Here is the current status of things in guardian, for people who will be locking latter tonight:
Today I attempted to confirm that the X-end RGA was still functional after having been vented, removed from its protective nipple, re-installed and evacuated yesterday. I wasn't able to do this due to an inability to connect to it with the RGA-dedicated laptop. There is some sort of network issue that has been hit or miss for awhile now but, recently, has degraded to the point of being unusable. I'll try again after investigating and/or at the next window of opportunity.
Here is the current state of the H1 DAQ.
h1fw0 and h1fw1 have been completely stable for several weeks, and following the code fix on Wednesday 9/7 the frames written by both are 100% identical. These systems have the same hardware and are running the same code, but h1fw0 occasionally asks for retransmissions and h1fw0 never does. The OS install is slightly different between the machines, and we will try cloning fw0 from fw1 next week.
h1fw2 is a front end computer, running U12 with a local disk system for frames. It was running the original daqd code, but went unstable after the Sat 9/10 power outage. We upgraded it to the rcg3.2 daqd this afternoon to see if this improves stability. It is now connected to UPS power.
Now that h1fw2 has the new set of EPICS diagnostics channels I have expanded the DAQ overview medm screen to show these.
State of H1: locking and easilly getting to DC_READOUT
Activities:
Lisa, Sheila, Ed
Durring ER9, the intensity/jitter noise in DARM was worse than inO1, but not as bad as now. We looked at the PSL accelerometers, and see that the table is moving about a factor of 5 more over a broad range of frequencies from 150-550 Hz. The coherence of DARM with PSL table acclerometers and the ISS PD is also increased compared to July. The last panel shows the coherence of the table accelerometer with the ISS PD. Ed trended PSL pressures, and found that the flow rate durring ER9 was 0.5lpm, compared to 0.7lpm now.
Past 78 days chiller trends showing mainly flowrates.
So we asked Peter to reduce the flow rate, thinking that the increased flow rate wrt July contributes to the higher intensity noise we observe now. This will be done on Monday morning.
Albert, Chandra 9:00am local 20 sec. to overfill CP3 with 1/2 turn open on bypass LLCV. Left bypass exhaust valve open. Exhaust line is still frosty, but the ice balls near exhaust valve have melted. CP3 show 'n tell for Albert Lazzarini.
This morning I opened up the ISS photodiode box. The beam incident into the box is attenuated by a polarising beamsplitter
and half waveplate combination. The rejected beam is dumped into an absorbing Schott glass filter. (see ISSBeamDump.jpg)
The power going into the photodiode box was measured to be 597 mW. For ease of installation I inserted a neutral density
filter labelled "ND=0.3" at the input to the photodiode box. (see NeutralDensityFilter.jpg). The output of PDA went from
-8.43 +/- 0.07 Vdc to -2.22 +/- 0.01 Vdc, which is somewhat more than one would expect from the label. The half waveplate
was adjusted to bring the output of PDA to -5.22 +/0 -0.03 Vdc (see BeforeWaveplate.jpg and AfterWaveplate.jpg).
rpn1.jpg shows the free-running spectra as measured by PDA and PDB after the installation of the neutral density filter.
As previously, both agree quite well. Attached are a few other recorded spectra with the differences being the gain slider
setting. The transfer functions for a few gain slider settings were measured (see ISSTF.jpg).
It is not obvious to me why the in-the-loop noise would increase as the gain increases. See for example g3.jpg and
g6.jpg. So this might be an electronics problem?
Also attached is the power noise as seen by the PDA55 photodiode that monitors the power going into the input modecleaner.
0.307 mV was the output of the photodiode when the ISS was locked. However it might be that the bandwidth of this photodiode
was only 60 kHz, depending on how the gain selector switch was set. Beyond about 1.6 kHz, the measurement is limited by the
dark noise of the photodiode.
Click here for the datasheet for this obsolete photodiode from Thorlabs
Photos of photodiodes removed from damage OMC. Photos taken by GariLynn B. at Caltech.
Photos have been posted on E1600268-v1, "Initial Inspection of OMC Bench at CIT Aug 2016"
https://dcc.ligo.org/LIGO-E1600268
Naming conventions are from E1600262-v3, "aLIGO OMC: Spare OMC production plan"
https://dcc.ligo.org/LIGO-E1600262
Diode pictures attached to this alog entry:
It seems that since the last power glitches, the DBB is back to operational status... hmmmm. That is to say, before the power glitches it was unusable due to the mysterious tripping of the (dbb) interlock when trying to open the beam path into the box. In the past week I've been trying to get good measurements. This was a abit daunting due to the ISS 3rd loop work that was breaking the IMC lock consistently and yanking on the LASER frequency. Below is what I could muster over the course of a couple of days. ISS_RPN will come later as Peter is currently in the enclosure and working on it.
Comments:
RPN - 35W path looks suspiciously good accross the spectrum by a factor af a few...hmmm? 200W path looks tyo be below requirement from 2Hz to ~25Hz and then above to about a factor of a few through high(er) frequencies.
FRQ - 35W path is quite noisy at low freq (below 15Hz). Other than that the rest looks good save for the usual resonances. 200W path looks ok across the spectrum except for the usual resonances.
PNT - 35W path pointing looks pretty good. 200W path shows seperation of the 1x/y and 2x/y by a factor of a couple.
MSC - 35W path shows TEM 20 +02 an order of magnitude higer than reference. 200W path shows TEM 02+02 mode higher by a factor of a couple.
...and this should wrap it up.
Morning Meeting:
FRS: reviewed
Currently: no locking with Peter in the PSL modifying the ISS beam path
SEI - JimW loading new code into the SEI guardian
Fire Department here to test the fire hydrants - should be here most of the morning, but probably not this afternoon.
Activities:
As of 17:45:
Tonight there was a flurry of activity with H1. There were issues with locking at the beginning of the shift, but it got better....we had a range reminiscent of H1 during S6: 19Mpc! :) Slowly but surely, we're gettin' there.
Locking Notes:
At the beginning of the shift, had issues locking ALS. This was traced down to the wrong polarizations.
FOM Note on Nuc5:
Cheryl mentioned this was booted during the day, and then it died & no DMT Viewer/DTT came up on their own. I rebooted nuc5 and conversely to last night, the DMT seismic trends looked fine & current.
I was actually loking at the "Fiber trans wrong pol (%)" numbers after a number of subsequent locking attempts and noticed a change of ≈ 4% over the course of a couple of hours. Y arm was worse than X by about 4%. Perhaps we should make this one of our weekly FAMIS checks to keep folks up-to-speed on how to deal with this? Some of us have done it numerous times and I'm willing to bet some of us have NEVER done it.
Matt, Terra
Status of PI after TCS work and ring heater change transient (for at least the first three hours of 50 W lock):
3 hours into the lock and all PI's are stable. Leaving PI control with Matt for the night.
Another 3 hour lock and similar PI story.
These modes shift very little in frequency during a lock stretch (see attached). I've tightened up the individual band passes quite a bit (~1 Hz) to see if that helps next lock. Still need to check lock to lock variation.
WP 6087 The GDS software has been updated to branch 2.17.4 to address Bugzilla 897, 936, and 952. Release notes (T1600007) document has been updated. This affects Ubuntu 12 and Ubuntu 14 workstations.
as of monday morning, fw2 has been running 2.8 days. Looks like Jonathan's code has made it stable. It is still a mystery why the power outage apparently caused the instability of the old code (it has been power cycled since then).