(Eric A., Corey G.) Over the last few days, we've bypassed our original BNC feed-thrus on our Interface Board, and are now using actual BNC feed-thrus. The issue here is that the original BNC feed-thrus proved to be troublesome and sacrificed the performance of our Sensors. We've found that using the "real" feed-thrus (and also buying better/expensive feed-thrus) have made our Sensors perform much better. Switching to the new feed-thrus has changed the DC value of of our sensors. At "zero" our Sensors were all under ~1000 counts, but with the new feed-thru the zeroes have increased to around 10,000 counts. Because of this we had to re-gap the Sensors. This sort of change required us to re-do a few more Sensor Tests (according to document T1000329). New Power Spectra for the Displacment Sensors have been run. The measurement is located here: /opt/svncommon/seisvn/seismic/HAM-ISI/X1/Data/unit_1/dtt/unit_1/dtt/20100720_all_disp_spectra.xml Attached are plots of spectra with BOTH Sensor Mini-Racks ON, and with only one ON.