After the Calibration Observing drop, H1 had a lockloss about 15min later (with ETMy ring up a second before the lockloss).
Taking Observatory Mode to MICROSEISM since it is still close to 1µm/s.
DRMI looked misaligned, so have went to CHECK MICH FRINGES (but then had another lockloss....but DOWN did NOT ramp PSL Power down from 10W....I'm only noting because this also happened yesterday afternoon).
At 1542utc (742am local), H1 was dropped out of OBSERVING due to the gain for H1:CAL-INJ_CW_GAIN going to 0. I reverted to 1 and returned H1 to OBSERVING within 2min. (and then we had a lockloss 15min later) :(
TITLE: 12/14 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 157Mpc
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
SEI_ENV state: USEISM
Wind: 10mph Gusts, 8mph 3min avg
Primary useism: 0.04 μm/s
Secondary useism: 0.93 μm/s
QUICK SUMMARY:
Looks like H1 relocked itself overnight, even with the gargantuan microseism of late! (H1's been locked 8.25hrs and also see that L1 has been up for the last 6+hrs).
If I look at the microseism, it does indeed looks like it has dropped down a little bit from its peak state 12hrs ago to inch down below 1µm/s.
Saturday Docket:
Local weather note: light snows were up the Yakima Vallley last night, but nothing down here last night, but on the drive in there was definitely ice on the roads (eventually took the old Jeep to 4H and drove slower; parking lot was also slippery).
AND as I was typing there was an OBSERVATION drop due to the Calibration Injection CW GAIN being taken to 0.0. I reverted back to 1.0 to return to observing.
TITLE: 12/14 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Lock Acquisition
INCOMING OPERATOR: Oli
SHIFT SUMMARY:
IFO is LOCKING at CARM_OFFSET_REDUCTION
Shift was dominated by microseism (hovering at 1 um/s) with a 6.4 magnitude EQ sprinkled in. H1 couldn't lock for the majority of the shift, managed to get there for 1hr and 9 mins before a lockloss.
Lockloss assumed to be due to the environment - lockloss alog 81820
Other than that, there were 2 SDF diffs that were erroneously accepted, then manually reverted by Sheila and I. We then accepted those. SDF screenshot included. The relevant SDFs are the two top ones (dated 21:18)
LOG:
| Start Time | System | Name | Location | Lazer_Haz | Task | Time End |
|---|---|---|---|---|---|---|
| 16:53 | OPS | LVEA | LVEA | YES | LVEA IS LASER HAZARD | 21:13 |
| 03:49 | PCAL | Francisco | PCAL Lab | Local | Turning off equipment | 03:58 |
Lockloss likely due to ground motion. Microseism is high and in the two minutes pre-lockloss, ASC channels were noisy.
No tags on LL Tool.
Closes FAMIS 26347. Last checked in alog 81655.
All trends look well. No fans above or near the (general) 0.7 ct threshold. Screenshots attached.
Anamaria took some time with me today to help me calibrate the sideband PRGs the same way she calculated them here for Livingston (LLO:69872). This is mostly with the intention of providing calibrated data for the finesse modeling effort that can help us better understand IFO behavior during TCS changes/powering up.
Reminder: I ran a check of the carrier PRG calibration this week (81752), which is based on the IM4 trans and end transmission values, which confirms that at 2W, our PRG is about 55.
Measure the power in the PRC at DRMI lock and compare it to the power in the PRC at 2W lock. Knowing the gamma of the 9 and 45 MHz sideband, estimate how much of the input power is sideband power. Use the predicted sideband PRG from an IFO model with perfect mode matching, and scale the expected sideband power in the PRC by those PRGs. Compare this expected value to the actual PRC power, and then scale the sideband PRGs accordingly.
