Displaying reports 861-880 of 85528.Go to page Start 40 41 42 43 44 45 46 47 48 End
Reports until 16:31, Thursday 25 September 2025
LHO General
ryan.short@LIGO.ORG - posted 16:31, Thursday 25 September 2025 (87148)
Ops Day Shift Summary

TITLE: 09/25 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 147Mpc
INCOMING OPERATOR: Oli
SHIFT SUMMARY: H1 has been locked the whole day so far; current lock stretch is up to almost 11.5 hours. The ETMY roll mode is slightly more rung up than it has been in the past few days, but maybe not egregiously so; Oli will keep an eye on it this evening. Other than commissioning time this morning, it's been a quiet shift.

LOG:

Start Time System Name Location Lazer_Haz Task Time End
16:48 VAC Pump LVEA N AIP pumping on HAM6 Ongoing
17:52 SEI Jim, Mitchell MX N Looking at 3IFO HEPI parts 18:36
21:29 FIT Matt Y-arm N On a run 22:12
H1 CAL (ISC)
elenna.capote@LIGO.ORG - posted 15:59, Thursday 25 September 2025 (87146)
Effect of DHARD gain change on sensing function

Today I adjusted the DHARD gain and asked Ryan to rerun the simulines measurements to see if changing the DHARD gain has any measureable effect on the sensing function. This tests possible L2A2L coupling present in the sensing function.

Based on recent OLGs of the HARD loops, I decided we could probably lower each gain by 3 dB. For DHARD P, this takes the gain from -30 to -18 and DHARD Y from -40 to -24.

We ran one simulines with DHARD P gain lowered and one with DHARD Y gain lowered. These measurements can be compared with this morning usual calibration measurement.

The results seem a bit inconclusive to me. There might be some change in the phase below 10 Hz. First I compare the three sensing measurements, then I take the ratio of each measurement with the "nominal settings" measurement.

Non-image files attached to this report
H1 General
oli.patane@LIGO.ORG - posted 15:59, Thursday 25 September 2025 (87147)
Ops EVE shift start

TITLE: 09/25 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Observing at 150Mpc
OUTGOING OPERATOR: Ryan S
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 19mph Gusts, 14mph 3min avg
    Primary useism: 0.06 μm/s
    Secondary useism: 0.16 μm/s 
QUICK SUMMARY:

Observing at 150 Mpc and have been Locked for almost 11 hours. Wind has increased recently, but shouldn't get too bad. ETMY roll mode has been slowly increasing over the past ~few hours, so I'll be keeping an eye on it.

H1 General
ryan.short@LIGO.ORG - posted 14:16, Thursday 25 September 2025 - last comment - 11:38, Wednesday 01 October 2025(87145)
Automatic Injections Start Times Updated

Since maintenance will start early next Tuesday, I've updated the PEM_MAG_INJ node to start its injections at 6:00am (13:00 UTC) and the SUS_CHARGE node will start at 6:25am (13:25 UTC). These should all be finished by 6:45am (13:45 UTC).

Comments related to this report
ryan.short@LIGO.ORG - 11:38, Wednesday 01 October 2025 (87247)OpsInfo

I've put both Tuesday morning automatic injection start times back to their usual with magnetic injections starting at 7:20am and in-lock charge at 7:45am (both in local time). Guardians have been loaded.

H1 PSL
keita.kawabe@LIGO.ORG - posted 12:21, Thursday 25 September 2025 (87143)
PSL beam shape and EOM inspection

I don't have time to analyze these for now, so I'm just dumping the pictures here.

On Tuesday Sep/23 RyanS and I went to the PSL room to do two things.

