Reports until 14:22, Thursday 15 May 2014
H1 SUS
arnaud.pele@LIGO.ORG - posted 14:22, Thursday 15 May 2014 (11915)
TMSs in vacuum testing for acceptance review

Last week I took some in chamber, in vacuum acceptance measurements for TMSx and TMSy. ISI tables were both isolated during the measurement.

Attached are the 6 Dofs transfer functions measurements as well as the osems spectra for both suspended tables :

1. TMSX TF compared with the model
2. TMSX spectra for each osem, as well as in the euler basis
3. TMSY TF compared with the model
4. TMSY spectra for each osem, as well as in the euler basis
5. TF comparison of both H1 TMSs and L1 TMSs

To note :

The difference seen in the TF of H1 TMSx is explained in previous alogs
TMSx spectra shows excess 60Hz noise in F1 F2 F3 and LF osems (see 2nd attachment)
Damping for the yaw degree of freedom, for both TMSs is increasing the motion at the resonnance (see last page of 2nd and fourth attachment). I will turn the yaw damping off for both TMSs for now.

Non-image files attached to this report
H1 AOS
jason.oberling@LIGO.ORG - posted 14:11, Thursday 15 May 2014 - last comment - 08:46, Friday 16 May 2014(11917)
SRM Surrogate X, Y, Z Alignment Complete
IAS: J. Oberling
SUS: J. Bartlett
 
Took a first look at the SRM surrogate (SRM-s) today and lo-and-behold, no adjustment was necessary.  cool  The cookie cutters did their job and placed the suspension within spec for X, Y, and Z axis positions.  The position errors are below:

I have begun and mostly completed the setup for the rough pitch/yaw alignment of the SRM-s.  This will be completed tomorrow morning and the rough pitch/yaw alignment will follow.

Comments related to this report
betsy.weaver@LIGO.ORG - 16:29, Thursday 15 May 2014 (11923)

I ran quick V and R TFs and the sus looks healthy to me compared to old references that Arnaud left me.  I need to dig up a good L reference TF and run that in the morning, but then we should be good to go on the pitch/yaw tuning.

betsy.weaver@LIGO.ORG - 08:46, Friday 16 May 2014 (11934)

And L looks good, too!

H1 CDS (DAQ, ISC, SEI)
david.barker@LIGO.ORG - posted 13:30, Thursday 15 May 2014 (11916)
ISI ITMY and ASC model restarts, DAQ restart

Fabrice and Dave:

We successfully made a new h1isiitmy model using the latest isi2stagemaster model changes. We restarted h1isiitmy with the new code

Chris, Sheila, Dave:

I found that I had mistakenly restarted an incorrect h1asc model on Tuesday. After the first DAQ restart that day I noticed that the DAQ picked up a newer INI fle for h1asc and I assumed (which we should never do) that the last "make install" was correct. It appears that later changes were not installed. 

I did a clean "make" and "make install" on h1asc and restarted the model.

Dave:

Finally I did a DAQ restart to support the above changes (remembering to wait into the next 10min period to flush out the second trends)

Here is the restart sequence in detail:

2014_05_15 13:16 h1isiitmy
2014_05_15 13:19 h1asc

2014_05_15 13:22 h1broadcast0
2014_05_15 13:22 h1dc0
2014_05_15 13:22 h1fw0
2014_05_15 13:22 h1fw1
2014_05_15 13:22 h1nds0

H1 General
andres.ramirez@LIGO.ORG - posted 09:13, Thursday 15 May 2014 (11913)
Installation Meeting Minutes
8:15-8:30

Fred, Richard, Cyrus, Filiberto, Aaron, Bubba, Gerardo, Jeff B, Bubba, Travis, Hugh, Mitchell, Justin, Mike, Thomas, Jeff K, Fabrice, Alexa, Steban, Robert, Jamie, etc.


•	SRM alignment on HAM5 - (Jeff B/Jason)
•	TCS tables work continues. – (David, Thomas)
•	Getting things ready to transport the OFI - (Gerardo)
• 	Craning activity in LVEA / Moving clearoom from west bay -(Apollo)
• 	ICS rack inventory in LVEA – (Richard/Filiberto)
•	Commissioning work continues (Commissioners)
•	Testing and assembly operations continue in Staging building –(Jim/Hugh/Scott)
•	Guardian work in progress -(Jamie)


        LVEA is Laser SAFE
H1 ISC
alexan.staley@LIGO.ORG - posted 08:48, Thursday 15 May 2014 (11912)
Fiber Polarization

Richard warned me that the Fiber polarization had probably drifted at EY due to some work done around the table. Indeed it had drifted to up to 33% for the wrong polarization. However, Stefan had increased the threshold to 36. I returned the threshold down to 30 and adjusted the controller. I also had to adjust the EX polarization when I turned the controller on, which forced both polarizations to drift. All is good now.

