Displaying report 1-1 of 1.
Reports until 11:33, Friday 10 February 2023
H1 TCS
camilla.compton@LIGO.ORG - posted 11:33, Friday 10 February 2023 - last comment - 18:51, Tuesday 14 February 2023(67327)
ETM Ring heaters differential changes tonight 8pm to 2am.

To test reducing the differential heating in the arms, I've set a script to reduce the ETMY ring heater from 1.4W to 1.0W and increase the ETMX ring heater from 0W to 0.4W for 6 hours starting at 8pm*, ending at 2am. Timing this so that the IFO will be mostly back to nominal by Saturday for Robert's work.

*If the IFO isn't at NLN at 8pm I'll have the script keep trying until 9pm. 

The awg lines will be turned on at the start of the test is they aren't already on. We haven't had big problems with the ETM violins for a while but Rahul plans to check on them ~10pm tonight.

ETMY 1.0W seems like a safe-ish value from a PI point of view. We used this April-August (67314) but increased from 1W to 1.2W power due to a PI lockloss at 40 hours locked at 280kW circulating power (64440). 

python3 userapps/../tcs/common/scripts/power_adj_scripts/ring_heater_schedule.py ETMY -s 1360123218 -d 6 -p 1.0 -ep 1.4

Comments related to this report
craig.cahillane@LIGO.ORG - 21:33, Friday 10 February 2023 (67349)
Regrettably, The ETM ring heater change appears to have caused a fast PI ringup and lockloss at around 9:20 pm this evening.

Attached is Sheila's picture of the OMC DCPDs (second PNG)

I reverted the EX RH settings now, so the transient is not too crazy for Robert in the morning.
Images attached to this comment
daniel.brown@LIGO.ORG - 21:45, Saturday 11 February 2023 (67353)

Looking at the frequency and intensity noise lines, they all seem to be reducing with this ring heater change, roughly a 25% improvement. The 222Hz intensity line is noisy so hard to say.

This suggests that we are reducing differential arm effects, which is good. X and Y arm powers seem to be balancing more and increase a small amount, though seems to start decreasing later. Corner build ups and PRG are not being affected, as you would hope with just an ETM change.

Given the PI ring up, if we want to pursue this further we'd need to lock and power-up with the ring heaters already changed and get to the other side of whatever HOM resonance is causing it. It does kind of look like the line reduction is levelling off before they got switched off or whatever happened, so how much more is to be won is hard to say.

We would probably need to tweak the CO2s to optimise the mode matching into the new arm modes too.

(Ignore the HWS COMM/DIFF subplot, it only shows the ITMs and forgot to remove it)

Non-image files attached to this comment
evan.hall@LIGO.ORG - 21:19, Sunday 12 February 2023 (67356)

This PI mode was at 10431.4 Hz. It is plausible that this is the same troublesome EY PI mode that has been seen in the past at 10430.5 Hz (50209). The ~1 Hz frequency difference can be explained by a ~1 K shift in test mass temperature, since the elastic modulus of fused silica has a fractional temperature shift of about 200 ppm/K.

Watching the 10.4 kHz region of DARM during the ring heating, there is evidently an evolution in the coupling of some technical noise at twice the arm transverse mode spacing (it is similarly seen at the transverse mode spacing itself).

Images attached to this comment
craig.cahillane@LIGO.ORG - 14:33, Monday 13 February 2023 (67375)
During this test, directly before the PI lockloss, I had measured the DARM plant in order to see if there was any change in the DARM spring.
There was not.  

It seems that the ETM ring heater change does not really affect the SRCL offset, or the corner in general.
Images attached to this comment
camilla.compton@LIGO.ORG - 17:13, Monday 13 February 2023 (67384)

You can see some change in the ETMX measured spherical power (orange trace) after the ring heater change but it increases. I would have expected spherical power to decrease with more EX ring heater. Strange, the TOTAL_PIXEL_VALUE is decreasing at this time which mean that the amount of light getting back to the CCD is changing, zoomed out plot here. It is hard to tell if there is still some change happening at the lockloss time.

Images attached to this comment
daniel.brown@LIGO.ORG - 23:28, Monday 13 February 2023 (67393)

Attached is a movie showing the 64 kHz OMC DCPD channel spectrum during the power up, thermalisation, and then this differential ring heater change. It focuses around the 2nd order mode frequencies in the arms, 10~11 kHz. This image highlights the main features we are looking at the in the movie.