Images attached to this report
H1 IOO (ISC, PSL)
jennifer.wright@LIGO.ORG - posted 10:51, Thursday 25 September 2025 - last comment - 09:27, Friday 03 October 2025(87140)
ISS measurements with new oscilloscope setting

Yesterday I went into the optics lab and re-measured the coupling between input beam motion and PD array. While taking measurements I noticed that the injection could not be seen on the AC readouts of the PDs (example) so I tuned the temperature of the laser via changing the resistance of the controller, I went from ~10kOhms up to 13 kOHms and down to 8kOhms and while I found places where the noise reduced see example of noisy trace here, I couldn't find anywhere with the controller where the trace renamed stably in the non-noisy state. I then decided to tune the pump current down from 130mA to ~100mA and eventually found a somewhat stable place. I still had to wait through some periods of noise to trigger the measurement of the PDs.

I alos increased the modulation ampltitude to 80 mVpp. The counts on the QPD LCD readout were 10672, see image.

When the laser is in its quiet state the AC PD traces should comfortably fit on the screen of the osclloscopes with a 5mV scale, with the laser noisy this is more like 100mV, I also use 100mV scale for the QPD, I didn't change it when I reduced the noise on the laser.

The noisy state for the QPD outputs is here, the quiet state is here.

For each measurement I used a capture range of 400ms on the time axis of the scope and 125 000 samples selected on the 'ACQUIRE' menu.

The final measurements are:

PD 1 - 4 measured at AC: T0012ALL.CSV

PD 5 - 8 measured at AC: T0014ALL.CSV

QPD X, Y and SUM channels measured at AC: T0013ALL.CSV

The two DC measurements are going to be averaged so I didn't wait for a quiet time to measure them.

PD 1 - 4 measured at DC: T0011ALL.CSV 

PD 5 - 8 measured at DC: T0010ALL.CSV

To save you need to click on the menu button and change the resolution to be'Full', the format to be CSV and the channels captured to be 'ALL', the file number will roll over every time you save so you don't need to enter it manually.

Images attached to this report
Comments related to this report
jennifer.wright@LIGO.ORG - 12:49, Monday 29 September 2025 (87200)

The code to produce this plot is in my optics lab code repo.

The graph of the TF from horizontal dither on the input mirror to horizontal dither across the array, shows that we are not getting much coherent modulation of the light intensity on the PDs at 100Hz which is the dither frequency. Either my code is wrong or I need to increase the dither amplitude for the mirror.

The maths to work this out was 

A time series = abs(dither in direction horizontal to bench on QPD in V)/ (motion horizontal to bench volts on QPD/mm moved horizontal to bench on QPD)

B time series = AC voltage on each PD / mean of DC voltage on each PD

 

TF = CSD (B, A) / PSD (A, A)

I used gwpy for the calculations this time.

The attached plot shows each PD in the array as a different colour with magnitude on the top and phase on the bottom.

Images attached to this comment
jennifer.wright@LIGO.ORG - 09:27, Friday 03 October 2025 (87275)

We realised in the analysis that we should be using :

H_amp on QPD = Xcos theta + Y sin theta where theta is the angle between the X axis of the QPD and the horizontal scan direction of the beam worked out from our previous calibration measurments, 14 degrees.

This gives a different value for the couplings. All the PDs other than PD6 did not have coherences 0.9 or over so I only attach the final TF for PD6. We increased the dither amplitude after this to improve the measurement of the other PDs.

 

Images attached to this comment
H1 AOS
camilla.compton@LIGO.ORG - posted 10:38, Thursday 25 September 2025 (87139)
SRM Offsets still look good at 5kHz HOM
We checked what the 5kHZ HOM peak looked like in +4dB Mid-SQZ, after adjusting the matrix for OMC-DCPD_524K_A2 as in 85937.  Checking as before our PSL power issues,  we adjusted SRM offsets in 86414 to minimize it. The 5kHz HOM peak still looked minimized as there was no 5kHz peak, see black trace in attached dtt. All other traces apart from "Today MidSQZ" black and "Today FDS" cyan are old. 
Dtt saved /ligo/home/camilla.compton/Documents/sqz/templates/dtt/20250925_5kHz.xml 
Images attached to this report
LHO VE
david.barker@LIGO.ORG - posted 10:24, Thursday 25 September 2025 (87142)
Thu CP1 Fill

