Displaying reports 74861-74880 of 76953.Go to page Start 3740 3741 3742 3743 3744 3745 3746 3747 3748 End
Reports until 16:34, Monday 06 February 2012
LHO VE
kyle.ryan@LIGO.ORG - posted 16:34, Monday 06 February 2012 (2165)
New ESD feedthroughs flanges leak
New 5-channel 4.5" CF ESD feedthrough flanges leak! -> Douglas D., Justin B. and I tested (2) each of these, one of which was to be installed and BSC8 today prior to pumping down the YBM.  Helium responses were large and fast but we did not record the actual peak numbers as we were in a "pass/fail" mindset-these definitely FAIL!  We may repeat test tomorrow and record leak rates for the record.
H2 SUS
betsy.weaver@LIGO.ORG - posted 16:02, Monday 06 February 2012 (2164)
ETMy SUS alignment work
Travis, Jason and I spent most of the day down at EY struggling with the mechanics of ETMy as the chains are not well set in yaw and are proving a pain in the a$$ to align.  In fact, we had to pull the reaction chain away from the main (via EQ stops) just to work on the main without reaction pushing on main.  After ~5 iterations of Top Stage blade tip alignment, we finally ended the day with the main chain hanging at some 500urad out.  Jason reports that this is roughly where we started this morning...

We also fought a differential pitch between the ERM and the PenRe above it - due to the loop.  We attempted to move the loop along the barrel a bit to see if that would help.  It didn't appear to, so that's when we opted to clamp it off away from the main anyways.
LHO VE
cheryl.vorvick@LIGO.ORG - posted 14:29, Monday 06 February 2012 (2162)
BSC8 Closeout - Pictures of Particulate
Pictures are in ResourceSpace:

https://ligoimages.mit.edu/?c=961

Original picture names describe what area the picture is capturing.
H2 SUS
jeffrey.kissel@LIGO.ORG - posted 14:24, Monday 06 February 2012 (2161)
H2 SUS ITMY Compensation Filters Updated
I've updated H2 SUS ITMY's compensation filters such that they contain all compensation for all stage's electronics. The coil driver compensation filters (found in the COILOUTF banks) are arranged in filter modules according to T1100507. The SatAmp whitening compensation filters (found in the OSEMINF) remain in FM1. 

I've also removed any old, erroneous, and/or unused filters in these banks. Note that the "to_nm" filter in FM5 of the OSEMINF banks is a calibration filter for the OSEM sensors (just a gain of ~40 nm/ct), but is untested and should not yet be used.

Details
----------------

The new filter file, 

/opt/rtcds/lho/h2/chans/H2SUSITMY.txt

has been copied over to the userapps repo,

/opt/rtcds/lho/h2/userapps/release/sus/h2/filterfiles/


Because the updated coil driver compensation scheme involves different filter modules than the previous scheme, I've also updated the burt snapshot files,

/opt/rtcds/lho/h2/userapps/release/sus/h2/burtfiles/h2susitmy_damped.snap
/opt/rtcds/lho/h2/userapps/release/sus/h2/burtfiles/h2susitmy_safe.snap

capturing them from the autoBurt.req file found in 

/opt/rtcds/lho/h2/target/h2susitmy/h2susitmyepics/
LHO VE
cheryl.vorvick@LIGO.ORG - posted 14:15, Monday 06 February 2012 (2153)
BSC8 Witness Plates:
Witness plates, which are 4" diam. silicon wafers, were installed in BSC8 and nearby beam tubes.  The first two attached images show the installation as of Feb. 1, and as installed on Feb. 2.  Locations match, but the second images includes the detail, that may or may not effect the results, that a witness plate was broken and replaced in one location.

Location for witness plates were chosen to give information about migration of particulate from beam tubes to the BSC8 chamber.

Additionally, I chose locations to look at specific areas of interest.
1) both sides of the bellows in the Y arm beam tube (West) and the beam tube that is on the North side of the chamber - (bellows effect on particulate migration)
2) gate valves - in North beam tube and East beam tube - (gate valve effect on particulate)
3) directly under the gap between the ACB and ITMY - (to measure what particulate is falling in front of the optic)
4) directly in the center of the chamber - requested by Jodi for ICC - (to measure overall BSC8 particulate)


I've chosen to aLog witness plate info under LHO, VE, and I've included the specific chamber in the title, i.e.