H1 General
andres.ramirez@LIGO.ORG - posted 08:45, Thursday 15 May 2014 (11911)
Crainig activity in LVEA
Craning activity in the LVEA at the moment. A cleanroom is being moved - Apollo
H1 CDS (DAQ)
david.barker@LIGO.ORG - posted 08:43, Thursday 15 May 2014 (11910)
CDS model and DAQ restart report, Wednesday 14th May 2014

model restarts logged for Wed 14/May/2014
2014_05_14 12:37 h1fw1

unexpected restart of h1fw1

H1 ISC (ISC)
stefan.ballmer@LIGO.ORG - posted 01:50, Thursday 15 May 2014 (11907)
3f work with arms locked
Kiwamu, Stefan,

Both ALS_DIFF and ALS_COMM were straight forward today, and we got them locked literally in minutes. I.e. the ESD is still working fine.

We misaligned the arms, and verified that 3f locking works just fine without the arms.

We locked the arms and parked them 1kHz off resonance. I this configuration the the RF27_I signal worked fine, but the REFL27_Q signal had significant contamination from the COMM noise at 70Hzish (periscope peak). Thus we designed a broad notch filter in the mich loop, which kept the BS from saturating.

However it seems that MICH is still too strongly contaminated by the low frequency COMM noise. As a result the MICH loop seems to drift out of its linear range.

Ideas:
 - divide REFL27_Q by POP_18 to expand the linear range
 - rephrase REFL27 to reduce the coupling

Either way, an earthquake told us to get some sleep.
H1 TCS
david.hosken@LIGO.ORG - posted 22:36, Wednesday 14 May 2014 (11906)
TCS/Hartmann
David H, Greg G, Thomas V

- Corner Hartmann table now has the Y-arm in-air optics mounted and positioned, ready for alignment using the visible laser. Components for the HWS were tracked down and staged.
- TCSX CO2 laser continued to run all day. Once going to laser hazard we realigned the CO2 beam that is transmitted through the output beam splitter. This transmitted beam passes through another 50/50 beam splitter, with the (second) transmitted beam aligned to the output power meter, while the reflected beam was expanded and aligned to the FLIR camera screen position. This required the repositioning of some cables and mounts located near the CO2 laser and RF driver.  

TSCX laser will continue to run overnight. Rotation stage set to 44 degrees (with interlock cable removed to prevent changes to this angle). Table doors are locked. 
TSCY chiller and power supplies continue to run, TCSY laser is turned off.
H1 SUS
keita.kawabe@LIGO.ORG - posted 19:06, Wednesday 14 May 2014 - last comment - 14:58, Thursday 15 May 2014(11905)
ETMX charged up? (Richard, Keita)

Richard showed me that there is a liner coupling from any of the electrodes to the OPLEV even when there's no bias anywhere, and the bias-less coupling is on the same order as the coupling with bias.

I started flipping the bias of the common ("DC") electrode while keeping all four quadrants at zero count output, and indeed there's a linear coupling, and the effect is not subtle.

In the attached, I changed the DC bias from 125000 to 0 to -125000, then to 0 and finally back to +125000. The ETM points up by 0.3urad in the first step, then further up 0.5urad in the second step (instead of going back to the original position). It's 0.8 urad peak to peak!

OTOH, when I set the four quadrants such that top two outputs -50900 counts and the bottom +50900 to tilt the optic legitimately while the DC bias is kept at 125000 counts, I can only move it by -0.5 urad peak to peak (second attachment).

This could still be electronics offset, so I looked at the ESD readback. The readback is not calibrated, but four quadrants always read from -75 to -200 counts when I'm outputing zero, the 300volt-ish thing reads 30000-ish counts, so the voltage offset for the quadrants should be 2V-ish at most. That's not large enough, I think.

Images attached to this report
Comments related to this report
richard.mccarthy@LIGO.ORG - 08:12, Thursday 15 May 2014 (11909)
I disconnected the ESD drive cables between the current limiting resistor and the flange.  Put large drive signals on the cables and did not see a response in the oplev. So it is not a direct coupling in the rack or cables up to the chamber.
brett.shapiro@LIGO.ORG - 09:56, Thursday 15 May 2014 (11914)
At LASTI we saw static charges on the test mass high enough to mimic a bias voltage of 125 V (very large!). John Miller put together a procedure and MATLAB script for measuring what this effective bias was. So the problems you are seeing with a non-symmetric response are within the limits of what has been observed before. 