Thu Sep 25 10:09:27 2025 INFO: Fill completed in 9min 23secs

 

Images attached to this report
H1 SQZ
camilla.compton@LIGO.ORG - posted 10:00, Thursday 25 September 2025 (87138)
Could not easily re-create low frequency SQZ noise with PSAMS

In 87071, we found that the low frequency SQZ noise could be removed by changing the ZM4 or ZM5 PSAMs and we left the ZM4 PSAMs at 6.2V strain (from 6.0V). 

Today I went back to the noisy area at 6.0V and lower on ZM4 and saw no increase in low frequency noise, plot attached. I did not try going higher than 6.2V yet, we could do that as 6.5V caused noise on Monday too. 

It is not surprising that this noise couldn't be re-created as it was coming and going in the past. 

Images attached to this report
H1 CAL
ryan.short@LIGO.ORG - posted 09:27, Thursday 25 September 2025 (87137)
Broadband and Simulines Calibration Sweeps

Following the procedure in the TakingCalibrationMeasurements wiki, at 15:30 UTC H1 dropped observing for the usual calibration sweeps.

Broadband - 15:30:52 to 15:36:02 UTC

Simulines - 15:36:37 to 16:00:00 UTC

Calibration monitor screenshot and calibration report attached.

File written out to: /ligo/groups/cal/H1/measurements/DARMOLG_SS/DARMOLG_SS_20250925T153638Z.hdf5
File written out to: /ligo/groups/cal/H1/measurements/PCALY2DARM_SS/PCALY2DARM_SS_20250925T153638Z.hdf5
File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L1_SS/SUSETMX_L1_SS_20250925T153638Z.hdf5
File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L2_SS/SUSETMX_L2_SS_20250925T153638Z.hdf5
File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L3_SS/SUSETMX_L3_SS_20250925T153638Z.hdf5

 

Images attached to this report
Non-image files attached to this report
LHO General
ryan.short@LIGO.ORG - posted 07:45, Thursday 25 September 2025 (87136)
Ops Day Shift Start

TITLE: 09/25 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Observing at 154Mpc
OUTGOING OPERATOR: Tony
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 5mph Gusts, 3mph 3min avg
    Primary useism: 0.02 μm/s
    Secondary useism: 0.16 μm/s 
QUICK SUMMARY: H1 has been locked for 2.5 hours. Looks like the lockloss last night was from yet another earthquake, this time likely a M5.7 from Mexico. Calibration and commissioning time scheduled today from 15:30 to 20:00 UTC.

LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 22:00, Wednesday 24 September 2025 (87135)
OPS Eve Shift Summary

TITLE: 09/25 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Earthquake
INCOMING OPERATOR: Tony
SHIFT SUMMARY:

IFO is in LOCKING in LOCKING_ARMS_GREEN

Very calm shift with a fully auto Lock acquisition from the M6.1 Venezuela EQ this afternoon (alog 87128).

We had one Lockloss due to another earthquake from Venezuela, M6.3. Lockloss alog 87134. I turned on the ASC High Gain before it happened, but we understandably still lost lock.

I've just set IFO to auto-lock, which it will do once the EQ rings down (I expect it to take another 30 mins to get to a lockable state).

LOG:

Start Time System Name Location Lazer_Haz Task Time End
16:48 VAC Pump LVEA N AIP pumping on HAM6 23:43
22:10 ISC Jennie Opt Lab Local ISS array work 23:31
22:54 VAC Gerardo LVEA N Checking on AIP 23:16
H1 PEM (Lockloss)
ibrahim.abouelfettouh@LIGO.ORG - posted 21:20, Wednesday 24 September 2025 (87134)
Lockloss 04:01 UTC

Lockloss due to another EQ from Venezuela - 6.3 Magnitude

H1 ISC
elenna.capote@LIGO.ORG - posted 17:09, Wednesday 24 September 2025 (87133)
DARM offset step results

I previously reported results from DARM offset step tests in October and last week. The overall goal here is to figure out what overall effect the power outage has had on the IFO, especially since we have lost 1% optical gain. For example, see this alog about comparing modulation depth tests.

