Evan, Kiwamu, Dan, Peter, Lisa The goal for tonight was to relock the interferometer on DC readout, properly calibrate the spectrum (see Kiwamu's entry with today measurements), and in the process see if theIO periscope PZT filters had improved the beam jitter noise that we think is responsible for the bump around 100 Hz in the sensitivity. The only other (known) change with respect to yesterday was the new DARM filter designed to give us better phase margin, with the goal (surprise!) of making the locking sequence more robust. However, we soon realized that we had to retune the transition from ALS COMM to CARM (common mode board input 1 changed from -16 dB to -7 dB) as the current gains were totally wrong (meaning that corresponding OLTF were extremely marginal). We fixed the gains up to the transition to DARM on AS 45 Q, to discover that there was no chance of moving forward with the CARM TF we measured (Evan will post the offending TF later). Even if we believe the new DARM filters had little to do with the problem we were seeing, we decided to go back to the old DARM filters and the Guardian CODE which was running last night, as a sanity check. Still, we had to keep the low CARM gain to transition to transmission signals for CARM, then we made a couple of attempts to full lock, but we would lose lock before transitioning to AS 45 Q, this time because the DARM gain was wrong.. So, the message is that the problems we see are really easy to identify (loop oscillation, nonsense TFs) and we could keep fixing the locking sequence, but we don't want to make very radical changes in the gain/filter tuning at this time. Old DARM filters and Guardian code are running (as last night).
This is a brief and preliminary update of the calibration activity from today. I calibrated the ITMX and ETMX reponses using the usual free-swing Michelson fringe.
If I believe the measurement, I must have underestimated the ESD response by a factor of 5.3 (!?) in the previous calibration which is hard to believe for me. I would like to repeat the measuerment perhaps with different conditions (e.g. opelv on/off, L2P off, linearization off/on, different bias, different frequencies and etc) and on ETMY as well.
(MICH free swing)
The method is the same as what Joe described in LLO alog 14135. To obtain the ASQ_pkpk value, I did not run the fancy matlab code or anything, but I just picked up a highest peak value and lowest one in H1:LSC-MICH_IN1_DQ. The alignment was adjusted beforehand by locking MICH. The pk-pk value was measured to be 27.0 counts. Using the relation, d (ASQ)/d(ITMX) = 2 * pi * ASQ_pkpk / lambda, I get
ASQ optical gain = 1.59 x 108 cnts/m
The input power to IMC was at 9.6 W, measured at the periscope bottom PD. ASAIR_ALF could get to 4550 counts at maximum and ASAIR_B_LF 1500 counts when MICH was freely swinging. The below are some dtails:
(ITMX L2 stage calibration using MICH)
After locking MICH, I shook ITMX L2 stage at H1:LSC-SUS_ITMX_L2_LOCK_L_EXC with a drive level as high as possible without DAC saturation. I did a swept sine measurement to check how high frequency I would be able to get without loosing good signal-to-noise ratio. It seems that exctiation above 20 Hz is hopeless -- the drive signal dives into sensor noise. From this measurement I picked up one frequency point, 13.05 Hz where the ITM response was measured to be
ITMX L2 response = 8.41 x 10-18 m/cnts @ 13.05 Hz
(ETMX calibration using X arm)
Keeping the 9.6 W incident power, I locked the IR laser to the X arm with a UGF of 100-ish Hz. I did a swept sine measurement on ITMX and ETMX at different times but in the same lock strech. On ITMX, the L2 stage was driven again with the same setting as that of the MICH locking. On ETMX, I had set up the suspension filters such that they are the same as the full locking condition (e.g. drive signal goes not only ESD but also L1 stage and so on). Neverthelss, since my swept sine measurement does not go below 10 Hz, the ETMX response essentially represents the ESD response (with a small effect from the L2 stage which is almost two orders of magnitude smaller than the ESD in terms of displacement).
Taking the ratio between the two actuators, I confirmed that the ratio goes as f^2 as expected in a frequency range from 10 to 60 Hz. The ETMX/ITMX ratio was measured to be
ETMX_L3 / ITMX_L2 = 1.70 x 102 @ 13.05 Hz
ETMX_L3 response = 1.43 x 10-15 m/cnts @ 13.05 Hz. This is almost 5.3 times stronger than what we have in the CAL-CS calibration.
I set up the BS bounce mode damping loop, tested it and confirmed that it works very fine. One should be now able to lower the peak height of the mode by a factor of 10 or so in a minute.
