jim.warner@LIGO.ORG - posted 10:27, Friday 03 October 2014 (14288)
Re: ITMX troubles from yesterday

Yesterday, Fabrice and I transferred the Livingston blend filters from the ITMY ISI to all the other BSC's. When I went to turn on ITMX, the ISI starting ringing up dramatically when I tried to engage the horizontal loops. Because we had to transfer the filters by hand, I immediately suspected I had screwed up a filter somewhere. I compared the blends between ITMY and ITMX, and found no difference. I did a diff between the a couple of archived foton files and the installed filters and found no differences beyond what I had to do to install the filters (i.e. no changes in any filters other than the blend filters I installed). When I started a rolling DTT spectra of the isolation loops (the ISO outs), I found that there was an instability in the Y isolation loop at 730 hz. When I looked at the filters I designedin matlab, I saw that there was indeed feature there, but I had sufficiently suppressed the feature. When I took an open loop measurement of the plant, however, the feature had evolved and was now a magnitued of 3 higher. I also found, looking at the foton digital filter, that there was again mismatch at 730 hz of about a factor of 5. At the time Fabrice and I decided that there must have been a change in the plant, added a notch and moved on.

This morning with, Jeff's help, we've found a few more interesting details:

        1. the foton filter does not agree with my matlab design (foton is in the first png, matlab design is my second figure). Up to about 400 hz, they agree well enough, but I have 2 high frequency notches that are not in the same place in foton. The notch I put at 730 hz has been shifted down to more like 660 hz.

         2. Looking more closely at the plant measurement I took on the 16th and the DTT measurement yesterday, we have less gain margin at high frequency than we thought we did. The comparison is show in my third plot(pdf), provided by JeffK. The green line is the measurement used in matlab, blue is the DTT measurement from yesterday. The difference is explained by a downsampling filter between the DQ channels we used to measure matlab tfs and the higher data rate channels we measured in DTT, shown on the bottom of Jeff's plot.

Images attached to this report
Non-image files attached to this report
H1 PSL (PSL)
peter.king@LIGO.ORG - posted 09:34, Friday 03 October 2014 (14287)
laser shutter closed
Whilst inserting/removing the second loop power stabilisation servo card, the laser shutter closed with
EPICSALARM highlighted on the PSL Beckhoff computer.  I reset the shutter and flow sensors from the
control room and opened the shutter (from the diode room).  Laser and servos came back by the time I
made it back to the control room.
LHO General
patrick.thomas@LIGO.ORG - posted 09:01, Friday 03 October 2014 (14286)
Morning meeting notes
SEI: TBD

SUS:
optical lever calibration
DRMI commissioning

Test stand:
review transfer functions
TMS builds

PSL:
modifications to 2nd loop ISS board

Commissioning:
DRMI
AS WFS
IR arm locking

CDS:
Running fibers
TCS X arm rotation stage
H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 08:26, Friday 03 October 2014 (14285)
CDS model and DAQ restart report, Thursday 2nd October 2014

model restarts logged for Thu 02/Oct/2014
2014_10_02 09:29 h1isibs
2014_10_02 09:33 h1hpiitmx
2014_10_02 09:33 h1hpiitmy
2014_10_02 09:35 h1isiitmy
2014_10_02 09:36 h1isiitmx
2014_10_02 09:42 h1hpibs
2014_10_02 09:44 h1isibs
2014_10_02 09:46 h1hpibs
2014_10_02 10:03 h1hpiham6
2014_10_02 10:03 h1isiham6
2014_10_02 10:10 h1hpiham1
2014_10_02 10:14 h1hpiham5
2014_10_02 10:14 h1isiham5
2014_10_02 10:16 h1hpiham2
2014_10_02 10:18 h1isiham2
2014_10_02 10:20 h1hpiham4
2014_10_02 10:20 h1isiham4
2014_10_02 10:27 h1isiham3
2014_10_02 10:29 h1hpiham3

2014_10_02 10:43 h1dc0
2014_10_02 10:45 h1broadcast0
2014_10_02 10:45 h1fw0
2014_10_02 10:45 h1fw1
2014_10_02 10:45 h1nds0
2014_10_02 10:45 h1nds1

2014_10_02 10:48 h1hpiham1
2014_10_02 11:53 h1ioppemmy
2014_10_02 11:53 h1pemmy

2014_10_02 11:53 h1sushtts
2014_10_02 16:38 h1pslfss
2014_10_02 16:55 h1psliss

2014_10_02 16:59 h1broadcast0
2014_10_02 16:59 h1dc0
2014_10_02 16:59 h1fw0
2014_10_02 16:59 h1fw1
2014_10_02 16:59 h1nds0
2014_10_02 16:59 h1nds1

2014_10_02 17:02 h1psliss

no unexpected restarts. Seismic and PSL model changes, with associated DAQ restarts. Recovery of h1pemmy after IO Chassis power supply swap.