However, after looking closer at the DARM offset measurement results, and especially investigating the effect of whether leaving the OMC ASC on during the measurement matters, I would like to revise the previously reported measurement results.

Some background: this test aims to measure the contrast defect light by changing the DARM offset in mA while injecting strong PCAL lines to capture the DARM optical gain. We expect the relationship between optical gain (mW/pm) and DARM offset power (mW) to be quadratic. We assume by fitting the data to a parabola, the resulting y-intercept can tell us the contrast defect light on the DCPDs, if we had zero DARM offset.

Some considerations:

First question, does leaving the OMC ASC on matter? Jennie has been doing a lot of these measurements, and her experience has been, yes, because the OMC ASC offsets are set at one DARM offset and are not reset at each different offset. We do not expect the OMC alignment to drift significantly during the 15 minutes of this measurement, so the best practice has been to turn the OMC ASC OFF for the measurement. However, I didn't do it the first time I measured last week, but I got significantly better results (using Craig's code) with the ASC ON than I did with the ASC OFF.

Second question, what is the best way to fit this data? This apparently has been a question for a long time, see Gabriele's comment to 30573 in 2016. I believe the model I described above suggests that the vertex of the resulting parabola should be centered around x=0, that is zero optical gain. However, I discovered this week that Craig's code, which I used to report the results in October and last week (see first two links), fits a nonzero vertex (i.e. y = a(x-x0)^2 + b instead of y = ax^2 + b). Of course, that will change the answer significantly! I don't understand the mechanism that would cause us to move the center of this vertex away from x=0, so I refitting the data to follow y = ax^2 + b.

Results:

I am revising the fitted contrast defects to be:

Frequency October 2024 (OMC ASC on) [mW] September 2025 (OMC ASC on) [mW] September 2025 (OMC ASC off) [mW]
255.0 Hz 1.088 +- 0.033 1.219 +- 0.032 1.256 +- 0.005
410.3 Hz 1.086 +- 0.030 1.214 +- 0.042 1.240 +- 0.008

Conclusions:

This measurement shows that the contrast defect is higher now than it was in October. I don't know if we can attribute all of this to the power outage; we lost 1% optical gain comparing kappa c from before and after the vent, giving us at least 2% less optical gain between October and now. It also shows that there is perhaps a small effect related to the OMC ASC being off: slightly higher contrast and a better overall fit.

This also increases our upper limit on the possible homodyne angle- in October I stated that the homodyne angle upper limit would be about 7 degrees, but with this revised value it should have been more like 8.8 degrees. Now it is 9.3 degrees.

LHO General
ryan.short@LIGO.ORG - posted 16:30, Wednesday 24 September 2025 (87130)
Ops Day Shift Summary

TITLE: 09/24 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Earthquake
INCOMING OPERATOR: Ibrahim
SHIFT SUMMARY: Very quiet day of H1 observing until an earthquake caused a lockloss in the late afternoon. Once the ground motion rings down, Ibrahim will start relocking.

