jodi.fauver@LIGO.ORG - posted 08:04, Monday 04 June 2012 (3023)
BSC1 ICC-HAM2 ISI Install Prep
On Friday, after spending some time with Hugh, Bubba and I made a brief checklist (See below) to coordinate our efforts in the BSC1-HAM2 area.

1. Friday pm
     -Second cleaning BSC1
2. Friday pm
     -Move A-frames and I-beams to HAM 2 area
3. Monday am
     -Move cleanroom from HAM4 to HAM5
     -Remove doors and dome from BSC1
     -Stage Optics extract equipment at BSC1 South
     -Install SEI Soft Roof and adjust cleanroom location at HAM2
4. Tuesday am
     -Move Garbing/Staging cleanroom from HAM2 to BSC1 South
     -Optics de-install BSC1 (Lead: Cheryl, Apollo help)
     -Move Garbing/Staging cleanroom from BSC1 to HAM2
5. Tuesday pm
     -BSC1 ISI de-install (Contact Dale)
6. Wednesday am
     -Set up A-frames, etc at HAM2 
7. Wednesday pm
     -Install HAM2 ISI



H2 SUS
jeffrey.garcia@LIGO.ORG - posted 22:41, Sunday 03 June 2012 (3022)
H2 SUS ETMY M0 R0 TFs before BSC 6 chamber close-out
The attached are the latest M0/R0 transfer function measurements on H2SUSETMY taken on the night of 06/02/2012.  Both chains appear to have no sign of rubbing and have most of their resonances coinciding with the predicted model.  The R0 chain appears to have a cross-coupled resonance in the Longitudinal DoF, but otherwise a very clean measurement.  This measurement suggests the top masses on H2SUSETMY are free-swinging without obstructing forces.  H2SUSETMY is ready for the BSC 6 pump-down. 
Non-image files attached to this report
H1 DAQ
david.barker@LIGO.ORG - posted 18:03, Friday 01 June 2012 (3021)
H1 frame writer h1fw0 up and running
Dan, Jim and Dave

We have built up the primary H1 frame writer system, which consists of

Sataboy QFS file system in LDAS server room.
Solaris QFS writer in MSR room.
h1fw0 frame writer, on the 10GigE DAQ network and writing to the solaris box over a private 1GigE NFS link.

h1fw0 is writing frames which are identical to h1fw1.
H2 CDS
david.barker@LIGO.ORG - posted 17:59, Friday 01 June 2012 (3020)
FMY crashed again at 5:55pm
h2susb78, the replaced front end computer for FMY, crashed again at 5:55pm after just over 6 hours operational. So it looks like the problem may be in the IO Chassis or the communications with it.
LHO General
patrick.thomas@LIGO.ORG - posted 17:39, Friday 01 June 2012 (3019)
plots of dust counts
Attached are plots of dust counts > .5 microns.
Non-image files attached to this report
H2 SUS
jeffrey.garcia@LIGO.ORG - posted 16:12, Friday 01 June 2012 (3018)
Ops Day Summary
Grouting and chamber door close at End-Y BSC 6.
Work on BSC 8 in the LVEA
X1 SEI
corey.gray@LIGO.ORG - posted 15:18, Friday 01 June 2012 (3016)
HAM ISI Work

(corey, eric, hugo, mitch)

#7

Started assembling Stiffener under Stage-0.  Managed to get about 75% done, and called it an early day due to high temperatures in Staging (swapping HVAC filters took temperature from 65deg to 72).  So, we didn't want to have more sweat than normal drop on Class-A parts.

Horizontal L4Cs

Helped Hugo troubleshoot issues with corner 3 L4Cs (we couldn't see their signal).  Hugo eventually tracked down the issue to all (3) BIO Interface  boxes & Corner (3) Anti-Alias boxes being turned OFF.

LHO VE
kyle.ryan@LIGO.ORG - posted 14:58, Friday 01 June 2012 (3015)
Moved Vertex purge connection from BSC1 dome to HAM5-BSC1 dome to be removed on Monday


			
			
H2 FMP
betsy.weaver@LIGO.ORG - posted 13:48, Friday 01 June 2012 (3014)
BSC6 Cleanroom Off for the weekend

On behalf of Bubba, I write:

The cleanroom at BSC6 seemed to be drying the new grout faster than desired, so the cleanroom is being turned off for the weekend.

H1 SEI
greg.grabeel@LIGO.ORG - posted 13:42, Friday 01 June 2012 (3013)
ShockWatch log of May Shipment GS-13s LHO to LLO
On the latest shipment of GS-13s to be repaired at LLO we included a G-View shock recorder. The G-View is 25g max and we had set the threshold for recording at 10g. On the graph you can see that it records 4 data points when any of the 3 axis (x y horizontal plane z vertical) 10g threshold is reached. According to LLO the battery had worked loose when it was received (on the 31st) but it doesn't look like too much data was lost as our last data point was the 30th.

