jeffrey.kissel@LIGO.ORG - posted 17:03, Wednesday 23 September 2015 - last comment - 19:18, Thursday 24 September 2015(21868)
DARMOLGTF and PCALY to DARM Sweeps taken for Calibration Validation
J. Kissel

I've taken new DARM open loop gain and PCALY to DARM transfer functions to validate the current calibration. During the PCALY to DARM transfer function, I take the transfer function from PCALY's RX PD (calibrated into [m] of ETMY motion) and the CAL-CS front-end's DELTAL_EXTERNAL (calibrated into DARM [m], which -- since we're driving ETMY -- is identical to [m] of ETMY motion). These two different methods agree to within 4% and 3 [deg] over the 15 [Hz] to 1.2 [kHz] band. The calibration discrepancy expands to a whopping 9% and 4 [deg] if we look a frequencies between 5 and 15 [Hz] ;-).

I think we're in great shape, boys and girls.

Details
--------------
- CAL-CS does not correct for any slow time depedence (optical gain, test mass actuation strength, etc), so any agreement you see with the current interferometer is agreement with the reference model taken on Sep 10th 2015 (LHO aLOG 21385).

- In the previous measurement, Kiwamu had to fudge the phase by ~90 [us] to get the phase to agree. Now that we've updated the cycle delay between sensing and actuation to 7 [16 kHz clock cycles] to better approximate the high-frequency frequency response of AA, AI, and the OMC DCPD signal chain, we no longer have to fudge the phase -- AND the phase between the two metrics agree. NICE.

- I've made sure to turn OFF calibration lines *during both of these measurements, but there should be ample data just before and just after with calibration lines ON, such that we can compare our results against theirs to help refine our estimates of systematic error.

- The measurements live in
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Measurements/DARMOLGTFs/2015-09-23_H1_DARM_OLGTF_7to1200Hz.xml
/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/O1/H1/Measurements/PCAL$/2015-09-23_PCALY2DARMTF_7to1200Hz.xml
and have been committed to the CalSVN. We'll process these results shortly, and perform a similar analysis as Darkhan has done in yesterday's aLOG 21827.
Images attached to this report
Comments related to this report
darkhan.tuyenbayev@LIGO.ORG - 14:41, Thursday 24 September 2015 (21898)

The parameter file for this measurement was committed to calibration SVN:

CalSVN/aligocalibration/trunk/Runs/O1/H1/Scripts/DARMOLGTFs/H1DARMparams_1127083151.m

Attached plots show components of DARM loop TF and their residuals vs. DARM model for O1.

Non-image files attached to this comment
kiwamu.izumi@LIGO.ORG - 19:18, Thursday 24 September 2015 (21926)CAL

It looks better. Very nice.

By the way, I wanted to measure the open loop without the MICH or SRCL feedforward because I wanted to demonstrate that the unknown shape in the residual in magnitude is not due to these feedforward corrections. Though this may be a crazy thought. Anyway, it would be great if you can run an open-loop measurement without the feedforwards at some point, just once.