LOG:

Start Time System Name Location Lazer_Haz Task Time End
16:48 VAC Pump LVEA N AIP pumping on HAM6 23:43
16:10 FAC Kim Opt Lab N Technical cleaning 16:26
17:02 SEI Jim, Mitchell Opt Lab N Taking parts out of clean bags 17:30
17:58 CAL Tony PCal Lab N Looking for a cable 18:34
20:08 ISC Jennie Opt Lab Local ISS array work 21:00
20:41 CAL Tony PCAL Lab N Dropping off supplies 21:25
20:49 AOS Betsy Opt Lab N Checking bag contents 21:16
21:26 SPI Jeff, RyanS Opt Lab N Looking at SPI optics 21:41
22:10 ISC Jennie Opt Lab Local ISS array work 00:09
22:54 VAC Gerardo LVEA N Checking on AIP 23:16
LHO General
ibrahim.abouelfettouh@LIGO.ORG - posted 16:22, Wednesday 24 September 2025 (87129)
OPS Eve Shift Start

TITLE: 09/24 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Earthquake
OUTGOING OPERATOR: Ryan S
CURRENT ENVIRONMENT:
    SEI_ENV state: LARGE_EQ
    Wind: 6mph Gusts, 3mph 3min avg
    Primary useism: 0.69 μm/s
    Secondary useism: 0.18 μm/s 
QUICK SUMMARY:

IFO is DOWN due to EARTHQUAKE

Just waiting for this 6.2 EQ from Venezuela to pass through and then back to LOCKING

 

H1 ISC
elenna.capote@LIGO.ORG - posted 18:17, Tuesday 23 September 2025 - last comment - 16:36, Wednesday 24 September 2025(87114)
Modulation Depth test results - March vs September

Back in March, I ran a modulation depth test, with several goals in mind related to providing useful calibration information for modeling. That alog is still unfortunately sitting in my drafts. However, I was able to use the results to make a side-by-side comparison with a modulation depth test Sheila ran last Thursday, after the power outage. We are still trying to understand what the overall effect of the power outage was on the IFO. Namely, we have lost 1% of optical gain, 86964.

Some background: the modulation depth test aims to measure the fraction of carrier, 9 MHz and 45 MHz power at each port. This is done by measuring the powers at the nominal settings, and then individiually stepping the 9 and 45 MHz modulation up or down by a known value. Using a measured calibration of V/dBm and rad/V (see alog 62883), and using the bessel functions, the power fraction of each field can be measured at each diode based on how much the total diode power changes at each step. (note: I still have a to do list item to better calibrate the modulation depth in radians using OMC scan data).

Procedure: I stepped both down and up in modulation depth, resulting in 5 different measurements (nominal, 9 down, 9 up, 45 down, 45 up). In March, I stepped down by 3 dBm and up by 2 dBm, but in September we were able to step both down and up by 3 dBm. I measured for 3 minutes at each step in March, and 1 minute at each step in September.

EDIT: I realized I made an error, and the first results I report below are actually from Feb 2025. The significant difference here is that in my February measurement I only stepped down by 3 dBm for 9 and 45 MHz each, so there is less data to fit. In my March measurement, I stepped up 2 dBm and down 3 dBm, getting 5 total different measurement times. There should be very little difference in the interferometer between February and March 2025, so the differences in the results I believe are due to the fact that more points (5 versus 3) gives you a much better fit to the data. I would compare March and now for a more accurate understanding of the differences.

February results, well before power outage, only fit from 3 data points:

Field Input POP REFL AS
carrier 0.9771 0.9842 0.9419 0.3386
9 MHz 0.01278 0.01527 0.02950 0.1787
45 MHz 0.01003 0.000474 0.02860 0.4827

March results, before power outage, fit from 5 data points:

Field Input POP REFL AS
carrier 0.9779 0.9831 0.9164 0.2134
9 MHz 0.01212 0.01332 0.04059 0.2761
45 MHz 0.009687 0.002302 0.04431 0.5413

September results, after power outage (and reduction of PSL power, attenuation at IMC REFL), fit from 5 data points:

Field Input POP REFL AS
carrier 0.9783 0.9833 0.9333 0.3556
9 MHz 0.01195 0.01325 0.02533 0.1852
45 MHz 0.009486 0.001909 0.04199 0.4641

EDIT: Including the March results (and trusting them more than the February results) changes the conclusion. The input ratios are very similar between all three measurements. This is also true for POP where carrier and 9 ratios are concerned. 45 MHz is the hardest field to measure because the 9 and carrier are so strong at POP. There may be less 45 MHz at POP, or this is just the measurement uncertainty. At REFL, there may be an increase in carrier light now after the power outage, and there may be half as much 9 MHz as in March.

