Displaying reports 81481-81500 of 84635.Go to page Start 4071 4072 4073 4074 4075 4076 4077 4078 4079 End
Reports until 21:31, Saturday 23 June 2012
H2 ISC
rana.adhikari@LIGO.ORG - posted 21:31, Saturday 23 June 2012 - last comment - 10:36, Monday 25 June 2012(3247)
ITMY Pitch gets the green beam back to ETMY

Using ~1/2 of the ITMY pitch bias range, the green beam gets back to the ETMY.

After enabling the ITMY HEPI + ISI, I scanned the ITMY bias sliders and looked at the ETMY face camera. The ITMY EUL2OSEM matrix has gains of ~5-10 (unlike the TMSY ?) and so the +/-15000 cts of the slider screen is enough to go full scale...and consequently, I tripped the watchdogs a couple times.

The yaw alignment was really good - the return beam to the ETM was nearly centered horizontally, but needed ~1/2 of the ful pitch range of the ITMY TOP stage.

With the illuminator off, it was easy to see the beam flashing around after the watchdog tripped. With the illuminator ON, but turned down to 0.5 Amps, I could see the stuff in the chamber as well as the green beam. Then I used the usual method of putting the beam on the 4 compass points of the ETMY cage and centering it on the mirror by averaging.

Moving the yaw slider to the right moves the beam on the cage to the right. Moving the pitch slider to the right moves the beam down. It takes +1000 slider units of torque to move the beam from top to bottom of ETMY.

The values of ITMY bias that center the beam on ETMY are P = +6300 & Y = +900.

So the ITMY alignment was off by ~6 mirror diameters =>  (0.210 m / 4000 m) / 2 = 25 micro radians.

Now that the beam goes both ways, the next thing to do is to align the ETMY to center the beam on ITMY using the baffle PDs as before. Then perhaps we'll be able to hear the cavity flashes if we can route the audio from the baffle PDs or the ALS green PDs into the control room.

Images attached to this report
Comments related to this report
rana.adhikari@LIGO.ORG - 23:05, Saturday 23 June 2012 (3248)

And...lots of YAW bias gets the ETMY aligned to get the beam back on to the ITMY baffle PDs. The ETMY bias for rough alignment is:

P = -2700

Y = +14200

I haven't seen any flashing yet that looks like a Fabry-Perot resonance, but at least all of the suspensions are close enough in alignment to proceed. Since the TMSY and ETMY both have large yaw biases, we can still hope that they can be mostly relieved by yawing the HEPI (assuming they need the same direction of yaw). Because of the double-loop nature of the quads, we should also get some ITMY pitch relief from pitching its HEPI.

From the flashing that can be seen from the ITMY spool cam, it seems like the return green beam is hitting the ITMY baffle at the edge around the mirror, at the 2 o'clock position look at the ITM from the spool.

richard.mccarthy@LIGO.ORG - 07:22, Sunday 24 June 2012 (3249)
While trying to get the video rerouted to the control room as well as EY I will look at hooking up a FiBox to the green light signal. Will talk to Keita about this.
peter.fritschel@LIGO.ORG - 10:50, Sunday 24 June 2012 (3250)

There is something funny about the numbers here. The claim is that the ITM alignment was off by about 25 micro-radians, and that correcting this uses about 1/2 of the pitch control range (I assume this is at the TOP stage). However, according to T1100595-v1, the control range of the quad, from the TOP stage, is 1.14 mrad in pitch; it is a little unclear as to whether this is +/- 1.14 mrad or a 1.14 mrad total swing; either way, it is much larger than 25 urad -- it should take less than 1/10 of the range to correct it.

rana.adhikari@LIGO.ORG - 17:50, Sunday 24 June 2012 (3251)

I think its funny too. Using ~50000 cts (as measured at the DAC outputs for the face OSEMs on TOP) seems like its close to the 217 limit for the DAC.

Should have checked the coil driver readbacks to find out how much current/voltage was being used to make sure there is no attenuation happening downstream.

peter.fritschel@LIGO.ORG - 09:32, Monday 25 June 2012 (3253)

Jeff K and I found the basic problem with the ITMY alignment numbers. There is a factor of 10 error in the observed pitch misalignment number. The 0.21 m above should be 2.1 m (6 * 0.34 m). The pitch offset is thus 250 micro-radians. We also noted that there seems to be a factor of 2 discrepancy between the designed input range for the TOP driver and the DAC output range, so that we can only access half the current range for each BOSEM: the TOP driver was designed to give +/- 200 ma for each coil, but we can only go to +/- 100 ma. So, the 250 micro-rad of offset is about half the pitch bias range we have available, which is about half of what we should have.

