Displaying reports 70441-70460 of 84471.Go to page Start 3519 3520 3521 3522 3523 3524 3525 3526 3527 End
Reports until 10:28, Monday 13 October 2014
H1 PSL
andres.ramirez@LIGO.ORG - posted 10:28, Monday 13 October 2014 (14424)
PSL Check: 10/13/2014
Laser Status: 
SysStat: Warning “VP program online” is red
Output power is 33.5 W (should be around 30 W)
FRONTEND WATCH is active
HPO WATCH is red

PMC:
It has been locked 5 days, 22 hr , and 54 minutes (should be days/weeks)
Reflected power is 2.0 Watts and PowerSum = 25.7 Watts.
(Reflected Power should be <= 10% of PowerSum)

FSS:
It has been locked for 1 d, 7 h, and 9 min (should be days/weeks)
Threshold on transmitted photo-detector PD = 2.26 V (should be at least 0.9V)

ISS:
The diffracted power is around 9.5% (should be 8-10%) 
Last saturation event was 2 d, 19 h and 0 minutes ago (should be days/weeks


NOTE: The ISS Diffracted Power was out of its range (8-10%) this morning. It was fixed by adjusting the ISS Reference Signal value.



H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 10:20, Monday 13 October 2014 (14423)
CDS model and DAQ restart report, Sunday 12th October 2014

model restarts logged for Sun 12/Oct/2014
2014_10_12 08:44 h1fw0
2014_10_12 18:27 h1fw1
2014_10_12 21:41 h1fw0

all restarts are unexpected

H1 SEI (DetChar, PEM)
krishna.venkateswara@LIGO.ORG - posted 10:07, Monday 13 October 2014 - last comment - 12:09, Monday 13 October 2014(14422)
Wind induced tilts

K. Venkateswara

As described by Sheila in 14416, high wind speeds were recorded at LHO on Saturday. BRS shows corresponding high ground tilts during this period.

The first plot shows the ASD of the GND STS, BRS RY and Stage 1 T240 X. There is excellent coherence between the GND seismometer and BRS RY below 60 mHz and the tilt-subtraction looks quite good - you can clearly make out the primary microseismic peak at 55-60 mHz. Also shown is the BRS reference mirror (~sensor noise). After Fridays calibration corrections, the fudge factor I had to use was smaller (0.79) to get the subtraction right.

Stage 1 shows slighlty better coherence with GND_STS and BRS, but it is still limited by some other source of noise below 0.1 Hz.

The second file shows a band-limited BRS RY (5-500 mHz) and wind speeds as measured by PEM anemometers and converted to X and Y directions using the 'direction' data. I might have the axes converted incorrectly, but visually it shows good correlation with the BRS data.

Non-image files attached to this report
Comments related to this report
krishna.venkateswara@LIGO.ORG - 12:09, Monday 13 October 2014 (14426)

In looking for source of low-frequency noise in Stage 1, I found something interesting: The first file shows the X,Y,Z and RY motion measured by the T240s on Stage 1. The next file shows the coherence between X, Y and Z sensors and RY.

There is large coherence between Y and RY below ~50 mHz which wasn't too surprising to me. Small misalignments of sensor axes can produce that. However, there is large coherence between Z and RY at the microseism which is odd. To get coherence with Z, one needs sensor calibration mismatch or actuator alignment/strength mismatch. It is also surpsisingly large.

Jim said he'd help me look into this further. But it does explain one thing Jeff had noted in 14047 about why RY looks very similar to X but is not coherent at the microseism. Because RY is measuring Z at the  microseism!

Non-image files attached to this comment
H1 General
andres.ramirez@LIGO.ORG - posted 09:06, Monday 13 October 2014 (14421)
Installation Meeting Minutes
Attendees: Bubba, Ed, Doug, Vern, Jodi, Keita, Jeff B, Jim, Hugh, Richard, Michael, Travis, John, Aaron, etc …

Today's Tasks: 