The most dramatic differences are at the AS port. Just comparing the February and March measurements, there may be considerable uncertainty in how much of each field is at AS in general. However, if we choose to believe the March results, this would suggest a significant increase in carrier at AS, and a significant decrease in 9 and 45 MHz. However, comparing February and now, the 9 MHz and carrier are nearly the same, and the 45 seems to have decreased.

Comments related to this report
elenna.capote@LIGO.ORG - 11:30, Wednesday 24 September 2025 (87123)

Rereading this alog, I see that I should say specifically which diodes measure these powers:

Input == IM4 trans

POP == POP A LF

REFL == REFL A LF

AS = AS_C DC NSUM

elenna.capote@LIGO.ORG - 16:36, Wednesday 24 September 2025 (87131)

Here are plots of various channels during the March and September mod depth tests. The shaded region indicates which step was being taken at the time, and the dotted line matching each shading color indicates the median of the channel at that time.

Non-image files attached to this comment
LHO VE
jordan.vanosky@LIGO.ORG - posted 12:45, Tuesday 23 September 2025 - last comment - 10:43, Wednesday 01 October 2025(87101)
HAM6 Annulus Ion Pump (AIP) Replacement

During Tuesday maintenance, we swapped the HAM6 AIP (Starcell). Note this annulus system is connected to HAM5 via the septum plate. We vented the lines with dry nitrogen and left a continuous nitrogen purge(~.3 psi) of the line during the pump swap. Nitrogen attached to HAM5 pump out port while HAM6 pump out port was left open to atmosphere.

No issues during the swap, annulus system is now pumping at both HAM5 & 6 ports with an aux cart and turbo pair. As of end of maintenance, the HAM6 cart was at ~3E-5 Torr, HAM5 cart at ~1E-4 Torr. These pumps will continue running until pressure is <1E-5 Torr at which point the ion pumps will be powered on.

Carts are placed on foam for isolation, and a piece of foam between the flex hose running up to HAM6 pump out port and HAM6 chamber. See attached pictures. 

Work permit will be closed once pumps are disconnected from chambers.

Images attached to this report
Comments related to this report
gerardo.moreno@LIGO.ORG - 17:10, Wednesday 24 September 2025 (87132)DetChar, VE

Update.

IFO was out of lock due to an earthquake, I went in to the LVEA to check on the aux-carts pumping down on the annuli for HAM5 and HAM6.  HAM5 aux-cart was good and pumping down on the annulus, however HAM6 aux-cart safety valve somehow managed to trip between yesterday and today, time is unknown as of now, I restored aux cart, and opened the valve.  Aux-cart for HAM6 was reporting a dubious pressure number of 1.26 x 10-07 Torr.

After restoring pumping to HAM6 annulus, both aux carts are reporting more believable numbers.

Images attached to this comment
gerardo.moreno@LIGO.ORG - 10:43, Wednesday 01 October 2025 (87245)VE

(Jordan, TJ, Gerardo)

Late entry.

TJ powered ON the ion pumps over the weekend, that allowed for the pumps to reach very good vacuum pressure on the shared annuli system, then on Tuesday morning, Jordan isolated the annuli system for HAM5 and HAM6 from the mechanical pumps and turned off the aux carts.

A couple of hours later we removed the small can turbos, flex hoses and aux carts from the HAM5/6 area, to conclude the replacement of the HAM5 annulus ion pump body.