DRMI lock values (measured after DRMI ASC converges):
| Channel | Value |
| POP A LF INMON | 12.5 ct |
| POPAIR RF18 | 151 ct |
| POPAIR RF90 | 27.25 ct |
| IM4 TRANS NSUM | 1.98 W |
2W full lock values (measured after full IFO ASC converges, including ADS):
| Channel | Value |
| POP A LF INMON | 283.8 ct |
| POPAIR RF18 | 154.5 ct |
| POPAIR RF90 | 28.5 ct |
| IM4 TRANS NSUM | 1.98 W |
Other useful numbers:
| Parameter | Value | Notes |
| Gamma 9 | 0.213 | 22.7 mW 9 MHz per W of input power, 62883 |
| Gamma 45 | 0.275 | 37.8 mW 45 MHz per W of input power, 62883 |
| PRM Transmission | 0.031 | https://galaxy.ligo.caltech.edu/optics/ |
| PR2 Transmission | 229 ppm | https://galaxy.ligo.caltech.edu/optics/ |
| M12 Transmission | 0.054 | Splitter on POP path, 63625 |
| 2W lock carrier PRG | 55 | Original measurement, checked again recently |
| PRG9 from model | 119 | Anamaria's Optickle model |
| PRG45 from model | 13 | Anamaria's Optickle model |
At 2 W lock, POP LF INMON measures 283.8 counts (this accounts for the dark offset). With an input power of 1.98 W and a PRG of 55, this should be 109 W. So we set a scale that 283.8 ct POP diode == 109 W PRC power.
At DRMI lock, we measure 12.5 counts in the POP diode (this accounts for the dark offset), so this gives us 4.8 W PRC power in DRMI lock.
At 1.98 W input power, 44.9 mW is 9 MHz power and 74.84 mW is 45 MHz power. We multiply by the perfect mode matching sideband PRGs and get that 5.3 W should be 9 MHz and 0.972 W should be 45 MHz in DRMI lock, so a total of 6.3 W expected PRC power in DRMI.
4.8 W / 6.3 W = 0.76, so 0.76 * 119 = 90.6 and 0.76 * 13 = 9.9, so we can say that our 2W PRG9 = 91 and PRG45 = 10.
Notice that this method ties the sideband PRGs to the measured carrier PRG. The method is perhaps not so good for the 45 MHz sideband, since we know there is so much less 45 MHz power than 9 MHz power. But these values are sensible.
You might notice we used the POP LF inmon value instead of the POP LF out value. There is a calibration in the POP A LF filter bank labeled "to_uW", which I assume means "microwatts of power on the POP diode". We started by using this calibrated value, and putting it into Watts of PRC power using the known transmission of the M12 splitter and PR2 (LLO calibrates POP A LF into Watts in the PRC). This resulted in a lower sideband PRG, (76.7 and 8.4 respectively). However, it also predicted that the carrier PRG should be 46. The carrier PRG calibration Craig and I performed does not rely on the POP diode at all, which is good because there is enough sideband power in the PRC to confuse the measurement. So, Anamaria and I decided to trust the independent carrier PRG measurement based on the arm transmission. Furthermore, we looked at how the H1:LSC-PR_GAIN_OUT16 channel is calibrated, and noticed that it ratios the H1:LSC-POP_A_LF_INMON value with IM4 trans. This motivated us to ignore the uW calibration filter (which then motivated us to correct for the dark offset of 3.8 ct).
If you have a good enough alog memory you might know that Craig and Jennie W have done a sideband PRG calibration in the past, using the modulation depth step test such as here: 70865 and 76358. Upon further inspection of the results, I am not sure they are so good. Mainly, the 45 MHz sideband PRG that results from this test is around 27-30, which is huge. I looked at these measurements again to try to understand why the values are so different. I noticed that when Craig steps down the 45 MHz sideband, the POP A LF value actually increases. Anamaria and I think this is because with lower 45 MHz modulation, the carrier input power increases, and the high carrier PRG (compared to the 45 MHz PRG) amplifies this excess power further. I can't find anywhere in Craig's code that accounts for this. For now, I am going to trust the method described in this alog over the mod depth step method.
These values will calibrate the POPAIR B RF18 I ERR and RF90 I ERR channels into sideband PRGs accordingly.
calibration factor = (2W sideband PRG value) / ( 2W POPAIR count value / IM4 trans)
POPAIR18 > PRG9 calibration factor = 91 / ( 154.5 / 1.98) = 1.17, and normalize by input power
POPAIR45 > PRG45 calibration factor = 10 / (28.5 / 1.98) = 0.7, and normalize by input power
Note that if you use the POPAIR B RF18/90 I NORM MON channels instead of the ERR channels, these are already normalized by input power.