•	SRM and Lab work – Richard/Aaron, Ed, etc
•	Hugh working on Guardian (Reboots). 
•	3IFO Quad Upper Structure work continues in LVEA – Travis/Betsy
•	Beam tube cleaning will continue today – Bubba and crew 
•	TMS Optical Table assembly (3IFO) in staging building – Jeff B
H1 PEM (PEM)
dale.ingram@LIGO.ORG - posted 08:57, Monday 13 October 2014 (14420)
1 -3 Hz seismic noise last week
The attached plot uses GMT on the horizontal axis.  The plot shows 1 - 3 Hz seismic noise last week.  Hanford contractors were hauling typical numbers of rubble loads on day shift and swing shift from Monday 10/6 through Thursday 10/9.  ERDF, the rubble pit, operated on Friday 10/10 but the number of trucks running to ERDF was low during the day on Friday and nearly absent on swing.  Commissioners can expect that the Monday-Thursday noise on this plot will continue through the fall.  Fridays will vary from week to week.  We're trying to learn the Friday haul estimates a day or two in advance.   

The big feature on the plot that shows during the weekend appears to be earthquake-related.
Images attached to this report
H1 AOS
andres.ramirez@LIGO.ORG - posted 08:24, Monday 13 October 2014 (14419)
ISS Difracted Power High this morning
The ISS Diffracted Power was over 12% this morning. It is been adjusted to 9.5 % 
H1 AOS
andres.ramirez@LIGO.ORG - posted 08:24, Monday 13 October 2014 (14418)
ISS Difracted Power High this morning
The ISS Diffracted Power was over 12% this morning. Is been adjusted to 9.5 % 
H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 08:41, Sunday 12 October 2014 (14417)
CDS model and DAQ restart report, Friday and Saturday 10th,11th October 2014

model restarts logged for Fri 10/Oct/2014
2014_10_10 00:22 h1fw1
2014_10_10 07:43 h1fw1

model restarts logged for Sat 11/Oct/2014
2014_10_11 05:33 h1fw1
2014_10_11 21:41 h1fw0

all restarts unexpected.

H1 SEI (DetChar, SEI, SUS)
sheila.dwyer@LIGO.ORG - posted 16:40, Saturday 11 October 2014 - last comment - 11:18, Friday 17 October 2014(14416)
high wind, high microseism

I'm leaving the interferometer with LSC feedback disabled, the the green lasers trying to lock to the arms.  We are currently having both high winds and high microseism, so anyone who is interested in looking at the motion of our optics or performance of ISIs when the ground motion is high could use this time, starting at 23:39 UTC October 11.

I've left the intent bit as undisturbed, could the first person to come in and commision or work in the LVEA toggle it to commisioning?

Comments related to this report
jim.warner@LIGO.ORG - 11:18, Friday 17 October 2014 (14503)

Tuesday, October the 14th, was also a bad wind day, starting about noon. See Sheila's log 14475.

H1 ISC
kiwamu.izumi@LIGO.ORG - posted 22:47, Friday 10 October 2014 (14415)
locking tonight

Alexa, Kiwamu,

We tried the DRMI tonight with the arm cavities held at a off resonance by ALSs. No exciting result yet.  

(Too high speed in DRMI)

We could lock the DRMI only twice and they did not last for more than a couple of minutes. According to the band-limited RMS plots in the control room, we are having a monontonically increasing micro-seismic tonight, starting 10 hours ago. (Presumably) due to the high seismic, the DRMI fringe tonight is much faster than usual.

(ESD-UIM crossover issue)

Also we experienced an issue in which the ESD-UIM had a bad cross-over. When we closed ALS DIFF, it stayed locked fine for the first 10 minutes or so. But we noticed that the DARM feedabck was slowly developing a 1-ish Hz oscillation which gradually started using up the ESD range. This eventaully caused saturation in either or both of the ESDs and we lost the lock. In order to avoid this issuem. we tried decreasing the ESD LOCK_L gain from 1 to 0.8 on both ETMX and ETMY. This reduced the amplitude of the oscillation and also the in-loop spectrum got back to how it had been. We are concluding that the cross-over between UIM and ESD changed from yesterday.

After adjusting the ESD gain, ALS DIFF reliably stayed locked for a long time, typically order of 20 minutes. Each lock ends usually due to misalignment in ETMX or ETMY.

(ETMX ESD check)

This is just an observation of ETMX ESD.