Hopefully we can gather some more data as we ship containers back and forth. 
Images attached to this report
H1 SEI
hugh.radkins@LIGO.ORG - posted 13:04, Friday 01 June 2012 (3012)
HEPI Welding Complete at HAM1--Moving BSC7 & 4 plumbing to BSCs 2 & 3
The welders are moving the plumbing already build at BSC7 to BSC2.  They've managed to disconnect without cutting.  Could be done with BSCs 2 & 3 pretty quick!

See photo for the modular design!
Images attached to this report
H2 SYS
hugh.radkins@LIGO.ORG - posted 12:53, Friday 01 June 2012 (3010)
BSC8 FeedThrus Upgraded & Cables strain relieved
I swapped out many of the 3" shrouds for 4" versions where needed and possible.  All cables are strain relieved to either the bolt through the shroud or to the Cable tray where necessary.

See the photos below for a typical cluster.
Images attached to this report
H2 SYS
betsy.weaver@LIGO.ORG - posted 12:52, Friday 01 June 2012 (3011)
PEEK cable clamps and ties in BSC6

For the record, this is the count of the PEEK cable ties and clamps installed in BSC6 on the SUS, ISI, and TMS:

 

(~13)  D1000600 sets on top side of ISI

(13) D1000600 sets on underside of ISI Stage 0

(6) D1000600 sets on underside of ISI Stage 2

(2) D1000600 sets on the TMS structure

 

(10) PEEK Zip ties on the ETMy QUAD structure

 

Each D1000600 set is comprised of 2 of the D1000600 PEEK clamps, so we used a total of

68 D1000600 individual clamps

H1 FMP
jodi.fauver@LIGO.ORG - posted 11:46, Friday 01 June 2012 (3008)
BSC2-HAM4 ICC
BSC2 flooring was moved into the LVEA. The crew went into HAM4 and removed the dust barriers. BSC2 flooring was installed. (There was a slight delay caused by a missing nut.) The doors will go back on this afternoon.
H2 CDS
david.barker@LIGO.ORG - posted 08:41, Friday 01 June 2012 - last comment - 11:59, Friday 01 June 2012(3006)
h2susb78 (SUS FMY) crashed again
h2susb78 remains unstable. It crashed between 2am and 3am this morning. It was removed from the watchdog system yesterday, and did not trip SEI BSC8 this time. I will power cycle it, and it should not crash SUS ITMY since it is no longer part of the Dolphin fabric.
Comments related to this report
david.barker@LIGO.ORG - 11:59, Friday 01 June 2012 (3009)
CPU was replaced with another H2 frontend with the same card configuration. New machine was restarted at 11:30 today. h2boot was modified to take the new MAC address.
H2 CDS
david.barker@LIGO.ORG - posted 22:12, Thursday 31 May 2012 (3005)
Qlogic switch relocated from H2 EE building to MSR to support H1 DAQ
We relocated one of the two Qlogic fiber channel switches from the H2 EE building (where it was part of a redundant CDS-LDAS network) to its new location in the MSR. It will connect the new H1 DAQ to LDAS, allowing the publishing of H1 frames and distribution of H1 data via the NDS2 server. Dan is setting up the QFS solaris machine for h1fw0 (we currently only have h1fw1 operational). Jim is building up h1fw0 from its cloned state.
H2 SUS
jeffrey.garcia@LIGO.ORG - posted 15:55, Thursday 31 May 2012 - last comment - 10:43, Friday 01 June 2012(2992)
ITMY Optical Lever signals added to QUAD model
D. Barker, J. Garcia


The optical lever photodiode signals were added to the H2 SUS ITMY Simulink model ('h2susitmy.mdl') at the "L3" stage.  The signals are routed to the second ADC on 'h2susb478', which is labeled "ADC_1" in the Simulink model.  The "h2susitmy.mdl" file was modified to incorporate the four quadrant signals from the ITMY Optical Lever photodiode into the "QUAD_MASTER.mdl" library part. 

In the 'h2susitmy.mdl' file, the four quadrant signals are on the ADC_1 channels: 
UL = ADC_1_1
UR = ADC_1_3
LL = ADC_1_2
LR = ADC_1_4

NOTE: this configuration will hold true for the One Arm Test and does NOT reflect the latest H2 SUS ITM, BS, FM wiring diagram, D1001725-v10.  

