Reports until 14:02, Wednesday 11 February 2015
H1 TCS (TCS)
aidan.brooks@LIGO.ORG - posted 14:02, Wednesday 11 February 2015 - last comment - 17:52, Sunday 22 February 2015(16654)
All RHs running within specifications. Passed installation acceptance testing

The attached PDFs show the resistance measurements for all RHs and the relative resistance vs time for the RHs as 1W is applied to each segment. All segments are operating within nominal parameters.

The resistances for the segments are:

Segment Resistance (Ohms)
ITMX UPPER 44.0
ITMX LOWER 42.4
ITMY UPPER 40.7
ITMY LOWER 42.5
ETMX UPPER 41.9
ETMX LOWER 41.0
ETMY UPPER 42.2
ETMY LOWER 43.6
Non-image files attached to this report
Comments related to this report
aidan.brooks@LIGO.ORG - 19:30, Wednesday 11 February 2015 (16672)

Just for reference, here is the transient thermal lens in ITMX, as measured by the HWS, from this morning's 2W RH test (~1800s of applied power).

Images attached to this comment
aidan.brooks@LIGO.ORG - 17:52, Sunday 22 February 2015 (16858)

A follow-up. The data from the RTD temperature sensors for each RH is plotted in the attaced PDF. 

The ETMY and ITMX RH RTDs are non-responsive.

Non-image files attached to this comment