Displaying reports 72341-72360 of 83026.Go to page Start 3614 3615 3616 3617 3618 3619 3620 3621 3622 End
Reports until 14:54, Wednesday 26 March 2014
H1 ISC
sheila.dwyer@LIGO.ORG - posted 14:54, Wednesday 26 March 2014 (11016)
RF preamplifier installed in ISC rack 1

I installed the spare H1 RF preamplifier (S1203936) in ISC R1, and routed the COMM BBPD signal through it. 

H1 ISC
stefan.ballmer@LIGO.ORG - posted 13:35, Wednesday 26 March 2014 - last comment - 14:17, Wednesday 26 March 2014(11013)
Spare Acromag BO chassis reconfigured
I reconfigured a spare Acromag binary output chassis for replacing the broken unit.

The new unit is:
D1100251, S1203287

IP addresses:
10.40.12.21 (top left, as seen from the front)
10.40.12.22 (top right, as seen from the front)
10.40.12.23 (bottom left, as seen from the front)
10.40.12.24 (bottom right, as seen from the front)

Attached are print outs of the settings of the new box (S1203287) and the old box (S1203246).
Non-image files attached to this report
Comments related to this report
stefan.ballmer@LIGO.ORG - 14:14, Wednesday 26 March 2014 (11014)
Turns out the ip address stickers on the old box were wrong. The right addresses for the new box (S1203287) are

IP addresses:
10.40.12.11 (top left, as seen from the front)
10.40.12.12 (top right, as seen from the front)
10.40.12.13 (bottom left, as seen from the front)
10.40.12.14 (bottom right, as seen from the front)

Note: this also means the attached file called OldBox.pdf is actually from a different unit...
stefan.ballmer@LIGO.ORG - 14:17, Wednesday 26 March 2014 (11015)
This seems to have fixed the REFLAIR_A_RF45_I dewhitening switching. Let's hope it didn't introduce a new bog in one of the other 383 channels...
H1 TCS (TCS)
greg.grabeel@LIGO.ORG - posted 09:46, Wednesday 26 March 2014 - last comment - 16:32, Wednesday 26 March 2014(11004)
ITMy lower ring heater installed
The install of the lower ITMy ring heater was completed Tuesday evening. There was a small issue with fit for SN#203 glass former assembly, the nichrome was a tight fit into the macor (not nearly as bad as SN#210). The diameter of the glass and nichrome near the end was 0.276", some chipping of the macor was seen after the glass was seated. The temperature sensor (RTD) has loosened some, apparently from the re-install efforts, about the epoxy but is still showing the same resistance as before (109.5 ohm). 
Images attached to this report
Comments related to this report
greg.grabeel@LIGO.ORG - 16:32, Wednesday 26 March 2014 (11017)TCS
A few more pictures. The first one shows a small amount of chipping on the macor that happened during assembly. This is significantly less than last time however. The second picture shows a casualty of uninstall, re-bake, and reinstall. The RTD has started to come loose.
Images attached to this comment
Logbook Admin Bug
corey.gray@LIGO.ORG - posted 08:15, Wednesday 26 March 2014 - last comment - 09:43, Thursday 27 March 2014(11008)
Problem Uploading Photos

Not sure what the deal is with my photos, but for some reason alog chooses to not let some of my photos upload to an entry.  Have used original names of images, and changed names of files.  Have tried from home and from work to no avail.  Will keep trying.  Attached is the error message I get afer I try uploading the file I select.

Entries made via MacBook Pro & with Firefox.

Images attached to this report
Comments related to this report
jonathan.hanks@LIGO.ORG - 09:43, Thursday 27 March 2014 (11031)
The issue was fixed by adjusting a parameter on the server.
H1 AOS (CDS, DAQ)
david.barker@LIGO.ORG - posted 08:06, Wednesday 26 March 2014 (11007)
CDS model and DAQ restart report, Tuesday 25th March 2014

model restarts logged for Tue 25/Mar/2014
2014_03_25 11:11 h1broadcast0
2014_03_25 12:07 h1dc0
2014_03_25 12:09 h1broadcast0
2014_03_25 12:09 h1fw0
2014_03_25 12:09 h1fw1
2014_03_25 12:09 h1nds0
2014_03_25 12:09 h1nds1

no unexpected restarts reported.

H1 SUS (SEI, SYS)
jeffrey.kissel@LIGO.ORG - posted 23:23, Tuesday 25 March 2014 (11003)
WD Tests in Support of ECR E1400102 / II 720
J. Kissel

As per request of the review committee responsible for Engineering Change Request E1400102, or Integration Issue 720, I've used the IFO down time to shake suspensions with quite vigor, on resonance, to prove the OSEM AC band-limited RMS sensor watchdog (the only portion of the SUS user that will remain after the ECR is complete) will sufficiently protect the suspension from danger.

