H1 General (Lockloss)
ryan.crouch@LIGO.ORG - posted 19:51, Friday 04 October 2024 - last comment - 21:00, Friday 04 October 2024(80472)
Lockloss 02:50 UTC

02:50 UTC lockloss

Comments related to this report
ryan.crouch@LIGO.ORG - 20:10, Friday 04 October 2024 (80474)ISC, Lockloss

Lockloss tool tagged FSS_OSCILATION

AS_A_DC and IMC-TRANS lost lock within 1 ms of eachother

Images attached to this comment
ryan.crouch@LIGO.ORG - 21:00, Friday 04 October 2024 (80476)

03:59 UTC Observing, automated relock with 1 round of check_mich, and 2 of PRMI and DRMI.

LHO General
corey.gray@LIGO.ORG - posted 16:26, Friday 04 October 2024 (80459)
Fri Ops DAY Shift Summary

TITLE: 10/04 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 147Mpc
INCOMING OPERATOR: Ryan C
SHIFT SUMMARY:

A bit of a rough day with H1.  Had two short locks and then it was rough going with locking, and then it got worse with winds in the afternoon.  BUT, at the end of the shift H1 was able to get to OMC WHITENING---the violin modes are elevated.
LOG:

H1 General
ryan.crouch@LIGO.ORG - posted 16:08, Friday 04 October 2024 - last comment - 16:34, Friday 04 October 2024(80470)
OPS Friday EVE shift start

TITLE: 10/04 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Wind
OUTGOING OPERATOR: Corey
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 29mph Gusts, 16mph 3min avg
    Primary useism: 0.05 μm/s
    Secondary useism: 0.22 μm/s
QUICK SUMMARY:

Comments related to this report
ryan.crouch@LIGO.ORG - 16:34, Friday 04 October 2024 (80471)

23:27 UTC Observing

H1 General
corey.gray@LIGO.ORG - posted 13:27, Friday 04 October 2024 - last comment - 15:21, Friday 04 October 2024(80468)
Mid Shift Status

H1 continues to have a rough day with short locks, and it's gotten worse with H1 dropping out at various states while trying to get to NLN.  Although H1, had been locking DRMI sort of decently, ran an Initial Alignment to see if this can help (it had been over 24hrs since the last IA was run).

For the last lock about 3hrs ago, OPO temperature was adjusted and this yielded a better range, but it was only locked 47min.

Comments related to this report
corey.gray@LIGO.ORG - 15:21, Friday 04 October 2024 (80469)

Locking woes continue.  Have had a couple of locks make it all the way past MAX POWER only to have locklosses, but most of our locklosses of late are early ISC LOCK stages.  This is some weather.  Winds are now in the 30mph range and microseism is above 50th percentile.

Images attached to this comment
H1 PSL (PSL)
marc.pirello@LIGO.ORG - posted 12:10, Friday 04 October 2024 - last comment - 17:59, Monday 11 November 2024(80467)
PSL FSS PZT Locked Comparison H1 and L1

In the process of investigating the locklosses due to FSS glitching and working on spare chassis for the FSS in the PSL, we compared the power spectrum of the PZT monitor between H1 and L1.  We found some difference in the power spectrum, plot attached.

Images attached to this report
Comments related to this report
ryan.short@LIGO.ORG - 17:59, Monday 11 November 2024 (81210)PSL

I discovered today that LLO is indeed doing some additional digital filtering to their FSS_FASTMON channel, which would very likely explain the difference in spectra Marc shows above. Just looking at the MEDM screen for the filter bank, it shows three filters in use (called "cts2V', "NPRO", and "toMHz") while LHO is using none; parameters of these are attached in a screenshot. I'm not entirely sure what the purpose of these are, but from what I can tell there is an additional pole at 10Hz, which would explain the 1/f-looking drop in noise towards higher frequencies.

Images attached to this comment
H1 OpsInfo (SQZ)
sheila.dwyer@LIGO.ORG - posted 10:40, Friday 04 October 2024 - last comment - 08:52, Wednesday 08 January 2025(80461)
OPO temperature adjustment instructions

Because we moved the OPO crystal position yesterday, 8045, the crystal absorption will be changing for a few weeks, and we will need to adjust the temperature setting for the OPO until it settles, as Tony and I did last night. 80455.  Here are two sets of instructions, one that can be used while in observing to adjust this, and one that can be done when we loose lock or are out of observing for some other reason. 