Title -> "specific chamber"
Section -> LHO
Task -> VE

I've also attached two in-chamber photos of witness plates placed on Feb. 1 - North Beam Tube and East Beam Tube.
Images attached to this report
H2 TCS
cheryl.vorvick@LIGO.ORG - posted 14:06, Monday 06 February 2012 (2160)
BSC8 TCS Ring Heater Pictures
One image attached here, and the rest found in ResourceSpace.

https://ligoimages.mit.edu/?c=959
Images attached to this report
H2 SUS
cheryl.vorvick@LIGO.ORG - posted 14:02, Monday 06 February 2012 (2159)
BSC8 SUS Cabling Pictures:
One image attached here, and the rest found in Resource Space:

https://ligoimages.mit.edu/?c=958
Images attached to this report
Logbook Admin Bug
jonathan.berliner@LIGO.ORG - posted 11:40, Monday 06 February 2012 (2155)
aLIGO Wiki / MoinMoinWiki Bug: Don't Click Through Links on Preview Pages

After saving a new version of a wiki page, you see the new version of the page with a message at the top that says:

Thank you for your changes. Your attention to detail is appreciated.

It is important to click Clear message underneath this message. If not, and you click through a link on the Wiki, you can't backtrack to the wiki page, lest you arrive at a preview page which often leads to either errors or resubmit requests.

For my instance in the Pcal wiki, I finish editing the wiki and arrive at this page:

https://awiki.ligo-wa.caltech.edu/aLIGO/pcalWiki#preview

If I click through a link and come back, I get popups.

But if I go through the normal wiki page:

https://awiki.ligo-wa.caltech.edu/aLIGO/pcalWiki

...everything works as expected.

H2 CDS
james.batch@LIGO.ORG - posted 10:54, Monday 06 February 2012 (2154)
Mac OS X (iMac Workstations) nds2-client changed
The default nds2-client software for Mac OS X workstations for H2 has been changed to use nds2-client-0.8.2.  Users of Matlab should execute matlab from a terminal window rather than clicking on the matlab icon in order to get the environment set up properly.

Note that there are a few machines (end Y, psl enclosure) that have not been configured to run matlab from the command line yet.  This will be done on a normal Tuesday maintenance schedule.
H2 DAQ
david.barker@LIGO.ORG - posted 08:49, Monday 06 February 2012 (2152)
h2fw1 set to partial mode while LDAS rebuild the QFS file system

Following the test whereby h2fw0 and h2fw1 had their QFS file systems switched: the Sataboy QFS in the LDAS server room continued to have problems. We have decided to rebuild the /ldas-h2-frames file system. Since h2fw1 is currently writing to this system, I have put h2fw1 into partial mode. It is running, but not writing any frames at all. The /ldas-h2-frames file system has been unmounted from h2fw1.

I've mounted the /cds-h2-frames file system to both NDS machines during the rebuild and repopulation.

LHO General
patrick.thomas@LIGO.ORG - posted 17:20, Friday 03 February 2012 (2151)
plots of dust counts
Attached are plots of dust counts > .5 microns.
Non-image files attached to this report
H2 SEI
hugh.radkins@LIGO.ORG - posted 17:04, Friday 03 February 2012 - last comment - 11:11, Tuesday 07 February 2012(2150)
H2 ITMY (BSC8) SEI State post chamber close-up
The BSC8 HEPI was placed on the fixed stops during the Dome install as the crew can't help but stepping on the Crossbeams.  Once the Dome bolts had been torqued, I suspended HEPI again and read dial indicators for a final delta from last alignment time.

Of note, the before & after dial indicator readings captures the ACB swung back and in place and I saw a 2mil(.002") lower reading to the west while the ACB had a little more leverage on the system--may be within the noise of DI reading and robustness.

Anyway, post final alignment numbers on Jan 19:
Level--There is a 5mil tilt down on the northside.  Our Dial Indicators are essentially at the Support Tubes so I put this tilt in the small category: 0.005"÷60"=80µrads
Position: I see a <6mil shift to the South & <10mil shift West.
Rotation: The dial indicators imply a CW (from above) of no more than 60µrads.  This is worse case as the SE corner shifted South 9mils and the NW corner shift South 1mil.  Averages would reduce the rotation amount.