In the process of checking the ESDs and etc, we checked the reponse of the ESD in the pitch direction by shaking it at 4.1 Hz and looking at it with oplev. We noticed that shaking it in pitch mostly gave us yaw. Shaking yaw gave a good yaw motion though.  We checked ETMY and it was not the case.

H1 ISC
alexan.staley@LIGO.ORG - posted 18:35, Friday 10 October 2014 - last comment - 20:15, Friday 10 October 2014(14412)
Slow feedback to ETMs

(Sheila, Alexa)

Today we worked on the slow/tidal feedback to the ETMs. In doing so, we saw that the L2P for the M0 state was not very good at low frequency. Using the same method as was done for the BS (alog  14022), we found that the gain of ETMX L2P (with FM1, FM2 ON), should be changed from -1.0 to -4.0 at 0.01Hz. So far it appears that this is OK for high frequency. Meanwhile, we found that the gain of ETMY L2P (with FM1,FM2 ON) should be changed from -1 to 22 at 0.01Hz. We are not sure if this is okay at higher frequencies.

The slow/tidal feedback for both arms now works, and no longer causes a drift in the alignment. We made the following the changes to the feedback servo (this is all taken care of in the guardian already):

It appears that the DIFF PLL VCO is now in range, but maybe that's a coincidence with ground motion today....

Comments related to this report
daniel.sigg@LIGO.ORG - 19:46, Friday 10 October 2014 (14413)

One can probably bleed off the dc term (f<10mHz) by feeding back to HEPI. The path is all set up and the LSC signals are received by a HEPI ISC input filter. All that should be required is an integrator in this path.

alexan.staley@LIGO.ORG - 20:15, Friday 10 October 2014 (14414)

We were having some problems with ALS DIFF earlier tonight, so we reverted the L2P gains back to -1. It is unclear at the moment if this was related at all.

H1 AOS
edmond.merilh@LIGO.ORG - posted 16:02, Friday 10 October 2014 (14397)
Ops Daily Summary

~Two more 5.2mag aftershocks this morning on the Southern East Pacific Rise near Easter Island~ 

08:00 Morning checklist: Reset ALH; vacuum system looks ok - HAM6 cold cathode pressure reporting 10-7;  CDS/DAQ - everything looks normal.

08:30 Morning meeting

08:42 Krishna to X-End to do some BRS testing/tweaking

09:20 Krishna back from X-End

09:50 Keita to End Y to align green WFS. He will be using ETMY/ITMY/BS/PR3

10:14 Dale into the control room with a tour group

12:04 Jason to End X to align OpLev

12:30 Adjusted ISS Diffracted power to ~9.5% by reducing the refsignal voltage from -2.01 to -2.05

12:37 Jason back from End X

14:45 Sudarshan into LVEA, crawling under HAM2

12:48 Keita back into control room from End Y

12:50 Sudarshan back from crawling

H1 SEI (CDS, DetChar)
krishna.venkateswara@LIGO.ORG - posted 15:47, Friday 10 October 2014 - last comment - 16:32, Friday 10 October 2014(14408)
H1 ETMX Ground sensor correction model

J. Kissel, K. Venkatewara

Jeff K. designed a model to use the GND_Supersensor (tilt-corrected STS) to do sensor correction on Stage 1.

To get the supersensor, we take the GND_STS and add the Tilt signal from BRS converted to velocity units by multiplying g/w. To further correct for the small amount of acceleration present in BRS, we further add a factor of F * V_STS * g/w^2 where F = M * d  /  I  (M=Mass of beam-balance, d = Center of mass offset, I = Moment of Inertia)

Jim and I will try to implement and test this next week when not interefering with commissioning activity.

