(not Bram)
Attached is a plot of the response of the laser crystal TEC when stepping 13mK up and then down at the slow input (assume -1K/V sensitivity). The response shows a spike by about +9mK and then -9mK (not clear how good either calibration is). The response time is roughly 5 sec which would indicate a ~30 mHz pole in the slow input of the laser.
B. Bland, J. Kissel [[ These results are extremely belated, but I want to post them for posterity. ]] After making adjustments to the static Roll of the optic, we've remeasured (as much as we remembered to of) H1 SUS MC2. Attached are the results. Several things that these results indicate: - The alleviation of the DC/static roll in the optic appears to have restored normalcy on the lowest L and T mode, i.e. they're no longer split in frequency. Huh! - Now all degrees of freedom match well with the model, and other degrees of freedom. - Spectra compared against in-chamber LLO suspensions indicate that excess noise seen at high-frequency is indeed most likely due to the noisy external environment on the optical tables. However, you'll notice that the noise has significantly reduced from when the SUS was in LHO staging building, so this confirms that the LHO assembly area is simply the noisiest environment where testing takes place, and it's not a problem fundamental to the suspension itself. As of these results, H1 SUS MC2 has passed Phase 2b testing and is ready for install into HAM3.
I re-locked the arm, after I had unlocked it earlier to take some measuremens with the ALS table PZTs.
I'm leaving it locked overnight.
I replaced the old ALS QPD output matrices (measured empirically by Elli and Thomas several weeks ago) with new ones based on Cartesian coordinates. I obtained these by ray-tracing on the ALS table. Then I measured the input matrices with a Matlab script (/svn/cdsutils/trunk/ALS).
Now the inputs of the IP_POS and IP_ANG filter modules should be calibrated in meters and radians.
These are the matrices:
INPIT =
-0.001934944012837 0.000295830132597
-0.000025683755141 -0.000141436038465
INYAW =
0.001338191005451 -0.000236296533156
-0.000010597031521 0.000352368394300
OUTPIT =OUTYAW=
1.0e+07 *
-0.746993867828206 2.240981603484618
0.048486940975147 -0.800820822925440
The loops have UGF of about 10 Hz with these gains:
IP_POS_PIT= -8; IP_ANG_PIT=-8; IP_POS_YAW=-8; IP_ANF_YAW=-4
Attached are some spectra measured with the loops either open or closed.
The TMS table relative lateral stability seems to comply with the requirements (=100urad RMS). On the other hand the angular stability seems to be a bit worse than desired (=1urad RMS).
Long term stability (12+ hrs) still has to be evaluated.
There seems to be a factor 2-3 gain peaking at 20Hz. The gain peaking seems to be responsible for most of the remaining rms. There are also lines at 60Hz and just below that are fairly large in the spectrum. Is this real motion or just sensor noise? Also: pdf of 2nd file
I multiplied the input matrices by 1e6 so that we read out um and urad at the input of the IP_POS and IP_ANG filter modules.
On all modules I enabled a filter called "cal" that divides that factor out.
Attached are plots of dust counts > .5 microns in particles per cubic foot.
[Michael R., Volker Q.]
After installing the new EOM and measuring the PM sidebands, see previous entry, we measured the RFAM on the core modulation frequencies 9.1MHz, 45.5MHz and 24.1MHz. All frequencies were driven with 10Vpp. The measurement was performed using a LZH aLIGO PSL locking PD (D1002163) mounted at the position of IO_AB_PD1.
The AC path of this diode has a 4x amplification with respect to the DC path and 50 ohm output impedance. The DC path also has 50 ohm output impedance.
The DC value was measured with a TDS 2024 into high impedance, the AC output was measured with an Agilent 4395a into 50 ohm. (Note, this gives another factor of 2 for the DC value.)
