Displaying reports 17301-17320 of 86757.Go to page Start 862 863 864 865 866 867 868 869 870 End
Reports until 15:38, Thursday 20 July 2023
H1 ISC
gabriele.vajente@LIGO.ORG - posted 15:38, Thursday 20 July 2023 - last comment - 15:49, Thursday 20 July 2023(71565)
Removed 1.34 Hz peak from DARM

Following up on the discovery that the 1.34Hz peak in DARM was injected by the SRCLFF, I implemented a more aggressive high-pass with a 1.34 Hz notch in the SRCLFF path.

The new high-pass and notch completely remove the 1.34 Hz peak from DARM, and improves the DARM noise in the 1-2 Hz region. The reduction in RMS is visible also in the time trace of the DARM error signal.

As expected, the new high-pass introduces a bit of phase rotation at 10-20 Hz, and this limits the amount of subtraction we can get now with the SRCL FF. I also tried to retune the SRCL FF with a new measurement, but I could not improve the subtraction. The reason is that I need to take a new measurement of the SRCLFF to DARM path with the new high-pass and notch. I did not have time to do it today because violin modes were damped and we needed to go back to Observe. But it should be an easy fix (10 minutes measurement + retuning + 10 minues test)

However, looking at the coherence between DARM and SRCL, I don't see a lot of it, so maybe we're fine with this reduced level of SRCL subtraction.

The new SRCLFF combination to be used is FM1 FM2 FM3 (all together). ISC-LOCK has been updated and reloaded

Images attached to this report
Comments related to this report
gabriele.vajente@LIGO.ORG - 15:49, Thursday 20 July 2023 (71567)

A comparison of the coherence between DARM and SRCL with the old FF (red) and the new FF (blue). The coherence is slightly higher below 20 Hz, as expected. Very similar everywhere else

Images attached to this comment
H1 General
anthony.sanchez@LIGO.ORG - posted 15:25, Thursday 20 July 2023 (71563)
SFD Acepted H1:LSC-SRCFF1

H1:LSC-SRCFF1 SDF Changes were Accepted after speaking with Gabriele.

ScreenShot attatched.

 

Images attached to this report
H1 SEI
jim.warner@LIGO.ORG - posted 15:00, Thursday 20 July 2023 (71561)
Bracing added to HEPI pump station pipe

Randy had pointed out a few months ago there was a section of piping coming out of the endstation HEPI pumps that should probably get some extra support. While the control room was relocking today, he went and added a bit of unistrut to the stand for the revervoir, attached phot shows how he attached to the foot of the reservoir stand. This was only needed for the end station pumps, the corner pumps have a much shorter run of pipe to the next support clamp.

Images attached to this report
H1 ISC (SUS)
camilla.compton@LIGO.ORG - posted 14:55, Thursday 20 July 2023 - last comment - 21:17, Monday 24 July 2023(71559)
DHARD_Y Transient during DHARD_WFS state, larger since June 29th.

Daniel, Camilla 

Following on from Ryan's 71420, we looked at the movement of ITMY (only has ASC control) during DHARD_WFS and see there is now larger spikes during locking than before 29th, see attached plot where H1:SUS-ITMY_L2_OSEMINF_{U,L}{L,R}_INMON spikes during locking are larger after the t-cursor (see bottom purple plot for when violins rang up).

Zooming in, attached plot, we can see when the Input to the DHARD_Y filter is turned on that the output to DHARD_Y filter has a large transient lasting ~1second, not seen in DHARD_Y inmon. The only filter that is on during that time FM6 and although it has a larger step response, there is a 5s TRAMP so Daniel doesn't see an issue with this.

Nothing obvious happened to DHARD on the 29th, only alog around that date shows a change that was reverted 70928.

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 14:09, Monday 24 July 2023 (71655)

During relock on Tuesday we plan to increase this DHARD_Y ramp time (currently 5s) to see if we can reduce this transient that appears to have increased by a factor of 10 around June 29th. See attached plots of relock before and after the violins rang up.