jeffrey.kissel@LIGO.ORG - 10:36, Monday 25 June 2012 (3254)
P. Fritschel, J. Kissel

The confusion lies in Rana's calculation of the total distance for beam travel:
1000 cts of torque = beam moves top to bottom of ETMY
the ETMY is 34cm in diameter.
So, a 6300 cts offset, is
6300 [ct]* 34/1000 [cm/ct] = 214 [cm] = 2.14 [m]
not ~0.210 [m] as Rana uses in his calculation (i.e. he's off by a factor of 10). So, correcting the number in his calculation, we get

pitch offset = torque [ct] / (2 L) % factor of two from "optical lever" effect see G1200698
             = (2.14 m / 4000 m) / 2 
             = 250 micro radians. 


This jives with the predicted range of the QUAD TOP driver and the current monitors:

Predicted Range
T0900232 says, in section 6.1: 
"As the input voltage is specified as 10V* and the voltage gain of the circuit is 1.2** the voltage swing of each output terminal is +/-12[V]. As each output resistor is 40 Ohms, and the [B]OSEM coil resistance is 40 Ohms, the total output resistance is 
     (40 [Ohms] + 40 [Ohms] + 40 [Ohms]) = 120 [Ohms]. 
The voltage across this resistance will be 24[V] maximim, so the [maxmimum] current will be
     24 [V] / 120 [V/A] = 200 [mA] "  

However,
* I've confirmed via direct measurement that the 18-bit DACs can only do +/- 10 [V] differentially NOT per leg. So the max possible voltage is
     DAC_MaxV = 10 [V] (differential output)
** The DC gain of the circuit is 
     CD_VoltageGain = (1+R3/R10) = 1.22 [Vin/Vout] 
(OK, 2% error, but just wanted to make sure we understood from where Ron's numbers came)
And just in case, to remove all suspicion:
From the QUAD Top Driver board schematic, D0902747:
-      Rout = (R1+R5)*(R90+R91)/(R1+R5+R90+R91) 
          = (22+22)*(220+220) / (22+22+220+220) [Ohms]
          = 40 [Ohms],
so that's correct, and Stuart just measured the BOSEMs to have a resistance of 
-      Rcoil = 42.7 [Ohms] , 
so that's correct as well.

So, the maximum output current of the circuit is actually

     DAC_MaxV * CD_VoltageGain / (2*Rout+Rcoil) = 10 [V] * 1.22 [V/V] / (40+40+42.7) [V/A] = 0.09943 [A] = 99 [mA]

which if all other numbers from T1100595 are correct, then this would produce a maximum displacement of

Pitch: 560 micro rad
Yaw: 580 micro rad


Current Monitors
The current monitors have a calibration of 

Amps across the BOSEM coil = ADC [V/ct] * Fast_I Mon Circuit Gain [V/V] / BOSEM Coil [V/A]
                           = 40/2^16 [V/ct] * 3/2 [(single ended Vout) / (differential Vin)] * / 42.7 [V/A]
                           = 2.1441e-05 [A/ct]

The current monitors for the primary Pitch OSEM (F1) on the ITM, with a 6400 [ct] torque offset from the sliders, show -2736 [ct] of current. (I use F1 and not either F2 or F3, because F2 and F3 also contain the Yaw offset). That means the current across the coil is

Icoil = 2736 [ct] * 2.1441e-05 [A/ct] = 0.059 [A] = 60 [mA] 

which is indeed about half the range of the coil driver, which jives with both the [corrected] calculation of 250 urads and the observation that Rana saw ~50000 cts (or half the range of the 2^17 ct DAC max).



Note that the original design intent was to have ~1 mrad of range from the TOP stage, so if we want to restore that intent, we should reduce the output resistance of each leg by a factor of 4, such that the range is

DAC_Max * CD_VoltageGain / (2*Rout+Rcoil) = 10 [V] * 1.22 [V/V] / (10+10+42.7) [V/A] = 0.19458 [A] = 194 [mA]

which could be done by changing the values of R1, R5, R90, and R91 to 10 Ohm resistors.


H2 SEI
rana.adhikari@LIGO.ORG - posted 18:23, Saturday 23 June 2012 (3246)
ITMY HEPI Filter suggestion

The attached plot shows the control filter, the existing boost filter, and a proposed filter addition to the HEPI controls.

As we had at LLO, I think we should have a true integrator on the HEPI loops to compensate for long term drifts introduced by hydraulic pressure, thermal expansion, etc. so that the HEPI platforms  can be considered to be stable alignment-wise.