The RFAM was calculated using this formula RFAM = (V_AC/4) / (V_DC/2) accounting for AC amplification and DC path impedance. V_AC denotes the Vrms as measured with the spectrum analyzer and V_DC the voltage read from the oscilloscope. The PM value below is the modulation index as measured previously. As a sanity check the V_AC measurements were done with the dBv, dBm and Volt settings of the spectrum analyzer to confirm that Vrms is displayed. The following table shows the measurements:
DC value in mV | PM | 45.5 ampl. (dBV) | 45.5 ampl. (dBm) | 45.5 ampl. (uV) | RFAM | Frequency | RFAM/PM | ||
930 | 0.31 | -79.8 | -66.8 | 103 | |||||
1.02E-04 | 1.02E-04 | 1.03E-04 | Vrms | 5.50158E-05 | 45.5 MHz | 1.77E-04 | |||
9.1 ampl. (dBV) | 9.1 ampl. (dBm) | 9.1 ampl. (uV) | |||||||
0.39 | -75.4 | -62.4 | 172 | ||||||
1.70E-04 | 1.70E-04 | 1.72E-04 | Vrms | 9.13034E-05 | 9.1 MHz | 2.34E-04 | |||
24.1 ampl. (dBV) | 24.1 ampl. (dBm) | 24.1 ampl. (uV) | |||||||
0.14 | -84.6 | -71.5 | 61 | ||||||
5.89E-05 | 5.95E-05 | 6.10E-05 | Vrms | 3.16583E-05 | 24.1MHz | 2.26E-04 |
The Agilent 4395a was set to BW = 3kHz.
Kyle, Gerardo 7/31/2012 With the leak detector backing a 50l/sec turbo which was pumping the HAM5 HAM6 annulus and the Vertex Volume being pumped by the Vertex MTP (@ 3.4 x 10-7 torr), we sprayed (~5sec bursts of audible flow) at each of the test ports of the HAM6 side of the HAM5/HAM6 septum. No response. Next, we moved the leak detector over to back the Vertex MTP such that 100% of the exhaust was being sampled and, at some point, noticed that the helium background had risen (from the ~ 2 x 10-9 torr*l/sec when backing annulus turbo) to 6.5 x 10-9 torr*L/sec 8/1/2012 The indicated helium background of the leak detector was 6.5 x 10-9 torr*L/sec when we left yesterday. It remained unchanged today. This value falls off rapidly as the 10" gate valve at the MTP inlet is closed and is behaving as if it is a real signal sourced on the VE side of the 10" gate valve. Any helium introduced via cross talk to a leaking metal joint yesterday would not remain unchanged for this many hours of 2000 l/sec MTP pumping. It is much more likely to be a reservoir permeated through the annulus viton from the previous days spraying -> Today we vented, then pumped, then vented, then pumped, then vented then pumped the HAM6 and HAM5 annulus space over the course of the afternoon in hopes that this might expedite the removal of helium permeated into the annulus viton. We observed during the initial annulus vent that the helium signal increased slightly while the annulus was vented. With the helium background too high for acceptance testing of new conflat or feed-through joints we decided to eliminated any gross leaks existing on HAM5 and HAM6. All conflat joints and electrical feed-throughs were sprayed with 10 second blasts of audible flow. The helium background rose slowly and steadily from 6.5 x 10-9 torr*l/sec to 1.3 x 10-8 torr*l/sec over the 1 hour period we were testing
[Michael R., Volker Q.]
Following the measurements at LLO we measured the frequency dependent sideband generation around the resonant frequencies. See here for the LLO measurements.
The table below shows the sideband height as measured with a OSA on the PSL table. The frequency is the modulation frequency in MHz. All three modulator inputs were driven with 10Vpp (in 50ohm).