Images attached to this comment
oli.patane@LIGO.ORG - 21:17, Monday 24 July 2023 (71672)

There were some SDF Diffs (71670) for H1:ASC-DHARD_Y_TRAMP and H1:ASC-DHARD_P_TRAMP that I accepted, which bumped the ramp times up to 15 seconds.

Images attached to this comment
H1 ISC
naoki.aritomi@LIGO.ORG - posted 14:24, Thursday 20 July 2023 - last comment - 15:01, Thursday 20 July 2023(71557)
Follow up of camera freeze on July 20th

As reported in alog71528, the ETM cameras were frozen last night, and the camera guardian went to WAIT FOR CAMERA state and came back in 1 min. This process was repeated three times as shown in the attached figure.

Looking back the guardian log, the ETMX camera (PIT2, YAW2) was frozen. I also looked at the VALID channel for ETMX/ETMY cameras mentioned in alog71460 (H1:VID-CAM25_VALID for ETMX, H1:VID-CAM27_VALID for ETMY). The VALID channels of ETMX/ETMY were 1 at that time, which means that the cameras were not actually frozen. So we probably can use this VALID channel to solve the short camera freeze issue for ETMX/ETMY. Unfortunately, the BS camera does not have the VALID channel since the ETMX/ETMY cameras are running on h1digivideo3, while the BS camera is running on h1digivideo2.

Images attached to this report
Comments related to this report
patrick.thomas@LIGO.ORG - 15:01, Thursday 20 July 2023 (71562)
It may be that the VALID channel was frozen, but I'm not certain how likely that is or how to best tell. Maybe if there was a way to put a source of dither on the light the camera is seeing, but then that would interfere with the servo. The best solution might be to put the cameras that are used for servos on Beckhoff. The issue there is that it is not well suited for providing a remote camera image display.

We could additionally add a keep alive signal that alternates between 0 and 1.
H1 SQZ
victoriaa.xu@LIGO.ORG - posted 14:01, Thursday 20 July 2023 - last comment - 16:32, Thursday 20 July 2023(71552)
Trying various FC detunings and generated sqz levels

We looked at some filter cavity detunings and slightly varied the generated squeezing level, to see if we could at all change the noise around 70-100 Hz. It's not entirely clear what's going on, no huge obvious changes, but perhaps subtle ones. At least, even at low frequencies < 100 Hz, it seems like higher NLG is slightly better, and that moving the FC detuning to -25 Hz is not obviously worse. To start, I just tried moving FC detuning up/down 5 Hz, and rotating sqz angle to see if anything was better, see the trends of the moves here.

From the screenshot, red + cyan are from hot om2 last time on 6/28. Other traces are from the recent heating of OM2; the recent no-sqz seems to have other injections going on at the same time. There may be a slight difference between brown FDS (before moving anything), the yellow FDS (moving FC detuning from -30 Hz to -25 Hz, and slightly increasing NLG), and the dark blue FDS (less NLG, same FC - 25Hz detuning, different sqz angle apparently).

Images attached to this report
Non-image files attached to this report
Comments related to this report
victoriaa.xu@LIGO.ORG - 16:32, Thursday 20 July 2023 (71564)

For now, I have left the squeezer in the "yellow" configuration from above (previously we were in "brown" configuration). Differences are 90uW OPO trans (compared to 85uW), and filter cavity detuning at -25 Hz (instead of -30 Hz). Squeeze angle is the same, 145 deg. In SDF, I've accepted the FC detuning change, and OPO temp change. I think the difference might be marginal at best, but interested how it goes when IFO is thermalized. Screenshot showing ~2 hours into lock in the black trace here, looks similar to the yellow.

Images attached to this comment
H1 AOS
robert.schofield@LIGO.ORG - posted 13:12, Thursday 20 July 2023 (71555)
Several viewports checked for scattered light and some shaking

While LLO was down, Genevieve, Lance and I worked on checking for scattered light at a few viewports and shaking at the ITMY cryobaffle in the 10 to 20 Hz region. We may have caused a lockloss when we were checking the top viewport of HAM2 on the +Y side.

H1 General (Lockloss)
anthony.sanchez@LIGO.ORG - posted 12:37, Thursday 20 July 2023 - last comment - 15:56, Thursday 20 July 2023(71553)
Thursday Lockloss report.