Also, why is there so much high frequency gain? It seems like the actuator watchdogs will trip due to acoustic noise or EMI with this kind of design. Perhaps a little low passing can be done without too much harm to the 10 Hz phase margin.

Non-image files attached to this report
H2 AOS
rana.adhikari@LIGO.ORG - posted 18:00, Saturday 23 June 2012 (3245)
ITMY Baffle PD preamp swapped

The SR570 (not 560) current preamp that we had been using since yesterday afternoon for channel 27 of the baffling diodes, ran out of batteries today at around 1330 today (so we know that a fully charged one will last just under 24 hours).

At that point it started producing glitches of ~30000 cts in channel 27 with a steady ~4 second period. Because of some grounding issue somewhere, it also produced smaller glitches in the other 3 channels. These glitches looked somewhat like a widly swinging green beam, so I spent a little while investigating the health of the various seismic filtering servos for the ETMY/ITMY before guessing that it was too regular to be mechanical.

Preamp has been switched back to the Melles Griot 13AMP003 "Large Dynamic Range" Amplifiers that the other channels are read out with. The SR570 on batteries was nicer than these; with a little bit of thought, I guess we could figure out how to adapt the ISC QPD readout board into one that can be used for this application (FET preamps, high transimpedance, some low pass fitlering) and then switched out to a lower gain one once we get into real interferometer locking.

After swapping, I used ezcaservo to remove the offsets on all the channels using commands like this one:

ezcaservo -r H2:PEM-CS_CHAN_27_OUT16 -g -0.21 H2:PEM-CS_CHAN_27_OFFSET -t 20
Images attached to this report
H2 General
rana.adhikari@LIGO.ORG - posted 21:04, Friday 22 June 2012 - last comment - 11:05, Monday 25 June 2012(3244)
Green Beam Found at ITMY!

Daniel, Keita, Rana, Aidan, Dale, Dick

Using the ITMY baffle PDs, we were able to get the green beam from EY to ITMY today! The aLIGO beam tubes are able to transmit light.

We spent hours today trying out various ideas about how the beam might be missing, but in the end we found it by moving the EY Transmon Suspension.

The MEDM screen for the TMS had a range of +/- 15000 cts. In fact, the TMS has an 18-bit DAC and so its full range is +/- 2^17 = +/- 131072 cts.

 

With the current ALS table alignment, the TMS biases which give the maximum voltage on each baffle PD channel are:

				chan name
				P
				Y
				CS_CHAN_27
				-10000
				+2900
				CS_CHAN_28
				-69000
				+1200
				CS_CHAN_29
				-48000
				+1000
				CS_CHAN_30
				-46700
				+17800

We used these numbers to "triangulate". Then we used the ITMY spool cam to further center the beam on the ITM. It is not at all visible on the ITM surface, but can barely be seen on the gate valve behind the ITM as well as a little glint on the edge of the hole in the baffle where the cavity axis is. Our random search pattern initially just gave flashes of ~50 cts, so the new digital filters helped after all. Eventually, the signal on each PD can separately be made to saturate at 32768 cts as we align the beam onto it.

The final TMS settings for "good alignment" to the ITM are P = -16000 and Y = +9250.  Correction: there were some digital offsets in the 'test' filter bank. After zeroing those out, the new numbers are P = -42800 & Y = +59200.yes

Vincent has turned on the ETMY HEPI with low frequency boosts (although no integrators yet) so the pointing should be more immune to the HEPI hydraulic pressure fluctuations. Still need to fix up the ITMY HEPI. There's some seismic measurements going on overnight, but we can next align the ITMY to get the beam back to EY and then lock the green laser to the arm.

Comments related to this report
keita.kawabe@LIGO.ORG - 10:44, Monday 25 June 2012 (3255)

Angle calibration of TMS OSEM:

When tilting TMS without tilting the ALS beam on the table, the beam leaving the TMS telescope would tilt by almost the same amount (19/20 or 21/20) as the TMS itself. Here I just ignore 5% error and assume that the beam tilts as much as the TMS.

Using the baffle diode (https://dcc.ligo.org/cgi-bin/private/DocDB/ShowDocument?.submit=Number&docid=d0901376) we can calibrate the TMS OSEM. Note that the PD locations are not the same for H1 and H2. We're looking at H2 ITMY (left hole on page 3 of the above document).

There are three PDs horizontally aligned. The top one (PD3) is about 6.25 in away from the middle (PD2). The bottom one (PD4) is 11in lower than the middle. We know (from experience) that increasing P offset moves the beam down. And there's a fourth diode (PD1) that is at the same height as the middle one and horizontally off to the left by 11 in. PD1-PD4 naming convention is explained in E1100867, page 15.