Carrier (V): | 2.15 | 2.10 | 2.14 | |||||||
Frequency | SB (mV) | m | Frequency | SB (mV) | m | Frequency | SB (mV) | m | ||
8.90 | 21 | 0.20 | 23.80 | 7 | 0.12 | 44.80 | 14 | 0.16 | ||
8.95 | 28 | 0.23 | 23.85 | 11 | 0.14 | 44.90 | 15 | 0.17 | ||
9.00 | 37 | 0.26 | 23.90 | 13 | 0.16 | 45.00 | 18 | 0.18 | ||
9.05 | 55 | 0.32 | 23.95 | 17 | 0.18 | 45.10 | 24 | 0.21 | ||
9.10 | 82 | 0.39 | 24.00 | 18 | 0.19 | 45.20 | 30 | 0.24 | ||
9.15 | 99 | 0.43 | 24.05 | 17 | 0.18 | 45.30 | 37 | 0.26 | ||
9.20 | 90 | 0.41 | 24.10 | 11 | 0.14 | 45.40 | 45 | 0.29 | ||
9.25 | 66 | 0.35 | 24.15 | 10 | 0.14 | 45.50 | 53 | 0.31 | ||
9.30 | 44 | 0.29 | 24.20 | 9 | 0.13 | 45.60 | 53 | 0.31 | ||
9.35 | 28 | 0.23 | 24.25 | 7 | 0.12 | 45.70 | 50 | 0.31 | ||
9.40 | 19 | 0.19 | 24.30 | 5 | 0.10 | 45.80 | 40 | 0.27 | ||
45.90 | 30 | 0.24 | ||||||||
46.00 | 22 | 0.20 | ||||||||
46.10 | 17 | 0.18 |
The sideband strength is well centered around the target frequencies of 9.1MHz and 45.5MHz. The 24.1MHz modulation is slightly off by 100kHz, but I did not want to risk to change the other two frequencies while trying to change the not so important 24.1MHz frequency.
[Michael R., Cheryl V., Volker Q.] The H1 modulator has been installed on the PSL table, the temporary H2 modulator got swapped out. The H2 modulator will be re-tuned and become the H1 spare. The beam after the modulator was not parallel to the grid on the optics table because IO_MB_M2 was not movable far enough to allow for the about 5 degrees of beam deviation going into the EOM. At LLO this problem was solved by using a New Focus Pedestal Base with Clamping Forks and moving mirror to an off-grid position. Here we decided to use a mirror mount with three adjustment screws and a modified blue base to mount IO_MB_M2. This provided a stable mounting the mirror mount and the needed flexibility to steer into the EOM. All IO PSL optics were realigned using pinhole 4 inch beam height apertures to keep the beam on the grid of the optical table. We also found that the reflections from the AR surfaces of the mode matching lenses IO_MB_L1 and IO_MB_L2 were surprisingly strong. (We followed various stray beams with 10W into the EOM) The immediate reflection from IO_AB_L1 hits the brushed aluminum surface of the modulator housing - the lens is turned slightly to not reflect the beam back into the modulator. This creates a wide horizontal "spray" of scattered light because of the brushed surface. We worked around this problem by placing a dog clamp beside the EOM aperture. See picture (dog_clamp_and_beam_dump.jpg), this works, but is certainly not a permanent solution. The next beam we blocked was a reflection from IO_MB_L2 partially hitting the rim of IO_MB_L1 and partially going trough hitting the dog clamp area. See the razor blade dump roughly in the center of the dog clamp picture. Another beam we blocked was going towards IO_MB_L2, see (L2_dump.jpg). (At the time of writing this we are not sure anymore if this beam was already blocked by the dump in front of IO_MB_L1, this needs verification). Finally we placed a razor blade dump close to IO_MB_M3 to catch the wrong polarization in the separated beam coming from the modulator crystal. See picture (wrong_pol_bd.jpg) Sideband strength and RFAM will be reported in a separate log entries.
35W beam
We briefly transitioned to high power yesteday for IO work. While transitioning I forgot to turn off the FE watchdog, so when I closed the FE shutter the laser tripped, which is the discontinuity you see.