-------

In addition to the above changes to the "custom" top level Simulink model ('h2susitmy'), the "QUAD_MASTER.mdl" generic library part was modified for several bug fixes.  Inside the 'QUAD_MASTER' diagram, the editing was performed mainly at the 'QUAD_MASTER/QUAD/L3/OPLEVWIT/' level.  Initially, SUM, Pitch, and Yaw outputs were test points without filter modules.  However, this prevented the DAQ from accepting these channels and writing them to frames, unbeknowst to CDS.  These library parts were replaced with actual filter modules instead of test points.  This resolved the ability to write these channels to frames.  

The Optical Lever photodiode channels being written to frames are:
H2:SUS-ITMY_L3_OPLEVWIT_P_OUT
H2:SUS-ITMY_L3_OPLEVWIT_Y_OUT
H2:SUS-ITMY_L3_OPLEVWIT_SUM_OUT

The 'h2susitmy' model has now been recompiled and installed on 'h2susb478'.  


Comments related to this report
jeffrey.garcia@LIGO.ORG - 10:43, Friday 01 June 2012 (3007)
In addition to the above changes to the Simulink model, there were two other modifications.  The inter-computer communication library part (PCI) was removed, because the Optical Lever signals are now fed directly to the H2 SUS ITMY machine 'h2susb478'.  Secondly, the input for the Laser Power monitor was grounded.  Currently, the laser power is not directly fed to the ADC on 'h2susb478'.  
H2 SUS
jeffrey.garcia@LIGO.ORG - posted 10:37, Thursday 31 May 2012 - last comment - 16:14, Friday 01 June 2012(2991)
H2 SUS ITMY, 2012-05-29 Measurements
J. Garcia, S. Steplewski

We made a set of TOP transfer function measurements on ITMY Tuesday.  This was the first time we tested ITMY since this chamber was pumped down.  Measurements were clean and the results match up reasonably well with the (fiber) model - although this is a wire rehang after the fiber break.

The .mat files are located in:
~/SusSVN/trunk/QUAD/H2/ITMY/SAGM0/Data/
~/SusSVN/trunk/QUAD/H2/ITMY/SAGR0/Data/

The .pdf files are located in:

~/SusSVN/trunk/QUAD/H2/ITMY/SAGM0/Results
~/SusSVN/trunk/QUAD/H2/ITMY/SAGR0/Results
Non-image files attached to this report
Comments related to this report
szymon.steplewski@LIGO.ORG - 14:44, Thursday 31 May 2012 (2995)
Attached is another plot that shows the ITMY TOP transfer function taken In-Vacuum compared to that taken In-Air.  Aslo plotted are the wire (for M0) and thincp (for R0) models for reference.
Non-image files attached to this comment
szymon.steplewski@LIGO.ORG - 16:14, Friday 01 June 2012 (3017)
Mark Barton, Szymon Steplewski 

We changed the parameters in the Mathematica model for the QUAD suspension for ITMY to reflect the fact that the glass test mass is suspended from steel wires after the fiber break.  These changes have resulted in noticeably better agreement between the measurements and model for several degrees of freedom.

The TOP Pitch to Pitch mode around 1.4 Hz has shifted to higher frequency (from about 1.31 Hz to 1.44 Hz), while the Pitch mode near 0.56 Hz is practically unchanged, which is where we want it according to the design requirements.  The TOP Longitudinal to Longitudinal plot also shows some nice improvement in the mode around ~1.5 Hz, which matches up a little more with the "wirerehang" model now.

There is now a new wirerehang Mathematica model in the SUS SVN under

~/trunk/Common/MathematicaModels/QuadLite2Lateral/mark.barton/20120601TMproductionTMrehang/

which was built from elements of the fiber model under

~/trunk/Common/MathematicaModels/QuadLite2Lateral/mark.barton/20120601TMproductionTM/

We have added 'wirerehang' to the list of case models for plotting in Matlab, so the different cases are now:

1.  'wire'
2.  'fiber'
3.  'erm'
4.  'thincp'
5.  'wirerehang'
Non-image files attached to this comment
H2 ISC
keita.kawabe@LIGO.ORG - posted 15:34, Wednesday 30 May 2012 - last comment - 10:37, Monday 04 June 2012(2985)
Note to self: Serial number of the broken channel in AI chassis

D070081 S1102755, the e-traveler of which is available here: https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?docid=58957

Comments related to this report
filiberto.clara@LIGO.ORG - 10:37, Monday 04 June 2012 (3024)
ISC Anti-Image Chassis (S1102755) found to have bad channel on one of the filter boards (D070081 Rev 5 SN S1106145) CH4. U1 output railing at 14V. Replaced filter boards: serial number S1106145(Rev5) and S1106146 (Rev5) with S1201597 (Rev6) and S1201598 (Rev6).