Jonathan, Satya, Jim, TJ, Dave:
This morning we installed the new x509 certificate on h1fescript0 to replace the expired cert. We were then able to restart the ext_alert code on h1fescript0. This runs under monit management.
I have created an external alert overview medm screen, it is called from the SITEMAP via the SYS->'GRB/SN External Alert' pull-down. I pulled out the ext_alert parts from the CAL overview screen to make the screen (screenshot attached).
TJ verified that the audible alarms at the operator station are announcing these alerts.
Updated Virgo logbook links at top of page per FRS# 6465.
Measurements of the pre-modecleaner high voltage monitor with the pre-modecleaner locked and unlocked (see plot legend).
Per FAMIS #6868, ran & checked ISI CPS signals for the BSCs & HAMs. On both measurements (attached), no high frequency noise was observed.
Lost lock 3 times briefly after reaching NLN. Possibly ASC related. No PI modes seen ringing up on striptool. Decided to try an initial alignment and have not been able to get past ENGAGE_DRMI_ASC since. At different times both H1:LSC-TR_X_QPD_B_SUM_OFFSET and H1:LSC-TR_Y_QPD_B_SUM_OFFSET got bad values that hindered finding and fine tuning IR. 07:18 UTC Terra turning on ETMY ring heater for PI test later today (Thursday) 09:15 UTC NLN 09:18 UTC Lock loss. EY saturated and ASC signals (mostly DHARD pitch?) started ringing up before lock loss. 10:02 UTC NLN 10:06 UTC Lock loss. Again ASC signals started ringing up before lock loss. 10:48 UTC NLN 10:52 UTC Lock loss. Again ASC signals started ringing up before lock loss. 10:56 UTC Starting initial alignment. 11:56 UTC Initial alignment done. Kept losing IMC lock. Had to redo input align. SRC also had trouble locking. 12:22 UTC Despite initial alignment had to check MICH fringes and lock on PRMI. Locking on PRMI took moving PRM around. PRMI to DRMI transition worked, but lost lock before DRMI_LOCK state. 12:36 UTC Stopped at DRMI_ASC. Signals fell and loss lock again. Happened again. Stopping at LOCK_DRMI_1F. Seems stable here. Definitely ASC. 13:54 UTC Still can't get past DRMI_ASC. Starting another initial alignment. 14:09 UTC Peter to LVEA to take dark noise measurements and needs PMC unlocked. Stopping initial alignment after INPUT_ALIGN_OFFLOADED and set ISC_LOCK to DOWN. 14:57 UTC Peter done
Jim W, Kiwamu,
As pointed out by Patrick, DRMI ASC was causing the locklosses. It was due to the PRC1 loops (POP A QPD -> PRM loop) which had too large misalignment. Here is what we did to work around it.
Though, we don't know why we ran into this situation at this point.
Unless otherwise noted all powers measured with the water-cooled 300 W Ophir power meter. The ISS was
unlocked but still diffracting light, whilst this does distort the beam it should be constant over the
time scale of the measurements taken.
Using the locking photodiode over a 5 minute observation time:
locked: -260 mV to -262 mV
unlocked: -1.001 V to -1.004 V
Calculated visibility using the locking photodiode is thus: (74.0 +/- 0.4)%
Using the power meter:
locked: 30.5 W to 30.6 W
unlocked: 121.6 W to 122.2 W
Calculated visibility using the power meter is (74.9 +/- 0.3)%
The power transmitted by the pre-modecleaner was 105.7 W to 106.1 W. Which suggests a pre-modecleaner
cavity transmission of (86.9 +/- 0.2)%.
The errors listed above are okay to the precision of the measurement and not its accuracy.
Bearing in mind that the accuracy of the power meter is +/- 3%, the visibility measurement and
transmission measurement are on the edge of agreement.
Taking accuracy into consideration the visibility is then (75 +/- 4)% and the transmission is
(87 +/- 5)%. The two measurements are in general agreement but are not very accurate nor precise.
Turning the ETMY ring heater on 0 --> 0.55 W top, 0 --> 0.55 W bottom at 7:25 UTC.
This is to revert alog 29702, bringing us back to a stronger optical mode overlap with the 47kHz ETMY mechanical mode so we can test the effectiveness of ESD damping at 50 W.