H1 SEI
fabrice.matichard@LIGO.ORG - posted 19:50, Thursday 02 October 2014 (14284)
SEI Status

SEI Status:

- Motion measured on ITMY (ISI and optical lever) using the LLO configuration was good again last night, so we duplicated this configuration to all the BSC-ISI units

- As we re-started the units, Jim found an oscilation in ITMX. He fixed it. It is not correlated with the new blend configuration. More details on that to follow.

- Hugh worked on the Guardian. It is now fully updated to account for the new config (decribed below). Thanks a lot to Jamie for the remote support. The Guardian is now fully updated and running.

- All the units are up and running. We leave the sensor correction OFF for the night.

 

Configuration in use:

- LLO blends installed everywhere (excact same filters copied/pasted from the filter text file)

- Stage 1 RZ off on all units

- Only Stage 1 X and Y isolated on Stage 2 (all units except the BS)

- Stage 2 damped only on the BS

- Sensor correction is off for the night

 

 

H1 ISC
sheila.dwyer@LIGO.ORG - posted 19:19, Thursday 02 October 2014 (14283)
work today

Other than this we have been disrupted by ITMX this evening. 

Images attached to this report
H1 SUS
sheila.dwyer@LIGO.ORG - posted 17:12, Thursday 02 October 2014 (14282)
changes to h1sushtts model

Jenne, Alexa, Sheila

Since the asc had IPC senders for signals to OM1 and OM2, but the only revciever in the h1sushtts model were from the OMC model, we added new recievers.  Anamaria told us where to find the livingston model, and confirmed that it was up to date, so we copied what they had done there, just summing the inputs from the OMC model and ASC. 

Complied, installed, seems to be working, and we can now use the DC signals from our AS WFS for a centering servo.

H1 PSL
gabriele.vajente@LIGO.ORG - posted 17:10, Thursday 02 October 2014 (14281)
ISS second loop closed

Peter K, Gabriele

In brief, we finally managed to close the ISS second loop. The attached plot compares the intensity noise measured on the second loop photodiodes in different configurations: second loop open (green and maroon), loop closed withe the additional integrator (red and blue), loop closed with the additional boost (cyan and magenta).

Clearly, the boost doesn't do a very good job. With the integrator on, we get a loop bandwidth of the order of 200 Hz, and a gain at 10 Hz of something like 10-15.  We don't have enough gain to cope with our present large power noise after the IMC.

Procedure to engage the second ISS loop

  1. All switches should be low (H1:PSL-ISS_SECONDLOOP_CLOSED, H1:PSL-ISS_SECONDLOOP_SERVO_ON, H1:PSL-ISS_SECONDLOOP_BOOST_ON, H1:PSL-ISS_SECONDLOOP_INT_ON)
  2. Move the second loop gain to the minimum (H1:PSL-ISS_SECONDLOOP_GAIN to 0)
  3. Switch on the second loop (H1:PSL-ISS_SECONDLOOP_SERVO_ON to high)
  4. Adjust the reference input H1:PSL-ISS_SECONDLOOP_REF_SIGNAL_ANA to move the output of the second loop as close as possible to zero (check H1:PSL-ISS_SECONDLOOP_SIGNAL_OUTMON)
  5. Switch on the additive path in the ISS inner loop (H1:PSL-ISS_SECONDLOOP_CLOSED to high)
  6. Increase the gain to maximum (H1:PSL-ISS_SECONDLOOP_GAIN to 40)
  7. Engage the integrator (H1:PSL-ISS_SECONDLOOP_INT_ON to high)

Some details

The input signals to the second loop servo board (the two sums of photodiodes) were not behaving correclty: they appeared to be choppen at the zero level, meaning that only the positive (or negative) part of the signal was present. We traced dow the problem to the OP27 which is used in the output stage of the transimpedance box. This OP27 was used to create a single ended monitoring of the box output, which was sent to the front panel and not used. For some reason we don't understand yet, this opamp was creating the problem. We removed it from the two boards and the sum output is now good.

