39, Friday 12 August 2016(29057)
NN Array L4C 7 and 8 swapped temporarily

Krishna,

Jenna and I had noticed that L4C 7 didn't appear to be working. L4C is close to the c-BRS, so it would have been good for comparison. In order to test if it was the instrument or the electronics, I temporarily swapped the cables with those of L4C 8 going in to the SUS rack at the LVEA. Now L4C "7" appears to be working and L4C "8" is not, indicating that the problem must be on the instrument side (or the pre-amp, I guess?). I'll swap it back in the afternoon.

Comments related to this report
krishna.venkateswara@LIGO.ORG - 13:39, Friday 12 August 2016 (29063)

Cables replaced in original configuration.

After looking at a different working L4C output (5), I think I was wrong about the instrument being faulty. It looks like the L4C was working when I swapped the cables, suggesting that the problem must be on the electronics side.

LHO General
patrick.thomas@LIGO.ORG - posted 00:13, Friday 12 August 2016 - last comment - 10:54, Friday 12 August 2016(29046)
Ops Evening Summary
Stefan and Sheila working on shutter control logic. I helped resolve problems with h1ecatc1. Had to restart EPICS IOC and copy updated PLC2 to target.

23:32 UTC Gerardo to HAM6 area to check on vacuum equipment
23:39 UTC Gerardo back
03:05 UTC Joe to Y2-5 to watch meteor shower with telescope
04:30 UTC Stefan and Sheila done working on shutter control logic. Kiwamu starting work.
06:17 UTC Kiwamu to end X to investigate green laser electronics
07:00 UTC Kiwamu back

I was able to get the Y arm green power to around 1 with LOCKED_NO_SLOW_NO_WFS, but if I put it on INITIAL_ALIGNMENT it would start to be driven away. Kiwamu was investigating the X arm.
Comments related to this report
kiwamu.izumi@LIGO.ORG - 00:25, Friday 12 August 2016 (29048)

Patrick, Kiwamu,

We are unable to engage the QPD servo for the X green injection beam. It seems to be an issue with the PZTs. We are leaving this issue until tomorrow.

Also, the readback of the POPX whitening settings don't match to the requested. This also needs to be fixed.

kiwamu.izumi@LIGO.ORG - 10:54, Friday 12 August 2016 (29058)

For the QPD loop issue -- after a through investigation today, it tuend out that it was not the PZTs but it was the QPD signals that were not functioninng. I drove to EX again today and found that the whitening chassis for this particular signal was physically switched off. So I switched it on by latching the switch to the on position on the back side of the chasis. The QPD servo now runs. Also, during the investigation I confirmed that the pzt mirros were functioning by sending some sinusoidal excitation and looking at the beam spots on the table.

H1 SUS
jeffrey.kissel@LIGO.ORG - posted 16:48, Thursday 11 August 2016 - last comment - 15:59, Friday 12 August 2016(29035)
HAM6 SUS Final Health Assessment -- SUS Dynamics Look Great After Pump-Down
J. Kissel

I took an opportunity between shutter testing and beckhoff problems to gather the final at-vacuum measurements of the HAM6 SUS. I'll process the data in complete detail tomorrow (I attach a screenshot of an example DOF from the OMC), but the message is that the TFs look just as good as they did before the vent: free, clear of rubbing, and all resonances are at the same frequency with roughly the same Q. 

Excellent!

Watch this space for full analysis and comparison with previous results.

/ligo/svncommon/SusSVN/sus/trunk/OMCS/H1/OMC/SAGM1/Data/
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_L_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_P_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_R_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_T_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_V_0p02to50Hz.xml
2016-08-11_2305_H1SUSOMC_M1_WhiteNoise_Y_0p02to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM1/SAGM1/Data/
2016-08-11_2249_H1SUSOM1_M1_WhiteNoise_L_0p03to50Hz.xml
2016-08-11_2249_H1SUSOM1_M1_WhiteNoise_P_0p03to50Hz.xml
2016-08-11_2249_H1SUSOM1_M1_WhiteNoise_Y_0p03to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM2/SAGM1/Data/
2016-08-11_2151_H1SUSOM2_M1_WhiteNoise_L_0p03to50Hz.xml
2016-08-11_2151_H1SUSOM2_M1_WhiteNoise_P_0p03to50Hz.xml
2016-08-11_2151_H1SUSOM2_M1_WhiteNoise_Y_0p03to50Hz.xml