As this is a close out, I've attached 17 pages of my log book mostly showing dial indicator readings and elevation surveys while we adjust to IAS requests and install and center HEPI Actuators.  I know it is a bit messy but it is a logbook.  If you need clarification just ask.  A couple notes:
DI readings are in .001"(mils).
Corner 1 is NE, 2 is NW, 3 is SW & 4 is SE---This is different from LLO established pier numbering.  I'll try to conform to this in the future.
V T & L corresponds to vertical, Transverse(N-S), & Longitudinal(E-W)
Non-image files attached to this report
Comments related to this report
hugh.radkins@LIGO.ORG - 11:11, Tuesday 07 February 2012 (2170)
Due to the stellar performance of an assembly tech, the final numbers I report here are actually with the Vertical Actuators locked down.  See a subsequent report on Tuesday 7 Feb that reports final floating position.
H2 SEI
hugh.radkins@LIGO.ORG - posted 15:52, Friday 03 February 2012 (2149)
ETMY ISI has Keel Ballast Mass (on Test Stand)
Jim & Corey (I helped a little) put the 600lbs on top of the H2 ETMY ISI Keel.  We passed the Keel Mass components (50lbs each) from the floor to the man on a four foot ladder.  In turn the mass was then passed to the man (Jim) standing on the Test Stand.  The mass was placed as per D1000897-v2.  This should put us very close to balanced when the mass is replaced after Cartridge install.
H2 SUS
jeffrey.garcia@LIGO.ORG - posted 15:11, Friday 03 February 2012 (2147)
ETMY R0, L1, L2 Open Light counts
ETMY R0, L1, L2 Open Light counts, offsets, and normalising gain values.  Offsets are calculated by dividing the open light counts by two and multiplied by (-1).  The gain values are calculated such that the open light counts are normalised to 30,000cts.
-----------------------------------------------------------
R0 
controls@cdsws1:~ 0$ tdsavg 30 H2:SUS-ETMY_R0_OSEMINF_F1_IN1 H2:SUS-ETMY_R0_OSEMINF_F2_IN1 H2:SUS-ETMY_R0_OSEMINF_F3_IN1 H2:SUS-ETMY_R0_OSEMINF_LF_IN1 H2:SUS-ETMY_R0_OSEMINF_RT_IN1 H2:SUS-ETMY_R0_OSEMINF_SD_IN1
29408.9
26594.2
28718
28591
27628.8
29841.9

OSEMINF OFFSETS:
F1 = -14705
F2 = -13297
F3 = -14359
LF = -14296
RT = -13815
SD = -14921

OSEMINF GAINS:
F1 = 1.020
F2 = 1.128
F3 = 1.045
LF = 1.049
RT = 1.086
SD = 1.005
-------------------------------

L1
tdsavg 30 H2:SUS-ETMY_L1_OSEMINF_UL_IN1 H2:SUS-ETMY_L1_OSEMINF_UR_IN1 H2:SUS-ETMY_L1_OSEMINF_LL_IN1 H2:SUS-ETMY_L1_OSEMINF_LR_IN1
30801.6
26947.8
31350.3
32767

OSEMINF OFFSETS:
UL = -15401
UR = -13474
LL = -15675
LR = -16384

OSEMINF GAINS:
UL = 0.974
UR = 1.113
LL = 0.957
LR = 0.916
------------------------------

L2
controls@cdsws1:~ 0$ tdsavg 30 H2:SUS-ETMY_L2_OSEMINF_UL_IN1 H2:SUS-ETMY_L2_OSEMINF_UR_IN1 H2:SUS-ETMY_L2_OSEMINF_LL_IN1 H2:SUS-ETMY_L2_OSEMINF_LR_IN1
23679.1
23181.5
26177.6
23786

OSEMINF OFFSETS:
UL = -11840
UR = -11591
LL = -13089
LR = -11893

OSEMINF GAINS:
UL = 1.267
UR = 1.294
LL = 1.146
LR = 1.261
H1 PSL
michael.rodruck@LIGO.ORG - posted 14:53, Friday 03 February 2012 (2148)
H1 PSL installation

Insulation work is finished. Next week the interior drywall will be attached, roof plywood installed, and feedthrus drilled.