First Lockloss of the Day: 18:14 UTC
Lockloss tool failed on analyze of this LockLoss. 
https://ldas-jobs.ligo-wa.caltech.edu/~lockloss/index.cgi?event=1373912078
Happened while the PEM crew were potentially working around HAM2's sensitive areas. Possibly could have been them.

Screen shot of lockloss select's ndscopes attached.

Images attached to this report
Comments related to this report
anthony.sanchez@LIGO.ORG - 15:56, Thursday 20 July 2023 (71568)Lockloss

The second Lockloss: 19:25 UTC Lock loss was certainly due to a button being pressed that then caused a lockloss.
https://ldas-jobs.ligo-wa.caltech.edu/~lockloss/index.cgi?event=1373916371

"When I set H1:LSC-MICHFF_GAIN back to 1 from zero, we lost lock
This is because I had FM7 and FM6 on in that filter bank, when I should only have had FM7."
~The button pusher.


The third lockloss was immediately afterwards: Was from OFFLOAD_DRMI and almost certainly cause by poor alignment.
https://ldas-jobs.ligo-wa.caltech.edu/~lockloss/index.cgi?event=1373916371

Images attached to this comment
H1 ISC
gabriele.vajente@LIGO.ORG - posted 11:58, Thursday 20 July 2023 (71548)
1.34 Hz peak in DARM is SRCL re-injected by the SRCLFF

The DARM error signal RMS is dominated by the large peak at 1.34 Hz. This is coherent with SRCL. I don't know what is the origin of the peak in SRCL, but today I did a quick test by turning off MICHFF and SRCLFF.

When SRCLFF is off, the peak is still present in SRCL, but it disappears from DARM. So the SRCLFF is injecting the peak in DARM.

One way to fix it (until we figure out why SRCL is moving that much at 1.34 Hz) would be to add a notch to the SRCLFF1 path. I have a design that should not change gain or phase of the SRCLFF abvoe 10 Hz in any significant way, while notching the 1.34 Hz peak by 20 db. And also a more aggressive high pass filter, that might work without spling the SRCL subtraction at 10Hz

Images attached to this report
H1 ISC
jenne.driggers@LIGO.ORG - posted 11:30, Thursday 20 July 2023 - last comment - 14:25, Thursday 20 July 2023(71549)
Potential discrepancy between GDS-CALIB_STRAIN_NOLINES and GDS-CALIB_STRAIN_CLEAN

Derek noticed (via the summary pages) that there seems to potentially be an occasional discrepancy between CALIB_STRAIN_NOLINES and CALIB_STRAIN_CLEAN.

The NonSENS subtraction should be off.  The first attachment shows that indeed the noise estimate output by the front end NonSENS is zeros to the 1e-27 level (these are in the same strain units as GDS-CALIB, and I just didn't bother scrolling in more to see how 'zero' I can make the y-axis) during a time when the discrepancies are visible on the summary page. 

A next step is likely to try to zoom in on the time axis of the spectrograms from the summary page to see when these things are happening, and see what other things might be changing around those times (eg, are those the beginning of Observe segments or going in/out of NLN_CAL_MEAS?).  Since the output of the noise estimate from the front end is zeros, _NOLINES and _CLEAN *should* be identially the same. 

Images attached to this report
Comments related to this report
derek.davis@LIGO.ORG - 14:25, Thursday 20 July 2023 (71558)DetChar

I investigated this issue further by looking at the data for the NOLINES and CLEAN strain channel between 6 and 8 UTC. This time period contained multiple times where the summary pages indicated that NOLINES and CLEAN were not identical (see attached plot).

When directly comparing the data for these channels saved in the H1_HOFT_C00 frames, I find no differences between the two channels. The difference noted on the summary pages appears to be a summary page issue/bug, and not an issue with the strain data. 

The earliest day these anamolous features appeared on the either the L1 or H1 summary pages is June 5, which is the same day as a summary page update that included changes to the strain channels.