During the calibration activities in this evening, I ran into a problem with the MICH locking where I was not able to lock it. It turned out that the bounce mode was too high so that the length control saturated the DACs on the BS suspension. I guess that the mode was excited due to several times of locking attempts as I was trying unusual locking settings for the calibration. The below is a screen shot of the bounce mode damping filter:
The design process was essentially the same as similar to what Jeff described for LLO's violin mode damping loos (LLO alog 15000) but I did not try to drive it as hard as DAC saturation. I am leaving the filter enabled for now. If someone find this filter doing something bad, please feel free to switch off at its output (because turning off at the input seems to create a huge impulse response). I did not explore the gain so much.
Replaced hard coded optic and site strings with macros, and fixed sme errors in the on/off indicator, both of which were introduced when copying the DARM_ERR damping stuff from L1:
/opt/rtcds/userapps/release/sus/common/medm/quad/SUS_CUST_QUAD_M0_DAMP_MODE.adl
/opt/rtcds/userapps/release/sus/common/medm/quad/SUS_CUST_L2_DAMP_MODE_FILTERS.adl
/opt/rtcds/userapps/release/sus/common/medm/quad/SUS_CUST_QUAD_OVERVIEW.adl
Attached is a plot of the residual (in-lock) DARM spectrum, calibrated in meters according to the calibration factor that is used in OAF: 2.86e-7 m/cnt for DARM_IN1. The residual is about 7e-14 m-rms. The residual is dominated by the 0.45 Hz quad pendulum mode. We should suppress this peak further. There is a resonant gain filter in DARM, but it is a little too low (0.43 Hz vs 0.45), and just is not enough. However, there is also a dip in the SUScomp filter at 0.45 Hz, which reduces the gain at that frequency by 10-12 dB. This compensates the damped pendulum response -- but what is the point of including here? It's not like we're going to try and have a ~1 Hz bandwidth for this loop. Seems like the thing to do is to remove this feature (some complex pole and zero pairs around 0.4-0.6 Hz) from the SUScomp filter, which will increase the gain at 0.45 Hz by 10-12 dB.
Next would be further squashing the noise from 1-5 Hz. The L1 DARM boosts have more gain in this band, so we could adopt that design. First we should make sure we are not injecting noise at these frequencies from residual OpLev feedback, or quad local (BOSEM) damping.
08:01 I changed the state bit from Undisturbed to Commissioning. However there is an entry on the card reader for Peter K. entering the LVEA at 6:50 AM. Karen and Cris also entered the LVEA after Peter and before 08:01. 08:04 Corey to the squeezer bay for 3IFO work 08:13 Alastair to the LVEA to check on the TCSY laser 08:49 Thomas and Elli to end X and end Y to reboot PCAL computers 09:34 Thomas and Elli back 09:34 Jim W. working on HAM4 in the control room 09:38 Corey done 09:43 Filiberto and Manny back from end X, were soldering cables 09:45 Filiberto to end X 10:08 Filiberto back ~11:18 Richard, Peter and Kiwamu working on installation of low pass filter for IOO periscope piezo controller 11:53 Richard back 12:42 Betsy running transfer functions on ETMX, optical lever damping turned off 12:49 Travis starting measurement on SRM 13:16 Karen opening outside rollup door in OSB shipping/receiving 13:16 Elli to LVEA to refocus ITM cameras 13:21 Kiwamu to IOT2L table to check if ghost beam is on photodiodes 13:32 Kiwamu back 15:49 Nutsinee to end X to change PCAL camera autofocus from manual to auto
The phase and gain margin in the DARM loop in the final configuration is quite small. In the interest of improving it, I started by comparing what is used on DARM for H1 and L1. The attached plots show the comparison; the differences are:
Attached plots:
I have modified the H1 DARM suspension compensation filter so that it is now just like the L1 version. Previously this filter (FM8) was called 'LLO', but I changed the name to 'SUScomp'. I also changed the 60 Hz notch (in FM7) to be a bandstop for the violin modes. This is not exactly the same filter as in L1 since the H1 violin mode frequencies are a bit lower (though more narrowly grouped) than L1. This filter is now called 'Violin', and is defined to be: ellip("BandStop",4,1,80,480,530)
Of course the new FM8 filter has more gain than the previous one above ~50 Hz. This could be a problem in terms of ESD range while using ALS-DIFF, so there may need to be some tailoring of other low-pass filters that are used in that state of the acquisition sequence.
