SRM-MC2-SR2 balanced - overnight measurements starting

[Jeff Arnaud]

Tonight we balanced the coils of the lower stages of MC2 SRM and SR2. The MC2 notes attached are explaining the procedure in detail. Below is summarized the ASD ratio after/before the exercise for each suspension and each level.
For some reason M3 stage of SR2 was harder to balance than the other ones, so we gave up on it for now and will come back to it later.

SUS STAGE        ASD RATIO before/after balancing
PITCH YAW
MC2 M2 19.71 2.63
M3 21.27 13.52
SRM M2 9.95 38.53
M3 6.96 3.8
SR2 M2    
M3 107.7 5.12

And the gains which were found optimum for each coils

SUS STAGE OSEM GAIN
MC2 M2 UL 0.988
LL -0.992
UR -1.007
LR 1.012
M3 UL 1.082
LL -0.948
UR -1.051
LR 0.922
SRM M2 UL 0.935
LL -1.024
UR -0.971
LR 1.063
M3 UL 0.972
LL -1.014
UR -0.972
LR 0.964
SR2 M2 UL  
LL  
UR  
LR  
M3 UL 1.003
LL -1.050
UR -0.951
LR 0.996

 

Measurements will be running overnight to measure the cross couplings transfer functions for SR2 SRM and MC2

Images attached to this report
Non-image files attached to this report
Comments related to this report
borja.sorazu@LIGO.ORG - 17:40, Tuesday 05 August 2014 (13223)
There is a discrepancy in the text. While it says that SR2 M3 was not balanced however in the results table it is SR2 M2 the one blank. I can confirm that SR2 M2 is the one not yet balanced.