[Mackenzie, Paul]
We took some initial sweeps of the PRC with the aux laser this afternoon with the PRMI locked on sidebands. We'll save most of the analysis for later, but just as a quick update, I've attached data from sweeps around two FSRs which are 20 FSRs apart, as well as a full FSR sweep. Just estimating the peak frequencies of these sweeps gives a rough FSR estimate of 2.6005MHz, or 57.6413m for the PRC length. We'll increase the frequency resolution on subsequent scans, and take more data from FSRs over a greater frequency range for the Schnupp measurement. In the end we'll do some fitting rather than just guessing peak frequencies too. More to come later.