I haven't made any changes to the boost/resonant gain filters yet. The spectrum of the residual (in-lock) DARM should be looked at to see if these should be modified.
Just to minimize confusion: - we know that the current calibrated spectra in H1:CAL-DELTAL_EXTERNAL_DQ is wrong, so don't look at this channel yet. There is an overall factor that must be wrong by at least a factor of 2, as the high frequency noise is too good given the current circulating power in the arms; - also, as explained here , we know that the floor of the high frequency part of the spectrum is shot noise, but there was something wrong about the power scaling and/or scaling with DARM offset in the high power data from last night, and that's why the calibrated spectrum was not scaling as expected. (More) Accurate calibration measurements are in progress.
No, the calibration is wrong.
Alan,
We think the calibration is off by an overall factor of a few. We are makeing another set of measurements to make it more accurate.
The piezo low-pass filter box (D1500001) S/N S1500001 was installed. Measured voltages at the 15-pin analog interface pin before after signal 1 0.110 0.016 V-MON-X 2 0.206 0.143 V-MON-Y 3 0.109 0.160 V-MON-1 4 0.206 0.143 V-MON-2 5 0.000 0.000 V-MON-3 6 0 4.664 Servo-1 OFF/ON 7 0 4.656 Servo-2 OFF/ON 8 GND 9 0.160 -0.476 SGS-MON-X 10 1.431 1.173 SGS-MON-Y 11 0.163 -0.472 SGS-MON-1 12 1.430 1.173 SGS-MON-2 13 -13.15 -13.12 SGS-MON-3 14 4.824 4.821 OFL1 15 4.821 4.819 OFL2 DIP switches 1 & 2 were switched from ON to OFF. To bring the beam back, most of the adjustment was done with the control slider for pitch, which went went a couple of hundreds on the slider to ~4000. The yaw slider was adjusted by a few hundred. Kiwamu, Richard, Peter
What does changing the DIP switches from ON to OFF do? Are these for closed loop operation using the built-in strain gauge sensors? And if so, were they operating closed loop before, and are now open, or vice versa? (They should not have been operating closed loop, because the strain gauge sensors inject noise.)
distance from the PSL table to the acoustic enclosure wall = 4' 10-11/16" distance from acoustic enclosure wall to periscope face plate = 2' 9" The face plate is labelled as IO ALS M5.
Sudarshan, Thomas, Darkhan, Travis, Rick
At 104.7. Hz the displacement DARM ASD was 1.26e-16 m/rtHz , the ASD according to the Pcal line was 2.07e-16 m/rtHz. The ratio is 1.64, with Pcal reporting a higher ASD.
We were running the Pcal calibration line at 104.7 Hz at LHO ENDY (25000 cts. excitation amplitude) during the lock stretch
After calibration of the power going to the test mass using the Working Standard at Yend and using the radiation pressure induced displacement equation
x= 2*P*Cos(theata)/M*c*(2*pi*f)^2
we have calculated (preliminary) the calibration value for the Transmission Monitor photodetector to be
x= 1.68e-12 (1/f^2)*(m/V)
Attached is the calibrated plot of Pcal ASD together with the DARM ASD in the region near the calibration line.
Found one of the channels for PEM AA chassis at EX (CH24) has an offset of about ~1500cts. Will pull unit on Tuesday to troubleshoot.
At GPS time 1107854779 [2015-02-13 09:26:03 UTC], H1 appeared to unlock but the ISC_LOCK guardian state didn't change. The attached plot shows the X-arm power and the guardian state to confirm (hopefully).
At the time of writing the guardian is still reporting 'DC_READOUT' as the current state, while there is no circulating power in the arms. The other guardian nodes (e.g. ISC_DRMI) did respond to the lockloss, however, so I presume it's just an oversight in the ISC_LOCK DC_READOUT state definition to check that the other guardians are still nominal, or that the arms are actually locked.
This should be fixed now.
Dan, Jeff, Sheila
Tonight we had some more fun noise hunting. Turning the power up to 8 Watts was easy, but didn't improve the noise so much since we are mostly not shot noise limited.
This gave us a lower noise floor, we sat here from 5:28:10 UTC Feb 13th to 5:41:29 UTC, with the intent bit undisturbed.