We also modified the ISS model to wire in the two additional switches ADD_GAIN and ADD_PD_58_SUM. We plan to modify  the additional gain module to get an additional factor of ten in the gain.

Images attached to this report
LHO VE
kyle.ryan@LIGO.ORG - posted 16:32, Thursday 02 October 2014 (14280)
Restarted IP5 with help of pump cart (has been off since July(?)) -> Valved-in IP5


			
			
H1 General
travis.sadecki@LIGO.ORG - posted 16:27, Thursday 02 October 2014 (14279)
Ops shift summary

8:45 SEI bootfest of all CS ISI and HEPI

9:04 Aaron to EY to retreive AA chassis

9:19 Jodi to MY

9:22 Betsy and Corey to LVEA on 3IFO cable hunt

9:29 Jodi back from MY molesting tiltmeter

9:55 Peter King and Gabriele to PSL looking at ISS

9:55 McCarthy to LVEA looking for tools

10:05 McCarthy out of LVEA

10:24 Jeff and Andres to EX

10:34 JeffK to LVEA testing seismometers

10:43 Framebuilder restarted

10:48 Betsy, Kyle, Bubba to LVEA crane fest

10:50 Cris to EX

11:11 Karen to EY

12:20 Craning in LVEA complete

14:02 Jason to realign ITMy OpLev

14:28 Jason out of LVEA

15:00 Rick and Sudarshan to EX to mount PCal camera housing

16:20 Robert to BTE between MY and EY

H1 AOS
fabrice.matichard@LIGO.ORG - posted 15:51, Thursday 02 October 2014 (14278)
ITMX Pitch excess motion

Commisioners have reported excess pitch motion in ITMX for the past few nights. The plot attached shows the Optical lever motion (Top left is ASD, bottom left is RMS) , and the ground motion (Top right is ASD, bottom right is RMS). Measurement are taken every morning at 1am on Sunday (Red), Monday (Blue), Tuesday (Green), Wednesday (Brown) and Thursday (Magenta).

 

The numbers below show that the pitch motion has varied by as much as a factor of 5 from one night to another, but it is not necessarily directly correlated with the amplitude of the ground motion (worst pitch motion was yesterday though the mircro-seism was not higher than during the previous nights).

 

- Optical Lever RMS motion at 100 mHz, Sunday through Thursday at 1 am PT:

18 nRad / 44nRad / 63nRad / 42nRad / 100 Rad

 

- Ground instruments RMS motion at 100 mHz in nm/s (must check the units, but it's fair to compare the day to day relative motion), Sunday through Thursday at 1 am PT:

113  / 106 /  197 / 189 / 140

 

- Ratio in nRad/(nm/s):

0.16   0.41   0.32  0.22   0.71

Images attached to this report
H1 AOS
filiberto.clara@LIGO.ORG - posted 12:15, Thursday 02 October 2014 (14276)
H1 PEM MId Y IO Chassis
Power supply for the PEM IO chassis was replaced. Unit powered up with no issues. Front end computer was also powered on.  
H1 SEI (CDS, DetChar, PEM, SYS)
jeffrey.kissel@LIGO.ORG - posted 11:49, Thursday 02 October 2014 - last comment - 14:22, Thursday 02 October 2014(14273)
All Corner Station HPI and ISIs Successfully Readout All GND STSs
J. Kissel, J. Warner, H. Radkins

This completes the changes needed at LHO for the GND STS Kerfuffle, Integration Issue 942, and completes activity for Work Permit 4876. We can't yet close ECR E1400386, or the integration issue, because Ben needs to update the system wiring diagrams, but those should be very coming soon.

After moving all STS distribution chassis outputs into the now-pre-defined AA chassis spigots (see drawing in E1400386 for reference, for now, until the drawings get updated), then completing all model changes to include all three corner-station, sensor-correction, STSs in every HPI and ISI last night (see LHO aLOG 14264), Hugh, Jim, and I took on the task of installing the new code. We