For these first few days, please follow the out of observing instructions when relocking, if this hasn't been done in the last few hours, and please try to do it in the last few hours of the evening shift so that the temperature is close to well tuned at the start of the owl shift.

Out of observing instructions (to be done while recovering from a lockloss, can be done while ISC_LOCK is locking the IFO) (screenshot) :

In observing instructions:

Both the OPO temperature and the SQZ phase are ignored in SDF and can be adjusted while we are in observing, but it's important to log the times of any adjustments made. 

Changes made to make this easier (Thanks Corey for beta testing the instructions):

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 08:52, Wednesday 08 January 2025 (82177)

Update to the In observing instructions:

Both the OPO temperature and the SQZ phase are ignored in SDF, but it's important go out of observing for the change and to log any adjustments made. 

  • To adjust the temperature in full lock, go to the purple SQZ_SCOPE button and choose the OPO temp scope, adjust the OPO temperature to maximize CLF_REFL_RF6_ABS (the green trace). 
  • After doing this it might be necessary to adjust the SQZ PHASE, which can be adjusted in steps of 5 degrees or so.  Set this to maximize the squeezing level, which you can do by watching the DARM FDS dtt template.  It is available from the SQZ DTT purple button, or on NUC33.  It is easiest to adjust this for high frequency squeezing (around 1kHz), this should get you close to good squeezing at 100 Hz. This step isn't need anymore as the SQZ_ANG_ADJUST servo should take care of this. 
H1 ISC
sheila.dwyer@LIGO.ORG - posted 10:05, Friday 04 October 2024 (80464)
one DRMI lockloss investigation

Corey has had a series of locklosses from DRMI/PRMI and ALS, I've looked into one of these: 1412094082

In this case it looks like there is a large glitch seen in all DRMI LSC signals (and ASC, but it seems to start in LSC), with a small glitch about 4 seconds before lockloss, which DRMI recovers from.  Then there is a larger glitch that happened about 1 second before lockloss, where POP18 drops below the trigger threshold, and the DRMI guardian goes to DOWN, at the same time the ALS Y arm looses green lock, which then unlocks the ALS X arm. 

This glitch doesn't seem to be cause by the DRMI ASC, although we were in the process of engaging that ASC when it happened.  It also isn't due to the BS ISI isolation coming on, which hadn't happened yet. 

Edit:  I looked into a second lockloss, from PRMI: 1412095134, this one is different in that it happens very quickly and PRMI and ALS signals drop out at close to the same time.  This happened while the DRMI guardian was waiting for the refl WFS centering to converge.

Images attached to this report
H1 PSL (PSL)
corey.gray@LIGO.ORG - posted 09:48, Friday 04 October 2024 (80463)
PSL Status Report (FAMIS #26303)

For FAMIS 26303:
Laser Status:
    NPRO output power is 1.818W (nominal ~2W)
    AMP1 output power is 64.55W (nominal ~70W)
    AMP2 output power is 137.7W (nominal 135-140W)
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN
    PDWD watchdog is GREEN

PMC:
    It has been locked 0 days, 22 hr 2 minutes
    Reflected power = 22.1W
    Transmitted power = 104.0W
    PowerSum = 126.1W

FSS:
    It has been locked for 0 days 0 hr and 7 min
    TPD[V] = 0.8089V

ISS:
    The diffracted power is around 2.4%
    Last saturation event was 0 days 0 hours and 7 minutes ago

Possible Issues:
    AMP1 power is low
    PMC reflected power is high

H1 SEI (SEI)
corey.gray@LIGO.ORG - posted 09:24, Friday 04 October 2024 (80462)
SEI ground seismometer mass position check - Monthly (#26494)

Monthly FAMIS Check (#26494)

T240 Centering Script Output:

Averaging Mass Centering channels for 10 [sec] ...
2024-10-04 09:08:54.264102
There are 14 T240 proof masses out of range ( > 0.3 [V] )!
ETMX T240 2 DOF X/U = -0.745 [V]
ETMX T240 2 DOF Y/V = -0.637 [V]
ETMX T240 2 DOF Z/W = -0.529 [V]
ITMX T240 1 DOF X/U = -1.51 [V]
ITMX T240 1 DOF Y/V = 0.325 [V]
ITMX T240 1 DOF Z/W = 0.447 [V]
ITMX T240 3 DOF X/U = -1.563 [V]
ITMY T240 3 DOF X/U = -0.773 [V]
ITMY T240 3 DOF Z/W = -1.934 [V]
BS T240 1 DOF Y/V = -0.394 [V]
BS T240 3 DOF Y/V = -0.307 [V]
BS T240 3 DOF Z/W = -0.49 [V]
HAM8 1 DOF Y/V = -0.475 [V]
HAM8 1 DOF Z/W = -0.786 [V]
All other proof masses are within range ( < 0.3 [V] ):
ETMX T240 1 DOF X/U = -0.083 [V]
ETMX T240 1 DOF Y/V = -0.1 [V]
ETMX T240 1 DOF Z/W = -0.159 [V]
ETMX T240 3 DOF X/U = -0.106 [V]
ETMX T240 3 DOF Y/V = -0.165 [V]
ETMX T240 3 DOF Z/W = -0.093 [V]
ETMY T240 1 DOF X/U = 0.009 [V]
ETMY T240 1 DOF Y/V = 0.116 [V]
ETMY T240 1 DOF Z/W = 0.173 [V]
ETMY T240 2 DOF X/U = -0.083 [V]
ETMY T240 2 DOF Y/V = 0.171 [V]
ETMY T240 2 DOF Z/W = 0.03 [V]
ETMY T240 3 DOF X/U = 0.182 [V]
ETMY T240 3 DOF Y/V = 0.029 [V]
ETMY T240 3 DOF Z/W = 0.109 [V]
ITMX T240 2 DOF X/U = 0.127 [V]
ITMX T240 2 DOF Y/V = 0.258 [V]
ITMX T240 2 DOF Z/W = 0.22 [V]
ITMX T240 3 DOF Y/V = 0.117 [V]
ITMX T240 3 DOF Z/W = 0.104 [V]
ITMY T240 1 DOF X/U = 0.078 [V]
ITMY T240 1 DOF Y/V = 0.083 [V]
ITMY T240 1 DOF Z/W = -0.061 [V]
ITMY T240 2 DOF X/U = 0.054 [V]
ITMY T240 2 DOF Y/V = 0.185 [V]
ITMY T240 2 DOF Z/W = 0.092 [V]
ITMY T240 3 DOF Y/V = 0.066 [V]
BS T240 1 DOF X/U = -0.162 [V]
BS T240 1 DOF Z/W = 0.087 [V]
BS T240 2 DOF X/U = -0.09 [V]
BS T240 2 DOF Y/V = 0.007 [V]
BS T240 2 DOF Z/W = -0.113 [V]
BS T240 3 DOF X/U = -0.206 [V]
HAM8 1 DOF X/U = -0.28 [V]

Assessment complete.

STS Centering Script Output:

 

Averaging Mass Centering channels for 10 [sec] ...
2024-10-04 09:22:39.265893
There are 1 STS proof masses out of range ( > 2.0 [V] )!
STS EY DOF X/U = -2.389 [V]
All other proof masses are within range ( < 2.0 [V] ):
STS A DOF X/U = -0.514 [V]
STS A DOF Y/V = -0.927 [V]
STS A DOF Z/W = -0.438 [V]
STS B DOF X/U = 0.323 [V]
STS B DOF Y/V = 0.945 [V]
STS B DOF Z/W = -0.432 [V]
STS C DOF X/U = -0.723 [V]
STS C DOF Y/V = 0.743 [V]
STS C DOF Z/W = 0.539 [V]
STS EX DOF X/U = -0.224 [V]
STS EX DOF Y/V = -0.05 [V]
STS EX DOF Z/W = 0.084 [V]
STS EY DOF Y/V = 0.007 [V]
STS EY DOF Z/W = 1.209 [V]
STS FC DOF X/U = 0.24 [V]
STS FC DOF Y/V = -1.04 [V]
STS FC DOF Z/W = 0.636 [V]

Assessment complete.

LHO VE
david.barker@LIGO.ORG - posted 08:21, Friday 04 October 2024 (80460)
Fri CP1 Fill

Fri Oct 04 08:09:58 2024 INFO: Fill completed in 9min 55secs

Images attached to this report
LHO General (PSL)
corey.gray@LIGO.ORG - posted 07:45, Friday 04 October 2024 (80458)
Fri Ops Day Transition

TITLE: 10/04 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: Ryan S
CURRENT ENVIRONMENT:
    SEI_ENV state: EARTHQUAKE
    Wind: 6mph Gusts, 3mph 3min avg
    Primary useism: 0.06 μm/s
    Secondary useism: 0.22 μm/s
QUICK SUMMARY:

Walked in to see H1 at MAX POWER after a 59min lock.  (There have been EQ spikes, in last few hours, but they do NOT line up with atleast the last 2 locklosses---for this current lock there was a M5.2 Iceland EQ which occured as H1 was stepping though ISC LOCK).  This last 1340utc lockloss also does not have an FSS tag; 1139utc lockloss had a FAIL; 0845, 840, 837, utc locklosses all have an FSS Oscillation tag.

