H1 Manager called on me to help with a bunch of SDF diffs.
- Some ASC POP A offsets, which were accepted in safe earlier today (85490), so I went ahead and accepted them for Observe as well
- IMC PZT offset changes, which were made earlier when working on realigning the IMC (85476)
- ALSX camera offsets, which were changed earlier (85472)
- the SR3 and SRM dackill bypass times were lengthened temporarily for part of the work today, which I realize now should have been reverted oops need to fix when we're out of Observing tomorrow
- SQZ OPO servo's common board excitation had been left switched on (switch on the right from EXC A), so I just opened the switch
TITLE: 07/01 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Lock Acquisition
INCOMING OPERATOR: Oli
SHIFT SUMMARY:
SQZr trouble at the start of the lock but recovered and got to Observing at 00:07 UTC.
H1 was locked fine for 4 hours and 35 Minutes.
Then all the sudden Unknown Lockloss at 2025-07-01 03:56:04 UTC.
I tried to allow H1 to relock itself but it wanted me to find ASL-Y arm by hand and then seemed to get trapped in a PRMI loop. So after about 30 Minutes I did an Initial Alignment.
This did allow for a quick lock of DRMI_1F.
H1 is currently at ENGAGE_SOFT_LOOPS and locking Quickly.
LOG:
Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
---|---|---|---|---|---|---|
00:25 | PCAL | F Llamas | PCAL Lab | yes | Settign up measurements | 01:13 |
Unknown lockloss @ 3:56:07 UTC
No noticeable Seismic activity.
No PI ring Ups.
No saturations aside from HAM6 from the Fast Shutter.
On Friday Jun 27th Francisco and I went to the the PCAL lab and did a PCAL TX module maintenance according to T1600436.
AOM alignment was touched up
Results:
Laser SN05 | |
Date | June-26-2025 |
Laser Shutter Check | Fail |
Max OFS Offset | 7.2V |
95% OFS Offset | 6.84V |
Operating OFS Offset | 3.42V |
Laser Output Power | 1.97W |
Rejected Laser Power | 3.3mW |
AOM Input Power | 1.92W |
Max Diffracted Power | 1.23W |
Un-Diffracted Power | 0.500W |
AOM Diffraction Efficiency | 64% |
AOM Rejected Power | 3.3mW |
TxPD Power | 1.4mW |
OFSPD Power | 1.73mW |
Outer Beam Power | 294mW |
Inner Beam Power | 292mW |
Output Beam Power Ratio | 0.9932 |
OFS Gain | 7.50V |
OFS Phase Margin | 43.7 |
ALOG |
TITLE: 06/30 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: Ryan C
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 10mph Gusts, 4mph 3min avg
Primary useism: 0.02 μm/s
Secondary useism: 0.08 μm/s
QUICK SUMMARY:
H1 Is Currently Locked at NLN {600} But not Observing due to a SQZ_SHG issue:
Error SQZ_SHG [LOCKED.run] USERMSG 0: PZT voltage limits exceeded.
Sheila is currently working on the issue.
Sheila Fixed the issue, I have attached soem pictures to help future Ops that find this issue:
Sitemap -> SQZ -> SQZ_OVERVIEW -> SQZT0 -> PZT -> Move H1:SQZ-SHG_PZT_OFFSET slider.
We Now have a "pump fiber rej power in ham7, nominal 35e-3, align fiber pol on sqzt0" error message on SQZ-OPO_LR. nopted will do tomorrow.
Then we took SQZ_ANG_ADJUST Guardian to ADJUST_SQZ_ANG_ADF [10] which kinda worked but didn't improve sqzing and reverted back.
I then accepted the SDFs and now we're back in Observing at 00:07:25 UTC.
An NDScope of what the GRD-SQZ_SHG Guardian was doing.
TITLE: 06/30 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Lock Acquisition
INCOMING OPERATOR: Tony
SHIFT SUMMARY: We've had to relock twice, the second time we struggled a bit with DRMI. We're currently locked at NLN but trying to fix a SQZ SHG PZT issue.
LOG:
Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
---|---|---|---|---|---|---|
14:50 | FAC | Randy | EndX | N | Chiller yard pump checks | 16:01 |
15:13 | FAC | Kim | MidY | MidY | Tech clean | 16:20 |
15:14 | FAC | Nellie | MidX | N | Tech clean | 16:20 |
16:24 | EE | Ken | Mezz | N | Electrical work | 18:56 |
16:49 | FAC | Chris | VAC Prep | N | Sanding | 17:39 |
19:34 | FAC | Randy | MidY | N | Inventory | 20:30 |
20:06 | FAC | Tyler | Mids | N | FAMIS checks | 20:36 |
22:39 | CAL | Tony | PCAL lab | LOCAL | Prep for tomorrows measurement | 22:58 |
20:46 UTC lockloss
19:26 UTC back to NLN
M. Todd, R. Crouch, E. Capote, S. Dwyer
After 5 minutes or so from stepping the CO2 lasers from 1.7W to 0.9W, we began to see the 13Hz ring-up again before losing lock.
We believe this to be PRCL losing gain because of the change in the CO2s. I'm also wondering if the CO2s may be miscentered and causing some roll modes via differential heating of the fibers -- the frequenices do not exactly match up I think but need to check on this.
15:40 UTC lockloss while we were running the CO2 stepper. alog85429
16:50 UTC back to NLN
TITLE: 06/30 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 145Mpc
OUTGOING OPERATOR: Ryan S
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 8mph Gusts, 4mph 3min avg
Primary useism: 0.01 μm/s
Secondary useism: 0.07 μm/s
QUICK SUMMARY:
TITLE: 06/29 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 146Mpc
INCOMING OPERATOR: Ibrahim
SHIFT SUMMARY: One lockloss, I had to run 2 alignments to get DRMI to lock.
LOG: No log.
16:38 UTC lockloss
19:42 UTC Observing
TITLE: 06/29 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 146Mpc
OUTGOING OPERATOR: Ryan S
CURRENT ENVIRONMENT:
SEI_ENV state: CALM
Wind: 2mph Gusts, 0mph 3min avg
Primary useism: 0.01 μm/s
Secondary useism: 0.06 μm/s
QUICK SUMMARY:
TITLE: 06/28 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 151Mpc
INCOMING OPERATOR: Ibrahim
SHIFT SUMMARY: Two locklosses, mostly automated relock on the first one. We are relocking at DRMI currently.
LOG: No log.