See the LASTI ilog
http://emvogil-3.mit.edu/ilog/pub/ilog.cgi?group=lasti&task=view&date_to_view=12/18/2009&anchor_to_scroll_to=2009:12:18:16:36:37-brett
u: reader
p: readonly

This script happens to be on the svn at
/ligo/svncommon/SusSVN/sus/trunk/QUAD/Common/MatlabTools/ChargeScript
It will need some modification to work at the sites, since it was written to work in the old LASTI file structure. Basically though, it automatically runs DTT from MATLAB, and measures the response at 11 Hz with a range of bias voltages.

Note,
Rai has put together a gizmo to discharge the test masses with ionized nitrogen. Not sure of its status. Venting to some torr might also help, not sure.
rainer.weiss@LIGO.ORG - 14:58, Thursday 15 May 2014 (11918)
The discharging rig Brett is referring to is described in LIGO T1100332. It currently resides at MIT
under clean conditions. It and the associated electronics can be shipped to LHO. It needs a 2 3/4 inch conflat gatevalve
attached to one of the ten inch blank flanges on the BSC chamber at the charged mirror height. The gas injected into 
the deionizer needs to be clean nitrogen such as the boil off from liquid nitrogen passed through another liquid 
nitrogen trap and warmed to room temperature before injection into the BSC.
RW
H1 SUS (CDS, ISC)
jeffrey.kissel@LIGO.ORG - posted 16:57, Wednesday 14 May 2014 (11903)
H1 SUS ETMX ESD restored, but not yet quantitatively confirmed
J. Kissel, for F. Clara, and R. McCarthy

After investigating further, Richard and Filiberto found that source of failure in the chassis was the mysterious loss of the +18 V power supply to the ESD driver chassis between rack power strip and the the back of the chassis. This was confirmed after disconnecting, to find the positive leg without continuity. However, after "pulling the cable apart" to find the source of the discontinuity, he was not able to reproduce it. After reassembling the cable, and restoring the connection, the driver came alive again. Further vigorous attempts to recreate the disconnect were unsuccessful. It appears the gremlins inside the driver chassis are creeping there way out to the power supply.

This 18V power cable came with the ESD driver from Strathclyde, so Richard plans to replace it with one commercially made in house.

As soon as I get the arm from fellow commissioners trying to finish gathering data for her thesis, I'll try to grab the data I'd originally set out to get this morning.
H1 SEI (SUS)
arnaud.pele@LIGO.ORG - posted 16:26, Wednesday 14 May 2014 (11902)
ETMX YAW motion from ISI : using Start Rz blends for now

While working on EX suspension, I realized the yaw motion was big on the oplev, so I tried switching the Rz blends on the ISI to see what's a good config. Moving the Rz blends of both stages from Tcrappy to Start seem to have helped (~0.6urad peak to peak -> ~0.2urad). Green curve in attached plot is the oplev yaw signal (urad)

Images attached to this report
LHO General
corey.gray@LIGO.ORG - posted 16:13, Wednesday 14 May 2014 (11881)
Ops DAY Summary

Day's Activities

 

H1 SUS (ISC, SYS)
jeffrey.kissel@LIGO.ORG - posted 16:01, Wednesday 14 May 2014 (11899)
H1 SUS ETMX Distribution Filter Changes from Successful ALS DIFF Locking
J. Kissel, for K. Izumi, S. Ballmer, and S. Dwyer

One of the tricks to get ALS DIFF working without saturation was to tweak the distribution filter design. I attach a comparison between the filters installed on 2014-05-08 vs. those tweaked last night (2014-05-14). Only two small changes: 
(1) To the UIM-TST LP, whose composition included 
    (a) two poles at 6 [Hz], three poles at 150 [Hz] 
    (b) elliptic filter of (freq,order,ripple,stopDB,zQ) = (9, 3, 1, 20, 1.4)
    (c) elliptic filter of (freq,order,ripple,stopDB,zQ) = (19, 3, 1, 20, 1.4),
now has the second elliptic filter's corner frequency much lower, at 6 [Hz], i.e. a new (c) of 
    (c) elliptic filter of (freq,order,ripple,stopDB,zQ) = (6, 3, 1, 20, 1.4),

(2) To the TST-UIM HP, whose composition included
    (a) (1 - UIM-TST LP) 
    (b) an elliptic high-pass of (freq,order,ripple,stopDB,zQ) = (.7, 2, 1, 10, 3)
now has an additional high-pass filter,
    (c) two zeros at 0 [Hz], two poles at 0.2 [Hz], normalized to unity at high frequency, i.e. zpk([0;0],[0.2;0.2],1,"n")gain(25)
Note that this extra low-pass has been installed, ignoring its affect on the complementarity with the UIM-TST LP. This needs to be fixed if we desire complementarity.