Here are some thoughts, observations:

  • Why do we see these peaks in DARM?
    • Not clear why, but the HOM resonances could be getting excited by some laser noise at high frequency, as the HOM resonances are different in each arm this excites couples differentially to AS.
  • After thermalisation there is approximately a 400Hz difference between the X and Y modes. As this is the 2nd order mode resonances, the mode-separation frequency difference between the arms is 200Hz
  • From plot we can see that having the mode-separation frequency equal between the arms is necessary but not sufficient to say we are mode-matching the arm eigenmodes
    • This shows that if we take some curvature off of one mirror, we can put it on the other and keep the same mode-separation frequency - as we were doing in tests in previous weeks
    • Assuming there are only small RoC differences, the best case with a 200Hz mode-separation frequency is the arms are about 0.1% mismatched, but could be worse.
  • How much do we expect the arm mode separation frequency to change during a power-up?
    • From FEA models, the surface RoC should change by -30uD/W of absorbed power
    • From Galaxy pages, the coating absorptions are about 0.5ppm in the ITM and 0.3ppm in the ETM. Assuming 380kW in the arms there should be 0.14W and 0.11W in the ITM and ETM respectively
    • Using the these absorbed powers and the spherical power actuation of uniform heating, going from cold RoCs to absorbing the above powers should result in around 500Hz of mode-separation frequency change in the arm
    • From the start of the movie we can see 4 distinct peaks from the 4 HG 2nd order modes from each arm (why these are so separated I do not know)
      • The average of the Y arm modes starts at ~9900Hz and after thermalising gets to ~10400Hz: a 500Hz change
      • The average of the X arm modes starts at ~10150Hz and ends up at 10750Hz: a 600Hz change
      • So order of magnitude, what we see here is sensible and what we would expect. Overall the X arm looks like it is absorbing more, which due to the ITM point absorbers isn't surprising.
  • During the differential ring heater change we see the X and Y arm modes move closer together. We can identify which modes belong to which arm as we know which direction each ETM RH was changed:
    • Increasing the RH power => decreases the mode separation frequency
    • Decreasing the RH power => increases the mode separation frequency
    • As the right most peaks move lower in frequency and we are turn up ETMX RH, we can infer that the HOM peaks at higher frequencies are from the X arm
  • When we decrease the ETMY RH we increase the Y arm mode separation frequency, this moves the two Y arm HOMs up further into the 10.4kHz acoustic modes and eventually causes a parametric instability to ring up and we lose lock.
  • We also see that this same Y arm HOM peak also excites the 10.2 kHz acoustic mode. We either moved through it quick enough it didn't have long enough to ring up, or the PI gain isn't large enough to be a problem.
  • It seems to be the same Y arm HOM which excites the 10.2 and 10.4 kHz modes

As posted in a previous comment in this alog the frequency/intensity noise coupling reduces during this differential RH test. From the movie we can see this RH change brings the X and Y HOMs closer together, which should mean we are bringing the arm eigenmodes closer together therefore reducing any differential coupling path for the laser noise.

 

Images attached to this comment
Non-image files attached to this comment
vladimir.bossilkov@LIGO.ORG - 07:59, Tuesday 14 February 2023 (67395)

Wow Dan, that is the most text book movie of how PI tunes in/out of modes. You could even see a 10200 mode get transiently excited as the HOM flew past, fortunately it moved beyond that mode before it got more excited.

This is the first time I am seeing HOMs in the OMC output. We don't get that at LLO, but we have a different (slightly older) OMC chassis, which you just changed at LHO very recently. Do you directly see any other HOMs at say ~5kHz, 15kHz or higher? (if you don't, that would indicate mode matching issues with the 20/02 modes) [edit: I see the 15kHz ones only in this recent alog]

Looks to me like a frontal ring heater can really help in micromangaing RoC in a shorter time frame to avoid these lock losses [link1, link2].

At what point will LHO recommision the ESD PI path? Seems to me thats the only sure fire way of handling this.

daniel.sigg@LIGO.ORG - 10:32, Tuesday 14 February 2023 (67405)

Even the old omcpi model had a 64KHz channel, H1:OMC-PI_DCPD_64KHZ_AHF, that got recorded to disk. This hasn't changed. What has changed is the availibility of a 512KHz test point, H1:OMC-PI_DOWNCONV_SIG_OUT.

For the current state of the PI models see alog 67402.

daniel.brown@LIGO.ORG - 18:51, Tuesday 14 February 2023 (67427)

You can see some 1st order mode movements around 5kHz, movie attached. I couldn't see anything interesting happening around 15kHz.

The Y 1st order peaks are easily seen and do not look as astigmatic as the 2nd order modes. The X arm modes are not as large and harder to see. Occasionally these HOM peaks jump up a bit and become visible. I would guess this coincides with some arm alignment change, maybe a glitch at this stage during the lock.

When thermalised, the Y arm is sitting at ~5200 Hz and the X arm ~5400 Hz which also agrees wiht the 2nd order mode peaks.

 

Non-image files attached to this comment
Displaying report 1-1 of 1.