Both MC2 and PR2 suspensions in HAM3 chamber are healthy as per the transfer function measurements I took this morning. Please see the plots attached below, the templates are stored at the following location,
/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/MC2/SAGM1/Data
2024-02-02_1500_H1SUSMC2_M1_WhiteNoise_L_0p02to50Hz.xml
2024-02-02_1500_H1SUSMC2_M1_WhiteNoise_P_0p02to50Hz.xml
2024-02-02_1500_H1SUSMC2_M1_WhiteNoise_R_0p02to50Hz.xml
2024-02-02_1500_H1SUSMC2_M1_WhiteNoise_T_0p02to50Hz.xml
2024-02-02_1500_H1SUSMC2_M1_WhiteNoise_V_0p02to50Hz.xml
2024-02-02_1500_H1SUSMC2_M1_WhiteNoise_Y_0p02to50Hz.xml
MC2 T_dof (see page 4) has a small unwanted peak at around 4.1Hz which I suspect is due to the purge air in the Chamber. However it looks better when I took a second measurement as shown here. So nothing to worry.
/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/PR2/SAGM1/Data
2024-02-02_1600_H1SUSPR2_M1_WhiteNoise_L_0p01to50Hz.xml
2024-02-02_1600_H1SUSPR2_M1_WhiteNoise_P_0p01to50Hz.xml
2024-02-02_1600_H1SUSPR2_M1_WhiteNoise_R_0p01to50Hz.xml
2024-02-02_1600_H1SUSPR2_M1_WhiteNoise_T_0p01to50Hz.xml
2024-02-02_1600_H1SUSPR2_M1_WhiteNoise_V_0p01to50Hz.xml
2024-02-02_1600_H1SUSPR2_M1_WhiteNoise_Y_0p01to50Hz.xml