These tweaks served to (1) to move the elliptic filter's gain peaking bump out of the way of the 10 [Hz] features in the input noise, and (2) roll off the low-frequency end of the test mass drive faster.

-------------
Details 
The New H1 SUS ETMX Filter Configuration:

H1:SUS-ETMX_M0_LOCK_L (same as before)
FM1 "invL2L1" (part one of plant inversion filter)
FM2 "invL2L2" (part two of plant inversion filter)
FM3 "top/tst" (gain only filter to compensate for the gain ratio)

FM5 "blendLP"

H1:SUS-ETMX_L1_LOCK_L 
FM6 "invL2LNEW" (part one of plant inversion filter)
FM7 "patch" (part two of plant inversion filter)
FM8 "LISOfit" (part three of plant inversion filter)
FM9 "uim/tst" (gain only filter to compensate for the gain ratio)

FM3 "blendHP" (UIM-TOP HP complement)
FM4 "blendLP-2" (UIM-TST LP complement, with lower frequency elliptic filter)

H1:SUS-ETMX_L3_LOCK_L
FM6 "MatchedinvL2L" (part one of plant inversion filter)
FM7 "patch" (part two of plant inversion filter)

FM4 "blendHP-M0" (TST-TOP HP complement)
FM5 "blendHP-L1_2" (TST-UIM HP complement)
FM9 "hp.2" (extra high-pass for TST-UIM HP complement)


In order to create these modifications to the UIM-TST LP, Kiwamu grabbed my design script from 
/ligo/svncommon/SusSVN/sus/trunk/QUAD/Common/FilterDesign/HierarchicalControl/DARMmodel_ALS_20140428.m
copied it to his home directory, 
/ligo/home/kiwamu.izumi/
to play around,
/ligo/home/kiwamu.izumi/Desktop/farm/matlab/
modified the filters mentioned above, and then used a fancy new script,
/ligo/home/kiwamu.izumi/Desktop/farm/matlab/fofon.m
which takes in a continuous zpk filter, and produces a foton design string to be copied and pasted into Foton. If the filter is a high-pass, the only other step is to use the "Gain" field to set the gain to be unity at high-frequency.

The modification to the TST-UIM HP was done later in the evening after I'd left, but by the simplicity of its design, I think it was just added directly without testing any stability or complementarity in Matlab prior to install. 

[[elliptic filters are generated using ${SeiSVN}/seismic/Common/MatlabTools/myellip_z2.m]]
[[hp elliptic filter is generated by ${SeiSVN}/seismic/Common/MatlabTools/myhpellip_z.m]]
Non-image files attached to this report
H1 SUS
jeffrey.bartlett@LIGO.ORG - posted 15:58, Wednesday 14 May 2014 (11901)
Install H1-SRM
Andres R., Margot P., & Jeff B.

   We installed H1-SRM into HAM5 today. The installation went smoothly. Will start on alignment work in the morning.  
Images attached to this report
H1 SEI (SEI, SUS)
sheila.dwyer@LIGO.ORG - posted 23:37, Tuesday 13 May 2014 - last comment - 08:08, Thursday 15 May 2014(11876)
ETMX ISI tripped

I think that the ISI trip happened before the suspension trip, this was while we were locking diff.  I also thought that the T240s were green before I untripped the watchdog, but maybe they weren't because the ISI tripped again as it was isolating. 

Images attached to this report
Comments related to this report
jameson.rollins@LIGO.ORG - 08:08, Thursday 15 May 2014 (11908)

I stongly suspect that the second trip is due to the bug Fabrice and I noticed in 11888

H1 SUS
arnaud.pele@LIGO.ORG - posted 17:47, Wednesday 16 April 2014 - last comment - 15:25, Thursday 15 May 2014(11399)
ETMY measurements for filter design

This afternoon, after Jeff balanced the coils, I took measurements on the main chain of ETMY for filter design.
Results attached are showing Length/Pitch/Yaw drive to Pitch/Yaw Response for the top mass, UIM and PUM. Y axis is in urad/cts

Templates live in the usual sus path H1/ETMY/SAGL1/Data/2014-04-16_H1SUSETMY_P2PY_WhiteNoise.xml
 

Non-image files attached to this report
Comments related to this report
arnaud.pele@LIGO.ORG - 15:25, Thursday 15 May 2014 (11919)

Corrected version (without mistakes !)

Non-image files attached to this comment
Displaying reports 65641-65660 of 77209.Go to page Start 3279 3280 3281 3282 3283 3284 3285 3286 3287 End