 

PD3=CHAN_28

(P.Y)=(-69000, +1200)

PD1=CHAN_30,

(P,Y)=(-46700, +17800)

PD2=CHAN_29,

(P,Y)=(-48000, +1000)

 

PD4=CHAN_27,

(P,Y)=(-10000, +2900)

For PIT, PD3-PD2 is about 6.25 in for 21000 counts, so it's 6.25*25.4e-3/21000/4000 = 1.9 nrad/count.

Also for PIT, PD2-PD4 is about 11 in for 38000 counts, 11*25.4e-3/38000/4000 = 1.8 nrad/cout.

For YAW, PD1-PD2 is about 11 inches  for 16800 counts, so 11*25.4e-3/16800/400 = 4.2 nrad/count.

There's no funny thing in the output matrix, PIT to coil matrix elements are just +-1, same for YAW, so you can multiply 131072 to get the range.

We can also get the initial alignment number using the offsets as of now (-42800 for PIT and +59200 for YAW)

  PIT YAW
Calibration

1.9E-9 rad/count

Positive offset moves the beam down.

4.2E-9 rad/count

Positive offset moves the beam towards the inside of L.

Range +-0.25E-3 rad +-0.55E-3 rad
Initial alignment

Off by +81E-6 rad

The beam was pointing down.

Off by -250E-6 rad

The beam was pointing outside the L.

keita.kawabe@LIGO.ORG - 11:05, Monday 25 June 2012 (3256)

Note that the order of the channel numbers (27, 28, 29, 30) is reversed. Right now PD1, 2, 3 and 4 = Chan30, 29, 28 and 17.

H2 SUS
szymon.steplewski@LIGO.ORG - posted 18:23, Friday 22 June 2012 (3242)
Characterizing AOSEMs
M. Barton & S. Steplewski

Today we characterized an advanced LIGO AOSEM (Serial Number 003, without an IR filter) using a 3-axis translation stage in the electronics shop. The voltage from the photodiode was recorded as a function of position in each of the three different axes. As shown in the attached graphic, the X axis goes through the circular coil (longitudinal to the OSEM) the Y axis is along the LED beam path, and the Z axis is vertical.

We found the expected sigmoidal curve by moving the flag in the longitudinal (X) direction, and computed a rough slope of about -19.4 Volts/mm in the central linear region. We also saw that moving the flag along the Z axis gives a flat response over about 1 mm of range, even when we moved to different X offset values. This means that the Z position can be slightly off-center and the OSEM will still work well as a longitudinal sensor. However, when the Z position of the flag is moved to a larger offset the longitudinal sensing is reduced, as can be seen in the first graph green and yellow curves. We also looked at the Y axis and found pretty flat responses, even when varying X.
Images attached to this report
Non-image files attached to this report
LHO General
cheryl.vorvick@LIGO.ORG - posted 16:37, Friday 22 June 2012 (3243)
Viewport Guard Lexan (AR Polycarbonate)
Three versions running around LHO:

D080366-00
D080366-v1
D080366-v2 - taller than -v1

All are AR Polycarbonate, AR just means abrasion-resistant.

None found without DCC and serial numbers, though the big stash that's rumored to be here also not found (at 4PM on a Friday, go figure).

FYI: MSDS I found online, so not delivered with any of our orders, but still AR Polycarbonate, shows:
Softening Point: 150 - 160 °C (302 - 320 °F)
Melting Point: 220 - 230 °C (428 - 446 °F)
H1 TCS
aidan.brooks@LIGO.ORG - posted 16:20, Friday 22 June 2012 (3241)
Hartmann sensor imaging of ETM

[Aidan, Jeff G, Thomas, Elli]

We've been running the HWS without the Hartmann plate in place to simply measure the position of the return beam from the ETM on the sensor. We have been yawing L1 on the ETM quad suspension with a 0.35Hz sine wave excitation of 5000 counts into H2:SUS-ETMY_L1_TEST_Y.

We measured the ratio of the 0.35Hz signals on the HWS to the excitation (in magnitude and phase). The intent is to move the HWS along the optical axis (or adjust F1) such that the ETM HR surface is imaged on the HWS CCD (if this is the case, then a yawing of the ETM HR surface should produce no motion on the HWS CCD).

With a ruler bolted next to the HWS the following measurements were taken. Smaller values of position are closer to the lens F2. The 18 degree phase delay corresponds to about 140ms time delay (most likely associated with recording images, analyzing them and writing HWS data to EPICS - which occurs at roughy 7Hz).