1445 Back to Observing.

H1 General
anthony.sanchez@LIGO.ORG - posted 22:03, Thursday 03 October 2024 (80457)
Thursday Eve Shift End

TITLE: 10/04 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 142Mpc
INCOMING OPERATOR: Ryan S
SHIFT SUMMARY:
Was a great shift, Locked the whole time. Dropped from opserving very breifly to adjust the SQZ OPO temp.


LOG:

Start Time System Name Location Lazer_Haz Task Time End
21:58 PEM Robert, guests Y-arm n CE equipment demonstration 00:49
21:58 PEM Anamaria, guests X-arm n CE equipment demonstration 01:58
00:36   Geoffrey YARM n Picking up guests 00:56
00:51 VAC Gerardo Mid Y N Getting parts. 01:00
H1 SUS
anthony.sanchez@LIGO.ORG - posted 21:50, Thursday 03 October 2024 (80456)
Weekly In Lock SUS Charge measurements Famis Task 28373

https://ligo-wa.accruent.net/LB_Request_UpdateElevate.asp?RequestID=28373

 

anthony.sanchez@cdsws25: ls -lt | head -n 6
total 507
-rw-r--r-- 1 test_user          controls   160 Sep 24 07:58 ETMX_12_Hz_1411225133.txt
-rw-r--r-- 1 test_user          controls   160 Sep 24 07:50 ETMY_12_Hz_1411224662.txt
-rw-r--r-- 1 test_user          controls   160 Sep 24 07:50 ITMY_15_Hz_1411224644.txt
-rw-r--r-- 1 test_user          controls   160 Sep 24 07:50 ITMX_13_Hz_1411224643.txt
-rw-r--r-- 1 test_user          controls   160 Sep 17 07:58 ETMX_12_Hz_1410620340.txt

These files are from the Prior week : https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=80324

Looks like the SUS Charge measurements didn't run this week. :(

 

Images attached to this report
H1 General (SQZ)
anthony.sanchez@LIGO.ORG - posted 21:35, Thursday 03 October 2024 (80455)
SQZ OPO Temp adjustment

Starting 3:46 UTC

I adjusted H1:SQZ-OPO_TEC_SETTEMP while Watching the H1:SQZ-CLF_REFL_RF6_ABS_OUTPUT ndscope ( OPO temo ndscope found on the SQZ scopes dropdown)  to change the OPO Tec Temp.
ending at 3:50 UTC


Sheila and I discussed doing a SQZ Angle adjustment but it seemed fine at the time.

 

Images attached to this report
H1 General (Lockloss)
ryan.short@LIGO.ORG - posted 15:55, Thursday 03 October 2024 - last comment - 11:47, Friday 04 October 2024(80453)
Lockloss @ 22:19 UTC

Lockloss @ 22:19 UTC - link to lockloss tool

No obvious cause, but it looks like the IMC lost lock at the same time as the arms. However, I don't suspect the FSS in this case.

Images attached to this report
Comments related to this report
oli.patane@LIGO.ORG - 11:47, Friday 04 October 2024 (80466)
Images attached to this comment
H1 General (Lockloss, PSL)
anthony.sanchez@LIGO.ORG - posted 16:29, Wednesday 02 October 2024 - last comment - 11:21, Friday 04 October 2024(80436)
Control room has Internet access again and an unrelated Lockloss

TITLE: 10/02 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 148Mpc
CURRENT ENVIRONMENT:
    SEI_ENV state: USEISM
    Wind: 21mph Gusts, 13mph 3min avg
    Primary useism: 0.08 μm/s
    Secondary useism: 0.45 μm/s
QUICK SUMMARY:

Sudden Lockloss @ 21:55 UTC very likley cause by a PSL FSS issue.

IMC Had a hard time relocking locking after the lockloss.

NLN Reached @ 22:54 UTC
Observing reached @ 22:56 UTC

Images attached to this report
Comments related to this report
oli.patane@LIGO.ORG - 11:21, Friday 04 October 2024 (80465)

Definitely looks like the FSS had a large glitch and lost lock before DARM saw the lockloss. This lockloss didn't have the FSS glitches happening before though.

Images attached to this comment