/ligo/svncommon/SusSVN/sus/trunk/HTTS/H1/OM3/SAGM1/Data/
2016-08-11_2257_H1SUSOM3_M1_WhiteNoise_L_0p03to50Hz.xml
2016-08-11_2257_H1SUSOM3_M1_WhiteNoise_P_0p03to50Hz.xml
2016-08-11_2257_H1SUSOM3_M1_WhiteNoise_Y_0p03to50Hz.xml
Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 15:15, Friday 12 August 2016 (29068)
J. Kissel

After looking at the results of OMC transfer functions in detail -- using the traditional Matlab plotting software -- I can doubly confirm that the OMC is as healthy as the previous installation of the OMC breadboard. In addition to the cross coupling information which looks quite normal (see 2016-08-11_2305_H1SUSOMC_M1_ALL_TFs.pdf ), a comparison between these measurements, the measurements taken just prior to the vent with the former OMC breadboard, plus a few measurements of LLO's OMC reveals the following:
- For the Vertical, Yaw, Roll, and Transverse DOFs, either all measurements agree with the model, or where they do not, the measured resonant frequencies are consistently offset from from the model implying the model is not accurately representing the physical system.
- For Longitudinal and Pitch, the first pitch mode modeled to be at 1.9 Hz, is different from the model and different from L1 to H1. However, the H1 OMC suspended with the former and current OMC breadboard show the same dynamics, so the discrepancy is likely not due this past week's work or re-suspension.
- Although it appears as though the Q is lower that previous results, it is an artifact of the measurement: in order to expedite testing, the frequency resolution has been reduced from 0.01 Hz to 0.02 Hz. 
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 15:59, Friday 12 August 2016 (29073)
J. Kissel

Attached are the full results for the OMs. Similar to the OMC, I can double confirm after complete analysis that the OMs are as healthy as they were prior to the vent. Nothing to see here!
Non-image files attached to this comment
H1 SEI (CDS, PEM, SEI)
krishna.venkateswara@LIGO.ORG - posted 15:14, Thursday 11 August 2016 - last comment - 13:29, Friday 12 August 2016(29029)
Compact-BRS relevant channels and latest spectra

Krishna

Here are the recorded c-BRS associated channels and their calibrations (I think it is accurate to <20%):

1. H1:NGN-CS_CBRS_RY_PZT1_CTRL_IN1_DQ  :   PhotoDiode 1 output  :  1.6e-10 rad/count

2. H1:NGN-CS_CBRS_RY_PZT1_MASTER_OUT_DQ : PZT1 output  :  2.8e-10 rad/count

3. H1:NGN-CS_CBRS_RY_PZT2_CTRL_IN1_DQ  :   PhotoDiode 2 output  :  1.23e-10 rad/count

4. H1:NGN-CS_CBRS_RY_PZT2_MASTER_OUT_DQ : PZT2 output  :  2.8e-10 rad/count


I've also attached more spectra recorded this morning after the clean room fans were turned off. Many of the noise peaks went away, especially above 100 Hz, but some lines remain which are common to the seismometers and c-BRS. The DIFF signal above 100 Hz, drops to ~10 picorad/rt(Hz), indicating that it's noise floor is indeed well below the SUM channel. I expect that the current SNR at ~10 Hz is between 10-30. I've attached Matlab code showing the analysis.

 

Non-image files attached to this report
Comments related to this report
krishna.venkateswara@LIGO.ORG - 13:29, Friday 12 August 2016 (29050)

Spectra from night time are attached.

Non-image files attached to this comment
Displaying reports 57361-57380 of 85646.Go to page Start 2865 2866 2867 2868 2869 2870 2871 2872 2873 End