It will. Removing all *other* components of the SUS User Watchdog, as requested, should be fine.

I attach 4 sets of time series of the excitations for three different suspension types (an HSTS [PR2], an HLTS [PR3], and a QUAD [ETMX]), driving in Longitudinal or Vertical (those degrees of freedom with the largest number of actuators), from the top mass (which have the strongest actuators). 
Upper Left -- the requested drive (the TEST_OUT in red) and the the sent drive (the MASTER_OUT, which is after the watchdog). 
Bottom Left -- the response to the drive, in [um] as measured by the OSEMs. (Top = Brown, Middle = Magenta, Bottom = Cyan)
Upper Right -- the watchdog state during the time series. (Top = Green, Yellow = Middle, Black = Bottom, Pink = USER DACKILL)

For green (top mass WD), a FIRSTTRIG bit value of 2 indicates that the Sensor OSEM AC WD caused the watchdog to trip. For all* studies, top mass OSEM AC watchdog trips, stopping the excitation as expected. (*For the 1st, 5th and 6th page, with PR2 drive in L, the FIRSTTRIG doesn't report a trip, but the excitation is turned off by guardian [nice!], because the lower stage watchdogs have tripped -- on the AC Band-Limited Sensor Signal.)

Also under debate is to remove the lower stage sensor band-limited OSEM AC triggering, but I've convinced myself through the few PR2 and ETMX trips that they're still worth it. HOWEVER, we should most certainly simplify the watchdog MEDM screen system to have one screen with a "RESET ALL" button, like the SEI subsystem has.

I did find an undesirable mode of the Guardian though -- if the SUS trips, and resonances are rung up such that the WD continues to trip, one cannot untrip the watchdog long enough for Guardian to get damping loops up and running again to cool off the system. This means you have to wait for the system to cool down itself or temporarily raise the WD threshold, neither of which are desirable for these high-Q suspensions.

Details
-------
Measurement templates live in
/ligo/svncommon/SusSVN/sus/trunk/Common/Data/20140325_WDTests/

I used AWGGUI to create the drive -- a 10000 [ct] (or 100000 [ct] where needed) sine wave, driven at the first resonant frequency for the SUS type, as determined by T1200404.
Non-image files attached to this report
H1 SUS
keita.kawabe@LIGO.ORG - posted 20:28, Tuesday 25 March 2014 - last comment - 08:41, Wednesday 26 March 2014(11002)
Undoing cable routing change for TMS (Corey, Jax, Keita)

When we went into the chamber we have found that the top cable rounting was done in such a way that

  1. Cables were loose and rubbing with each other,
  2. Cables were rubbing with the cable loops inside the top mass,
  3. One cable was touching the inside wall of the top mass hole from which the cables come out.

These are all bad, and I spent some time last week to divide four cables into two groups to make sure that none of the above was happening, but the cables were re-routed yesterday evening and they were bundled together without paying attention to the above.

Two of the four cables that go through the TMS top mass are for picomotors and are much thicker than BOSEM cables (the core diameter is bigger). These are stiff cables, and it's very difficult to route them such that they don't touch anything, and this is especially true for the pilot unit (TMSY) where the cable clamping inside the top mass is totally different from production units. How this is done has a big effect.

Attached are free swing spectra of three different cable routing. It's convenient to remember that there is a three-finger structure between 0.67 and 0.82 Hz.

Freeswing, this morning, ie cable rubbing (green).

Freeswing, after we made it the way we want it, no rubbing (blue).

We unbundled the cables, divided into three different routes (one route has two cables) and made sure that nothing is touching anything. This should be roughly the state of things from yesterday morning, except that the cables are divided into three instead of two (somehow it was very tough to do it right in just two routes today).

I believe that this is roughly the same state we were in during OAT days (https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=2587)

Freeswing, after we bundled cables together and used zip ties so that the cables at least don't rub against each other (red).

This is the state we're in now, as I wanted to see if zip ties help after we tested the blue configuration.

So, all in all, I think we need to go back to blue state tomorrow. Note that free swinging spectra are not transfer functions. This is a good tool to detect rubbing/interference.

Images attached to this report
Comments related to this report
corey.gray@LIGO.ORG - 08:41, Wednesday 26 March 2014 (11006)

(Corey, Jax, Keita)

Additional notes to Keita's entry for Tues work.

Alignment Check In Morning

Started off the morning aligning the TMS with Jax after Arnaud/Travis' cable change on Mon night.  Saw a pitch change and needed to pitch the TMS "up" (which required me to move the "screw balance weight" from one end of the Table to the other, AND adding THREE more (see attached photo), in hindsight, I could have slid masses under the table).