 

 

Images attached to this comment
H1 SQZ
sheila.dwyer@LIGO.ORG - posted 17:23, Monday 22 September 2025 - last comment - 12:18, Thursday 25 September 2025(87071)
squeezer adding noise, not filter cavity related, depends on psams settings.

Elenna Capote, Camilla Compton, Sheila Dwyer, Derek Davis

This afternoon we had a repeat of the bad low frequency noise that we have been suspecting was from filter cavity backscatter 86596.  We saw that the symptom of elevated noise in the filter cavity error signal was similar to previous incidents plot

We compared squeezing with and without the filter cavity, and no squeezing, and see that this noise is there when squeezing is injected no matter what the filter cavity state is.  plot and plot with mean sqz and anti squeezing

We repeated the fringe wrapping measurements, we saw a higher scattered amplitude when moving ZM5 than last week. (shelf is higher by 10dB). the ZM2 shelf is about the same. plot

We also did some 30 Hz excitations in ZM5 + ZM2, we can see a bilinear coupling of these but the background didn't change during this excitation. plot

Derek and Elenna looked at the glitches in DARM that showed up at the time of the noise.  Derek ran some hveto runs for times with frequency dependent squeezing and frequecy independent squeezing, and saw that filter cavity length signals are a good witness when the filter cavity is locked, when the filter cavity is not locked the giltches stay but aren't witnessed by the FC error signal. 

Camilla found that she could reproduceably make the noise go away by moving the ZM4 +5 PSAMs small amounts.  She moved the PSAMs and adjusted the alignment to get a good level of high frequency squeezing back.  She also tried to do this with alignment only. 

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 17:35, Monday 22 September 2025 (87078)

We also took SQZ_OPO_LR GRD to LOCKED_CLF no ISS to check that the pump AOM wasn't injecting any noise.

Ended up leaving ZM4 PSAMs at 6.2V on the strain gauge, old nominal was 6.0V sdf attached. This is only a 9V change on the 0-200V PSAMs, from 78V to 87V. Which doesn't seem big enough to cause such an effect.

  • Plot of repeatably changing ZM5 PSAMS along with ZM5 PIT alignment to make noise appear and go
  • Plot of repeatably changing ZM4 PSAMS (no ZM4 PIT alignment needed) to make noise appear and go.
    • Interestingly 6.0V would be noisy, 6.5V would be noisy, but 6.2V was not! This is confusing.
  • Plot of changing ZM4 PIT and YAW alignments slowly ~50urad each way, no difference in noise.
    • I also checked ZM5 more quickly and noticed no obvious change
    • For ZM4 I tried taking the PIT alignment back to the same osems the good 6.2V PSAMS setting has the osems at, the noise did not change.
Images attached to this comment
camilla.compton@LIGO.ORG - 11:47, Thursday 25 September 2025 (87141)

Sheila posted backscatter measurements in 86836, and opened an FRS ticket for this issue: FRS # 35457.

B:BS1 is a 99/1 BS (see D2000021 spreadsheet). The PD that the beam transmitting B:BS1 goes to H1:IOO-OFI_PD_A_DC_POWER, this doesn't see any increased noise at the noisy time, although it's only a 16Hz channel: time series and dtt attached. 

We do see the DC power on H1:IOO-OFI_PD_A_DC_POWER change with the YAW of ZM4, see attached. This could be a sign that there is some clipping... Could investigate this a little more. 
We moved B:PD1 in 65006 and checked that it's reflected beam was being dumped in 65066
Images attached to this comment
camilla.compton@LIGO.ORG - 12:18, Thursday 25 September 2025 (87144)

I could increase the power on B:PD1 by ~25% by moving ZM4 and ZM5 in yaw before we lost RF3 and SQZ went down, showing we are nominally clipping this PD, plot attached. We can repeat Sheila's backscatter measurements with a different amount of light on this PD to see if it's the culprit.

Images attached to this comment
Displaying reports 861-880 of 85528.Go to page Start 40 41 42 43 44 45 46 47 48 End