You are right that when the 45 MHz goes down, the carrier goes up. That's why POP LF and the arm transmissions all rise when we lower our mod depths. The mod depth code does account for this, otherwise it would not work at all, see Daniel's original work. It gives an estimate that the arm powers are 100% carrier, for instance, which comes from the fact the arm transmissions rise.
TITLE: 12/13 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 153Mpc
INCOMING OPERATOR: Ibrahim
SHIFT SUMMARY:
Locked for most of the shift, but throughout the day, microseism continued climbing and is clearly above the 95th percentile now. Eventually H1 lost lock (after a 10.5+ hr lock), but have struggled with very high microseism (+ we also had a 6.3 Chilean EQ). In this downtime, did some To Do items: (1) LOADed ISC_LOCK and SQZ_MANAGER nodes, and (2) Ran an Initial Alignment. Have taken OBSERVATORY MODE to MICROSEISM. For locking, H1's green arms were fine, but even after the alignment, DRMI struggled (but its flashes looked good); this is what I handed off to Ibrahim.
When H1 does get back to Observing this weekend, if there are periods of the low-range glitching we've had in the last week, Sheila has a request for tests she would like operators to run (see alog81817).
LOG:
TITLE: 12/14 Eve Shift: 0030-0600 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Microseism
OUTGOING OPERATOR: Corey
CURRENT ENVIRONMENT:
SEI_ENV state: USEISM_EARTHQUAKE
Wind: 11mph Gusts, 7mph 3min avg
Primary useism: 0.34 μm/s
Secondary useism: 1.16 μm/s
QUICK SUMMARY:
IFO is DOWN in Earthquake Mode.
Microseism is very high and we just got hit by a 6.4 EQ. Will resume locking when possible.
Today I made some temporary changes to the SQZ MANGER guardian, to turn off all the servos when the request is down. (Edited to add, now I've undone these changes but added a if False statement on line 270 in SQZ Manager's down state that can be toggled to put this back in). This isn't usually what we want because we can leave the FSS and PMC locked.
Derek pointed out in 81811 that some filter cavity length control channels are well correlated with the large glitches. We know that blocking the squeezer doesn't stop the glitches, and without the squeezer locked to the IFO the filter cavity is locked on green and the noise of these channels seems too high to show the issue.
Request for operators: If you see the glitches and low range happening again, please go out of observing. First do a test of turning off and on the HAM1 feedforward (81775) it would be go to do 5 minutes off, 5 minutes on, 5 minutes off, then on again. Then open SQZ_MANager, and edit line 270 so that it says if True: , load sqz manager and request down, wait 5 minutes, request FREQUENCY DEPENDENT SQUEEZING for 5 minutes, DOWN for another 5, and back to Frequency dependent squeezing. These things can be done at the same time that Robert walks around looking for noise if he's available when the glitches come back.
Follow up from Tuesday maintenance: Daniel upgraded the Beckhoff slow controls software to no longer error on a version mismatch for the new timing cards installed in h1susex and h1omc0 (for new LIGO DAC and variable Duo-Tone frequency respectively).
I have changed the CDS HW stat code to no longer check that the EX timing error was solely due to the version mismatch. I also reverted the SYS_TC3_TIMING.adl file, removing the temporary blurb I had put in there explaining the expected EX error.
I took the opportunity to upgrade the timing entry on the CDS Overview to a system block. This will turn RED if there are any timing errors, or MAGENTA is the timing system freezes. Clicking on it opens the timing overview, please see attached.
Fri Dec 13 10:10:07 2024 INFO: Fill completed in 10min 3secs
Jordan confirmed a good fill curbside.
The mystery glitching this team has been thoroughly investigating was present again for ~30 minutes starting at approximately 6 UTC on Dec 13, as seen in this range time series. HVeto runs over this time period highlight H1:SQZ-FC_LSC_DOF2_OUT_DQ as the most correlated channel (same as in 81587), this time with a significance of 200 (a significance of > 10 is of interest). After checking omega scans of example glitching in the strain data, there is clear glitching present again in this squeezer channel. Similar to previous occurrences of this noise, it promptly appeared and then disappeared.