Photos of Cable Set-ups

GREEN:  All four together from clamp to clamp

BLUE:  Two thick cables together & each thin cable separate

[alog doesn't like this set-up and doesn't let me upload a photo....yet] 
Since I can't upload any photos to the alog, here are links to BLUE photos in resourcespace:  overall, clamping at bottom

RED:  All four cables zip-tied together

We removed some Aluminum banding on a pair of cables running from the TMS table up to the Upper Mass.  This is where we currently are set up, but will move back to the blue set up.

[alog doesn't like this set-up and doesn't let me upload a photo....yet]

Here is an overall photo of cable.

Another Alignment Note:

During one of the Alignment checks, we needed to pitch down, so one of the screws which was added in the morning was removed (the Ag-plated one on the edge)...didn't take photo.

Images attached to this comment
H1 ISC
stefan.ballmer@LIGO.ORG - posted 19:56, Tuesday 25 March 2014 (11001)
PRMI locking
Dave, Stefan

Trying to lock PRMI, I first found a sign error in the PRCL loom gain - somehow this wasn't properly updated last Tuesday.

Next though I had trouble keeping PRX locked. we suspect the cause is an unusually high anthropogenic noise (1-3Hz) tonight - tonight is worse than a typical day noise.
H1 ISC
alexan.staley@LIGO.ORG - posted 17:14, Tuesday 25 March 2014 (10998)
Electronics Noise and ADC Noise Measurements

(Alexa, Sheila)

 

With a terminator at Input 1 of the LSC-REFL Servo CM board, we measured the amplitude spectrum (Vrms/sqrtHz):

With a terminator at the SlowMon of the DAQ channels, I measured the ADC Noise of LSC_REFL_SERVO_SLOW_OUT (/ligo/home/alexan.staley/Public/ADC_ElectronicsNoise_03252014.xml ref 4)

We also measured the amplitude spectrum from Output 2 of the CM board with the COMM PLL locked, and input 1 of the CM board at 0dB (CommPLL_NoiseSpecV_Out2.txt).

Non-image files attached to this report
H1 CDS
cyrus.reed@LIGO.ORG - posted 17:06, Tuesday 25 March 2014 (11000)
CDS SSID Cleanup

I have removed the disused 'CDS aLIGO' SSID from the CDS wireless access points now that I have transitioned all of the CDS workstations to the new SSID/subnet.  I have also cleaned up the corresponding WAP switchport configs.

H1 TCS (TCS)
aidan.brooks@LIGO.ORG - posted 17:04, Tuesday 25 March 2014 (10999)
TCS CO2X install - boxes received. Cabling continues

Aidan. Dave H. Thomas.

We continued cabling TCS CO2X on-table cables. The RF driver had a Conec connector installed on it. We finished receving and sorting all the equipment from yesterday's shipments.

Full status is in the attached PDF.

Non-image files attached to this report
H1 AOS (AOS)
david.ottaway@LIGO.ORG - posted 16:24, Tuesday 25 March 2014 - last comment - 15:18, Thursday 27 March 2014(10996)
H1 Output Faraday Isolator Performance
Dave O, Gerado,

The performance of the Output Faraday Isolator was tweaked and a transmission of 96.5 +/- 0.5% was recorded and an isolation of 9e-5 was measured. See attached report for more details.
Non-image files attached to this report
Comments related to this report
david.ottaway@LIGO.ORG - 15:18, Thursday 27 March 2014 (11038)
We also used the Shack Hartmann Sensor to put and limit on the spherical power introduced by the Output Faraday Isolator (See attached document) 
Non-image files attached to this comment
H1 SUS
jeffrey.bartlett@LIGO.ORG - posted 16:09, Tuesday 25 March 2014 (10995)
Phase 2b TFs & PS for H1-SR3
   After resolving rubbing and magnet attachment issues on H1-SR3, we took phase 2b transfer functions and power spectra. There are small unmodeled peaks in yaw at 1.57Hz, in pitch at 1.06Hz, and in transvers at 1.06Hz, which are suppressed with damping. There is some of the environmental noise we have seen in the other suspensions tested on the same optics table in the LVEA. Overall H1-SR3 compares favorably other LOH and LLO HLTS suspensions.
 