Images attached to this comment
H1 ISC
ryan.short@LIGO.ORG - posted 06:56, Thursday 20 July 2023 - last comment - 11:48, Thursday 20 July 2023(71538)
TOO - MICHFF Measurement

While LLO was relocking this morning, I dropped observing from 13:44 to 13:54 UTC to run Gabriele's MICH FF template with 35 averages.

The measurement is saved in /ligo/home/ryan.short/MICH_excitation_2023_07_20.xml

Images attached to this report
Comments related to this report
gabriele.vajente@LIGO.ORG - 07:58, Thursday 20 July 2023 (71540)

Using this measurement, I retuned the MICH FF. The new filter is FM6 '7-20-23'. The attached plots show the fit and a comparison of the performance of the new filter and the old. There's a lot of improvement above 60 Hz, and a few db improvement below 60 Hz

Images attached to this comment
ryan.short@LIGO.ORG - 08:14, Thursday 20 July 2023 (71542)

I've accepted the new FM6 filter in SDF as well as updated and loaded ISC_LOCK to use this filter when locking.

gabriele.vajente@LIGO.ORG - 08:25, Thursday 20 July 2023 (71544)

Here's a plot showing the DARM improvement with the new MICH FF, that gave us about +6 Mpc

Images attached to this comment
gabriele.vajente@LIGO.ORG - 09:30, Thursday 20 July 2023 (71547)

Here's a second iteration of the MICH FF, with much better performance below 60 Hz and marginally worse above 60 Hz. I think this is a better FF, so we'll leave it running as FM7 '7-20b-23'. ISC_LOCK has been updated

Images attached to this comment
jenne.driggers@LIGO.ORG - 11:48, Thursday 20 July 2023 (71550)

Here is a different view of the impact of Gabriele's MICH FF tuning this morning.  Seafoam and purple are the same as in alog 71529, and red is today after the MICH FF tuning.  All of these are with hot OM2 and no MICH ASC offsets. Seafoam is late June, purple is yesteray, and red is today.  Red only has 94 averages because there were other commissioning tasks ongoing, but the other two both have 200 avgs.

The MICH FF tuning got us all the way back to where we were in June, with the excellent ~70 Hz sensitivity.  There is still some jitter-like noise, but that is unchanged from yesterday.

Images attached to this comment
H1 ISC
jenne.driggers@LIGO.ORG - posted 13:08, Wednesday 19 July 2023 - last comment - 11:49, Thursday 20 July 2023(71510)
Range improves with AS36Q offset removed when OM2 is hot

Last night and this morning, we've turned the OM2 heater back on.  Our range didn't improve, however.  We still had in place the AS36Q yaw offset from last Thursday (alog 71309).  At the start of our Wed commissioning period, I turned that offset off.  This improved our range back to ~140 Mpc according to CAL-DELTAL on the wall, but our range is still not back to where it was the last time we had OM2 hot. We may need to think about more alignment changes / checks.

Comments related to this report
jenne.driggers@LIGO.ORG - 13:26, Wednesday 19 July 2023 (71511)

Naoki just reminded me that we probably need to revert the MICH LSC feedforward to the hot OM2 setting.  We'll do that once the PEM team is between tests.

jenne.driggers@LIGO.ORG - 13:52, Wednesday 19 July 2023 (71512)

I've just removed the AS36Q offset from ISC_LOCK's ADS_to_camera state and reloaded. I've accepted the SDF with the offset of 0.  I've also accepted the TRAMP of 10 sec, which is what it should come back to when the safe.snap is reverted.  So, hopefully operators won't see any SDF diffs that they need to address.

naoki.aritomi@LIGO.ORG - 14:04, Wednesday 19 July 2023 (71513)

The MICH FF change between cold and hot OM2 is written in alog71285. We switched the MICH FF filter from FM5 to FM3 while NLN. We also implemented it in the ISC LOCK guardian and reloaded it following alog71285.

jenne.driggers@LIGO.ORG - 14:21, Wednesday 19 July 2023 (71514)

Unfortunately, I'm not so sure that we like this version of MICH FF. In the attachment, the green dashed trace is what we have right now (LSC MICH FF is FM3 and FM10, the hot OM2 settings), and the red trace is what we had before Naoki reverted to the hot OM2 settings (FM5 and FM10, nominally the cold OM2 settings).  The hot OM2 FF has increased our coherence with LSC MICH today.