- HAM 2 alignment - Jason - HAM ISI transfer functions - Hugo - Reclean Optics table lab - Terry - Connecting H2 ITMY Ring Heater Cables - Fil - HAM 3 : Doug & Jason - HAM 6 leak checking - Kyle & 2:45 to 2:55 - RFAM measurements - Michael & Volker
Pulled H2 Ring Heater Cable from BSC8 along Y-Arm to H2 PEM/TCS Rack. Ring Heater Chassis power is connected, but unit was left off. IO Chassis still needs work, missing ADC card. Work was done around 10-11 am. Bram/Hugo/Control room was notified prior to start of work.
Bubba and I went walk-about in the LVEA. There is a chamber cleanroom in place over HAM 5/6. There is an iLIGO garbing room on the south side of the chambers and an aLIGO garbing/staging cleanroom on the north side. As far as cleanroom positioning goes, we are in good shape for a HAM install next week. (I forgot to check the sock situation but I'll follow-up tomorrow.)
Cheryl and I went down to Xend and removed the ETM from the SUS Cage and put it in a cake tin. Cheryl returned the cake tin to the corner station Optics Lab. The Apollo crew removed the passive SEI stack from the chamber, then shimmed it and wrapped it for short-term storage. It is sitting on the near-side termination slab with CAUTION tape around it. Apollo will continue to stage for ICC.
The iLigo H1 ETMX was loaded into the cake tin, and transported to the OSB optics lab. Optics Carrier: D961460-SN008, base and cover. Optic: ETM01-A I also put a label on the cover which reads "4K ETMX, ETM-01-A (first dash a mistake, but you get the idea), Verified 8/1/2012, CV & JF." The SN of the cake tin, and the label with the "verified" contents, allows us to track optics without opening the cake tins, and exposing the optic to more contamination and risk of damage. All cake tins in the OSB optics lab have "verified" labels/contents.
At EY, I checked the voltages on the laser controller diagnostic cable (H2:ISC_WSCBSC_81), which runs from the Laser controller to the Backhoff interface chassis. They seem to be possible so I connected it up to the Beckhoff chassis. Now we have more meaningfull values on the ALS_CUST_LASER.adl screen.
The attached is the latest transfer function measurements taken on H1 SUS PR2 SAGM1 stage. The data was taken 07-29-2012 and is plotted with the predicted model and a previous measurement from the metal build in the Staging Building. The pdfs are committed to the SUS SVN under: '~/sus/trunk/HSTS/Common/Data/allhstss_2012-07-29_*' Data committed in: '~/sus/trunk/HSTS/H1/PR2/SAGM1/Data/2012-07-29_H1SUSPR2_M1_WhiteNoise_*DoF*_0p1to50Hz.txt' DTT templates used had the same name but with an "xml" extension.
Great - since these TFs do not show any indication of rubbing, and the SUS is still healthy, we can begin the metal-to-glass swap.
After digging through the old file hoping to compare them against thelatest data, I found that this data set was saved wihout the "_HHMM_" tag in the date and time portion of the file name. I've svn moved and committed all associated .xmls and .txt files such that they now have a "_1838_" (the HHMM of the Longitudinal TF), and then re-ran plotHSTS_dtttfs.m, saving new plots and .mat files, so that they can be properly used in the plotallhsts_tfs.m for future reference. So, new files are: ${SusSVN}/sus/trunk/HSTS/H1/PR2/SAGM1/ .xmls of raw data: Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_L_Op1to50Hz.xml Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_P_Op1to50Hz.xml Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_R_Op1to50Hz.xml Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_T_Op1to50Hz.xml Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_V_Op1to50Hz.xml Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_Y_Op1to50Hz.xml .txts of exported data: Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_L_0p1to50Hz.txt Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_P_0p1to50Hz.txt Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_R_0p1to50Hz.txt Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_T_0p1to50Hz.txt Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_V_0p1to50Hz.txt Data/2012-07-29_1838_H1SUSPR2_M1_WhiteNoise_Y_0p1to50Hz.txt .mat of data processed by plotHSTS_dtttfs.m: Results/2012-07-29_1838_H1SUSPR2_M1.mat and attached is the individual SUS's set of plots which include off-diagonal TFs.