   I have attached the plots below. All scripts and data files have been committed to the SVN repository.  
Non-image files attached to this report
H1 General
andres.ramirez@LIGO.ORG - posted 16:00, Tuesday 25 March 2014 (10994)
Ops Shift Summary
7:50-9:14 Heading to End Y for cleaning - Karen
8:00 Crystal chiller refilled (Maintenance) - Justin
8:45-10:30 Heading to EndX/EndY to scan electronics – Filiberto
8:50-12:05 Working in LVEA – David H.
8:58 End Y transitioned to Laser Hazard- Jax
9:00-10:18 Searching for TCS parts in LVEA – Thomas
     LN2 delivery to CP1 – Praxiar
9:10-13:01 Going to End Y for TMS alignment – Corey/Jax
9:15-12:00 Chassis work on ISC racks in LVEA – Aaron
9:40-10:06 Heading to End X to check leaking valve – Hugh
9:42 Working at Mid Y – Jodi
10:00 Replacing RAM modules in h1broadcast0 - James
10:14 Rebooting H1 boot server – James
10:35-12:00 Cable work by HAM4 (LVEA) – Filiberto
10:45 Removing iLIGO RGA from BSC10 – Kyle
11:05 Heading into the LVEA – Betsy
11:42 Working on HAM5 – Hugh/Jim
12:07 DAC restart – Dave
12:44 Back to HAM4 – Filiberto
13:13 Monolithic welding of ITMx and ITMy at LVEA - Travis
13:15 Moving Laser barrier from HAM2 to BSC3 area – Justin
13:48 Heading into H2 PSL enclosure (Laser Hazar) – Gerardo
13:55 Going to End X - Cris
LHO VE
john.worden@LIGO.ORG - posted 15:44, Tuesday 25 March 2014 - last comment - 09:26, Wednesday 26 March 2014(10992)
Y1 Beam Tube Accumulation

My estimate of N2 leak/outgassing for the Y1 module at LHO. The accumulation took place over 4.8 days.

Total pressure at the end according to the CC gauges was 5e-8 torr and the total RGA ion current was 2e-10 amps yielding 0.004 amps/torr(H2).

Ignoring the gauge factors I get an N2 leak/outgassing rate of 4e-9 tl/sec. There may also be a contribution from CO generated by the CC gauges.

Also, the RGA setup was coupled to the tube at an oring angle valve so

there are a number of small orings in close proximity to the RGA. Kyle did perform a mild bake on this setup. In addition, included in the accumulation

is the LN2 pump at YMID.

I would say we have no measureable air leak on Y1.

The RGA data is shown on a log plot for better visibilty - on a linear plot the curves appear straight.

Images attached to this report
Non-image files attached to this report
Comments related to this report
john.worden@LIGO.ORG - 09:26, Wednesday 26 March 2014 (11011)

For Rai and Mike,

Here is the CC gauge data in ascii. Two gauges PT124(first data set in file) and PT243 are recorded.

The times are as follows:

RGA start record at 3/19/2014 13:25 PDT = 20:25 UTC =  2456736.3704

RGA end record at 3/24/2014 9:15 PDT = 16:15 UTC  =  2456741.1958

The temperature data is here:

https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=10984

 

Average temperature was ~50F during the accumulation.

Non-image files attached to this comment
H1 ISC
stefan.ballmer@LIGO.ORG - posted 14:03, Monday 24 March 2014 - last comment - 15:49, Tuesday 25 March 2014(10961)
REFLAIR_A_RF45_I whitening stage 1 diagnostic
Jax, Stefan

We verified that the whitening stage 1 of REFLAIR_A_RF45_I is indeed not switching on the analog side. We checked that the command bit (pin 25 / GND=pin 20) makes it to the board. Thus Jax will pull the board tomorrow and debug it..

Details:
Rack: ISC R2 slot 32
S/N: S1101590
Assembly: D1002559 (board2)
Comments related to this report
stefan.ballmer@LIGO.ORG - 15:49, Tuesday 25 March 2014 (10993)
Turns out we confused I and Q: in the electronics the Q channel comes before the I channel.

Bottom line: the whitening board works fine. The Beckoff system does not switch line 33 on cable H1:ISC 56. I left a scope in the electronics room hooked up to that line.

 
H1 SUS
arnaud.pele@LIGO.ORG - posted 16:48, Monday 15 April 2013 - last comment - 16:24, Tuesday 25 March 2014(6014)
B&K PRM PR2 PR3 MC1 MC3

B&K hammer test was made monday and tuesday last week with Betsy and Travis on HAM2 and HAM3 chamber for MC1-MC3-PR3-PRM-PR2 suspensions

The measurements have been taken with :

Pictures attached are showing position of the tri-axis accelerometer, and hammer hit locations for : 

The following pdf document shows the different measurements, with comparisons between VA on / off for PR3 and PRM, and comparisons between LLO and LHO measurements.

As required for the ISI tables, resonnances of the structure below 150Hz (marked by the black dotted line) are damped by the vibration absorbers (page 9 to 12 to see their effects on PRM and PR3)

Data have been exported and plotted following the operations manual wikipage.

Images attached to this report
Non-image files attached to this report
Comments related to this report
arnaud.pele@LIGO.ORG - 16:24, Tuesday 25 March 2014 (10997)

Label on the first picture should be MC1 and not MC3.

Displaying reports 72341-72360 of 83026.Go to page Start 3614 3615 3616 3617 3618 3619 3620 3621 3622 End