It's possible that OM2 isn't all the way thermalized yet and that matters, or more likely, we have some other unknown alignment change that has caused this to be less effective than it was in the past.

Images attached to this comment
oli.patane@LIGO.ORG - 16:15, Wednesday 19 July 2023 (71524)

LSC-MICHFF diff accepted

Images attached to this comment
jenne.driggers@LIGO.ORG - 16:42, Wednesday 19 July 2023 (71529)

Here are two quick plots comparing today's hot OM2 configuration versus late June when we had excellent 70 Hz noise. The traces and data are the same, just the x-axis is different.

The seafoam is June 28th at 08:00:00 UTC, hot OM2, no deliberate ASC offsets.  Brown is today's hot OM2 (19 July 2023 18:15:27 UTC) before I removed the AS36Q offset, and purple is after removing that offset (19 July 2023 19:41:09 UTC).  So, purple should be the same as seafoam since in principle the configuration is all the same, but clearly it's not.

One suggestion is that perhaps there are some squeezer adjustments that could affect the 70 Hz region; I believe that Sheila and Naoki were discussing some potential tests. 

We should double check, but I don't think the residual LSC MICH coherence explains the whole difference. 

Its' certainly interesting to note that the brown trace (hot OM2, but with the AS36Q MICH ASC offset that we had found while OM2 was cold) seems to have very little jitter coupling, and also a relatively small amount of laser noise at high frequency.  It does however have lots of excess noise below 60 Hz, which is consistent with what Marissa had been seeing earlier today.

Images attached to this comment
jenne.driggers@LIGO.ORG - 11:49, Thursday 20 July 2023 (71551)

Indeed most of the sensitivity around 70 Hz was recoveredy by MICH FF: See alog 71550.

H1 General (TCS)
oli.patane@LIGO.ORG - posted 12:30, Wednesday 19 July 2023 - last comment - 12:59, Thursday 20 July 2023(71506)
Pushed out of Observing by TCS-ITMX SDF Diff

At 18:50, we were briefly pushed out of observing when the ITMX CO2 laser lost lock, and the resulting SDF differences in H1:TCS-ITMX_CO2_CHILLER_SERVO_GAIN & H1:TCS-ITMX_CO2_PZT_SERVO_GAIN (see attached). The SDF differences disappeared on their own about a minute later as the CO2 laser locked back on, and we observed that the power in H1:TCS-ITMX_CO2_LSRPWR_HD_PD_OUTPUT went down slightly from 46.7W before it lost lock to 46.4W after it locked back on.

At 18:53 we put the detector back into Observing

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 12:59, Thursday 20 July 2023 (71554)

This is the third time this has happened in 1 month, see 7121370910. We should investigate what the issue is. We had seen similar behavior with CO2Y in the past and blamed the CO2 chiller, 54980

H1 DetChar (DetChar)
taylor.starkman@LIGO.ORG - posted 10:24, Wednesday 19 July 2023 - last comment - 10:07, Thursday 20 July 2023(71501)
Effect of violin mode amplitude on narrow lines near the violin modes

We conducted a study using data from the Fscan pipeline that shows when the violin modes are rung up, there is a dramatic increase in the number of narrow lines (of the type problematic for CW searches) in the regions around the violin modes and harmonics. This increase in the number of narrow lines is visible both in daily and weekly plots since June 29th when the violin modes and its harmonics are known to have rung up. 

Figure 1 shows how the number of lines jumped significantly in the 300-600Hz and 900-1200Hz bands from June 29th to June 30th, and again from July 9th to July 10th without much change in the other frequency bands. Figures 2 and 3 show the averaged spectrum for the 29th and 30th respectively, demonstrating just how dramatic the increase in the number of lines is. Figure 4 shows how visible and persistent the lines are in the weekly data. 