- used guardian to ramp down each chamber, taking the manager to OFFLINE, where applicable, and where not, bringing each ISI and HPI to the READY state and turning OFF the master switch.
- turned OFF all HPI and ISI sensor correction filters by zeroing the MATCH filter bank gain and turning OFF the output switch.
- captured a new safe.snap, using
      ]$ makeSafeBackup ${subsystem} ${modelName}
  and committed to the appropriate "burtfiles" directory in the userapps repo
- *** Made sure all filter files are softlinks to the userapps repo, and repo versions were up-to-date and committed
      -- we found that the majority of the HPI and ISI filter files (both .txt and .fir) were NOT softlinks to the userapps repo. To fix, we
          - Copied any non-softlink files to the appropriate filterfiles directory in the userapps repo
          - Committed the files
          - removed the files from the chans directory
          - created the softlink in the chans directory to the userapps repo, e.g.
              ]$ cd /opt/rtcds/lho/h1/chans/
              ]$ ln -s /opt/rtcds/userapps/release/${subsystem}/h1/filterfiles/${filterfilename} ./${filterfilename}
- re-compiled and installed the new model code
- restarted the front-end code
- untripped all watchdogs
- cleared any IPC errors on HPI/ISI/SUS by hitting the "diag reset" button on everyone's GDS_TP screen.
- used guardian manager to bring up each chamber, takingthe manager to ISOLATED, where applicable, and where not HPI to ROBUST_ISOLATED and ISI to HIGH_ISOLATED
- Restarted the frame builder / data concentrator
    ]$ telnet h1dc0 8087
    daqd> shutdown
- Confirmed that every corner station HPI and ISI reads out the correct GND STS on the correct channel using dataviewer and me zydeco dancing in proximity to each sensor successively.


Note that *** may need some attention when loading in new filters. It's unclear yet what quack (the matlab method of installing new filters that really only the SEI team uses) will do when loading filters. It currently installs new filters into the /opt/rtcds/${site}/${ifo}/chans/ directory. Will it over write the soft-link? Will it follow the softlink? I think this has been checked that it won't break anything, but it's something we should pay attention to in the future. It's especially interesting to see if the .fir filters can handle it.

Laissez les Bon Temps Roulet!
Comments related to this report
keith.thorne@LIGO.ORG - 14:22, Thursday 02 October 2014 (14277)CDS
You can also use the semi-automated archive at https://daqsvn.ligo-la.caltech.edu/websvn/ that posts date-time-stamped changes to all filter files when loaded by front-ends.  
H1 SEI
hugh.radkins@LIGO.ORG - posted 11:09, Thursday 02 October 2014 (14275)
Guardian Restart Problem -- Connect Error

Issue likely applies to other subsystems but SEI has a hierarchical guardian and this may be more a problem for these types of guardians.

Before restarting the front ends, the guardian was used to bring things down by moving the request from Isolated to offline. When the BS (and all other) guardian (Chamber Manager) was turned from OFFLINE to ISOLATED after the frontend restart, we had to RELOAD the guardian.  This was expected and maybe normal.

However on HAM4, the chamber manager reload was not sufficient.  The Chamber Manager reported "ERROR in state INIT; see log for more info; MODE=>RELOAD to reset"

The log file reported "NameError: name 'Connect' is not defined"

Until I reloaded the ISI and HPI subordinate guardians (can't tell if just one or actually both needed to be reloaded), reloading the Chamber guardian was insufficient.

After reloading the subordiantes, reloading the Manager brought things back as usual.

H1 SEI
jim.warner@LIGO.ORG - posted 10:32, Thursday 02 October 2014 (14271)
HAM5 ISI watchdog weirdness

Jeff enlisted Hugh and I to help with a mass snap, model compile, model restart -fest this morning in the corner station for all SEI platforms. The BSC's went fairly smoothly, but when I did HAM5, the watchdogs behaved strangely. I couldn't reset the DACkill or Rogue Exc WD's, but when I randomly tried turning on the master switch, I was then able to reset all WD's. The first couple tries, this immediately resulted in DACkill trips, but the 4th one stayed up. The chamber is now isolating again, but it sound like this should be investigated.

H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 08:31, Thursday 02 October 2014 (14270)
CDS model and DAQ restart report, Wednesday 1st October 2014

model restarts logged for Wed 01/Oct/2014
2014_10_01 10:33 h1broadcast0
2014_10_01 10:33 h1dc0
2014_10_01 10:33 h1fw0
2014_10_01 10:33 h1fw1
2014_10_01 10:33 h1nds0
2014_10_01 10:33 h1nds1

2014_10_01 10:33 h1psliss
2014_10_01 16:13 h1ioppemmy
2014_10_01 16:13 h1pemmy
2014_10_01 16:35 h1ioppemmy
2014_10_01 16:35 h1pemmy

2014_10_01 19:33 h1fw1
2014_10_01 23:13 h1fw1

unexpected restarts of h1fw1. New psliss model with associated DAQ restart. Recovery of h1pemmy after IO Chassis problem.