We then made a single destructive attempt to reduce the ESD bias voltage, we'll have to come back to this. :)
On the next lock we decided to try increasing the input power. This was suprisingly easy, we have now been sitting at 8 Watts input power since 7:03:43 UTC. We have been adjusting the BS alignment by hand every 10-15 minutes or at 10 Watts. We have watched the AS36Q WFS signals while we do this, and can see that they are not always correct (zero does not always give us the best AS90). BS yaw dramatically improves the intensity noise coupling. It seems as though the IFO would stay locked like this all night if we sat here and adjusted the BS once in a while. We are leaving it locked and undisturbed.
In the attached plot, the green trace is earlier today, The blue trace is with 1 stage of DCPD whitening on and the ISS second loop on, which is injecting noise at around 700 Hz. The red trace is now, with 8 Watts input power, BS YAW oplev damping gain reduced, and the ISS second loop off. You can see that we are not really shot noise limited, as increasin the power by a factor of 2.8 only gives us a small improvement at high frequencies. Something is clearly wrong with the calibration.
The attached plot illustrates eight hours of progress today. (Times are approximate.)
At 3:30pm we were using the in-vacuum POP photodiode for MICH, PRCL, and SRCL.
At 9:30pm the violin damping ninjas had allowed us to enable one stage of whitening on the DCPDs. The ISS second loop was on for this lock stretch. During this lock the intent bit was set to OBSERVATION for about ten minutes (see times above), and ISC_LOCK Guardian was in the 'DC READOUT' state.
At 11:30pm we had increased the power from 2.8W --> 7.8W. But, we forgot to turn on the ISS second loop. Tweaking the BS alignment reduced the beam jitter coupling at 100-300Hz and around 700Hz. There is a lot of coherence above 1kHz with the LSC DOFs, we believe that we are not dark noise or shot noise limited in this region.
As we increased the power we changed the DARM offset, gain, and OMC scale factor & gain to maintain good phase margin. At 7.8W we could not incerase the DARM offset to provide ~16mA in the carrier mode at the OMC, the DCPDs would saturate from the violin mode. The last lock stretch had a DARM offset of 2e-5 and a DCPD sum of ~12mA.
We noticed the 1/f^3 noise between 30 and 70Hz was breathing on a several-second timescale. The noise here is somewhat coherent with MICH, we wonder if it is also due to ESD DAC noise or DAC zero crossing glitches. We have left the IFO locked in 'DC READOUT' with the intent bit set to observe. DetChar, please examine the data for glitches associated with major-carry transitions.
Glitch investigation: Keep in mind that there are about a dozen IPC erros per second on the controls signal sent to the ETMs.
The interferometer unlocked around 9:25UTC, so it stayed locked for more than 2 hours at 8 W. The plot shows a 14 hour trend from yesterday, with several hours of lock overall. Most of the unlocks were commissioning-related. (P.S.: still very quiet seismic environment).
Shot noise dominates the noise floor in kHz range, there was something wrong about the power scaling and/or scaling with DARM offset.
The first attachment shows that, in calibrated spectrum, the Pcal line in 7.8W (red) was a factor of 1.6 larger than in 2.8W (blue).
There should have been something wrong about power scaling somewhere, and if you want to make a fair comparison between 7.8W and 2.8W, 7.8W should be brought down by a factor of 1.6.
If you take this into account, 7.8W noise floor would be about a factor of 1.7 smaller than 2.8W for, say, f>7000Hz. (The top panel in the second attachment shows that 7.8W data is 5% or so smaller than 2.8W, and 1.6/0.95 = 1.7.)
sqrt(7.8W/2.8W)=1.7.
Further, in the same plot, the bottom and the middle panel shows that both in 7.8W (red, blue) and 2.8W (green, brown) the noise floor of DCPD SUM is very close to the NULL stream except for many structures, and PDB/PDA coherence is very low except these structures. (Also, in the middle pane of the second attachment, the black trace is the dark noise with high Z and 1 stage whitening, which is a factor of 5 smaller than 7.8W noise.)
Therefore we're confident that the noise floor for kHz range is already dominated by the shot noise.
Overall scaling is another problem which should be fixed separately.
Sheila, Dan, Elli, Kiwamu, Lisa, Peter, Robert
Here is a calibrated DARM spectrum from today's DC-readout lock, using the new CAL-CS channel described in LHO#16698.
According to GWINC, the BNS inspiral range of this spectrum is 8 Mpc.
The calibration must be wrong, ignore this for now.