Laser Status:
NPRO output power is 1.832W
AMP1 output power is 70.23W
AMP2 output power is 137.1W
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PDWD watchdog is GREEN
PMC:
It has been locked 2 days, 16 hr 5 minutes
Reflected power = 23.33W
Transmitted power = 106.0W
PowerSum = 129.4W
FSS:
It has been locked for 0 days 5 hr and 21 min
TPD[V] = 0.7857V
ISS:
The diffracted power is around 2.8%
Last saturation event was 0 days 5 hours and 21 minutes ago
Possible Issues:
PMC reflected power is high
TITLE: 12/13 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 35Mpc
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
SEI_ENV state: USEISM
Wind: 2mph Gusts, 0mph 3min avg
Primary useism: 0.03 μm/s
Secondary useism: 0.73 μm/s
QUICK SUMMARY:
Wet drive in with H1 locked over 3.75hrs after Oli had a wake-up call for SDF diffs overnight. H1 range is hovering around 150Mpc. Over the last 24hrs, µseism has creeped up from ~95th percentile to above it. Currently have had INVALID dust alarms for Vacuum Prep Lab. There was a YELLOW SDF Diff for the CDS DIFF box on the CDS Overview, but it greened up as I was typing about it.
NOTE: next time we are out of OBSERVING, hit LOAD for ISC_LOCK (for RyanS/TJ change)
CDS SDF Diffs:
Reminder that there is a cronjob which automatically accepts any CDS SDF differences related to lock-loss-alert (LLA) settings. Any other diffs, e.g. WIFI Is Activate, are not touched.
The cronjob runs every 10 minutes. If a CDS SDF LLA diff persists longer that this, please contact me.
Sheila found that H1:TCS-ITMX_CO2_ISS_CTRL2_OUT_DQ stepped up from -0.2 to 0 on 2024/12/05 21:40:07 UTC (13:40 PST) Plot attached. This is of interest as this channel has ben a witness to our noisy/low range periods in DARM but is not connected to anything. I see no reason for this step up, the only person in the LVEA at the time was Robert setting up VP measurements (near HAM3 not CO2X) 81628, the CO2 laser remained l locked and we were not touching the CO2 chiller around the time 81634.
Since this step up, this channel has not been a good witness of our DARM noise, maybe the cable wasn't grounded and something changed to ground it on 12/05. Plot of it being a witness to the noise on 12/02 and not on 12/11.
Before and after the step up, this CO2 channel is still a witness to the CO2 rotation stage moving, attached. Both CO2X and CO2Y ISS CRTL2 channels see the rotation stages move, some crosstalk in chassis? CO2Y signal is orders of magnitude larger. Jason was looking into these channels and the chassis
I don't see any change in the H1:TCS-ITMX_CO2_RIN_INLOOP_OUTMON channel at that time, but the reason for the step in the CRTL2 output is from a turned off digital offset in the H1:TCS-ITMX_CO2_AOM_SET_POINT bank. I turned this off when we were looking at it and I forgot to alog it, apologies.
Curious that we lost sensitivity to whatever this is when an offset was removed, but I think this is a good clue.
I've just put the offset back in just to see if we get our "monitor" back. Accepted in safe and observe snaps, but only one screenshot.
It seems that with the offset on again, this channel is again a witness of the noisy times.
Ansel reported that a peak in DARM that interfered with the sensitivity of the Crab pulsar followed a similar time frequency path as a peak in the beam splitter microphone signal. I found that this was also the case on a shorter time scale and took advantage of the long down times last weekend to use a movable microphone to find the source of the peak. Microphone signals don’t usually show coherence with DARM even when they are causing noise, probably because the coherence length of the sound is smaller than the spacing between the coupling sites and the microphones, hence the importance of precise time-frequency paths.
Figure 1 shows DARM and the problematic peak in microphone signals. The second page of Figure 1 shows the portable microphone signal at a location by the staging building and a location near the TCS chillers. I used accelerometers to confirm the microphone identification of the TCS chillers, and to distinguish between the two chillers (Figure 2).
I was surprised that the acoustic signal was so strong that I could see it at the staging building - when I found the signal outside, I assumed it was coming from some external HVAC component and spent quite a bit of time searching outside. I think that this may be because the suspended mezzanine (see photos on second page of Figure 2) acts as a sort of soundboard, helping couple the chiller vibrations to the air.