HWS position - arbitrary Ratio [mm/excitation count] Phase
99.5mm 16.6E-9 18 degrees
92.5mm 17.9E-9 18 degrees

This implies that we need to move the HWS to the position 189mm to be imaging the ETM HR surface. Alternatively, we could move the lens F1 by apprxoimately 3.5mm (increasing the distance between it and F0).

H1 TCS
aidan.brooks@LIGO.ORG - posted 16:20, Friday 22 June 2012 (3240)
Hartmann sensor imaging of ETM
LHO General
michael.rodruck@LIGO.ORG - posted 16:08, Friday 22 June 2012 (3239)
Ops summary
LHO VE
kyle.ryan@LIGO.ORG - posted 16:02, Friday 22 June 2012 (3238)
Briefly soft-closed GV17 for Rana and Keita to temporarily expose viewport on BSC6


			
			
H2 INS
cheryl.vorvick@LIGO.ORG - posted 16:00, Friday 22 June 2012 (3237)
Illuminator at EY has some foil over it.
I've put a loose foil cover over the illuminator at EY, where John took the Lexan out yesterday.  The official dust cover for the lexan slit that we do have, D0902791-v1, doesn't fit tightly over the slit, so flops back and forth leaving a gap, so not an improvement.  That part is on -v2 (D0902791-v2), which I believe will fit tightly, and modifying -v1 would be a good short-term fix.  However, the better plan is to beef up the dust cover into a part that fits into the slit, and thus holds itself in place, which can then be secured with a couple screws.  In the mean time, I was uncomfortable leaving the slit uncovered, and folded a piece of foil to prevent something small, like a screw, from falling through the slit and hitting the viewport.  I called John and talked it over, so he's aware of the very temporary solution.
LHO VE
kyle.ryan@LIGO.ORG - posted 13:34, Friday 22 June 2012 (3231)
Y-end RGA scan. Also, changed IP11 voltage step current
Attached is an RGA scan of the Y-end taken this morning (RGA parameters are not optimized for current conditions but, rather, are left unchanged from initial scans for comparison purposes).   

I changed the programmed pump current value at which point IP11 "steps" from 7000V to 5000V from 5x10-4 amps to 1.5x10-3 amps.  Prior to this change the pump current was 1.4x10-3 amps @ ~8.5x10-8 torr.  The result is that IP11 is now at 5000V.
Non-image files attached to this report
X1 SUS
stuart.aston@LIGO.ORG - posted 11:37, Friday 22 June 2012 - last comment - 07:20, Monday 25 June 2012(3229)
PR2 (HSTS) Phase 1b testing M1-M1 transfer funtions plus M1, M2 and M3 power spectra
[Stuart A, Betsy B, Deepak K, Andres R, G2]

After taking an initial set of M1-M1 transfer functions earlier in the week on PR2 (HSTS), it was observed by Jeff B that somehow a magnet had become detached from the M2 mass (see LHO aLog entry 3157). 

Thanks to the efforts of the assembly team, this magnet was rapidly re-attached and left to cure for a period of 24 hrs. PR2 was re-suspended yesterday, and a small pitch offset corrected. AOSEMs were then adjusted to their final operating positions and aligned. The canopy/cover was fitted over PR2. Following this re-work, it was prudent to take another M1-M1 transfer function with damping loops OFF for all degrees of freedom, which can be found below (see 2012-06-21_1700_X1SUSPR2_M1_ALL_TFs.pdf). 

Damping loops were then turned ON and a further complete set of M1-M1 transfer functions taken overnight. All the transfer functions obtained have now been plotted against all other HSTS suspensions previously measured on both LLO and LHO test-stands (see allhstss_2012-06-22_AllHSTS_ALL_ZOOMED_TFs.pdf).

n.b. 
Yellow trace = X1 PR2 M1 (2012−06−21_1700) with damping loops OFF
Purple trace = X1 PR2 M1 (2012−06−21_2120) with damping loops ON

The transfers functions obtained again demonstrate good agreement with the model and the spread of all HSTS measurements obtained thus far. 

Power spectra have been taken with damping loops both ON and OFF for each stage (012-06-22_0800_X1SUSPR2_M*_ALL_Spectra.pdf).

Power spectra plots, with both damping ON and OFF have been produced, which compare LHO PR2 and LHO MC2 measurements (allhstss_2012-06-22_ALL_Spectra_Don.pdf and allhstss_2012-06-22_ALL_Spectra_Doff.pdf).

In addition, power spectra for specific degrees of freedom (L, P and Y) can be more conveniently compared across multiple stages (M1, M2 and M3) of the same suspension in the final plots found below (allhstss_2012-06-22_X1SUSPR2_M1M2M3_Spectra_ALL_Don.pdf).