H1 ISC
jenne.driggers@LIGO.ORG - posted 00:26, Thursday 02 October 2014 - last comment - 01:01, Thursday 02 October 2014(14268)
DRMI work tonight
Alexa, Kiwamu, Sheila, Jenne

This is sheila, accidentally logged in as Jenne

We made some attempts tonight to transition DRMI to 3F.  We see that we don't have much signal in the 3F detectors at 1 Watt, so we have been transitioning manually to 10 Watts after lock is acquired.  We have transitioned PRCL successfully several times, with a gain of 1.5 in the input matrix, this is handled by guardian.  We have tried srcl, and used 60% of the 3F with 40% of the 1F.  Our next steps are to get some asc loops running, so that we don't spend as much time manually aligning DRMI, and also to try transitioning mich after prcl.  

The rotation stage has a lot of hysteresis, we will ultimately want some sort of servo to make sure that the input power is what we requested.

Kiwamu has measured the modulation depth: yesterday AS RF 90 was 33 counts, today (with the 19 dB RF amplifier and 6dB attenuator installed) it is 270.  This was consistent with the OMC scan which were the height of the 45 MHz sideband with DRMI locked was 0.3, today it was 3.  (These both mean that the modulation depth increased by roughly a factor of 3).  The gains in the guardian have been adjusted to take this into account.  

The good news is that we have locked DRMI many times tonight, it normally locks within a few minutes.  Also, the attached screen shot shows that DRMI was locked on 1F for about 10 hours last night.  
Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 01:01, Thursday 02 October 2014 (14269)

A fortuitus lock as we were about to leave let us transition to 3F (at 1 watt input power) , which was stable for about 15 minutes before it dropped (we were aligning) at 7:57 UTC october 2nd.  For srcl we used -2 in refl 135 I, for prcl we used 1.5 in refl 27 I, and for mich we used -0.43 in refl 135 Q.  ( we transitioned prcl, then mich, then srcl).  the rms at the coil output for srm m2 is 3*10^4 counts. 

H1 PEM (PEM)
sudarshan.karki@LIGO.ORG - posted 14:40, Wednesday 01 October 2014 - last comment - 10:58, Thursday 02 October 2014(14250)
EY PEM AA Chassis

Dave, Sudarshan, Richard

LIGOCAM reported that three microphone at EY were disconnected.

I found last week that these microphone would work as normal when switched from their original channel position to some other channel on the AA chassis so, we restarted the  H1IOPISCEY yesterday to see if the problem was in the DAC. This was an unsuccessful attempt as mentioned in alog 14225.

Today, I injected 3Vpp signal on all the channels using a function generator and found out the following:

Channel 1-16 are working as normal (I happened to switch those microphones to one of these channels during my first attempt).

Channel 17-30 show nothing on the output end of medm screen monitor.

 

Summary: The problem is with the AA chassis board. Richard has been notified and this AA Chassis will be switched sometime next week.

          

 

Comments related to this report
aaron.sevigny@LIGO.ORG - 10:58, Thursday 02 October 2014 (14272)
Internal cables of the AA chassis were unplugged.  They have been plugged in and the chassis has been reinstalled in working condition.
H1 SUS
betsy.weaver@LIGO.ORG - posted 14:24, Wednesday 01 October 2014 - last comment - 14:43, Friday 03 October 2014(14247)
QUAD comparison plot for interested parties

For those interested in looking closer at QUAD model parameters, attached are plots comparing all of the QUAD Main Chains when suspended with wires and also when suspended with fibers.   Note, if QUAD data is missing for one of these configurations it's because there was no clean data available.  Between the 2 plotted configurations, all 12 (H1, L1, and 3IFO) QUADs are represented.  Note, I tried to chose data sets that had the same or similar environmental conditions, but it was difficult due to the fact that some QUADs were reworked on test stands and some were reworked in chamber.  In all cases they were mounted on Solid Stack Test Stands or Locked ISIs and in-air.

 

Data is committed to the svn and can be found at:

/ligo/svncommon/SusSVN/sus/trunk/QUAD/Common/Data/

There does not seem to be a pattern in the data of the 2nd pitch mode peak which are clustered by a specific type of suspension (ETM vs ITM, or wire segment hang vs wire loop hang).