Figures 5, 6, and 7 show the max amplitude of the violin modes vs number of lines per the square root number of SFTs in three different frequency bands surrounding the violin modes and its harmonics. To remove the effect of the duty cycle on the number of lines counted, the total number of lines counted is divided by the square root of the number of SFTs produced in a day because the number of SFTs produced in a day is directly proportional to the amount of time spent observing on that particular day. From here on out, I will refer to the number of lines per square root of the number of SFTs as the number of lines.

Looking at all the O4 data so far, it is obvious that the number of lines increases with the max amplitude violin modes in the bands surrounding the violin modes and its harmonics. The most dramatic effect is visible in the 300-600Hz band (Figure 5) where the number of lines seems to increase approximately linearly with the max amplitude of the violin modes between amplitudes of 10-20-10-18 strain/√Hz. Between 10-18-10-17 strain/√Hz, the number of lines increases significantly with no clear pattern. Multiple days after the June 29th ring up appear in this region, showing that these long ring ups can greatly increase the number of lines and decrease CW data quality for a significant period of time. Around the 1000Hz (Figure 6) and 1500Hz (Figure 7) harmonics, the number of lines increases approximately linearly with the max amplitude of the violin modes without displaying the jump present in the 300-600Hz band. This might be because the amplitude of the violin modes has a much smaller range of 10-19-10-18 strain/√Hz during the ring up. 

 

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 10:07, Thursday 20 July 2023 (71545)OpsInfo, SUS

Nice alog Taylor. This shows how important it is for us to understand and stop these violin ring ups, which started 29th June, links in alog71404. They decreased over July 5/6/7th as we had a long 40+hr lock where we were able to nicely damp them (overflows plot from 71129). Tagging OpsInfo and SUS. 

This could also be a good opportunity, while the lines are large, for CW to tag these lines as caused by violins if not yet done.  

H1 General (Lockloss)
camilla.compton@LIGO.ORG - posted 21:21, Thursday 06 July 2023 - last comment - 14:33, Thursday 20 July 2023(71126)
Lockloss @ 04:01UTC

Lockloss after 46h25 in NLN. 1372737707.

No obvious cause, DCPD saturation tag. DARM was the first loop to change attached plot.

NOISE_CLEAN reloaded as requested in 71124.

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 23:25, Thursday 06 July 2023 (71129)ISC, SUS

Back to NLN and Observing at 06:24UTC.

This lockloss has again rung up the violins again, 71063. 1h25m in OMC_WHITENING. ITMY was slowest to damp.

The 500Hz violins were low before the lockloss, 1000Hz and higher harmonics still a little elevated, DARM at 03:50UTC attached.
The OMC channels H1:FEC-8_ADC_OVERFLOW_0_{12,13} were still overflowing, see attached ndscope and compare zero overflows on the "-8 days ago" lock to todays with ~500/second. Is this expected?

The last lock we had with low violins was the 42hr lock after Tuesday maintenance, 06/27 21UTC to 06/2915UTC. Search for things that changed during that time:

  • During Tuesday maintenance: We changed OM2 TSAMs 70849 but no arm alignment changes seen 70886, ISS gain was increased 70895, OMC single bounce measurement was taken 70866
  • During the long lock: Jenne adjusted the LSC input matrix 70919
  • During 06/29 lock acquisition: We had issues with ALSX PLL 70951, 70959 and the revCav alignment was tweaked 70944.
Images attached to this comment
rahul.kumar@LIGO.ORG - 07:44, Friday 07 July 2023 (71135)SUS

All the modes (both 500Hz and kHz) have been damping down nicely in the last 6hrs of so (EY20 is still higher than expected and we don't have setting for it - have tried finding one few times but it needs more effort).

camilla.compton@LIGO.ORG - 14:33, Thursday 20 July 2023 (71560)

Daniel and Sheila note that the H1:FEC-8_ADC_OVERFLOW_0_{12,13} are from before the ADC was updated so the real OMC channels are not overflowing, this is just a relic and can be used to see OMC_DPDC being closer to saturating level, could equally see using  H1:OMC-DCPD_{A,B}_WINDOW_{MIN,MAX}. 

We checked that OMC gain settings were not changed on the 29th June.

Displaying reports 17301-17320 of 86757.Go to page Start 862 863 864 865 866 867 868 869 870 End