X1 SUS
andres.ramirez@LIGO.ORG - posted 13:33, Friday 22 March 2013 - last comment - 14:51, Friday 22 March 2013(5865)
Phase1b I1-MC3 TFs Results
Here are the Tfs and Spectra Test results for I1-MC3.

Non-image files attached to this report
Comments related to this report
jeffrey.bartlett@LIGO.ORG - 14:08, Friday 22 March 2013 (5867)
These plots were made using incorrect Burt settings in MEDM. We will load the correct Burt file and rerun the transfer functions. 
mark.barton@LIGO.ORG - 14:51, Friday 22 March 2013 (5868)
Further to previous, the problem was that HLTS settings had been loaded instead of HSTS ones. Thus the R, P and Y curves are shifted vertically by the square of the ratio of the respective lever arms (mostly noticeable in R and P). Allowing for that, the curves look very good, and we simply need to retake the data.