Non-image files attached to this report
Comments related to this report
betsy.weaver@LIGO.ORG - 14:30, Wednesday 01 October 2014 (14248)

And now with some cursors and in a second format for Brett.

Non-image files attached to this comment
betsy.weaver@LIGO.ORG - 15:31, Wednesday 01 October 2014 (14254)

As suggested, I looked at the stiffnesses of the Top Mass blades to see if there is a correlation with the second pitch mode frequency shifts.  I don't see it.  In order of the peaks on the P to P plot, starting with the lowest frequency to the highest the blade sets used in each QUAD are:

H1ETMx - SET 9  (~1.28Hz)

L1ETMy - SET 13

L1 ETMx - ?

L1ITMx - SET 14/15

L1ITMy - SET 12

Q8 ETM - SET 8

Q9 ETM - SET 2

Q6 ITM - SET 10 (~1.531 Hz)

 

The blade sets go in order of stiffness from highest to lowest, so SET 2 is stiffer than SET 15.  SET 14/15 is a mixed SET with blades still of adjascent stiffness.

brett.shapiro@LIGO.ORG - 16:27, Wednesday 01 October 2014 (14257)

I took the two wireloop quads that have the highest and lowest 2nd pitch mode frequencies and made a fit to them. These measurements and their respective made-to-fit models are shown in the attached plot. QUAD06 (H1 QUADTST) is the highest,  X1 ETMX is the lowest.

I previously did a fit for QUAD06, see log 14235. The fit for X1 ETMX was made simply by taking the QUAD06 fit and subtracting 3 mm from dn, which works quite well.

Since the outliers are 3 mm apart on dn, the other quads seem to have an even spread between those, and no correlation with spring stffness is evident, then a possible explanation is that our tolerance on positioning the top mass blade tip height is +-1.5 mm.

Non-image files attached to this comment
brett.shapiro@LIGO.ORG - 11:14, Thursday 02 October 2014 (14274)

Attached is a prediction of what +-1.5 mm on dn would look like for the fiber quads.

The black is a model of H1ETMY (which has been the default fiber model for some time) where dn=1.78 mm; blue is the same model but with dn=0 mm; red is again the same model but with dn = 3 mm. Some data is included as well. The H1ETMY measurement is in orange, which matches well because of the previous fitting of H1ETMY. In purple is H1ETMX. I think H1ETMX corresponds to the wireloop quad X1ETMX, which was the low outlier on dn for the wireloop configuration. In that configuration a dn of 0 mm worked quite well to the fit model to the data. Here the same 0 mm dn makes almost as good of a fit. There is not data matching the dn=3mm. +3 mm was found to work well for the high dn outlier wireloop QUAD06, which is not yet a fiber quad.

So it seems that for the existing fiber quads, +-1 mm on dn explains the spread well. However, the most recent 3rd IFO quads, still with wireloops, are the stiffest yet in pitch, so they would be expected to bring this to +-1.5 mm and line up with the dn=3mm red curve.

Non-image files attached to this comment
brett.shapiro@LIGO.ORG - 14:43, Friday 03 October 2014 (14290)

Posting some notes from recent email converstions looking into the large apparent shifts in dn (top mass blade tip height) and d2 in the all metal build (PUM wire loop prism).

Attachement PUMCOMDetails.pdf is from Eddie Sanchez and is a drawing showing that the position of the PUM wire loop break off in the all metal build is basically the same as where it should be in the final fiber build. However, the model fitting suggests the actual break off is about 1.8 mm lower. So Betsy took some photos of this prism on a suspended metal quad. See image files 1445.jpg to 1447.jpg. Since the prism is round, it could be the wire does not have a clean break off. The pictures seem to indicate the wire has a significant length of a line contact. The 1.8 mm shift could be within this line contact.

The last image, 1449.jpg, shows a picture of the top mass blade spring tip in a suspended top mass. The spring looks pretty well centered, not consistent at all with +3 mm of apparent shift in dn for this quad. Quoting some numbers from Betsy:

"The top surface of the blade, as close to the tip as possible, is supposed to be at 9.6mm down from the top of the bridge notch.  The notch is 14.6mm wide, the blade is 5mm wide, therefore the bottom of the blade should line up with the bottom notch.  No gauge blocks needed.  From the picture, this looks very close to lining up."

PUMCOMDetails.pdf
Images attached to this comment
Non-image files attached to this comment
Displaying reports 63341-63360 of 77237.Go to page Start 3164 3165 3166 3167 3168 3169 3170 3171 3172 End