Images attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 16:32, Friday 10 October 2014 (14411)PEM
J. Kissel

A reminder to add the damping control signal monitor (currently fed into ADC0 CH 30, see D1400077, as of LHO aLOG 14351) to the h1isietmx model's channel list. In fact, it's already included and sent into a filter bank called "ETMX_SPARE" in the GND_BRS block, so let's just change it from a filter bank to an epics output (like the drift monitor) and call it "ETMX_DAMPCTRLMON." Once added to the front end, we should add it to the MEDM screen.
H1 SEI (CDS, DetChar)
krishna.venkateswara@LIGO.ORG - posted 15:45, Friday 10 October 2014 - last comment - 16:19, Friday 10 October 2014(14409)
H1 BRS calibration changed

J. Kissel, K. Venkateswara

We realized we had the wrong factor of 1/2 as described in alog entry 13448.  1/2 [V_sing / V_diff] was not needed/

The filters have been changed to account for the new calibration factor which is:

3.5e-6 [rad/pixel] * 1 [pixels/V_sing] * 1/2 [V_sing / V_diff] * 40/2^16 [V_diff/ct] * 1e9 [nrad/rad] =  2.1362 [nrad/ct]

 

Comments related to this report
jeffrey.kissel@LIGO.ORG - 16:19, Friday 10 October 2014 (14410)PEM
J. Kissel, K. Venkateswara,

Krishna copied, pasted, and posted too fast and still included my erroneous single-ended to differential conversion in the piece-by-piece equation, even though the answer quoted is correct. 
The calculation to achieve the correct calibration is
3.5e-6 [rad/pixel] * 1 [pixels/V_sing] * 40/2^16 [V_diff/ct] * 1e9 [nrad/rad] =  2.1362 [nrad/ct]

Also, I've committed the updated filter file to the userapps repo.

Also, also, we need to
(a) Add the reference mirror filter bank and damping motor control signal* to the MEDM screen
(b) Capture a new safe.snap with the reference mirror calibration filter turned ON as it is now.

*The damping motor control signal isn't yet present in the user model, it's only available as a raw IOP test point. We should add this to the h1isietmx user model when we're making the front-end code modifications to create the tilt-free ground super sensor. I'll post this as a comment to the other suggested model changes so we don't forget.
LHO General
edmond.merilh@LIGO.ORG - posted 15:03, Friday 10 October 2014 (14407)
Adjustment of ISS Diffracted Power

At ~12:34PT I adjusted the diffracted power in the ISS to ~9.5% as instructed by Rick Savage. The refsignal is currently set to -2.05V.

H1 AOS (DetChar)
sheila.dwyer@LIGO.ORG - posted 10:25, Thursday 09 October 2014 - last comment - 09:07, Friday 14 November 2014(14381)
etmx oplev glitches

Jason, Sheila

There have been some glitches on the ETMX oplev, which don't seem to correspond to motion of the optic, the attached screen shot shows an example.  I think that I have seen glitches like this that were much larger in amplitude, but am not sure. 

One question for detchar is if they have a tool that can search for glitches like these, and give us some information about how often they are happening, how large they are, and maybe monitor to see if they are happening on other op levs

Another issue is that sometimes the lasers fail, which is often foretold by a several seconds oscillation in the oplev sum.  Can detchar set up some kind of monitoring for that?

Of course the real question is how can we fix it......

Images attached to this report
Comments related to this report
shivaraj.kandhasamy@LIGO.ORG - 14:40, Friday 10 October 2014 (14406)DetChar

Hi Jason and Sheila, here at LLO  Olmo Cerri, a summer student from UMISS, looked at the possible causes of glitches in OPLEVs. He worked with Suresh. Suresh earlier found that the the temperature variation could cause changes in the cavities of OPLEV laser and thereby changing the laser intensity which would look like glitches. Olmo looked at week long strecthes of data from a few OPELVs and charactersized how many glicthes we see (he wrote a matlab script to find the such gltiches).  They also implemetned a simple temperature stabilization method (LLO a-log) to reduce the gltiches. It seemed to work well. Here is a presentation by  Olmo on this work which shows data before and after the temperature stabilization setup. He is now looking into implemneting an online version of the code.

cody.arceneaux@LIGO.ORG - 09:07, Friday 14 November 2014 (15053)
I set Olmo's code to run on a cron job with the results here:

https://ldas-jobs.ligo-wa.caltech.edu/~codyca/MHoF_results/

We still need to put together a better way to present the results.  The code is set to remove glitches caused by human interaction by looking at the OPTICALIGN offsets.  Also, the results for LLO can be found here:

https://ldas-jobs.ligo-la.caltech.edu/~codyca/MHoF_results/
Displaying reports 70441-70460 of 84471.Go to page Start 3519 3520 3521 3522 3523 3524 3525 3526 3527 End