TITLE: 10/19 Eve Shift: 23:00-07:00 UTC (16:00-00:00 PST), all times posted in UTC
STATE of H1: Commissioning
INCOMING OPERATOR: Patrick
SHIFT SUMMARY: Struggled to lock a few times but eventually made it to NLN. Noise hunting continues on the commissioning side.
Sheila Kiwamu Jenne Daniel
We measured the OLG of the PMC loop with the interferometer unlocked, and saw that it is aroun 500Hz while it is supposed to be at around 5kHz. Plotting the measured OLG as closed loop supression, we predict that this loop should have gain peaking approximately around the frequency of our lump in DARM. (second attachment). We tried increaseing and decreasing the gain by 6dB, and didn't see much of a change in DARM.
However, a driven measurement using the newly amplified HV mon as a readback predicts that this noise is about a factor of 2 below DARM in our lump. The third and fourth attachments are the same noise injections that Jenne and I posted on monday for MICH SRCL and PZT jitter, with projections based on PMC PZT HV mon. Kiwamu found an alog from april, 26538 indicating that the gain should be set to 30 dB (it has been 16 dB for the last several months). The third attachment shows the noise projection with the PMC gain at 16dB, while the 4th one shows 30dB. The 5th screenshot shows the difference in the DARM spectrum with the increased gain.
People are still invesitgating the coupling mechanism, we think that intensity noise (which we think was the explanation for the simliar noise at LLO in 2014 16186) is ruled out by the intensity noise injection, although it is interesting to note that the spectrum of this PMC HV lines up fairly well with the ISS control signal.
According to various signals when the PMC HV was excited, we are concluding that this is not a coupling through intensity or frequency of the light. We don't know how the HV noise couples to DARM.
In the attached screen shot, the right two panels show various signals with and without a broad band excitation in the HV. The PMC control gain was at 30 dB throughout the measurements. The upper right panel shows an increase in the PDA spectrum (which has been used as the sensor for the inner loop), indicating that the ISS witnesses increase in the HV noise somehow. However, the second loop sensors don't really show increase in their noise level below 1 kHz. This means that RIN at OMC DCPD should be at 1e-10 RIN/sqrtHz which is a factor of 10 lower than shot noise because our RIN to RIN coupling from the interferometer input to OMC DCPD is roughly -40 dB. So this does look like an intensity noise coupling.
As for frequency noise, the situation seems similar to intensity. The CARM loop sees higher noise level in frequency according to REFL_CTRL_OUT in the lower right panel. However POP 9I, which is an out-of-loop frequency noise sensor, did not show any elevated noise at all below 1 kHz. Based on the coupling of POP 9I measured the other day (30610), POP 9I should show higher noise level by a factor of a few in order to explain the increased noise level in DARM by frequency noise. So it does not seem to be frequency noise coupling either.
I compared nominal PMC locking gain (red, green) and 6dB lower (blue, brown).
Due to gain peaking at 240Hz the feedback signal doesn't decrease below 400Hz.
Anyway, higher than 400Hz, I see some reduction in DARM when the gain was lower, but it seems as if the reduction was mainly at around the peak of the three bumps (440, 580 and 700 Hz). For example it seems as if there's no reduction of noise at 520Hz even though the feedback signal to PMC PZT was reduced by a factor of 4.
We can also see that DARM got worse at 240Hz by reducing the gain due to gain peaking.
Changing the PMC filter to allow us to lock at much lower UGF would help but the featureless bump might stay. The second plot is the same as the first one but the PMC PZT (dashed) is put on top of the DCPD (solid). PMC PZT is arbitrarily scaled so that the DCPD with high bandwidth PMC lock (red, green) looks like the SUM of DCPD with low bandwidth PMC lock (blue, brown) and PZT with high bandwidth PMC lock(pink, orange).
Last night the water level was at 7.7cm so I didn't add any water. Today I went out and it fell to 6.9 cm. That's less than 1 cm! Will continue to keep an eye on it.
1. Noise
When DBB HPO shutter was closed, DARM noise decreased. I did open/close/open/close and this was quite repeatable (first attachment, red/green are shutter closed, blue/brown are open). Some of the jitter-ish peaks subsided and the DARM looked much smoother when the shutter was closed.