A BURT snapshot has been taken of the current functioning environment "20120622_x1sushxts27_PR2.snap", which has been stored in the following directory:-
opt/rtcds3/tst/x1/cds_user_apps/trunk/sus/x1/burtfiles. This BURT snapshot directory has also been tidied to remove old or incomplete snapshots.

All of the above data, plots, scripts, and snapshots have been committed to the SUS svn as of this entry.

This should now provide sufficient measurements to complete Phase 1b testing of the PR2 suspension and allow it to progress to Phase 2 (chamber-side) testing.
Non-image files attached to this report
Comments related to this report
stuart.aston@LIGO.ORG - 15:01, Friday 22 June 2012 (3232)
The M3 stage watchdog was observed to be perpetually tripping and could not be reset. This is due to watchdogs being triggered for the OPELV_RMS and OPLEV_SUM, which are not visible/settable in the medm screens. 

To rectify this I have manually set the following:- 
caput X1:SUS-HXTS_M3_WD_OPLEV_RMS_MAX 5
caput X1:SUS-HXTS_M3_WD_OPLEV_SUM_MIN -10

For now, this prevents the watchdog from tripping. 
jeffrey.kissel@LIGO.ORG - 07:20, Monday 25 June 2012 (3252)
These results look excellent. 

The only thing that concerns me (where the level of concern (from 1 = "it's awesome. no worries" to 10 = "OMG take it apart and rebuild it") is a 5.5) is the cross-coupling I see in the individual comparison with the model, i.e. in
2012-06-21_1700_X1SUSPR2_M1_ALL_TFs.pdf.
In there, I see cross-coupling between degrees of freedom which we don't normally expect to be there:

1st Yaw Mode (@ 1.09 Hz) into T and L
2nd Yaw Mode (@ 2.04 Hz) into T and L
2nd Roll/Pitch Mode (@ 1.51 Hz) into Y
2nd Vert or 3rd Long Mode (@ 2.80 Hz) into Y

I'm not terribly concerned, because -- as usual -- the cross coupling is reduced to barely visible with damping loops ON. But, it's something to watch out for with this guy as he progresses through the testing (i.e. we'll look with more scrutiny after the optic is swapped). To refresh one's memory -- the reason why we care:  
(1) When the model doesn't match measurements, we can't trust the model to accurately predict other transfer functions which we can't normally measure
(2) If the transfer functions (which are representative of the "plant" upon which we design control loops) are *different* between suspensions by some (as-yet-to-be-quantified) significant amount, then it will make copying and pasting control systems more difficult and time consuming to design.
I think with this suspension, and all others, we're doing fine in both these departments -- but again, we won't really know until we start locking some cavities and really try to push the SUS's to their limit.
H2 AOS
eleanor.king@LIGO.ORG - posted 10:32, Friday 22 June 2012 (3228)
Common mode servo adjustment

Elli, Rana.   We connected a low-pass filter to the common mode servo input, to simulate the feedback response of the green laser inside the cavity. The transfer function of this loop confirmed the location of the pole and zero of the first boost filter. We removed common mode servo B and will change out some of the boost filter components in order to improve the servo transfer function.

Images attached to this report
H2 AOS
rana.adhikari@LIGO.ORG - posted 09:40, Friday 22 June 2012 - last comment - 11:11, Friday 22 June 2012(3223)
Amplitude Spectrum of Baffle PD signals (60 Hz)

This PDF shows that the RMS of the PD signals is dominated by 60 Hz + harmonics as we suspected. Unfortunately, it looks like we would have to notch out many, many harmonics to appreciably reduce the RMS. Perhaps we should use a low pass in combo with some comb filters?

Non-image files attached to this report
Comments related to this report
rana.adhikari@LIGO.ORG - 11:11, Friday 22 June 2012 (3230)

In order to reduce the noise in the channel, I added a Comb filter into the CS_CHAN_NN filter banks to knock down the first 8 harmonics. Also added a 2nd order Butterworth low pass at 11 Hz to get rid of the HF harmonics. The RMS was reduced by 100x.

Restarted the rotating lissajous search pattern on the ALS table PZTs, but still now blips seen in the baffle PD times series...sad

Non-image files attached to this comment
H2 ISC
keita.kawabe@LIGO.ORG - posted 19:19, Thursday 21 June 2012 - last comment - 15:11, Friday 22 June 2012(3214)
One arm test Day One: No light found at IY (Thomas, Elli, Aidan, Richard, Rana, Keita)

We set up a CCD camera looking at ITMY and the entire baffle and some more, sent a green beam from EY, modulated its angle by using PZTs, and found nothing on the camera. No luck with the baffle diode, either.

