Reports until 16:08, Friday 06 May 2016
H1 CAL (CAL, ISC, SUS)
evan.goetz@LIGO.ORG - posted 16:08, Friday 06 May 2016 - last comment - 11:45, Thursday 12 May 2016(27053)
ETMY ESD driver electronics measured

Jeff K, Evan G

We (finally!) measured the ESD driver quadrant control voltage path transfer functions, which was needed because of the change made on Feb 9 (see ECR E1500341 and LHO aLOG 25468) for calibration/compensation use. Measurements are stored under 

/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PreER9/H1/Measurements/Electronics/2016-05-06

Measurement notes attached. Analysis to come...

Non-image files attached to this report
Comments related to this report
kiwamu.izumi@LIGO.ORG - 17:15, Wednesday 11 May 2016 (27129)CAL

The anaylsis is done.

I wrote a matlab analysis script which automatically fits all the measured transfer function by calling LISO. The script can be found in SVN at

/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PreER9/H1/Scripts/model_ETMY_LVLN_driver_20160511.m

The results are attached and also can be found in SVN at

/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PreER9/H1/Results/Electronics/2016-05-11_H1SUSETMY_ESD_LVLNDriver_*.pdf

Non-image files attached to this comment
kiwamu.izumi@LIGO.ORG - 11:45, Thursday 12 May 2016 (27142)

An update:

Evan and Jeff pointed me out that I could have also subtracted the reference transfer function ( = the transfer function of the measurement setup, including SR785 and diff-to-single-ended-convertor). So I edited the code so that it subtracts the reference out of each measurement. This impacted mostly on the gain of each measruement which are now 1.88. The resulting pdfs are attached.

I have moved the analysis code to a slighlty place at:

/ligo/svncommon/CalSVN/aligocalibration/trunk/Runs/PreER9/H1/Scripts/Electronics/model_ETMY_LVLN_driver_20160511.m

The resulting pdfs are saved at the same location as the original ones.

Non-image files attached to this comment