Any direct vibrational coupling can be solved by vibrationally isolating the chillers. This may even help with acoustic coupling if the soundboard theory is correct. We might try this first. However, the safest solution is to either try to change the load to move the peaks to a different frequency, or put the chillers on vibration isolation in the hallway of the cinder-block HVAC housing so that the stiff room blocks the low-frequency sound.
Reducing the coupling is another mitigation route. Vibrational coupling has apparently increased, so I think we should check jitter coupling at the DCPDs in case recent damage has made them more sensitive to beam spot position.
For next generation detectors, it might be a good idea to make the mechanical room of cinder blocks or equivalent to reduce acoustic coupling of the low frequency sources.
This afternoon TJ and I placed pieces of damping and elastic foam under the wheels of both CO2X and CO2Y TCS chillers. We placed thicker foam under CO2Y but this did make the chiller wobbly so we placed thinner foam under CO2X.
Unfortunately, I'm not seeing any improvement of the Crab contamination in the strain spectra this week, following the foam insertion. Attached are ASD zoom-ins (daily and cumulative) from Nov 24, 25, 26 and 27.
This morning at 17:00UTC we turned the CO2X and CO2Y TCS chiller off and then on again, hoping this might change the frequency they are injecting into DARM. We do not expect it to effect it much we had the chillers off for a ling period 25th October 80882 when we flushed the chiller line and the issue was seen before this date.
Opened FRS 32812.
There were no expilcit changes to the TCS chillers bettween O4a and O4b although we swapped a chiller for a spare chiller in October 2023 73704.
Between 19:11 and 19:21 UTC, Robert and I swapped the foam from under CO2Y chiller (it was flattened and not providing any damping now) to new, thicker foam and 4 layers of rubber. Photo's attached.
Thanks for the interventions, but I'm still not seeing improvement in the Crab region. Attached are daily snapshots from UTC Monday to Friday (Dec 2-6).
I changed the flow of the TCSY chiller from 4.0gpm to 3.7gpm.
These Thermoflex1400 chillers have their flow rate adjusted by opening or closing a 3 way valve at the back of the chiller. for both X and Y chillers, these have been in the full open position, with the lever pointed straight up. The Y chiller has been running with 4.0gpm, so our only change was a lower flow rate. The X chiller has been at 3.7gpm already, and the manual states that these chillers shouldn't be ran below 3.8gpm. Though this was a small note in the manual and could be easily missed. Since the flow couldn't be increased via the 3 way valve on back, I didn't want to lower it further and left it as is.
Two questions came from this:
The flow rate has been consistent for the last year+, so I don't suspect that the pumps are getting worn out. As far back as I can trend they have been around 4.0 and 3.7, with some brief periods above or below.
Thanks for the latest intervention. It does appear to have shifted the frequency up just enough to clear the Crab band. Can it be nudged any farther, to reduce spectral leakage into the Crab? Attached are sample spectra from before the intervention (Dec 7 and 10) and afterward (Dec 11 and 12). Spectra from Dec 8-9 are too noisy to be helpful here.
TJ touched the CO2 flow on Dec 12th around 19:45UTC 81791 so the flowrate further reduced to 3.55 GPM. Plot attached.
The flow of the TCSY chiller was further reduced to 3.3gpm. This should push the chiller peak lower in frequency and further away from the crab nebula.
The further reduced flow rate seems to have given the Crab band more isolation from nearby peaks, although I'm not sure I understand the improvement in detail. Attached is a spectrum from yesterday's data in the usual form. Since the zoomed-in plots suggest (unexpectedly) that lowering flow rate moves an offending peak up in frequency, I tried broadening the band and looking at data from December 7 (before 1st flow reduction), December 16 (before most recent flow reduction) and December 18 (after most recent flow reduction). If I look at one of the accelerometer channels Robert highlighted, I do see a large peak indeed move to lower frequencies, as expected. Attachments: 1) Usual daily h(t) spectral zoom near Crab band - December 18 2) Zoom-out for December 7, 16 and 18 overlain 3) Zoom-out for December 7, 16 and 18 overlain but with vertical offsets 4) Accelerometer spectrum for December 7 (sample starting at 18:00 UTC) 5) Accelerometer spectrum for December 16 6) Accelerometer spectrum for December 18
1821utc Back To Observing
This is post-Initial Alignment (with some ALSx minor grief), several DRMI locklosses, and continued high microseism).