It seems like the beam is not getting to the corner station at all.

One (or more) of three things should be true:

  1. The ETMY is not pointing to the ITMY. The beam from ALS table is very close to retro-reflection by the ETM when TMS and MCL PZTs are in a neutral position.
  2. Because of a factor of 20 reduction in angle in the telescope, we're not wiggling the angle enough (however I doubt this).
  3. Mode matching is so bad that, by the time the beam reaches the corner, the beam diameter is 4m or something.

Tomorrow we should use HEPI and/or ISI to move everything by a milli radian or so in YAW.

The reason why I think we're already wiggling enough is because, when we wiggle MCL PZTs, I can see on the CCD camera monitoring ETMY that the motion is so big the beam falls off of the primary.  It's not impossible that this is all in translation and not in angle, but we used both PZTs that are placed at different Gouy phase positions and we didn't see anything at IY.

Comments related to this report
keita.kawabe@LIGO.ORG - 20:24, Thursday 21 June 2012 (3216)

As a side note, 1/20 angle reduction only applies to the tilt upstream of TMS telescope (e.g. PZT on the ALS table).

For the rotatoin of TMS itself, it should be 19/20 or 21/20. However, we (Rana and I) felt as if TMS rotation is not producing enough beam rotation. We should calibrate the TMS angle to know how much we're actually tilting it by measuring the displacement of the retro reflected beam on the ALS table.

Also, we need to keep the optical axis fixed to the TMS even during the alignment, which (among other things) means that the TMS angle and the ALS beam angle on the ALS table should always be moved by the same amount. Elli and Thomas are working on the QPD servo for this.

rana.adhikari@LIGO.ORG - 00:24, Friday 22 June 2012 (3217)

Some details on the sensitivity of the search:

* For the ETMY camera, we are only a few meters from the TM. The iris is opened up as far as it can go and still stay nearly focused on the mirror (as seen w/ illuminator on).

* For the ITMY camera, we are ~30m away at the spool position. A large maglight flashlight injected from there can easily be seen on the baffle, the chamber, and the optic.

* The light from the flashlight shows up as ~10000 cts on the baffling photodiodes (mounted near the ITM). Wiggling the green beam from the end produces no signal as seen on the camera and if it produces any signal on the baffle PDs, it must be below the noise level of ~50 cts-rms. Given that the flashlight puts out ~500-1000 lumens, how big of a signal should the green beam make if its the right size?

* Assuming that we inject ~10 mW into the ETMY and that it has a transmission of ~1.5% for 532 nm, we should get ~100 uW of light at ITMY. By comparison, when we hunted for the iLIGO beams, we were injecting ~100 mW at the PRM and ~50-100 uW would get through to the ETM before lock (in the single arm config). The silicon CCDs should pick up the green light much better than the IR, so we should be seeing some beam on the chamber walls, etc. We were injecting non-integer frequency sine waves simultaneously into both pitch and yaw for the PZT, so I think its unlikely that we were unluckily sitting in some dark space.

* We need an SEI & SUS expert to inspect the HEPI/ISI & the ETMY angle biases to ensure that they are pointed in the right direction as determined by the IAS.

rana.adhikari@LIGO.ORG - 09:37, Friday 22 June 2012 (3222)

Video of the swinging beam on the ITM (left) and ETM (right)

Non-image files attached to this comment
rana.adhikari@LIGO.ORG - 10:01, Friday 22 June 2012 (3224)

Images of the ETMY as seen from the spool cam: illuminator on (left) and off (right)

Images attached to this comment
aidan.brooks@LIGO.ORG - 15:11, Friday 22 June 2012 (3235)

I calculated roughly what the intensity of light should be on the baffle PDs. Rana's 500 lumen flashlight was emitting around 10W of light. I assumed this was filling an area roughly 1m in diameter - this yields an intensity of 13W/m^2. The corresponding intensity for the green beam at the ITM (assuming 100uW is transmitted through the ETM and the beam radius is ~5cm at the ITM) is about 12mW/m^2. So the response of the PDs to the green beam should be ~1000x less than the response to the flashlight.

H1 AOS
betsy.weaver@LIGO.ORG - posted 18:47, Thursday 21 June 2012 - last comment - 15:05, Friday 22 June 2012(3213)
ITM-08 (LLO ITM) Ears bonded successfully this week

On behalf of Gerardo and Danny:

Both ears were silicate bonded to the LLO ITM (ITM-08) this week.  The mass has been stowed back in it's cake tin and await shipment out to LLO. 

