Preliminary REFL beam size measurements on ISCT1

Today I used the nanoscan to measure the ITMX direct reflected beam size that makes it through the REFL path onto ISCT1.

I had previously done some more preparation for this measurement (since aLOG entry 9633), measuring the ITMX direct reflected beam power again on ISCT1 to be 31.6uW. This time the green beam was misaligned at the TMS, so was not coming through the X-arm. Also, ITMY was misaligned, and ETMX was misaligned, so there shouldn't  have been anything that was adding to the measured power. It seems that either the green light, or X-arm flashes were contributing to the ~50uW I measured last time.

At the same time, I measured the PRM direct reflected beam size using the nanoscan. However, the capture settings I used for logging the data were probably not ideal for this measurement, so I would like to retake this data at some point. In any case, the preliminary measured PRM direct reflected beam sizes were wx=1949um and wy=2094um. These can be compared to the numbers I expect from my model: wx=2170um and wy=2138um.

At this time I tried taking a measurement of the ITMX direct reflected beam, but the nanoscan software kept complaining that the detector was "saturating". At 30uW this seemed unlikely to me, but it prompted me to turn off frame averaging. As soon as I turned off the averaing I saw that the beam was swinging around on the nanoscan, with peak to peak motion of roughly a beam size. This seemed like too much motion to get a decent measurement so I decided to wait until we had some quieter optics in the path between ITMX and ISCT1.

Today I went back into ISCT1 and looked again at the nanoscan data for the ITMX direct reflected beam. The ALS beam was misaligned at TMS, ITMY, ETMX  and PRM were all misaligned. The beam motion seemed significantly less for the most part, although I did see it go through quiet and loud times. I logged data at 10Hz for around a minute, which I then averaged offline to get ITMX direct reflected beam sizes of wx=2525um and wy=2510um. The attached plot shows the beam sizes at this location that I expect from my model, for a range of different ITMX lens cases and PR2-PR3 distances. In the cold state, I expect that the ITMX direct reflected beam sizes should be on the cyan line. A naive interpretation at this stage is therefore either that the PR2-PR3 distance is longer than designed, or the ITMX on-thermal lens is weaker than we thought, but there are many variables which still need to be accounted for.

Non-image files attached to this report
Comments related to this report
paul.fulda@LIGO.ORG - 22:45, Monday 03 February 2014 (9775)

I did a little more analysis on the nanoscan data, and saw that the beam position data and beam radius data appeared correlated. This might be expected if the beam swings across the aperture during a scan, thus "smearing" the intensity over a wider distribution. In this case I would expect to see a 90deg phase lag between position data and beam size data, although that might be neglecting some specifics of how the slit is scanned across the PD. The two attached screenshots show the Gaussian fits for the PRM direct reflected beam and ITMX direct reflected beam. I also attached time series of the y-position and radius (which shows the correlation quite clearly).

Finally I did some frequency analysis on the data, as shown in the attached ASDs.

In pitch there seem to be two main peaks, at 0.717Hz and 0.95Hz. Another peak appears at 1.167 Hz. The closest suspension pitch mode I can find to the 0.717Hz peak is 0.745Hz pitch mode from the HLTS. HSTS pitch mode at 0.6725Hz is also in the ballpark. The 0.95Hz peak is quite close to the HLTS yaw mode at 0.99Hz, and 1.167Hz mode is close-ish to the HSTS yaw mode at 1.09Hz, but I don't see any pitch modes matching up there.

In yaw there seem to be two main peaks. One again at 0.95Hz, and one at 2.267Hz. The 2.267Hz peak matches up pretty closely with the 2.253Hz BS yaw mode. I didn't see any Quad modes matching the peaks, so it seems that ITMX is not causing the motion.