These days DBB shutter is open most of the time (it seems to have been in local mode most of the time for jitter FF). Now we're in half-working remote mode where the MEDM screen says "local" all the time but we can close/open the shutter by pressing buttons.
This should be some kind of scattering problem, e.g. optical feedback into HPO and/or even to the frontend.
Inside DBB the reflection from DBBPMC is received by photo diodes, there's one transmission received by CCD camera but the DBBPMC was not resonant. There's a thing called "TFP low power attenuator" that seem to attenuate the power going into DBBPMC, which is between DBB breadboard and the DBB HPO shutter.
Anyway, I wonder how many other significant scattering sources are there on the PSL table. There are other TFP low power attenuators as well as thermal type power meters which might receive some non-negligible power. I don't know which one is essential, but non-essential ones I'd like to temporarily block using black glasses.
2. Power
Stupid thing about this excercize was that the PSL power increased by about 3W when closing the DBB shutter (second attachment, left bottom is PSL power, left mid is the light the DBB REFL diode receives).
Even if you think the scatter causes some problem to the intensity, it should be taken care of by ISS.
It turns out that this doesn't have anything to do with the scatter, it's probably shoddy electronics (design or implementation).
The 1st loop PD actually changes even though the 1st loop is DC-coupled. REF signal is not changing. TRANSFER1_A is just the DCPD scaled and reference voltage added, i.e. PDA_CALI_DC/5+REF.
The "error signal" of the 1st loop is actually TRANSFER1_B (CH10), which is the sum of TRANSFER1_A and TRANSFER2_B.
TRANSFER2_B is the second loop output measured by the 1st loop board, and this should be proportional to the secondloop output measured by the second loop board (CH13). But the second loop was AC-coupled during this measurement.
If you look at all these signals, it's clear that the analog signal downstream of the second loop AC coupling point is pulled by a tiny amount by opening/closing of the DBB shutter (because of the current the driver has to supply?), amplified by the boost stage of the second loop, injected into the first loop, and the first loop dutifully responded by changing the power.
This is not a huge deal as far as we always close the shutter but it's disappointing.
Added 150ml to the crystal chiller.
Diode chiller didn't complain. No water added.
The CW injections have been restarted remotely (after checking with the control room). I discovered that because of a sign convention change in actuation functions between S6 and O1, the CW injections with explicit actuation correction were being injected with a default sign flip. That sign flip has now been removed. In addition, the amplitudes of pulsar injections 0, 1, 2, 4, 7, 9 and 14 have now been doubled over what they were in O1. (This amplitude change was supposed to take effect several months ago, but I had neglected to change a symlink to make it so.) In detail, the above amplitude changes were accomplished by changing the symlink RELEASE pointer in the hinj account from O2test to O2_H1_test2 (it was supposed to be O2_H1_test1 from early summer until now.) The removal of the sign flip was accomplished by adding an argument --actuationScale=1.0 to the call to lalapps_Makefakedata_v4 for each of the 15 injections. Previously, the default value for actuationScale was -1.0, a vestige of iLIGO sign conventions.
Correction: 6 of the 7 injection amplitudes listed above were indeed doubled, but the amplitude of the highest-frequency signal (pulsar 14 at 1991 Hz) was tripled. The attached spectra of the excitation channel from last night and tonight, 1 sidereal day apart, show the increases of amplitude for all injections above 200 Hz.
I tried for a while today to go back toward our July 16th alignments, as suggested by Sheila in alog 30648. I put offsets in POP_A and the SOFT loops to try to make the witnesses/oplevs go back toward their July16th values.
I only made the DARM spectrum worse, not better. In the attached screenshot, green is the spectrum before I started doing anything, pink is after my alignment work, and red is after taking the offsets out again.
I also attach a screenshot showing that the buildups and the recycling gain all deteriorated while trying to go back to July's alignment. To be fair, I was not able to get all optics' witnesses back to their July values, so it's still not really the same alignment, but I think it's closer than what we usually run with. I don't really think that this is a fruitful direction to continue to persue unfortunately, especially in light of what has just been discovered about the noise improvement when we close the DBB.
https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=30631 Satisfies FAMIS asset 364 maintenance
....just when I thought I understood FAMIS. I still need to "create request" to create a task associated with this asset created which already includes a schedule, procedure, and personnel assignment.