Comments related to this report
gerardo.moreno@LIGO.ORG - 15:05, Friday 22 June 2012 (3233)

(GMoreno, DSellers, BBland, GMorenoJr.)

ITM08 ears were bonded succesfully, ears with SN 59 and SN 64 were used.

SN 59 was used for side S3, and SN 64 for side S4.

After setting the mass on the base of cake tin, a full inspection of ears and mass was performed, no problems found.

Attached are photos of ears on ITM08.  Also a photo of the mass as I was putting it away in its cake tin, ready for shipping.

Images attached to this comment
X1 SEI
hugo.paris@LIGO.ORG - posted 11:37, Thursday 21 June 2012 - last comment - 16:17, Friday 22 June 2012(3198)
HAM-ISI GS13: inspection of Defective Pod (#94) - Two Broken Flexures and Debris

One of the GS13s of HAM-ISI unit #6 (Pod #94) appeared to be defective this week (see aLog 3183).

Greg and I opened it this morning. Two flexures were broken.

One can see wear on the non-broken side of D0901319-v4, s/n199  (Pic. 3,4). The pattern is very similar to what Jim reported in the SEI Log a few months ago.

We also found a small flexure debris. This debris was identified as being part of D0901318-v3 s/n 124 (see black line on pic. 5).

The pod was tested at reception on a leveled surface. Nothing abnormal could be seen on its response at that time. It was then installed in the ISI. No major event was reported (drop/shock/…). Spectra were taken with the ISI tilted and the pod appeared defective. Any spectra, even with the ISI leveled showed it defective after that.

As it already happened, it is possible that the shocks, endured by the instrument during its shipment, damaged the flexures. In these conditions, tilting the pod back and forth to install it, and/or tilting the ISI for testing purposes, could have lead to the breakage.

We have shockwatch data for this shipment. Greg will post it later on.

Images attached to this report
Comments related to this report
greg.grabeel@LIGO.ORG - 12:46, Thursday 21 June 2012 (3201)
This shipping log was less eventful, especially as the threshold had been set higher, than the previous LHO-LLO. A big difference in this shipment was the large vertical shocks whereas the previous one had seen mostly horizontal. 3 of the 4 events also happened in Louisiana going by the time stamps.
Images attached to this comment
hugo.paris@LIGO.ORG - 13:48, Thursday 21 June 2012 (3202)

It appears that pod #94 was already shiped back to LLO in March to be reworked after a pre-amp issue. see alog #2369.

Nothing was noticed on its flexures then. If the flexures were not changed at LLO, they saw at least 3 Livingston-Hanford trips .

In his aLog, Greg also mention a similar issue experienced on Pod #71: Working at reception. Stopped working afterwards. Opened and broken flexure found.

DCC document Q1100073 sensor issues tracking was updated.

hugo.paris@LIGO.ORG - 16:17, Friday 22 June 2012 (3236)

I went through the history/time-series/spectra available for the Pod #94 (H1).

06/05 the batch of GS13s the Pod #94 belongs to, is sent from LLO.

06/12 Pod #94, and all the pods of that batch, are huddle tested OK (Plot 1). None of their spectra show partiicularities (H2 channel corresponds to the horizontal GS13 #58 which was already in the ISI).

06/13 Pod #94 is installed in HAM-ISI Unit #6 with the Pod #66(H) from the same batch. Interfaces chassis are turned ON. All instruments are tested OK (Plot 2).

06/14 Plot.3 shows that Pod #94 was functional, with no particularities on its spectra, until the interfaces chassis were turned OFF around 3.45 PM

//Barrell nuts appeared to be missing on the walls behind the horizontal GS13s. Interfaces chassis were turned OFF at 3.45pm and GS13s were pulled out of the ISI. Barell nuts were added and GS13s were re-installed. GS13 walls were mounted and interfaces chassis were turned back ON//

06/19 Interfaces chassis are turned back ON around 10am. Pod #94 appears defective right away (Plot 4). It seems like the flexure breakage happened during the re-installation of the GS13s, even though no major event was reported (shock, drop, ...).

Time series before/after re-installation of the GS13s are attached (Plot 5).

 

I can not notice any feature on spectra/time series that would have warned us of upcoming flexure breakages.

While talking about it at today's SEI call, we agreed on gently tilting the GS13s back and forth before huddle testing them at reception, so we can discard the GS13s with flexures that are about to break.

Non-image files attached to this comment
Displaying reports 81481-81500 of 84635.Go to page Start 4071 4072 4073 4074 4075 4076 4077 4078 4079 End