Images attached to this report
H2 SUS
betsy.weaver@LIGO.ORG - posted 14:06, Friday 03 February 2012 - last comment - 12:36, Monday 06 February 2012(2146)
ETMy-to-ISI Alignment
This afternoon, after a quick consult between IAS/SUS (Dennis, Calum, Hugh, Gerardo, Travis, Doug, Jason, me), it was decided that we would proceed with translating the QUAD to bring the longitudinal measurement back into tolerance.  So, we will mobilize to push the suspension ~4mm towards the corner station.  Doug has confirmed that on Monday (and probably Tuesday) IAS/SUS together will:

- fix pitch R0
- translate QUAD longitudinally, monitoring yaw, x, y

 --Need to insert possible "fix" to R0 roll/side shift if indeed we need to.  Norna is checking on tolerances of these DOFs.

- make IAS gap measurement
- Adjust R0 to fix gap and yaw if needed.

OSEM general health has been checked, and all open light voltages have been measured and set.
Comments related to this report
betsy.weaver@LIGO.ORG - 12:32, Monday 06 February 2012 (2156)
Shortly after my prompt alog, Dennis emailed the following:
>
Betsy et. al.,
I have discussed the 4mm longitudinal position error with Peter Fritschel. Peter and I both recall that the tolerance for the arm cavity length is more like centimeters. (The +/- 3 mm position may apply to the recycling cavity lengths.) Given this fact plus a possible change in longitudinal position related to the move into the WBSC6 chamber, we do not think it is worthwhile to correct this 4 mm positional error on the test stand.
Once the cartridge is installed into WBSC6, a new IAS measurement of the longitudinal positional error will be made. It is likely that this error will be within the static correction range for HEPI, or at the very least well within the true allowable arm cavity length tolerance.

Dennis Coyne
>
The longitudinal position tolerance can be relaxed for all arms, not just the H2 Y-arm for the ALS test.

Dennis Coyne


betsy.weaver@LIGO.ORG - 12:34, Monday 06 February 2012 (2157)
Further, we solicited for roll and side-shift tolerances which need to be fixted, and received the following email from Ken Strain:

Hi,

a roll error mixes the coordinates, a transverse error of a few mm 
has little effect on the force (as the pattern still faces the TM), 
but the resulting torque changes a little.
Due to the large ratio between the radius of the mirror an the ESD 
gap, and the relatively benign nature of the changes in torque, the 
effects are very small compared to the permitted variation in 
force/torque among the 4 quadrants due to relative pitch and yaw 
wedge angles (between TM and RM).
With errors described, of order 2mm, I can't see any consequences 
worse than (or even comparable to) those that have already been 
permitted.
Ken
betsy.weaver@LIGO.ORG - 12:36, Monday 06 February 2012 (2158)
While the emails above appear to clear the plate of many adjustments, we are currently working on the suspension to fix the yaw of both chains and height of the reaction chain.  
H2 SUS
jeffrey.kissel@LIGO.ORG - posted 08:25, Tuesday 31 January 2012 - last comment - 15:32, Monday 06 February 2012(2109)
Update on SUS Sign Conventions and Status
Earlier this month, we had made a complete assessment of all aLIGO SUS signs and flips across suspension types (see T1200015) and currently functional builds of those types (see SignStatus_120111.pdf). 

In doing so we identified:

- Some suspensions had a few of their magnets flipped
- The BSFMs had some mistakes in their Basis Transformation Matrices
- Some suspensions had there OSEMINF sat-amp compensation filters defined with more poles than zeros
 

For the most part, we have now rectified these problems. I attach a new assessment. The remaining action items are

- Fix H2 ITMY M0's F2 and F3 magnet sign; to fix when we swap out for real ITM Optic
- Figure out what's wrong with X2 PR3's SD OSEM; currently under investigation.
 
Non-image files attached to this report
Comments related to this report
jeffrey.kissel@LIGO.ORG - 15:32, Monday 06 February 2012 (2163)
In regards to the issue with PR3's SD OSEM: 

Because of the arrangement and tight quarters in HAM2, it has been decided (RODA pending!) that some suspensions have their OSEMs Sensor/Coils over on "opposite side" (as opposed to the "side"), i.e. -Trans side of the suspension -- see E1100109 for which is which. It turns out PR3 is one of those suspensions. 

We therefore expect the magnet polarity to be [N] not [S], which means the COILOUTF gain should be [+] not [-]. Once this COILOUTF gain was flipped, the Transverse damping loop closed and was stable, just the like the rest of the DOFs. Mystery solved!

*achoo!*
Displaying reports 74861-74880 of 76953.Go to page Start 3740 3741 3742 3743 3744 3745 3746 3747 3748 End